15 resultados para Multicommuted
Resumo:
A photometric procedure for the determination of ClO(-) in tap water employing a miniaturized multicommuted flow analysis setup and an LED-based photometer is described. The analytical procedure was implemented using leucocrystal violet (LCV; 4,4', 4 ''-methylidynetris (N, N-dimethylaniline), C(25)H(31)N(3)) as a chromogenic reagent. Solenoid micropumps employed for solutions propelling were assembled together with the photometer in order to compose a compact unit of small dimensions. After control variables optimization, the system was applied for the determination of ClO(-) in samples of tap water, and aiming accuracy assessment samples were also analyzed using an independent method. Applying the paired t-test between results obtained using both methods, no significant difference at the 95% confidence level was observed. Other useful features include low reagent consumption, 2.4 mu g of LCV per determination, a linear response ranging from 0.02 up to 2.0 mg L(-1) ClO(-), a relative standard deviation of 1.0% (n = 11) for samples containing 0.2 mg L(-1) ClO(-), a detection limit of 6.0 mu g L(-1) ClO(-), a sampling throughput of 84 determinations per hour, and a waste generation of 432 mu L per determination.
Resumo:
In this work a downscaled multicommuted flow injection analysis setup for photometric determination is described. The setup consists of a flow system module and a LED based photometer, with a total internal volume of about 170 mu L The system was tested by developing an analytical procedure for the photometric determination of iodate in table salt using N,N-diethyl-henylenediamine (DPD) as the chromogenic reagent. Accuracy was accessed by applying the paired r-test between results obtained using the proposed procedure and a reference method, and no significant difference at the 95% confidence level was observed. Other profitable features, such as a low reagent consumption of 7.3 mu g DPD per determination: a linear response ranging from 0.1 up to 3.0 m IO(3)(-), a relative standard deviation of 0.9% (n = 11) for samples containing 0.5 m IO(3)(-), a detection limit of 17 mu g L(-1) IO(3)(-), a sampling throughput of 117 determination per hour, and a waste generation 600 mu L per determination, were also achieved. (C) 2010 Elsevier B.V. All rights reserved.
Resumo:
A new procedure for spectrofluorimetric determination of free and total glycerol in biodiesel samples is presented. It is based on the oxidation of glycerol by periodate, forming formaldehyde, which reacts with acetylacetone, producing the luminescent 3,5-diacetyl-1,4-dihydrolutidine. A flow system with solenoid micro-pumps is proposed for solution handling. Free glycerol was extracted off-line from biodiesel samples with water, and total glycerol was converted to free glycerol by saponification with sodium ethylate under sonication. For free glycerol, a linear response was observed from 5 to 70 mg L(-1) with a detection limit of 0.5 mg L(-1), which corresponds to 2 mg kg(-1) in biodiesel. The coefficient of variation was 0.9% (20 mg L(-1), n = 10). For total glycerol, samples were diluted on-line, and the linear response range was 25 to 300 mg L(-1). The detection limit was 1.4 mg L(-1) (2.8 mg kg(-1) in biodiesel) with a coefficient of variation of 1.4% (200 mg L(-1), n = 10). The sampling rate was ca. 35 samples h(-1) and the procedure was applied to determination of free and total glycerol in biodiesel samples from soybean, cottonseed, and castor beans.
Resumo:
Liquid-liquid microextraction without phase segmentation was implemented in a multicommuted flow system for determination of the anti-hypertensive diltiazem. The procedure was based on ion pair formation between the drug and the dye bromothymol blue at pH 3.5. The detection was performed without phase separation in a glass tube coupled to a fiber-optics spectrophotometer. The total volume of chloroform was reduced to 50 mu L in comparison with 10 mL consumed in batch. A linear response was observed between 9 and 120 mu mol L(-1), with a detection limit of 0.9 mu mol L(-1) (99.7% confidence level). The coefficient of variation (n = 10), sampling rate and extraction efficiency were estimated as 0.6%, 78 determinations per hour and 61%, respectively. About 30 mu g of bromothymol blue was consumed and the waste volume was 380 mu L per determination. The results for pharmaceutical samples agreed with those obtained by the reference procedure at the 95% confidence level. (C) 2011 Elsevier B.V. All rights reserved.
Resumo:
A flow system designed with solenoid micro-pumps is proposed for the determination of paraquat in natural waters. The procedure involves the reaction of paraquat with dehydroascorbic acid followed by spectrophotometric measurements. The proposed procedure minimizes the main drawbacks related to the standard chromatographic procedure and to flow analysis and manual methods with spectrophotometric detection based on the reaction with sodium dithionite, i.e. high solvent consumption and waste generation and low sampling rate for chromatography and high instability of the reagent in the spectrophotometric procedures. A home-made 10-cm optical-path flow cell was employed for improving sensitivity and detection limit. Linear response was observed for paraquat concentrations in the range 0.10-5.0 mg L-1. The detection limit (99.7% confidence level), sampling rate and coefficient of variation (n = 10) were estimated as 22 mu g L-1, 63 measurements per hour and 1.0%, respectively. Results of determination of paraquat in natural water samples were in agreement with those achieved by the chromatographic reference procedure at the 95% confidence level. (c) 2008 Elsevier B.V. All rights reserved.
Resumo:
The physical and chemical characteristics of peat were assessed through measurement of pH, percentage of organic matter, cationic exchange capacity (CEC), elemental analysis, infrared spectroscopy and quantitative analysis of metals by ICP OES. Despite the material showed to be very acid in view of the percentage of organic matter, its CEC was significant, showing potential for retention of metal ions. This characteristic was exploited by coupling a peat mini-column to a flow system based on the multicommutation approach for the in-line copper concentration prior to flame atomic absorption spectrometric determination. Cu(II) ions were adsorbed at pH 4.5 and eluted with 0.50 mol L(-1) HNO(3). The influence of chemical and hydrodynamic parameters, such as sample pH, buffer concentration, eluent type and concentration, sample flow-rate and preconcentration time were investigated. Under the optimized conditions, a linear response was observed between 16 and 100 mu g L(-1), with a detection limit estimated as 3 mu g L(-1) at the 99.7% confidence level and an enrichment factor of 16. The relative standard deviation was estimated as 3.3% (n = 20). The mini-column was used for at least 100 sampling cycles without significant variation in the analytical response. Recoveries from copper spiked to lake water or groundwater as well as concentrates used in hemodialysis were in the 97.3-111 % range. The results obtained for copper determination in these samples agreed with those achieved by graphite furnace atomic absorption spectrometry (GFAAS) at the 95% confidence level. (C) 2009 Elsevier B.V. All rights reserved.
Resumo:
A flow system exploiting the multicommutation approach is proposed for spectrophotometric determination of tannin in beverages. The procedure is based on the reduction of Cu(II) in the presence of 4,4`-dicarboxy-2,2`-biquinoline, yielding a complex with maximum absorption at 558 nm. Calibration graph was linear (r=0.999) for tannic acid concentrations up to 5.00 mu mol L-1. The detection limit and coefficient of variation were estimated as 10 nmol L-1 (99.7% confidence level) and 1% (1.78 mu mol L-1 tannic acid, n=10), respectively. The sampling rate was 50 determinations per hour. The proposed procedure is more sensitive and selective than the official Folin-Denis method, also minimizing drastically waste generation. Recoveries within 91.8 and 115% were estimated for total tannin determination in tea and wine samples. (C) 2007 Elsevier B.V. All rights reserved.
Resumo:
A simple, rapid, and automated assay for hydrogen peroxide in pharmaceutical samples was developed by combining the multicommutation system with a chemiluminescence (CL) detector. The detection was performed using a spiral flow-cell reactor made from polyethylene tubing that was positioned in front of a photodiode. It allows the rapid mixing of CL reagent and analyte and simultaneous detection of the emitted light. The chemiluminescence was based on the reaction of luminol with hydrogen peroxide catalyzed by hexacyanoferrate(III). The feasibility of the flow system was ascertained by analyzing a set of pharmaceutical samples. A linear response within the range of 2.2-210 μmol l-1 H2O2 with a LD of 1.8 μmol l-1 H2O2 and coefficient of variations smaller than 0.8% for 1.0×10-5 mol l-1 and 6.8×10-5 mol l-1 hydrogen peroxide solutions (n=10) were obtained. Reagents consumption of 90 μg of luminol and 0.7 mg of hexacyanoferrate(III) per determination and sampling rate of 200 samples per hour were also achieved. Copyright © Taylor & Francis Group, LLC.
Resumo:
In this work, a LED (light emitting diode) based photometer for solid phase photometry is described. The photometer was designed to permit direct coupling of a light source (LED) and a photodiode to a flow cell with an optical pathlength of 4 mm. The flow cell was filled with adsorbing solid phase material (C-18), which was used to immobilize the chromogenic reagent 1-(2-thiazolylazo)-2-naphthol (TAN). Aiming to allow accuracy assessment, samples were also analyzed employing ICP OES (inductively coupled plasma optical emission spectrometry) methodology. Applying the paired t-test at the 95% confidence level, no significant difference was observed. Other useful features were also achieved: linear response ranging from 0.05 to 0.85 mg L-1 Zn, limit of detection of 9 mu g L-1 Zn (3 sigma criterion), standard deviation of 1.4% (n = 10), sampling throughput of 36 determinations per h, and a waste generation and reagent consumption of 1.7 mL and of 0.03 mu g per determination, respectively.
Resumo:
Neste trabalho é proposto um fotômetro baseado em LED (diodo emissor de luz) para fotometria em fase sólida. O fotômetro foi desenvolvido para permitir o acoplamento da fonte de radiação (LED) e do fotodetector direto na cela de fluxo, tendo um caminho óptico de 4 mm. A cela de fluxo foi preenchida com material sólido (C18), o qual foi utilizado para imobilizar o reagente cromogênico 1-(2-tiazolilazo)-2-naftol (TAN). A exatidão foi avaliada empregando dados obtidos através da técnica ICP OES (espectrometria de emissão por plasma indutivamente acoplado). Aplicando-se o teste-t pareado não foi observada diferença significativa em nível de confiança de 95%. Outros parâmetros importantes encontrados foram faixa de resposta linear de 0,05 a 0,85 mg L-1 Zn, limite de detecção de 9 µg L-1 Zn (n = 3), desvio padrão de 1,4 % (n = 10), frequência de amostragem de 36 determinações por h, e uma geração de efluente e consumo de reagente de 1,7 mL e 0,03 µg por determinação, respectivamente.
Resumo:
The determination of uric acid in urine shows clinical importance, once it can be related to human organism dysfunctions, such as gout. An analytical procedure employing a multicommuted flow system was developed for the determination of uric acid in urine samples. Cu(II) ions are reduced by uric acid to Cu(I) that can be quantified by spectrophotometry in the presence of 2,2`-biquinoline 4,4`-dicarboxylic acid (BCA). The analytical response was linear between 10 and 100 mu mol L(-1) uric acid with a detection limit of 3.0 mu mol L(-1) (99.7% confidence level). Coefficient of variation of 1.2% and sampling rate of 150 determinations per hour were achieved. Per determination, 32 mu g of CuSO(4) and 200 mu g of BCA were consumed, generating 2.0 mL of waste. Recoveries from 91 to 112% were estimated and the results for 7 urine samples agreed with those obtained by the commercially available enzymatic kit for determination of uric acid. The procedure required 100-fold dilution of urine samples, minimizing sample consumption and interfering effects. In order to avoid the manual dilution step, on-line sample dilution was achieved by a simple system reconfiguration attaining a sampling rate of 95 h(-1). (C) 2009 Elsevier B.V. All rights reserved.
Resumo:
In the current work a Green Analytical Chemistry (GAC) procedure for photometric determination of orthophosphate in river water at mu g L-1 concentration level is described. The flow system module and the LED-based photometer were assembled together to constitute a compact unit in order to allow that a flow cell with optical path-length of 100mm was coupled to them. The photometric procedure based on the molybdenum blue method was implemented employing the multicommuted flow injection analysis approach, which provided facilities to allow reduction of reagent consumption and as well as waste generation. Aiming to prove the usefulness of the system, orthophosphate in river and tap waters was determined. Accuracy was ascertained by spiking samples with orthophosphate solution yielding recoveries ranging from 96% up to 107%. Other profitable features such as a wide linear response range between 10 to 800 mu g L-1 [image omitted]; a detection limit (3 sigma criterion) of 2.4 mu g L-1 [image omitted]; a relative standard deviation (n=7) of 2% using a typical water sample with concentration of 120 mu g L-1 [image omitted]; reagent consumption of 3.0mg ammonium molybdate, 0.3mg hydrazine sulfate, and 0.03mg stannous chloride per determination; a waste generation of 2.4mL per determination; and a sampling throughput of 20 determination per hours were also achieved.
Resumo:
A multicommuted method for determination of chlorine in water samples using a 100-cm cell was developed. In this method, orto-Tolidine reacts with chlorine and the product was monitored at 438 nm. The analytical curve for chlorine was linear in concentration range from 1.34x10-6 to 2.01x10-5 mol L-1 with a detection limit of 9.40x10-8 mol L-1. A sampling rate of 45 h-1and a RSD of 1.0 % (n = 15) were obtained. The method was applied with success for chlorine determination in six water samples.
Resumo:
In this manuscript, an automatic setup for screening of microcystins in surface waters by employing photometric detection is described. Microcystins are toxins delivered by cyanobacteria within an aquatic environment, which have been considered strongly poisonous for humans. For that reason, the World Health Organization (WHO) has proposed a provisional guideline value for drinking water of 1 mu g L-1. In this work, we developed an automated equipment setup, which allows the screening of water for concentration of microcystins below 0.1 mu g V. The photometric method was based on the enzyme-linked immunosorbent assay (ELISA) and the analytical signal was monitored at 458 nm using a homemade LED-based photometer. The proposed system was employed for the detection of microcystins in rivers and lakes waters. Accuracy was assessed by processing samples using a reference method and applying the paired t-test between results. No significant difference at the 95% confidence level was observed. Other useful features including a linear response ranging from 0.05 up to 2.00 mu g L-1 (R-2 =0.999) and a detection limit of 0.03 mu g L-1 microcystins were achieved. (C) 2011 Elsevier B.V. All rights reserved.
Resumo:
A iodação do sal de mesa é considerada o caminho mais eficiente para controlar os Distúrbios por Deficiência de Iodo. Em países tropicais, o elemento pode ser adicionado na forma de KIO3. Para garantir que os níveis ideais do ânion estejam disponíveis ao consumidor, o controle de qualidade do sal consiste numa estratégia fundamental. Sistemas em fluxo com multicomutação representam uma alternativa versátil para o desenvolvimento de procedimentos simples, rápidos e limpos, minimizando o consumo de reagentes e a geração de resíduos. Nesse contexto, um procedimento analítico utilizando sistema com multicomutação e detecção espectrofotométrica foi desenvolvido para a determinação de iodato em sal de mesa. A reação empregada foi baseada na formação de um composto roxo (540 nm) entre iodato (IO3-) e p-aminofenol (PAP) em meio ácido. O tempo de residência da zona de amostra no percurso analítico foi explorado a fim de favorecer a reação lenta e a frequência de amostragem para a melhoria do desempenho analítico. Foram selecionados 2 pulsos para inserção de amostra, 3 pulsos para reagente (PAP 0,25% (m/v) em HCl 0,025 mol L-1), 7 ciclos de amostragem, 200 pulsos de carregador (água), bolha de ar de 1 s (40 µL), reator de 70 cm (3 mm d.i.) e parada de fluxo de 480 s. Resposta linear foi observada entre 2,28x10-5 e 3,65x10-4 mol L-1, descrita pela equação A = 0,2443 + 2030 C, r = 0,997. Limite de detecção (99,7% de confiança), coeficiente de variação (n = 20) e frequência de amostragem foram estimados em 8,2x10-6 mol L-1, 0,42% e 70 determinações por hora, respectivamente. Houve consumo de 1,05 mg de PAP e geração de 0,70 mL de resíduos por determinação. As principais espécies concomitantes presentes na amostra não interferiram na determinação de iodato em concentrações até 8 vezes maiores que as usualmente encontradas. Estudos de adição e recuperação de iodato foram realizados pelo procedimento proposto, obtendo porcentagens de recuperação entre 88 e 104%. O procedimento analítico desenvolvido apresenta sensibilidade adequada para a determinação de iodato em amostra de sal de mesa e elevada frequência de amostragem quando comparado com procedimentos descritos na literatura