Desenvolvimento e validação de método por cromatografia líquida de alta eficiência para determinação simultânea das impurezas timina e timidina na matéria-prima estavudina
| Data(s) |
01/01/2008
|
|---|---|
| Resumo |
A HPLC method was developed to quantify thymine and thymidine impurities in stavudine bulk drug. The separation was carried out in isocratic mode using methanol/water (20:80) as mobile phase, a C18 column and UV detection at 266 nm. The method provided selectivity based on peak purities and resolution among peaks. It was linear over the range of 0.5-5.0 µg/mL. The quantitation limits were 0.021 µg/mL for thymine and 0.134 µg/mL for thymidine. The average accuracies of three concentrations ranged from 97.06 to 102.61% and precision was close to 1%. The method showed robustness, remaining unaffected by deliberate variations in relevant parameters. |
| Formato |
text/html |
| Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422008000700016 |
| Idioma(s) |
pt |
| Publicador |
Sociedade Brasileira de Química |
| Fonte |
Química Nova v.31 n.7 2008 |
| Palavras-Chave | #HPLC #stavudine #impurities |
| Tipo |