5 resultados para Pachymetry

em BORIS: Bern Open Repository and Information System - Berna - Suiça


Relevância:

10.00% 10.00%

Publicador:

Resumo:

BACKGROUND/AIMS: Rebound tonometry (RT) is performed without anaesthesia with a hand held device. The primary aim was to compare RT with Goldmann applanation tonometry (GAT) and to correlate with central corneal thickness (CCT). The secondary aim was to prove tolerability and practicability of RT under "study conditions" and "routine practice conditions." METHODS: In group 1 (52 eyes/28 patients), all measurements were taken by the same physician, in the same room and order: non-contact optical pachymetry, RT, slit lamp inspection, GAT. Patients were questioned about discomfort or pain. In group 2 (49 eyes/27 patients), tonometry was performed by three other physicians during routine examinations. RESULTS: RT was well tolerated and safe. Intraocular pressure (IOP) ranged between 6 mm Hg and 48 mm Hg. No different trends were found between the groups. RT tended to give slightly higher readings: n = 101, mean difference 1.0 (SD 2.17) mm Hg; 84.1% of RT readings within plus or minus 3 mm Hg of GAT; 95% confidence interval in the Bland-Altman analysis -3.2 mm Hg to +5.2 mm Hg. Both RT and GAT showed a weak positive correlation with CCT (r2 0.028 and 0.025, respectively). CONCLUSIONS: RT can be considered a reliable alternative for clinical screening and in cases where positioning of the head at the slit lamp is impossible or topical preparations are to be avoided.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

BACKGROUND: Several conversion tables and formulas have been suggested to correct applanation intraocular pressure (IOP) for central corneal thickness (CCT). CCT is also thought to represent an independent glaucoma risk factor. In an attempt to integrate IOP and CCT into a unified risk factor and avoid uncertain correction for tonometric inaccuracy, a new pressure-to-cornea index (PCI) is proposed. METHODS: PCI (IOP/CCT(3)) was defined as the ratio between untreated IOP and CCT(3) in mm (ultrasound pachymetry). PCI distribution in 220 normal controls, 53 patients with normal-tension glaucoma (NTG), 76 with ocular hypertension (OHT), and 89 with primary open-angle glaucoma (POAG) was investigated. PCI's ability to discriminate between glaucoma (NTG+POAG) and non-glaucoma (controls+OHT) was compared with that of three published formulae for correcting IOP for CCT. Receiver operating characteristic (ROC) curves were built. RESULTS: Mean PCI values were: Controls 92.0 (SD 24.8), NTG 129.1 (SD 25.8), OHT 134.0 (SD 26.5), POAG 173.6 (SD 40.9). To minimise IOP bias, eyes within the same 2 mm Hg range between 16 and 29 mm Hg (16-17, 18-19, etc) were separately compared: control and NTG eyes as well as OHT and POAG eyes differed significantly. PCI demonstrated a larger area under the ROC curve (AUC) and significantly higher sensitivity at fixed 80% and 90% specificities compared with each of the correction formulas; optimum PCI cut-off value 133.8. CONCLUSIONS: A PCI range of 120-140 is proposed as the upper limit of "normality", 120 being the cut-off value for eyes with untreated pressures or=22 mm Hg. PCI may reflect individual susceptibility to a given IOP level, and thus represent a glaucoma risk factor. Longitudinal studies are needed to prove its prognostic value.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

PURPOSE: To compare the performance of dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT) in measuring intraocular pressure in eyes with irregular corneas. METHODS: GAT and DCT measures were taken in 30 keratoconus and 29 postkeratoplasty eyes of 35 patients after pachymetry and corneal topography. Regression and correlation analyses were performed between both tonometry methods and between tonometry methods and corneal parameters. Bland-Altman plots were constructed. RESULTS: DCT values were significantly higher than GAT values in both study groups: +4.1 +/- 2.3 mm Hg (mean +/- SD) in keratoconus and +3.1 +/- 2.5 mm Hg after keratoplasty. In contrast to DCT, GAT values were significantly higher in postkeratoplasty eyes than in keratoconus. The correlation between the 2 tonometry methods was moderate in keratoconus (Kendall correlation coefficient, tau = 0.34) as well in postkeratoplasty eyes (tau = 0.66). The +/-1.96 SD span of the DCT-GAT differences showed a considerable range: -0.42 to +8.70 mm Hg in keratoconus and -1.87 to +7.98 mm Hg in postkeratoplasty eyes. In the keratoconus group, neither DCT nor GAT correlated significantly with any of the corneal parameters. In the postkeratoplasty group, both DCT and GAT measures showed a moderate positive correlation with corneal steepness, but only DCT had a significant negative correlation with the central corneal thickness (tau = -0.33). CONCLUSIONS: DCT measured significantly higher intraocular pressures than GAT in keratoconus and postkeratoplasty eyes. DCT and GAT measures varied considerably, and DCT was not less dependent on biomechanical properties of irregular corneas than GAT.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

The planning of refractive surgical interventions is a challenging task. Numerical modeling has been proposed as a solution to support surgical intervention and predict the visual acuity, but validation on patient specific intervention is missing. The purpose of this study was to validate the numerical predictions of the post-operative corneal topography induced by the incisions required for cataract surgery. The corneal topography of 13 patients was assessed preoperatively and postoperatively (1-day and 30-day follow-up) with a Pentacam tomography device. The preoperatively acquired geometric corneal topography – anterior, posterior and pachymetry data – was used to build patient-specific finite element models. For each patient, the effects of the cataract incisions were simulated numerically and the resulting corneal surfaces were compared to the clinical postoperative measurements at one day and at 30-days follow up. Results showed that the model was able to reproduce experimental measurements with an error on the surgically induced sphere of 0.38D one day postoperatively and 0.19D 30 days postoperatively. The standard deviation of the surgically induced cylinder was 0.54D at the first postoperative day and 0.38D 30 days postoperatively. The prediction errors in surface elevation and curvature were below the topography measurement device accuracy of ±5μm and ±0.25D after the 30-day follow-up. The results showed that finite element simulations of corneal biomechanics are able to predict post cataract surgery within topography measurement device accuracy. We can conclude that the numerical simulation can become a valuable tool to plan corneal incisions in cataract surgery and other ophthalmosurgical procedures in order to optimize patients' refractive outcome and visual function.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

PURPOSE To determine whether the scleral stroma is affected as much as the corneal stroma in keratoconus. SETTING University Eye Clinic, Bern, Switzerland. DESIGN Comparative case-control study. METHODS Eyes with keratoconus (keratoconus group) and eyes of age-, sex-, and axial length-matched controls (control group) were analyzed. Corneal videokeratometry and pachymetry were performed using a Scheimpflug tomographer (Pentacam). For measurements of the peripheral cornea and the anterior sclera, a spectral-domain anterior segment optical coherence tomography device (Spectralis) was used. RESULTS The study group comprised 51 eyes and the control group, 50 eyes. The mean central corneal thickness in the keratoconus group was statistically significantly lower than in the control group (447.8 μm ± 57.8 [SD] versus 550.5 ± 35.5 μm) (P < .0001). No significant difference in the mean anterior scleral thickness was found between the keratoconus group and the control group (479.1 ± 43.7 μm versus 474.2 ± 43.0 μm) (P =.57). CONCLUSION Although corneal thinning was observed in keratoconus patients, the anterior scleral stroma thickness in these patients seemed to be similar to that in healthy control eyes.