In this work, the scattered X-ray beams produced by a mammography unit with a Mo/Mo, Mo/Rh and W/Rh anode/filter combinations were applied in the evaluation of the Hp(10, 0) and mean conversion coefficients from air kerma to the personal dose equivalent ((C) over barH(p(10,0 degrees))). The higher values of H(p)(10,0 degrees) are related to the Mo/Rh combination whereas the lower ones are for the W/Rh target/filter. (C) over barH(p(10,0 degrees)) values are in the range 0.19-0.54 Sv/Gy, where the higher values comprise the W/Rh combination. (C) 2009 Elsevier Ltd. All rights reserved.
Resumo:
In order to validate the Geant4 toolkit for dosimetry applications, simulations were performed to calculate conversion coefficients h(10, alpha) from air kerma free-in-air to personal dose equivalent Hp(10, a). The simulations consisted of two parts: the production of X-rays with radiation qualities of narrow and wide spectra, and the interaction of radiation with ICRU tissue-equivalent and ISO water slab phantoms. The half-value layers of the X-ray spectra obtained by simulation were compared with experimental results. Mean energy, spectral resolution, half-value layers and conversion coefficients were compared with ISO reference values. The good agreement between results from simulation and reference data shows that the Geant4 is suitable for dosimetry applications which involve photons with energies in the range of ten to a few hundreds of keV. (C) 2008 Elsevier Ltd. All rights reserved.
Resumo:
The protective shielding design of a mammography facility requires the knowledge of the scattered radiation by the patient and image receptor components. The shape and intensity of secondary x-ray beams depend on the kVp applied to the x-ray tube, target/filter combination, primary x-ray field size, and scattering angle. Currently, shielding calculations for mammography facilities are performed based on scatter fraction data for Mo/Mo target/filter, even though modern mammography equipment is designed with different anode/filter combinations. In this work we present scatter fraction data evaluated based on the x-ray spectra produced by a Mo/Mo, Mo/Rh and W/Rh target/filter, for 25, 30 and 35 kV tube voltages and scattering angles between 30 and 165 degrees. Three mammography phantoms were irradiated and the scattered radiation was measured with a CdZnTe detector. The primary x-ray spectra were computed with a semiempirical model based on the air kerma and HVL measured with an ionization chamber. The results point out that the scatter fraction values are higher for W/Rh than for Mo/Mo and Mo/Rh, although the primary and scattered air kerma are lower for W/Rh than for Mo/Mo and Mo/Rh target/filter combinations. The scatter fractions computed in this work were applied in a shielding design calculation in order to evaluate shielding requirements for each of these target/filter combinations. Besides, shielding requirements have been evaluated converting the scattered air kerma from mGy/week to mSv/week adopting initially a conversion coefficient from air kerma to effective dose as 1 Sv/Gy and then a mean conversion coefficient specific for the x-ray beam considered. Results show that the thickest barrier should be provided for Mo/Mo target/filter combination. They also point out that the use of the conversion coefficient from air kerma to effective dose as 1 Sv/Gy is conservatively high in the mammography energy range and overestimate the barrier thickness. (c) 2008 American Association of Physicists in Medicine.