Preconcentración y atomización con tubo en la llama para la determinación de Cadmio en agua de mar
| Data(s) |
01/01/2011
|
|---|---|
| Resumo |
A flow injection on-line pre-concentration system coupled to thermospray flame furnace atomic absorption spectrometry (TS-FF-AAS) for cadmium determination at sub μg L-1 levels in seawater samples was developed. The on-line system was evaluated by analysing cadmium containing in a synthetic seawater matrix (2.5% m/v NaCl, 0.5% m/v MgCl2 and 0.8% m/v CaCl2). A sample volume of 2 mL allows determining Cd with a detection limits of 30 ng L-1 (3* σblank/slope), pre-concentration factor of 34 and repeatability of 1,8% (calculated as RSD, N=8 and containing 200 ng L-1 of Cd ). |
| Formato |
text/html |
| Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422011000400025 |
| Idioma(s) |
es |
| Publicador |
Sociedade Brasileira de Química |
| Fonte |
Química Nova v.34 n.4 2011 |
| Palavras-Chave | #cadmium #seawater #TS-FF-AAS |
| Tipo |
journal article |