On the basis of its electrochemical behaviour a new flow-injection analysis (FIA) method with amperometric detection has been developed for quantification of the herbicide bentazone (BTZ) in estuarine waters. Standard solutions and samples (200 µL) were injected into a water carrier stream and both pH and ionic strength were automatically adjusted inside the manifold. Optimization of critical FIA conditions indicated that the best analytical results were obtained at an oxidation potential of 1.10 V, pH 4.5, and an overall flow-rate of 2.4 mL min–1. Analysis of real samples was performed by means of calibration curves over the concentration range 2.5x10–6 to 5.0x10–5 mol L–1, and results were compared with those obtained by use of an independent method (HPLC). The accuracy of the amperometric determinations was ascertained; errors relative to the comparison method were below 4% and sampling rates were approximately 100 samples h–1. The repeatability of the proposed method was calculated by assessing the relative standard deviation (%) of ten consecutive determinations of one sample; the value obtained was 2.1%.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

A flow injection analysis (FIA) system comprising a cysteine selective electrode as detection system was developed for determination of this amino acid in pharmaceuticals. Several electrodes were constructed for this purpose, having PVC membranes with different ionic exchangers and mediator solvents. Better working characteristics were attained with membranes comprising o-nitrophenyl octyl ether as mediator solvent and a tetraphenylborate based ionic-sensor. Injection of 500 µL standard solutions into an ionic strength adjuster carrier (3x10-3 M) of barium chloride flowing at 2.4mL min-1, showed linearity ranges from 5.0x10-5 to 5.0x10-3 M, with slopes of 76.4±0.6mV decade-1 and R2>0.9935. Slope decreased significantly under the requirement of a pH adjustment, selected at 4.5. Interference of several compounds (sodium, potassium, magnesium, barium, glucose, fructose, and sucrose) was estimated by potentiometric selectivity coefficients and considered negligible. Analysis of real samples were performed and considered accurate, with a relative error to an independent method of +2.7%.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

A flow injection analysis (FIA) system having a chlormequat selective electrode is proposed. Several electrodes with poly(vinyl chloride) based membranes were constructed for this purpose. Comparative characterization suggestedthe use of membrane with chlormequat tetraphenylborate and dibutylphthalate. On a single-line FIA set-up, operating with 1x10-2 mol L-1 ionic strength and 6.3 pH, calibration curves presented slopes of 53.6±0.4mV decade-1 within 5.0x10-6 and1.0x10-3 mol L-1, andsquaredcorrelation coefficients >0.9953. The detection limit was 2.2x10-6 mol L-1 and the repeatability equal to ±0.68mV (0.7%). A dual-channel FIA manifold was therefore constructed, enabling automatic attainment of previous ionic strength andpH conditions and thus eliminating sample preparation steps. Slopes of 45.5±0.2mV decade -1 along a concentration range of 8.0x10-6 to 1.0x10-3 mol L-1 with a repeatability ±0.4mV (0.69%) were obtained. Analyses of real samples were performed, and recovery gave results ranging from 96.6 to 101.1%.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

A procedure for the determination of seven indicator PCBs in soils and sediments using microwave-assisted extraction (MAE) and headspace solid-phase microextraction (HS-SPME) prior to GC-MS/MS is described. Optimization of the HS-SPME was carried out for the most important parameters such as extraction time, sample volume and temperature. The adopted methodology has reduced consumption of organic solvents and analysis runtime. Under the optimized conditions, the method detection limit ranged from 0.6 to 1 ng/g when 5 g of sample was extracted, the precision on real samples ranged from 4 to 21% and the recovery from 69 to 104%. The proposed method, which included the analysis of a certified reference material in its validation procedure, can be extended to several other PCBs and used in the monitoring of soil or sediments for the presence of PCBs.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

O trabalho descrito compreende o desenvolvimento de um anticorpo plástico (MIP, do inglês Molecularly Imprinted Polymer) para o antigénio carcinoembrionário (CEA, do inglês Carcinoembriogenic Antigen) e a sua aplicação na construção de dispositivos portáteis, de tamanho reduzido e de baixo custo, tendo em vista a monitorização deste biomarcador do cancro do colo-retal em contexto Point-of-Care (POC). O anticorpo plástico foi obtido por tecnologia de impressão molecular orientada, baseada em eletropolimerização sobre uma superfície condutora de vidro recoberto por FTO. De uma forma geral, o processo foi iniciado pela electropolimerização de anilina sobre o vidro, seguindo-se a ligação por adsorção do biomarcador (CEA) ao filme de polianilina, com ou sem monómeros carregados positivamente (Cloreto de vinilbenziltrimetilamónio, VB). A última fase consistiu na electropolimerização de o-fenilenodiamina (oPD) sobre a superfície, seguindo-se a remoção da proteína por clivagem de ligações peptídicas, com o auxílio de tripsina. A eficiência da impressão do biomarcador CEA no material polimérico foi controlada pela preparação de um material análogo, NIP (do inglês, Non-Imprinted Polymer), no qual nem a proteína nem o monómero VB estavam presentes. Os materiais obtidos foram caracterizados quimicamente por técnicas de Infravermelho com Transformada de Fourier (FTIR, do inglês, Fourier Transform Infrared Spectroscopy) e microscopia confocal de Raman. Os materiais sensores preparados foram entretanto incluídos em membranas poliméricas de Poli(cloreto de vinilo) (PVC) plastificado, para construção de sensores (biomiméticos) seletivos a CEA, tendo-se avaliado a resposta analítica em diferentes meios. Obteve-se uma boa resposta potenciométrica em solução tampão de Ácido 4-(2-hidroxietil)piperazina-1-etanosulfónico (HEPES), a pH 4,4, com uma membrana seletiva baseada em MIP preparada com o monómero carregado VB. O limite de deteção foi menor do que 42 pg/mL, observando-se um comportamento linear (versus o logaritmo da concentração) até 625 pg/mL, com um declive aniónico igual a -61,9 mV/década e r2>0,9974. O comportamento analítico dos sensores biomiméticos foi ainda avaliado em urina, tendo em vista a sua aplicação na análise de CEA em urina. Neste caso, o limite de deteção foi menor do que 38 pg/mL, para uma resposta linear até 625 pg/mL, com um declive de -38,4 mV/década e r2> 0,991. De uma forma geral, a aplicação experimental dos sensores biomiméticos evidenciou respostas exatas, sugerindo que os biossensores desenvolvidos prossigam estudos adicionais tendo em vista a sua aplicação em amostras de indivíduos doentes.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

This work proposes different kind of solid-contact graphite-based electrodes for the selective determination of sulphonamides (SPHs) in pharmaceuticals, biological fluids and aquaculture waters. Sulfadiazine (SDZ) and sulfamethoxazole (SMX) were selected for this purpose for being the most representative compounds of this group. The template molecules were imprinted in sol–gel (ISG) and the resulting material was used as detecting element. This was made by employing it as either a sensing layer or an ionophore of PVC-based membranes and subsequent potentiometric transduction, a strategy never reported before. The corresponding non-imprinted sol–gel (NISG) membranes were used as blank. The effect of plasticizer and kind/charge of ionic lipophilic additive was also studied. The best performance in terms of slope, linearity ranges and signal reproducibility and repeatability was achieved by PVC membranes including a high dielectric constant plasticizer and 15 mg of ISG particles. The corresponding average slope was −51.4 and −52.4 mV/decade, linear responses were 9.0 × 10−6 and 1.7 × 10−5 M, and limits of detection were 0.74 and 1.3 μg/mL for SDZ and for SMX, respectively. Good selectivity with log Kpot < −0.3 was observed for carbonate, chloride, fluoride, hydrogenocarbonate, nitrate, nitrite, phosphate, cyanide, sulfate, borate, persulphate, citrate, tartrate, salicylate, tetracycline, ciprofloxacin, sulphamerazine, sulphatiazole, dopamine, glucose, galactose, cysteine and creatinine. The best sensors were successfully applied to the analysis of real samples with relative errors ranging from −6.8 to + 3.7%.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

The indiscriminate use of antibiotics in food-producing animals has received increasing attention as a contributory factor in the international emergence of antibiotic-resistant bacteria (Woodward in Pesticide, veterinary and other residues in food, CRC Press, Boca Raton, 2004). Numerous analytical methods for quantifying antibacterial residues in edible animal products have been developed over years (Woodward in Pesticide, veterinary and other residues in food, CRC Press, Boca Raton, 2004; Botsoglou and Fletouris in Handbook of food analysis, residues and other food component analysis, Marcel Dekker, Ghent, 2004). Being Amoxicillin (AMOX) one of those critical veterinary drugs, efforts have been made to develop simple and expeditious methods for its control in food samples. In literature, only one AMOX-selective electrode has been reported so far. In that work, phosphotungstate:amoxycillinium ion exchanger was used as electroactive material (Shoukry et al. in Electroanalysis 6:914–917, 1994). Designing new materials based on molecularly imprinted polymers (MIPs) which are complementary to the size and charge of AMOX could lead to very selective interactions, thus enhancing the selectivity of the sensing unit. AMOX-selective electrodes used imprinted polymers as electroactive materials having AMOX as target molecule to design a biomimetic imprinted cavity. Poly(vinyl chloride), sensors of methacrylic acid displayed Nernstian slopes (60.7 mV/decade) and low detection limits (2.9 × 10−5 mol/L). The potentiometric responses were not affected by pH within 4–5 and showed good selectivity. The electrodes were applied successfully to the analysis of real samples.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

In the present work, the development of a genosensor for the event-specific detection of MON810 transgenic maize is proposed. Taking advantage of nanostructuration, a cost-effective three dimensional electrode was fabricated and a ternary monolayer containing a dithiol, a monothiol and the thiolated capture probe was optimized to minimize the unspecific signals. A sandwich format assay was selected as a way of precluding inefficient hybridization associated with stable secondary target structures. A comparison between the analytical performance of the Au nanostructured electrodes and commercially available screen-printed electrodes highlighted the superior performance of the nanostructured ones. Finally, the genosensor was effectively applied to detect the transgenic sequence in real samples, showing its potential for future quantitative analysis.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

A presente dissertação teve como objetivo o desenvolvimento e caracterização de sensores potenciométricos com base em polímeros de impressão molecular (MIP, do inglês, Molecularly Imprinted Polymer) para a determinação da molécula alvo, a acetilcolina. A acetilcolina (ACh) é um neurotransmissor que está associado à doença de Alzheimer. Os materiais biomiméticos desenvolvidos para a interação com a ACh foram obtidos por polimerização em bulk, recorrendo a uma combinação de nanotubos de carbono com monómeros de anilina, dispersos em solvente plastificante oNFOE e PVC. Para aferir sobre o efeito da impressão de ACh na resposta dos materiais MIP, foram igualmente preparados e avaliados materiais de controlo, ou seja, materiais sem impressão molecular (NIP). O controlo da constituição química destes materiais foi realizado recorrendo a Espectroscopia de Raman e Espectroscopia de Infravermelho com transformada de Fourier (FTIR, do inglês Fourier Transformed Infrared Spectroscopy). Os materiais desenvolvidos foram integrados em membranas seletivas de ião, preparadas com ou sem aditivo iónico lipófilo, de carga negativa ou positiva. A avaliação das características gerais das membranas baseou-se na comparação das caraterísticas dos diversos elétrodos. Estas caraterísticas foram obtidas a partir de curvas de calibração, conseguidas para valores de pH diferentes. Em meio ácido, mais precisamente para pH 4, as membranas com materiais impressos e aditivo aniónico foram as que apresentaram as melhores características analíticas, quer em termos de sensibilidade (+83,86 mV década-1) quer em gama de linearidade (de 3,52×10-5 a 1,73×10-3 M). O estudo de seletividade realizado aos sensores revelou que os elétrodos cuja membrana possuía aditivo aniónico apresentavam menores valores de log KPOT. A presença desse constituinte fez com que a seletividade aumentasse nesses mesmos elétrodos. A espécie menos interferente foi a creatina e a mais interferente a creatinina. Os elétrodos foram, ainda, aplicados em amostras de soro sintético. A qualidade dos resultados obtidos dependeu do nível de concentração em estudo, sendo possível identificar uma região onde os resultados foram exatos e precisos. De uma forma geral, os biossensores com MIP e aditivo aniónico apresentaram um desempenho adequado à prossecução deste estudo em amostras reais.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

A methodology based on microwave-assisted extraction (MAE) and LC with fluorescence detection (FLD) was investigated for the efficient determination of 15 polycyclic aromatic hydrocarbons (PAHs) regarded as priority pollutants by the US Environmental Protection Agency and dibenzo(a,l)pyrene in atmospheric particulate samples. PAHs were successfully extracted from real outdoor particulate matter (PM) samples with recoveries ranging from 81.4±8.8 to 112.0±1.1%, for all the compounds except for naphthalene (62.3±18.0%) and anthracene (67.3±5.7%), under the optimum MAE conditions (30.0 mL of ACN for 20 min at 110ºC). No clean-up steps were necessary prior to LC analysis. LOQs ranging from 0.0054 ng/m3 for benzo( a)anthracene to 0.089 ng/m3 for naphthalene were reached. The validated MAE methodology was applied to the determination of PAHs from a set of real world PM samples collected in Oporto (north of Portugal). The sum of particulate-bound PAHs in outdoor PM ranged from 2.5 and 28 ng/m3.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

This work proposes a novel approach for a suitable orientation of antibodies (Ab) on an immunosensing platform, applied here to the determination of 8-hydroxy-2′-deoxyguanosine (8OHdG), a biomarker of oxidative stress that has been associated to chronic diseases, such as cancer. The anti-8OHdG was bound to an amine modified gold support through its Fc region after activation of its carboxylic functions. Non-oriented approaches of Ab binding to the platform were tested in parallel, in order to show that the presented methodology favored Ab/Ag affinity and immunodetection of the antigen. The immunosensor design was evaluated by quartz-crystal microbalance with dissipation, atomic force microscopy, electrochemical impedance spectroscopy (EIS) and square-wave voltammetry. EIS was also a suitable technique to follow the analytical behavior of the device against 8OHdG. The affinity binding between 8OHdG and the antibody immobilized in the gold modified platform increased the charge transfer resistance across the electrochemical set-up. The observed behavior was linear from 0.02 to 7.0 ng/mL of 8OHdG concentrations. The interference from glucose, urea and creatinine was found negligible. An attempt of application to synthetic samples was also successfully conducted. Overall, the presented approach enabled the production of suitably oriented Abs over a gold platform by means of a much simpler process than other oriented-Ab binding approaches described in the literature, as far as we know, and was successful in terms of analytical features and sample application.