95 resultados para very high pressure liquid chromatography (VHPLC)

em Deakin Research Online - Australia


Relevância:

100.00% 100.00%

Publicador:

Resumo:

Using a column packed with fully porous particles, four methods for controlling the flow rates at which gradient elution runs are conducted in very high pressure liquid chromatography (VHPLC) were tested to determine whether reproducible thermal conditions could be achieved, such that subsequent analyses would proceed at nearly the same initial temperature. In VHPLC high flow rates are achieved, producing fast analyses but requiring high inlet pressures. The combination of high flow rates and high inlet pressures generates local heat, leading to temperature changes in the column. Usually in this case a post-run time is input into the analytical method to allow the return of the column temperature to its initial state. An alternative strategy involves operating the column without a post-run equilibration period and maintaining constant temperature variations for subsequent analysis after conducting one or a few separations to bring the column to a reproducible starting temperature. A liquid chromatography instrument equipped with a pressure controller was used to perform constant pressure and constant flow rate VHPLC separations. Six replicate gradient separations of a nine component mixture consisting of acetophenone, propiophenone, butyrophenone, valerophenone, hexanophenone, heptanophenone, octanophenone, benzophenone, and acetanilide dissolved in water/acetonitrile (65:35, v/v) were performed under various experimental conditions: constant flow rate, two sets of constant pressure, and constant pressure operation with a programmed flow rate. The relative standard deviations of the response factors for all the analytes are lower than 5% across the methods. Programming the flow rate to maintain a fairly constant pressure instead of using instrument controlled constant pressure improves the reproducibility of the retention times by a factor of 5, when plotting the chromatograms in time.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

A series of experiments was conducted to evaluate the degree of band broadening in very high pressure LC due to column connections. Different column manufacturers use slightly different designs for their column fittings. If the same column connections are repeatedly used to attach columns of different origins, different void volumes form between capillary tubes and column inlets. An Agilent Ultra Low Dispersion Kit (tubing id 75 μm) was installed on an Agilent Infinity 1290 ultra HPLC and used to connect successively an Agilent, a Phenomenex, and a Waters column. A series of uracil (unretained) samples were injected and eluted at a wide range of flow rates with a water/acetonitrile mixture as eluent. In order to determine the variance contribution from column connections as accurately as possible a nonretained probe compound was selected because the variance contribution from the column is the smallest for analytes, which have very low k values. Yet, this effect still has an impact on the resolution for moderately retained compounds (k > 2) for narrow-bore columns packed with fine particles, since variance contributions are additive for linear chromatographic systems. Each injection was replicated five times under the same experimental conditions. Then NanoViper column connections (tubing id 75 μm) were used and the same injections were made. This system was designed to minimize connection void volumes for any column. Band variances were calculated as the second central moment of elution peaks and used to assess the degree of band broadening due to the column connections. Band broadening may increase from 3.8 to 53.9% when conventional metal ferrules were used to join columns to connection sites. The results show that the variance contribution from improper connections can generate as much as 60.5% of the total variance observed. This demonstrates that column connections can play a larger role than the column packing with respect to band dispersion.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Various species in genus Hibiscus are traditionally known for their therapeutic attributes. The present study focused on the phytochemical analysis of a rather unexplored species Hibiscus caesius (H. caesius), using high-pressure liquid chromatography coupled with mass spectrometry (HPLC-MS). The analysis revealed five major compounds in the aqueous extract, viz. vanillic acid, protocatechoic acid, quercetin, quercetin glucoside and apigenin, being reported for the first time in H. caesius. Literature suggests that these compounds have important pharmacological traits such as anti-cancer, anti-inflammatory, anti-bacterial and hepatoprotective etc. however, this requires further pharmacological investigations at in vitro and in vivo scale. The above study concluded the medicinal potential of H. caesius.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

An algorithm was developed for 2DHPLC that automated the process of peak recognition, measuring their retention times, and then subsequently plotting the information in a two-dimensional retention plane. Following the recognition of peaks, the software then performed a series of statistical assessments of the separation performance, measuring for example, correlation between dimensions, peak capacity and the percentage of usage of the separation space. Peak recognition was achieved by interpreting the first and second derivatives of each respective one-dimensional chromatogram to determine the 1D retention times of each solute and then compiling these retention times for each respective fraction ‘cut’. Due to the nature of comprehensive 2DHPLC adjacent cut fractions may contain peaks common to more than one cut fraction. The algorithm determined which components were common in adjacent cuts and subsequently calculated the peak maximum profile by interpolating the space between adjacent peaks. This algorithm was applied to the analysis of a two-dimensional separation of an apple flesh extract separated in a first dimension comprising a cyano stationary phase and an aqueous/THF mobile phase as the first dimension and a second dimension comprising C18-Hydro with an aqueous/MeOH mobile phase. A total of 187 peaks were detected.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

To achieve the greatest peak capacity in two-dimensional high performance liquid chromatography (2D-HPLC) a gradient should be operated in both separation dimensions. However, it is known that when an injection solvent that is stronger than the initial mobile phase composition is deleterious to peak performance, thus causing problems when cutting a portion from one gradient into another. This was overcome when coupling hydrophilic interaction with reversed phase chromatography by introducing a counter gradient that changed the solvent strength of the second dimension injection. It was found that an injection solvent composition of 20% acetonitrile in water gave acceptable results in one-dimensional simulations with an initial composition of 5% acetonitrile. When this was transferred to a 2D-HPLC separation of standards it was found that a marked improvement in peak shape was gained for the moderately retained analytes (phenol and dimethyl phthalate), some improvement for the weakly retained caffeine and very little change for the strongly retained n-propylbenzene and anthracene which already displayed good chromatographic profiles. This effect was transferred when applied to a 2D-HPLC separation of a coffee extract where the indecipherable retention profile was transformed to a successful application multidimensional chromatography with peaks occupying 71% of the separation space according to the geometric approach to factor analysis.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

In-silico optimisation of a two-dimensional high performance liquid chromatography (2D-HPLC) separation protocol has been developed for the interogation of methamphetamine samples including model, real world seizure, and laboratory synthesised samples. The protocol used Drylab® software to rapidly identify the optimum separation conditions from a library of chromatography columns. The optimum separation space was provided by the Phenomonex Kinetex PFP column (first dimension) and an Agilent Poroshell 120 EC-C18 column (second dimension). To facilitate a rapid 2D-HPLC analysis the particle packed C18 column was replaced with a Phenomenex Onyx Monolithic C18 withought sacrificing separation performance. The Drylab® optimised and experimental separations matched very closely, highlighting the robust nature of HPLC simulations. The chemical information gained from an intermediate methamphetamine sample was significant and complimented that generated from a pure seizure sample. The influence of the two-dimensional separation on the analytical figures of merit was also investigated. The limits of detection for key analytes in the second dimension determined for methamphetamine (4.59 × 10-⁴ M), pseudoephedrine (4.03 × 10-4 M), caffeine (5.16 × 10-⁴ M), aspirin (9.32 × 10-4 M), paracetamol (5.93 × 10-4 M) and procaine (2.02 × 10-3 M).

Relevância:

100.00% 100.00%

Publicador:

Resumo:

This paper reports an investigation into the temporal stability of aqueous solutions of psilocin and psilocybin reference drug standards over a period of fourteen days. This study was performed using high performance liquid chromatography utilising a (955% vlv) methanol: 10 mM ammonium formate,
pH 3.5 mobile phase and absorption detection at 269 nm. It was found that the exclusion of light significantly prolonged the useful life of standards, with aqueous solutions of both psilocin and psilocybin being stable over a period of seven days.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

A simple and sensitive HPLC method was developed to simultaneously determine CPT-11 and its major metabolite SN-38 in culture media and cell lysates. Camptothecin (CPT) was used as internal standard (I.S.). Compounds were eluted with acetonitrile–50 mM disodium hydrogen phosphate buffer containing 10 mM sodium 1-heptane-sulfonate, with the pH adjusted to 3.0 using 85% (w/v) orthophosphoric acid (27/73, v/v) by a Hyperclon ODS (C18) column (200 mm × 4.6 mm i.d.), with detection at excitation and emission wavelengths of 380 and 540 nm, respectively. The average extraction efficiencies were 96.9–108.3% for CPT-11 in culture media and 94.3–107.2% for CPT-11 in cell lysates; and 87.7–106.8% for SN-38 in culture media and 90.1–105.6% for SN-38 in cell lysates. Within- and between-day precision and accuracy varied from 0.1 to 10.3%. The limit of quantitation (precision and accuracy <20%) was 5.0 and 2.0 ng/ml for CPT-11 and 1.0 and 0.5 ng/ml for SN-38 in culture media and cell lysates, respectively. This method was successfully applied to quantitate the cellular accumulation and metabolism of CPT-11 and SN-38 in H4-II-E, a rat hepatoma cell line.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

We have combined high-performance liquid chromatography (HPLC) separations using a monolithic column with acidic potassium permanganate and tris(2,2′-bipyridyl)ruthenium(II) chemiluminescence detection in a rapid and highly sensitive method to monitor the process of extracting opiate alkaloids from Papaver somniferum. Due to the high flow rates allowed with the monolithic column and the inherent selectivity of the chemiluminescence reactions, the four predominant alkaloids – morphine, codeine, oripavine and thebaine – were determined in less than 2 min. The results obtained with numerous process samples compared favourable with those of the standard HPLC methodology. Limits of detection were 1 × 10−10 M, 5 × 10−10 M, 5 × 10−10 M and 1 × 10−9 M, for morphine, codeine, oripavine and thebaine, respectively.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Adrenergic amines found in extracts of Citrus aurantium (bitter orange) evoke analytically useful chemiluminescence with acidic potassium permanganate in the presence of polyphosphates. From corrected chemiluminescence spectra, the wavelength of maximum intensity for these reactions was 680 ± 5 nm and, using flow injection analysis methodology, limits of detection for synephrine, octopamine, tyramine and hordenine were found to be between 1 × 10−9 and 1 × 10−8 M. We have applied this method of detection to the rapid determination of synephrine in dietary supplements using monolithic column chromatography.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

A high-performance liquid chromatography (HPLC) method for the determination of urea that incorporates automated derivatisation with xanthydrol (9H-xanthen-9-ol) is described. Unlike the classic xanthydrol approach for the determination of urea, which involves the precipitation of dixanthylurea (N,N′-di-9H-xanthen-9-ylurea), the derivatisation procedure employed in this method produces N-9H-xanthen-9-ylurea, which remains in solution and can be quantified using fluorescence detection (λex = 213 nm; λem = 308 nm) after chromatographic separation from interferences. The limit of detection for urea was 5 × 10−8 M (0.003 mg L−1). This method was applied to the determination of urea in human and animal urine and in wine.