Cloud point extraction for the determination of copper in environmental samples by flame atomic absorption spectrometry
| Data(s) |
01/01/2008
|
|---|---|
| Resumo |
A simple cloud point extraction procedure is presented for the preconcentration of copper in various samples. After complexation by 4-hydroxy-2-mercapto-6-propylpyrimidine (PTU), copper ions are quantitatively extracted into the phase rich in Triton X-114 after centrifugation. Methanol acidified with 0.5 mol L-1 HNO3 was added to the surfactant-rich phase prior to its analysis by flame atomic absorption spectrometry (FAAS). Analytical parameters including concentrations for PTU, Triton X-114 and HNO3, bath temperature, centrifugation rate and time were optimized. The influences of the matrix ions on the recoveries of copper ions were investigated. The detection limits (3SDb/m, n=4) of 1.6 ng mL-1 along with enrichment factors of 30 for Cu were achieved. The proposed procedure was applied to the analysis of environmental samples. |
| Formato |
text/html |
| Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422008000100015 |
| Idioma(s) |
en |
| Publicador |
Sociedade Brasileira de Química |
| Fonte |
Química Nova v.31 n.1 2008 |
| Palavras-Chave | #4-hydroxy-2-mercapto-6-propylpyrimidine #cloud point extraction #copper ion |
| Tipo |
journal article |