A simple cloud point extraction procedure is presented for the preconcentration of copper in various samples. After complexation by 4-hydroxy-2-mercapto-6-propylpyrimidine (PTU), copper ions are quantitatively extracted into the phase rich in Triton X-114 after centrifugation. Methanol acidified with 0.5 mol L-1 HNO3 was added to the surfactant-rich phase prior to its analysis by flame atomic absorption spectrometry (FAAS). Analytical parameters including concentrations for PTU, Triton X-114 and HNO3, bath temperature, centrifugation rate and time were optimized. The influences of the matrix ions on the recoveries of copper ions were investigated. The detection limits (3SDb/m, n=4) of 1.6 ng mL-1 along with enrichment factors of 30 for Cu were achieved. The proposed procedure was applied to the analysis of environmental samples.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

The microbiological bioassay, UV-spectrophotometry and HPLC methods for assaying gatifloxacin in tablets were compared. Validation parameters such as linearity, precision, accuracy, limit of detection and limit of quantitation were determined. Beer's law was obeyed in the ranges 4.0-14.0 μg/mL for HPLC and UV-spectrophotometric method, and 4.0-16.0 μg/mL for bioassay. All methods were reliable within acceptable limits for antibiotic pharmaceutical preparations being accurate, precise and reproducible. The bioassay and HPLC are more specific than UV-spectrophotometric analysis. The application of each method as a routine analysis should be investigated considering cost, simplicity, equipment, solvents, speed, and application to large or small workloads.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

In the present work, the development of a method based on the coupling of flow analysis (FA), hydride generation (HG), and derivative molecular absorption spectrophotometry (D-EAM) in gas phase (GP), is described in order to determine total antimony in antileishmanial products. Second derivative order (D²224nm) of the absorption spectrum (190 - 300 nm) is utilized as measurement criterion. Each one of the parameters involved in the development of the proposed method was examined and optimized. The utilization of the EAM in GP as detection system in a continuous mode instead of atomic absorption spectrometry represents the great potential of the analytic proposal.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

A new method is described for the determination of the herbicide bispyribac-sodium in surface water, especially from river and irrigated rice water samples. The method involves extraction in solid phase and quantification by high performance liquid chromatography with diode array detection (HPLC-DAD). After optimization of the extraction and separation parameters, the method was validated. The method presented average recoveries of 101.3 and 97.7%, under repeatability and intermediate precision conditions, respectively, with adequate precision (RSD from 0.9 to 7.5%). The method was applied for the determination of bispyribac-sodium in surface water samples with a limit of detection of 0.1 μg L-1.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

A derivative UV spectrophotometric method for determination of estradiol valerate in tablets was validated. The parameters specificity, linearity, precision, accuracy, limit of detection and limit of quantitation were studied according to validation guidelines. The first-order derivative spectra were obtained at N = 5, Δλ = 4.0 nm, and determinations were made at 270 nm. The method showed specificity and linearity in the concentration range of 0.20 to 0.40 mg mL-1. The intra and interday precision data demonstrated the method has good reproducibility. Accuracy was also evaluated and results were satisfactory. The proposed method was successfully applied to a pharmaceutical formulation.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Simultaneous determination of moxifloxacin (MOX) and H2-antagonists was first time developed in bulk and formulations. Purospher STAR C18 (250 x 4.6 mm, 5 μm) column was used. The mobile phase (methanol: water: ACN, 60:45:5 v/v/v, pH 2.7) was delivered at a flow rate of 1.0 mL min-1, eluent was monitored at 236, 270 and 310 nm for cimetidine, famotidine and ranitidine, respectively. The proposed method is specific, accurate (98-103%), precise (intra-day and inter-day variation 0.098-1.970%) and linear (r>0.998). The LOD and LOQ were 0.006-0.018 and 0.019-0.005 μg mL-1, respectively. The statistical parameters were applied to verify the results. The method is applicable to routine analysis of formulations and interaction of MOX with H2-antagonist.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

A dispersive liquid-liquid microextraction procedure coupled to spectrophotometry is described for the determination of the trace levels of Sudan Blue II. Analytical parameters, such as pH, volume of extraction solvent (carbon tetrachloride), volume of dispersant (ethanol), volume of sample, and extraction time, were optimized. Matrix effects were also investigated. Preconcentration factor was found to be 200. Detection limit and relative standard deviation (RSD) were 0.55 µg L-1 and 3.9%, respectively. The procedure was successfully used for the determination of trace levels of Sudan Blue II in food, ink, antifreeze, and industrial waste-water samples.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

A method for the determination of trace amounts of palladium was developed using homogeneous liquid-liquid microextraction via flotation assistance (HLLME-FA) followed by graphite furnace atomic absorption spectrometry (GFAAS). Ammonium pyrrolidine dithiocarbamate (APDC) was used as a complexing agent. This was applied to determine palladium in three types of water samples. In this study, a special extraction cell was designed to facilitate collection of the low-density solvent extraction. No centrifugation was required in this procedure. The water sample solution was added to the extraction cell which contained an appropriate mixture of extraction and homogeneous solvents. By using air flotation, the organic solvent was collected at the conical part of the designed cell. Parameters affecting extraction efficiency were investigated and optimized. Under the optimum conditions, the calibration graph was linear in the range of 1.0-200 µg L-1 with a limit of detection of 0.3 µg L-1. The performance of the method was evaluated for the extraction and determination of palladium in water samples and satisfactory results were obtained. In order to verify the accuracy of the approach, the standard addition method was applied for the determination of palladium in spiked synthetic samples and satisfactory results were obtained.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Silica gel chemically modified with 2-Aminotiazole groups, abbreviated as SiAT, was used for preconcentration of copper, zinc, nickel and iron from kerosene, normally used as a engine fuel for airplanes. Surface characteristics and surface area of the silica gel were obtained before and after chemical modification using FT-IR, Kjeldhal and surface area analysis (B.E.T.). The retention and recovery of the analyte elements were studied by applying batch and column techniques. The experimental parameters, such as shaking time in batch technique, flow rate and concentration of the eluent (HCl- 0.25-2.00 mol L-1) and the amount of silica, on retention and elution, have been investigated. Detection limits of the method for copper, iron, nickel and zinc are 0.77, 2.92, 1.73 and 0.097 mg L-1, respectively. The sorption-desorption of the studied metal ions made possible the development of a preconcentration method for metal ions at trace level in kerosene using flame AAS for their quantification.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

A flow-injection system with sample and reagent addition by the synchronous merging zones approach for calcium determination in milk by flame AAS is proposed. Main parameters were optimized using a factorial design with central point. The optimum conditions were 2.5% (m/v) for La concentration, 8 mL min-1 for the carrier flow-rate, 20 cm for coiled reactor and 250 ìL for sample volume. Different sample preparation procedures were evaluated such as dilution in water or acid and microwave-assisted decomposition using concentrated or diluted acids. The optimized flow system was applied to determine Ca in eleven commercial milk samples and two standard reference materials diluted in water. Similar calcium levels were encountered comparing the results obtained by the proposed method (dilution in water) with those obtained using microwave-oven digestion. Results obtained in two standard reference materials were in agreement at 95% confidence level with those certified. Recoveries of spiked samples were in the 93% - 116% range. Relative standard deviation (n = 12) was < 5.4% and the sample throughput was 150 measurements per hour, corresponding to a consumption of 250 µL of sample and 6.25 mg La per determination.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

National Health Surveillance Agency (ANVISA) established in the decree number 54 maximum allowed levels for Ni and Pb in mineral and natural waters at 20 µg L-1 and 10 µg L-1, respectively. For screening analysis purposes, the high-resolution continuum source flame atomic absorption spectrometry technique (HR-CS FAAS) was evaluated for the fast-sequential determination of nickel and lead in mineral waters.Two atomic lines for Ni (232.003 nm - main and 341.477 nm - secondary) and Pb (217.0005 nm - main and 283.306 nm - secondary) at different wavelength integrated absorbance (number of pixels) were evaluated. Sensitivity enhanced with the increase of the number of pixels and with the summation of the atomic lines absorbances. The main figures of merit associated to the HR-CS FAAS technique were compared with that obtained by line-source flame atomic absorption spectrometry (LS FAAS). Water samples were pre-concentrated about 5-fold by evaporation before analysis. Recoveries of Pb significantly varied with increased wavelength integrated absorbance. Better recoveries (92-93%) were observed for higher number of pixels at the main line or summating the atomic lines (90-92%). This influence was irrelevant for Ni, and recoveries in the 92-104% range were obtained in all situations.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

A simple spectrophotometric method for the determination of cefotaxime, ceftriaxone, cefadroxil and cephalexin with variamine blue is presented. The determination is based on the hydrolysis of β-lactam ring of cephalosporins with sodium hydroxide which subsequently reacts with iodate to liberate iodine in acidic medium. The liberated iodine oxidizes variamine blue to violet colored species of maximum absorption at 556 nm. The absorbance is measured within the pH range of 4.0-4.2. Beer's law is obeyed in the range of 0.5-5.8 µg mL-1, 0.2-7.0 µg mL-1, 0.2-5.0 µg mL-1 and 0.5-8.5 µg mL-1 for cefotaxime, ceftriaxone, cefadroxil and cephalexin respectively. The analytical parameters were optimized and the method is successfully applied for the determination of cefotaxime, ceftriaxone, cefadroxil and cephalexin in pharmaceuticals.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

A novel sensitive and relatively selective kinetic method is presented for the determination of V(V), based on its catalytic effect on the oxidation reaction of Ponceau Xylydine by potassium bromate in presence of 5-sulfosalicylic acid (SSA) as activator. The reaction was monitored spectrophotometrically by measuring the decrease in absorbance of Ponceau Xylydine at 640 nm between 0.5 to 7 min (the fixed time method) in H3PO4 medium at 25ºC. The effect of various parameters such as concentrations of H3PO4, SSA, bromate and Ponceau Xylydine, temperature and ionic strength on the rate of net reaction were studied. The method is free from most interferences, especially from large amounts of V(IV). The decrease in absorbance is proportional to the concentration of V(V) over the entire concentration range tested (1-15 ng mL−1) with a detection limit of 0.46 ng mL-1 (according to statistical 3Sblank/k criterion) and a coefficient of variation (CV) of 1.8% (for ten replicate measurement at 95% confidence level). The proposed method suffers few interferences such as Cr(VI) and Hg(II) ions. The method was successfully applied to the determination of V(V) in tap water, drinking water, bottled mineral water samples and a certified standard reference material such as SRM-1640 with satisfactory results. The vanadium contents of water samples were also determined by FAAS for a comparison. The recovery of spiked vanadium(V) was found to be quantitative and the reproducibility was satisfactory. It was observed that the results of the SRM 1640 were in good agreement with the certified value.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

A simple and rapid spectrophotometric method for the determination of nevirapine is described. The method is based on the reaction of nevirapine with tetrathiocyanatocobalt(II) ion in buffer of pH 4 to form the corresponding complex. Beer's law is obeyed in the range of 0.2 - 2.0 µg mL-1 for nevirapine. The optical parameters such as molar absorptivity, Sandell's sensitivity, detection limit and quantitation limit were found to be 1.16× 10(4) Lmol-1cm-1, 2.09 X 10-3 µg cm-2, 0.073 µg mL-1 and 0.222 µg mL-1 respectively. The optimum reaction conditions and other analytical parameters were evaluated. The statistical evaluation of the method was examined by determining intra-day and inter-day precision. The proposed method has been successfully applied for the determination of nevirapine in pharmaceutical formulations.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

The objective of this study was to optimize and validate the solid-liquid extraction (ESL) technique for determination of picloram residues in soil samples. At the optimization stage, the optimal conditions for extraction of soil samples were determined using univariate analysis. Ratio soil/solution extraction, type and time of agitation, ionic strength and pH of extraction solution were evaluated. Based on the optimized parameters, the following method of extraction and analysis of picloram was developed: weigh 2.00 g of soil dried and sieved through a sieve mesh of 2.0 mm pore, add 20.0 mL of KCl concentration of 0.5 mol L-1, shake the bottle in the vortex for 10 seconds to form suspension and adjust to pH 7.00, with alkaline KOH 0.1 mol L-1. Homogenate the system in a shaker system for 60 minutes and then let it stand for 10 minutes. The bottles are centrifuged for 10 minutes at 3,500 rpm. After the settlement of the soil particles and cleaning of the supernatant extract, an aliquot is withdrawn and analyzed by high performance liquid chromatography. The optimized method was validated by determining the selectivity, linearity, detection and quantification limits, precision and accuracy. The ESL methodology was efficient for analysis of residues of the pesticides studied, with percentages of recovery above 90%. The limits of detection and quantification were 20.0 and 66.0 mg kg-1 soil for the PVA, and 40.0 and 132.0 mg kg-1 soil for the VLA. The coefficients of variation (CV) were equal to 2.32 and 2.69 for PVA and TH soils, respectively. The methodology resulted in low organic solvent consumption and cleaner extracts, as well as no purification steps for chromatographic analysis were required. The parameters evaluated in the validation process indicated that the ESL methodology is efficient for the extraction of picloram residues in soils, with low limits of detection and quantification.