373 resultados para CASTE DETERMINATION
Resumo:
A sol-gel derived ceramic-carbon composite electrode is used for fabrication of a new type of optical fiber biosensor based on luminol electrochemiluminescence (ECL). The electrode consists of graphite powder impregnated with glucose oxidase in a silicate network. In this configuration, the immobilized enzyme oxidizes glucose to liberate hydrogen peroxide and graphite powder provides percolation conductivity for triggering the ECL between luminol and the liberated hydrogen peroxide. Both of the reactions occur simultaneously on the surface of the composite electrode, thereby the response of the biosensor is very fast. The peak intensity was achieved within only 20 s after glucose injection. In addition, the electrode could be renewed by a simple mechanical polishing step in case of contamination or fouling. The linear range extends from 0.01 to 10 mM for glucose and the detection limit is about 8.16 muM. The renewal repeatability and stability of the biosensor are also investigated in detail.
Resumo:
Background: Capillary electrophoresis (CE) with tris(2,2'-bipyridyl)ruthenium(II) [Ru(bpy)(3)(2+)]-electro-generated chemiluminescence (ECL) detection is a promising method for clinical analysis. In this study, a method combining CE with Ru(bpy)(3)(2+) ECL (CE-ECL) detection that can be applied to amine-containing clinical species was developed, and the performance of CE-ECL as a quantitative method for determination of sulpiride in human plasma or urine was evaluated. Methods: Sulpiride was separated by capillary zone electrophoresis in uncoated fused-silica capillaries [510 cm x 25 mum (i.d.)] filled with phosphate buffer (pH 8.0 and a driving voltage of +15 kV, with end-column Ru(bpy)(3)(2+) ECL detection. A platinum disc electrode was used as working electrode. Sulpiride in human plasma or urine samples (100 muL) was extracted by a double-step liquid-liquid extraction procedure, dried under nitrogen at 35 degreesC in a water bath, and reconstituted with 100 muL of filtered water. The extraction solvent was ethyl acetate-dichloromethane (5:1 by volume). Results: Under optimum conditions (pH 8.0 phosphate buffer, injection for 6 s at 10 kV, and +1.2 V as detection potential), separation of sulpiride was accomplished within 4 min. The calibration curve was linear over a concentration range of 0.05-25.0 mumol/L, and the limit of detection was 2.9 x 10(-8) mol/L for sulpiride. Intra- and interday CVs for ECL intensities were <6%. Extraction recoveries of sulpiride were 95.6-101% with CVs of 2.9-6.0%. The method was,clinically validated for patient plasma and urine samples. Conclusions: CE combined with Ru(bpy)(3)(2+) ECL is reproducible, precise, selective, and enables the analysis of sulpiride in human plasma and urine. It thus is of value for rapid and efficient analysis of amine-containing analytes of clinical interest.
Resumo:
A new chemiluminescence(CL) system for the determination of ascorbic acid has been established. By the fast reduction reaction between chromium(VI) and ascorbic acid, chromium(M was generated to react with luminol and hydrogen peroxide in alkaline aqueous solution and hydrogen peroxide to produce CL. The CL emission intensity was correlated with ascorbic acid concentration in the range 8.0 x 10(-9) to 1.6 x 10(-4) mol/L, and the detection limit was 8.0 x 10(-9) mol/L ascorbic acid. The relative standard deviation (n = 11) for 1.0 x 10(-6) mol/L ascorbic acid is 0.9%. The method has been applied to the determination of ascorbic acid in vitamin C tablets with satisfactory results.
Resumo:
The molecular weight of recombinant hirudin ( rHV-2) was determined rapidly by matrix-assisted laser desorption/ionization time of fight mass spectrometry (MALDI-TOF-MS). The effects of the three types of matrixes were compared and discussed, alpha-cynao-4-hydroxycinnamic acid was proved to be the best matrix. It showed that MALDI-TOF-MS was superior to the traditional method of molecular weight determination of the biological macromolecules. The mass spectrum data proved that the primary structure of rHV-2 was correct and there was no amino acid deletion, mutation and modification in its expression, refolding and purification.
Resumo:
The acid-base equilibrium of self-assembly monolayer(SAM) of mercaptopropionic acid on gold electrode was studied by capacitance titration. The change of capacitance was due to protonation and deprotonation of the modified electrode surface. This method can be used to measure pH value in solution and capacitance of solid/liquid on SAMs surface synchronously. The shift of pK(a) was observed during the experiment as previous reports. The factors that affect the capacitance measurement were discussed in detail.
Resumo:
A method was developed for the determination of trace and ultratrace amounts of REE. Cd. In. Tl. Th. Nb, Ta. Zr and Hf in soils and sediments. With NaOH-Na2O2 as the flux. Ti(OH)(4)-Fe(OH)(3) co-precipitation as the preconcentration technique and inductively coupled plasma mass spectrometry (ICP-MS) for measurement, the whole procedure was concise and suitable for batch analysis of multi-element solutions. An investigation was carried out of the Ti(OH)(4)-Fe(OH)(3) co-precipitation system, and the results obtained showed that the natural situation of Ti tightly coexisting with Nb. Ta, Zr and Hf in geological samples plays a very important role in the complete co-precipitation of the four elements. The accuracy of this procedure was established using six Chinese soil and sediment certified reference materials (GSS and GSD). and the relative errors between the found and certified values were mostly below 10%.
Resumo:
chemiluminescence suppression method for the determination of ascorbic acid based on Luminol-KIO4-H2O2-ascorbic acid system was established. The linear range for ascorbic acid is 1.0 x 10(-7) similar to 1.0 x 10(-5) mol/L and the detection limit is 6.0 x 10(-8) mol/L. The relative standard deviation (n = 11) is 1.0% for 8.0 x 10(-7) mol/L ascorbic acid. The method has been used to determine the content of ascorbic acid in tablets and injections with satisfactory results.
Resumo:
An amperometric biosensor for monitoring phenols in the organic phase was constructed by the silica sol-gel immobilization of tyrosinase on a glassy carbon electrode. The organic-inorganic hybrid materials with different sol-gel precursors and polymers were optimized, and the experimental conditions, such as the effect of the solvent, operational potential and enzyme loading were explored for the optimum analytical performance of the enzyme electrode. The biosensor can reach 95% of steady-state current in about 18 s, and the trend in the sensitivity of different phenols is as follows: catechol > phenol >p-cresol. In addition, the apparent Michaelis-Menten constants (K-m(app)) and the stability of the enzyme electrode were discussed. (C) 2000 Elsevier Science S.A. All rights reserved.
Resumo:
Three beta -blockers (propranolol, timolol, acebutolol) were separated by capillary electrophoresis (CE) and detected by end-column electrochemical detection (EC). In the present work, a carbon fiber (33 mum) electrode was used as the working electrode. The effect of the buffer concentration, buffer pH, detection potential and separation voltage on the separation of analytes and behavior of electrochemical detection was systematically investigated. The optimum conditions determined were as following: 40 cm length, 25 mum i.d. capillary; 17.5 kV separation voltage; 2 s injection at 15 kV; 70mM phosphate buffer, pH 3.5; detection potential + 1.2V (vs. Ag/AgCl). Under these conditions, the linear ranges of beta -blockers were over three orders of magnititude and the low detection limit of 10(-8)M was obtained. This method was also applied to detect the simulated urine sample.
Resumo:
Inductively coupled plasma mass spectrometry (ICP-MS),a highly sensitive inorgnic analytic technique,fits to determine ultra-nace rare-earth elements in human plasma. Under the optimized conditions detection limits for 15 rare-earth elements are in the range of 0.7 (for Eu)-5.4 (for Gd) ng.L-1. Indium as an internal standard element is used to compensate for matrix suppression effect and sensitivity drift. Three kinds of preparation methods, diluted with 1% HNO3, digested with HNO3-H2O2 and with HNO3-HClO4, are checked and compared,and the former is the simplest way to be measured. The samples diluted with 1% HNO3, stored in 4 degrees C, are very steady for 16 days. With the method, 11 healthy plasma samples in Changchun area of China are analysed.
Resumo:
Capillary electrophoresis with amperometric detection is evaluated for the simultaneous determination of 2-aminothiazole (A), 2-amino-benzothiazole (AB), 2-mercaptobenzothiazole (AM). The cyclic voltammogram, hydrodynamic voltammogram, effect of pH, concentration of buffer and separation voltage on the separation and the detection were studied. The conditions were optimized as follows: 50 mM phosphate buffet; pH 6.0, 2s at 17.5 kV sample injection, separation at 17.5 kV, 1.2 V as detection potential. The method provided low detection limit as 0.5 mu M, 0.05 mu M and 0.01 mu M, wide linear range 2-200 mu M, 10-200 mu M and 0.025-100 mu M for A, AB, and AM, respectively. The variations in peak current and migration time for 15 consecutive injections of a standard containing 5 mu M each compound were 3.7, 2.1, and 3.9%, and 1.2, 0.8, and 1.2%, for A, AB and AM, respectively. This method was employed to analyze river water.
Resumo:
A capillary electrophoresis-amperometric detection system was developed for the determination of propranolol (PRO) at a 33 mu m carbon fiber microdisk electrode (CFE). The cyclic voltammogram, the hydrodynamic voltammograms and the effect of pH were studied. Under the optimum conditions: separation Voltage 15 kV; injection 3 s at 15 kV; 10 mM pH 7.5 phosphate buffer, 1.15 V (vs. Ag/AgCl) detection potential, the detection limit (LOD) for PRO was 0.05 mu M (S/N = 3). The response for PRO was linear over two orders of magnitude with a linear correlation coefficient of 0.994. The feasibility of this method was demonstrated by the detection of PRO in urine sample.
Resumo:
Capillary electrophoresis (CE) with end-column electrochemical detection (EC) of barbituric acid (BA) and 2-thiobarbituric acid (TA) has been described. Under optimum condition, BA and TA were separated satisfactorily, and a response of high sensitivity and stability was obtained at a detection potential of 1.25 V versus Ag/AgCl. Optimized end-column detection provides detection limit as low as 0.5 and 0.1 mu M for BA and TA, respectively. The calibration graph was linear over three orders of magnitude. The relative standard deviations (n = 10) of peak currents and migration times obtained for both BA and TA were 3.4, 3.7, and 1.7, 1.2%, respectively. The proposed method has been applied to analyze water sample with satisfactory results. (C) 2000 Elsevier Science B.V. All rights reserved.
Resumo:
A novel method for the sensitive and selective determination of peroxydisulfate was proposed. It was based on an electrochemiluminescent method with Cr(bpy)(3)(3+). The applied potential was set at -0.5 V in Cr(bpy)(3)(3+) system which was much more positive than those in Ru(bpz)(3)(2+) and Ru(bpy)(3)(2+) systems. A linear calibration was obtained over the range of 7-100 mu mol l(-1) with a correlation coefficient of 0.993. The relative standard deviation was 2% for 10 replicate injections of 0.01 mol l(-1) S2O82- solution. The detection limit (S/N=3) was 1 mu mol l(-1). The method was successfully applied to determination of peroxydisulfate in phase transfer catalysis. (C) 2000 Elsevier Science B.V. All rights reserved.
Resumo:
The self-assembled monolayer(SAM) of 11-mercaptoundecanoic acid [HS(CH2)(10)COOH] was formed on a gold electrode and the effect of the charge of end group on the electrochemical response of Fe(CN)(6)(3-) at the SAM modified electrode was studied by cyclic voltammetry. At high pH, when the -COOH groups are dissociated, the current of Fe(CN)(6)(3-) is suppressed; as the solution pH is lowered, the current of Fe(CN)(6)(3-) increases. The electrochemical titration curve was obtained by correlating the currents of Fe(CN)(6)(3-) to the different pH values of electrolyte, from which the surface pK(a) was obtained to be 3. 0+/-0. 2. Furthermore, the reason of small pK(a) value was explained using SAMs of different surface coverage.