Development of an FIA system with amperometric detection for determination of bentazone in estuarine waters


Autoria(s): Cerejeira, Rodrigo P. A. G.; Delerue-Matos, Cristina; Vaz, M. Carmo V. F.
Data(s)

27/11/2013

27/11/2013

2002

Resumo

On the basis of its electrochemical behaviour a new flow-injection analysis (FIA) method with amperometric detection has been developed for quantification of the herbicide bentazone (BTZ) in estuarine waters. Standard solutions and samples (200 µL) were injected into a water carrier stream and both pH and ionic strength were automatically adjusted inside the manifold. Optimization of critical FIA conditions indicated that the best analytical results were obtained at an oxidation potential of 1.10 V, pH 4.5, and an overall flow-rate of 2.4 mL min–1. Analysis of real samples was performed by means of calibration curves over the concentration range 2.5x10–6 to 5.0x10–5 mol L–1, and results were compared with those obtained by use of an independent method (HPLC). The accuracy of the amperometric determinations was ascertained; errors relative to the comparison method were below 4% and sampling rates were approximately 100 samples h–1. The repeatability of the proposed method was calculated by assessing the relative standard deviation (%) of ten consecutive determinations of one sample; the value obtained was 2.1%.

Identificador

DOI: 10.1007/s00216-002-1344-5

1618-2642

1618-2650

http://hdl.handle.net/10400.22/2989

Idioma(s)

eng

Publicador

Springer-Verlag

Relação

Analytical and Bioanalytical Chemistry; Vol. 373, Issue 4-5

http://link.springer.com/article/10.1007%2Fs00216-002-1344-5

Direitos

openAccess

Palavras-Chave #Amperometry #Bentazone #FIA system #Herbicides #Estuarine waters
Tipo

article