An ex vivo comparison of root canal length determination by three electronic apex locators at positions short of the apical foramen
Contribuinte(s) |
UNIVERSIDADE DE SÃO PAULO |
---|---|
Data(s) |
19/10/2012
19/10/2012
2010
|
Resumo |
Objective. The aim of this study was to evaluate the precision of working length determination of 3 electronic apex locators (EALs): Root ZX, RomiApex D-30, and Ipex at 0.0 mm, at the apical foramen (AF), and at 1.0 mm short of the AF. Methodology. Thirty-eight mandibular premolars had their real lengths previously determined. Electronic measurements were determined at 1.0 mm, followed by measurements at 0.0 mm, performed in triplicate. Results. Precision of devices at 1.0 mm and 0.0 mm were: 94.7% and 97.4%, respectively (Root ZX); 78.9% and 97.4% (RomiApex D-30); and 76.3% and 97.4% (Ipex). Although no statistical differences were observed between the EALs at 0.0, at 1.0 mm Root ZX performed significantly better than the others. Conclusion. The EALs had acceptable precision when measuring the working length at the AF. However, when used at levels short of the AF, only Root ZX did not suffer a significant negative effect on precision. (Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2010;110:e57-e61) |
Identificador |
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, v.110, n.2, p.E57-E61, 2010 1079-2104 http://producao.usp.br/handle/BDPI/25918 10.1016/j.tripleo.2010.03.005 |
Idioma(s) |
eng |
Publicador |
MOSBY-ELSEVIER |
Relação |
Oral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology |
Direitos |
restrictedAccess Copyright MOSBY-ELSEVIER |
Palavras-Chave | #IN-VITRO EVALUATION #WORKING LENGTH #ACCURACY #ZX #ABILITY #DEVICE #TEETH #Dentistry, Oral Surgery & Medicine |
Tipo |
article original article publishedVersion |