This work optimized the HPLC conditions for the simultaneous determination of luteolin, apigenin, myricetin, quercetin and kaempferol in aglycone form, as well defined the best conditions for hydrolysis/extraction of these flavonoids in fruits, using the statistical central composite design and response surface analysis. A reverse phase method was developed using a gradient of methanol/water acidified with 0.3% formic acid as mobile phase and a photodiode array detector. The samples were extracted with methanol/water (50:50 v/v) at 90 ºC. The optimum time and HCl concentration varied for the different fruits investigated, demonstrating the necessity of optimizing these conditions for each fruit analyzed. Good recovery (87.1 to 96.3%), repeatability and linearity were obtained.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

A HPLC method was developed to quantify thymine and thymidine impurities in stavudine bulk drug. The separation was carried out in isocratic mode using methanol/water (20:80) as mobile phase, a C18 column and UV detection at 266 nm. The method provided selectivity based on peak purities and resolution among peaks. It was linear over the range of 0.5-5.0 µg/mL. The quantitation limits were 0.021 µg/mL for thymine and 0.134 µg/mL for thymidine. The average accuracies of three concentrations ranged from 97.06 to 102.61% and precision was close to 1%. The method showed robustness, remaining unaffected by deliberate variations in relevant parameters.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

A simple liquid chromatographic method for the simultaneous determination of creatinine, hippuric acid, mandelic acid, phenylglyoxylic acid and o, m and p-methylhippuric acids was developed and validated. Sample preparation was only dilution with water (1:10), followed by centrifugation. Analysis was performed in a reversed phase column (Lichrospher RP 8ec), 250 x 4.0 mm, with isocratic elution with phosphate buffer pH 2.3 and acetonitrile (90:10, v/v). The method presents adequate linearity, precision and accuracy and allows the simultaneous determination of the biomarkers of exposure to toluene, xylene and styrene together with creatinine, reducing cost and laboratory time.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Bupivacaine (S75-R25, NovaBupi®) is an amide type local anesthetic widely used. The present work consists of the development and validation of analytical methodology for evaluation of NovaBupi® content in the poly-lactide-co-glycolide nanospheres (PLGA-NS) by high performance liquid chromatography. The separation was made using the reversed-phase column LC-18, acetonitrile/phosphate buffer 85:15 v/v as mobile phase and detection at 220 nm. The results obtained show that the analytical methodology is accurate, reproducible, robust and linear over the concentration range 10-220.0 g/mL of NovaBupi®. The method was applied to determine the encapsulation efficiency and evaluate the release profile of NovaBupi®, showing good results.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

A high performance liquid chromatography (HPLC) method has been developed for a rapid determination of nimesulide in dissolution studies. Nimesulide was analyzed using 5 µm Lichrospher® RP-18 column (125 x 4 mm i.d.) and mobile phase acetonitrile: phosphate buffer pH=6.0 (55:45) at a flow-rate of 1.0 mL min-1. Detection was carried out at 300 nm at 25 ºC. The method was applied to analysis of nimesulide in in vitro release studies and showed a rapid and efficient analytical alternative for evaluation of dissolution profile of nimesulide.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

This paper describes the development and validation of simple and selective analytical method for determination of 3.4-methylenedioxymethamphetamine (MDMA) in Ecstasy tablets, using high performance liquid chromatography with fluorescence detection. Analysis was performed in a reversed phase column (LiChrospher 100 C18, 150 x 4.6 mm, 5 µm), isocratic elution with phosphate buffer 25 mmol/L pH 3.0 and acetonitrile (95:5, v/v). The method presents adequate linearity, selectivity, precision and accuracy. MDMA concentration in analyzed tablets showed a remarkable variability (from 8.5 to 59.5 mg/tablet) although the tablet weights were uniform, indicating poor manufacturing control thus imposing additional health risks to the users.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Valproic acid (VA) is a drug used to control seizures in several epileptic conditions. In VA pharmacotherapy, therapeutic drug monitoring is recommended to obtain adequate seizure control and avoid toxicity. The aim of this study was to validate a method for the determination of valproic acid in serum, employing high performance liquid chromatography with diode array detection (HPLC-DAD), after derivatization with phenacyl bromide. The calibration curve (y=0.0133x-0.0025) presented good linearity with r²=0.9999. Accuracy (101-115%), intra-assay precision (4.53-8.15%) and inter-assay precision (3.15-6.77%) were acceptable. The quantification limit was 2.0 µg/mL. The method presented similar results to enzyme immunoassay.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

PEGylation has become a widely applied technique to enhancing in vitro and in vivo stability of therapeutic proteins and to increasing materials biocompatibility. PEG branched structures have proven useful for protein and peptide modification. Furthermore, they may be better than linear structures for many purposes. This paper describes an improved procedure for obtaining 2-arms PEG based on L-lysine. The efficiency of the synthesis was not related to moisture of the raw materials. This procedure does not use hazardous reagents as previous protocols do. It implemented a purification process for obtaining the desired structure with high purity ( > 99%). Finally, the procedure described here allows the obtaining of others PEGylation reagents.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

The high cost of sensitivity commercial calorimeters may represent an obstacle for many calorimetric research groups. This work describes the construction and calibration of a batch differential heat conduction calorimeter with sample cells volumes of about 400 μL. The calorimeter was built using two small high sensibility square Peltier thermoelectric sensors and the total cost was estimated to be about US$ 500. The calorimeter was used to study the excess enthalpy of solution of binary mixtures of liquids, as a function of composition, for the following binary systems of solvents: water + 1,4-dioxane or + dimethylsulfoxide at 298,2 ± 0,5 K.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

A method using ultrasonication extraction for the determination of 17 polycyclic aromatic hydrocarbons (PAHs), selected by the USEPA and NIOSH as "consent decree" priority pollutants, in soil by High Performance Liquid Chromatography (HPLC) was studied. Separation and detection were completed in 20 min with a C18 columm, acetonitrile-water gradient elution and ultraviolet absorption and fluorescence detections. The detection limits, for a 10 µL of solution injection, were less than 9,917 ng/g in UV detection and less than 1,866 ng/g in fluorescence detection. Several organic solvents were tested for extraction of the 17 PAHs from soils. Acetone was the best solvent among the three solvents tested, and the order of the extraction efficiencies was: acetone>methanol>acetonitrile. Ultrasonication using acetone as solvent extraction was used to evaluate the biodegradation of those compounds in contaminated soil during a vermicomposting process.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

The validation of analytical methods is an important step in quality control. The main objective of this study is to propose an HPLC experiment to verify the parameters of validation of chromatographic methods, based on green chemistry principles, which can be used in experimental courses of chemistry and related areas.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

In this work a new experiment using HPLC is proposed in order to explore the role of acidity and the organic modifiers in the determination of methylxanthines in tea and coffee. Multivariate and univariate optimizations of the experimental conditions were used.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Different phenolic compound, 5- caffeoylquinic acid (5-CQA), caffeic acid (AC) and rutin (Ru) contents of yerba-mate (Ilex paraguariensis) Brazilian samples of 06 different regions of São Mateus - Paraná, during natural and accelerated industrial storage, were evaluated. For quantification, a reverse phase HPLC isocratic method was developed and validated using methanol:water (35:65 v/v) acidified with 0.5% acetic acid as mobile phase and a photodiode array detector. The six sample global average contents were (34.90 and 36.10 mg g-1) for 5-CQA, (0.18 mg g-1 and 0.23 mg g-1) for AC and (7.12 and 7.18 mg g-1) for Ru, respectively, for the natural and accelerated storage systems. The results showed that the 5-CQA and Ru content are kept constant during the storage while AC content increase only during accelerated storage.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

An analytical method has been developed and validated for the determination of an association of ampicillins in a lyophilized powder for injection by HPLC. The advantage of chromatographic method other than the microbiological one is that, it is possible to monitor precisely, out-of-specification results in quality control processes and also during stability studies, in which an association of ampicillins is present. The proposed HPLC method was developed by using forced degraded samples, in order to reach a selective analysis of ampicillins when in the presence of their degradation products. It was possible to detect benzatine and through indirect calculation, to determine the ampicillin sodium in the drug sample. The method showed to be selective, accurate, precise, robust and linear (from 45.92 to 36.04 μg mL-1 of total ampicillin and from 14.53 to 43.28 μg mL-1 of benzatine). The accuracy determined from recovery test, gave results in the range of 99.41% of total ampicillin to 100.31% of benzatine. Hence, it can be concluded that the proposed HPLC method is applicable for ampicillins determination.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

The objective of this work was to use the high-pressure homogenizer (HPH) to prepare stable oil/water nanoemulsions presenting narrow particle size distribution. The dispersions were prepared using nonionic surfactants based on ethoxylated ether. The size and distribution of the droplets formed, along with their stability, were determined in a Zetasizer Nano ZS particle size analyzer. The stability and the droplet size distribution in these systems do not present the significant differences with the increase of the processing pressure in the HPH). The processing time can promote the biggest dispersion in the size of particles, thus reducing its stability.