Aptamer-based label-free method for hemin recognition and DNA assay by capillary electrophoresis with chemiluminescence detection
Data(s) |
2007
|
---|---|
Resumo |
An aptamer-based label-free approach to hemin recognition and DNA assay using capillary electrophoresis with chemiluminescence detection is introduced here. Two guanine-rich DNA aptamers were used as the recognition element and target DNA, respectively. In the presence of potassium ions, the two aptamers folded into the G-quartet structures, binding hemin with high specificity and affinity. Based on the G-quartet-hemin interactions, the ligand molecule was specifically recognized with a K (d)approximate to 73 nM, and the target DNA could be detected at 0.1 mu M. In phosphate buffer of pH 11.0, hemin catalyzed the H2O2-mediated oxidation of luminol to generate strong chemiluminescence signal; thus the target molecule itself served as an indicator for the molecule-aptamer interaction, which made the labeling and/or modification of aptamers or target molecules unnecessary. This label-free method for molecular recognition and DNA detection is therefore simple, easy, and effective. |
Identificador | |
Idioma(s) |
英语 |
Fonte |
Li T;Li BL;Dong SJ.Aptamer-based label-free method for hemin recognition and DNA assay by capillary electrophoresis with chemiluminescence detection,ANALYTICAL AND BIOANALYTICAL CHEMISTRY,2007 ,389(3):887-893 |
Palavras-Chave | #FUNCTIONALIZED AU NANOPARTICLES #PEROXIDASE-ACTIVITY #TELOMERASE ACTIVITY #IMMUNOASSAY #PROTEINS #DNAZYME #COMPLEX #CELLS |
Tipo |
期刊论文 |