The capabilities of a high-resolution (HR), accurate mass spectrometer (Exactive-MS) operating in full scan MS mode was investigated for the quantitative LC/MS analysis of drugs in patients' plasma samples. A mass resolution of 50,000 (FWHM) at m/z 200 and a mass extracted window of 5 ppm around the theoretical m/z of each analyte were used to construct chromatograms for quantitation. The quantitative performance of the Exactive-MS was compared with that of a triple quadrupole mass spectrometer (TQ-MS), TSQ Quantum Discovery or Quantum Ultra, operating in the conventional selected reaction monitoring (SRM) mode. The study consisted of 17 therapeutic drugs including 8 antifungal agents (anidulafungin, caspofungin, fluconazole, itraconazole, hydroxyitraconazole posaconazole, voriconazole and voriconazole-N-oxide), 4 immunosuppressants (ciclosporine, everolimus, sirolimus and tacrolimus) and 5 protein kinase inhibitors (dasatinib, imatinib, nilotinib, sorafenib and sunitinib). The quantitative results obtained with HR-MS acquisition show comparable detection specificity, assay precision, accuracy, linearity and sensitivity to SRM acquisition. Importantly, HR-MS offers several benefits over TQ-MS technology: absence of SRM optimization, time saving when changing the analysis from one MS to another, more complete information of what is in the samples and easier troubleshooting. Our work demonstrates that U/HPLC coupled to Exactive HR-MS delivers comparable results to TQ-MS in routine quantitative drug analyses. Considering the advantages of HR-MS, these results suggest that, in the near future, there should be a shift in how routine quantitative analyses of small molecules, particularly for therapeutic drugs, are performed.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

A sensitive and specific ultra performance liquid chromatography-tandem mass spectrometry method for the simultaneous quantification of nicotine, its metabolites cotinine and trans-3'-hydroxycotinine and varenicline in human plasma was developed and validated. Sample preparation was realized by solid phase extraction of the target compounds and of the internal standards (nicotine-d4, cotinine-d3, trans-3'-hydroxycotinine-d3 and CP-533,633, a structural analog of varenicline) from 0.5mL of plasma, using a mixed-mode cation exchange support. Chromatographic separations were performed on a hydrophilic interaction liquid chromatography column (HILIC BEH 2.1×100mm, 1.7μm). A gradient program was used, with a 10mM ammonium formate buffer pH 3/acetonitrile mobile phase at a flow of 0.4mL/min. The compounds were detected on a triple quadrupole mass spectrometer, operated with an electrospray interface in positive ionization mode and quantification was performed using multiple reaction monitoring. Matrix effects were quantitatively evaluated with success, with coefficients of variation inferior to 8%. The procedure was fully validated according to Food and Drug Administration guidelines and to Société Française des Sciences et Techniques Pharmaceutiques. The concentration range was 2-500ng/mL for nicotine, 1-1000ng/mL for cotinine, 2-1000ng/mL for trans-3'-hydroxycotinine and 1-500ng/mL for varenicline, according to levels usually measured in plasma. Trueness (86.2-113.6%), repeatability (1.9-12.3%) and intermediate precision (4.4-15.9%) were found to be satisfactory, as well as stability in plasma. The procedure was successfully used to quantify nicotine, its metabolites and varenicline in more than 400 plasma samples from participants in a clinical study on smoking cessation.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Posaconazole (POS) is a new antifungal agent for prevention and therapy of mycoses in immunocompromised patients. Variable POS pharmacokinetics after oral dosing may influence efficacy: a trough threshold of 0.5 ?g/ml has been recently proposed. Measurement of POS plasma concentrations by complex chromatographic techniques may thus contribute to optimize prevention and management of life-threatening infections. No microbiological analytical method is available. The objective of this study was to develop and validate a new simplified ultra-performance liquid chromatography tandem mass spectrometry (UPLC-MS/MS) method and a sensitive bioassay for quantification of POS over the clinical plasma concentration range. The UPLC-MS/MS equipment consisted of a triple quadrupole mass spectrometer, an electrospray ionization (ESI) source, and a C(18) analytical column. The Candida albicans POS-hypersusceptible mutant (MIC of 0.002 ?g/ml) ?cdr1 ?cdr2 ?flu ?mdr1 ?can constructed by targeted deletion of multidrug efflux transporters and calcineurin genes was used for the bioassay. POS was extracted from plasma by protein precipitation with acetonitrile-methanol (75%/25%, vol/vol). Reproducible standard curves were obtained over the range 0.014 to 12 (UPLC-MS/MS) and 0.028 to 12 ?g/ml (bioassay). Intra- and interrun accuracy levels were 106% ± 2% and 103% ± 4% for UPLC-MS/MS and 102% ± 8% and 104% ± 1% for bioassay, respectively. The intra- and interrun coefficients of variation were 7% ± 4% and 7% ± 3% for UPLC-MS/MS and 5% ± 3% and 4% ± 2% for bioassay, respectively. An excellent correlation between POS plasma concentrations measured by UPLC-MS/MS and bioassay was found (concordance, 0.96). In 26 hemato-oncological patients receiving oral POS, 27/69 (39%) trough plasma concentrations were lower than 0.5 ?g/ml. The UPLC-MS/MS method and sensitive bioassay offer alternative tools for accurate and precise quantification of the plasma concentrations in patients receiving oral posaconazole.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

The objective of this work was to combine the advantages of the dried blood spot (DBS) sampling process with the highly sensitive and selective negative-ion chemical ionization tandem mass spectrometry (NICI-MS-MS) to analyze for recent antidepressants including fluoxetine, norfluoxetine, reboxetine, and paroxetine from micro whole blood samples (i.e., 10 microL). Before analysis, DBS samples were punched out, and antidepressants were simultaneously extracted and derivatized in a single step by use of pentafluoropropionic acid anhydride and 0.02% triethylamine in butyl chloride for 30 min at 60 degrees C under ultrasonication. Derivatives were then separated on a gas chromatograph coupled with a triple-quadrupole mass spectrometer operating in negative selected reaction monitoring mode for a total run time of 5 min. To establish the validity of the method, trueness, precision, and selectivity were determined on the basis of the guidelines of the "Société Française des Sciences et des Techniques Pharmaceutiques" (SFSTP). The assay was found to be linear in the concentration ranges 1 to 500 ng mL(-1) for fluoxetine and norfluoxetine and 20 to 500 ng mL(-1) for reboxetine and paroxetine. Despite the small sampling volume, the limit of detection was estimated at 20 pg mL(-1) for all the analytes. The stability of DBS was also evaluated at -20 degrees C, 4 degrees C, 25 degrees C, and 40 degrees C for up to 30 days. Furthermore, the method was successfully applied to a pharmacokinetic investigation performed on a healthy volunteer after oral administration of a single 40-mg dose of fluoxetine. Thus, this validated DBS method combines an extractive-derivative single step with a fast and sensitive GC-NICI-MS-MS technique. Using microliter blood samples, this procedure offers a patient-friendly tool in many biomedical fields such as checking treatment adherence, therapeutic drug monitoring, toxicological analyses, or pharmacokinetic studies.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Triple quadrupole mass spectrometers coupled with high performance liquid chromatography are workhorses in quantitative bioanalyses. It provides substantial benefits including reproducibility, sensitivity and selectivity for trace analysis. Selected Reaction Monitoring allows targeted assay development but data sets generated contain very limited information. Data mining and analysis of non-targeted high-resolution mass spectrometry profiles of biological samples offer the opportunity to perform more exhaustive assessments, including quantitative and qualitative analysis. The objectives of this study was to test method precision and accuracy, statistically compare bupivacaine drug concentration in real study samples and verify if high resolution and accurate mass data collected in scan mode can actually permit retrospective data analysis, more specifically, extract metabolite related information. The precision and accuracy data presented using both instruments provided equivalent results. Overall, the accuracy was ranging from 106.2 to 113.2% and the precision observed was from 1.0 to 3.7%. Statistical comparisons using a linear regression between both methods reveal a coefficient of determination (R2) of 0.9996 and a slope of 1.02 demonstrating a very strong correlation between both methods. Individual sample comparison showed differences from -4.5% to 1.6% well within the accepted analytical error. Moreover, post acquisition extracted ion chromatograms at m/z 233.1648 ± 5 ppm (M-56) and m/z 305.2224 ± 5 ppm (M+16) revealed the presence of desbutyl-bupivacaine and three distinct hydroxylated bupivacaine metabolites. Post acquisition analysis allowed us to produce semiquantitative evaluations of the concentration-time profiles for bupicavaine metabolites.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Most commercial recombinant proteins used as molecular biology tools, as well as many academically made preparations, are generally maintained in the presence of high glycerol concentrations after purification to maintain their biological activity. The present study shows that larger proteins containing high concentrations of glycerol are not amenable to analysis using conventional electrospray ionization mass spectrometry (ESI-MS) interfaces. In this investigation the presence of 25% (v/v) glycerol suppressed the signals of Taq DNA polymerase molecules, while 1% (v/v) glycerol suppressed the signal of horse heart myoglobin. The signal suppression was probably caused by the interaction of glycerol molecules with the proteins to create a shielding effect that prevents the ionization of the basic and/or acidic groups in the amino acid side chains. To overcome this difficulty the glycerol concentration was decreased to 5% (v/v) by dialyzing the Taq polymerase solution against water, and the cone voltage in the ESI triple-quadrupole mass spectrometer was set at 80-130 V. This permitted observation of a mass spectrum that contained ions corresponding to protonation of up to 50% of the ionizable basic groups. In the absence of glycerol up to 85% of the basic groups of Taq polymerase became ionized, as observed in the mass spectrum at relatively low cone voltages. An explanation of these and other observations is proposed, based on strong interactions between the protein molecules and glycerol. For purposes of comparison similar experiments were performed on myoglobin, a small protein with 21 basic groups, whose ionization was apparently suppressed in the presence of 1% (v/v) glycerol, since no mass spectrum could be obtained even at high cone voltages. Copyright (C) 2003 John Wiley Sons, Ltd.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Das Time-of-Flight Aerosol Mass Spectrometer (ToF-AMS) der Firma Aerodyne ist eine Weiterentwicklung des Aerodyne Aerosolmassenspektrometers (Q-AMS). Dieses ist gut charakterisiert und kommt weltweit zum Einsatz. Beide Instrumente nutzen eine aerodynamische Linse, aerodynamische Partikelgrößenbestimmung, thermische Verdampfung und Elektronenstoß-Ionisation. Im Gegensatz zum Q-AMS, wo ein Quadrupolmassenspektrometer zur Analyse der Ionen verwendet wird, kommt beim ToF-AMS ein Flugzeit-Massenspektrometer zum Einsatz. In der vorliegenden Arbeit wird anhand von Laborexperimenten und Feldmesskampagnen gezeigt, dass das ToF-AMS zur quantitativen Messung der chemischen Zusammensetzung von Aerosolpartikeln mit hoher Zeit- und Größenauflösung geeignet ist. Zusätzlich wird ein vollständiges Schema zur ToF-AMS Datenanalyse vorgestellt, dass entwickelt wurde, um quantitative und sinnvolle Ergebnisse aus den aufgenommenen Rohdaten, sowohl von Messkampagnen als auch von Laborexperimenten, zu erhalten. Dieses Schema basiert auf den Charakterisierungsexperimenten, die im Rahmen dieser Arbeit durchgeführt wurden. Es beinhaltet Korrekturen, die angebracht werden müssen, und Kalibrationen, die durchgeführt werden müssen, um zuverlässige Ergebnisse aus den Rohdaten zu extrahieren. Beträchtliche Arbeit wurde außerdem in die Entwicklung eines zuverlässigen und benutzerfreundlichen Datenanalyseprogramms investiert. Dieses Programm kann zur automatischen und systematischen ToF-AMS Datenanalyse und –korrektur genutzt werden.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

A universal and robust analytical method for the determination of Δ9-tetrahydrocannabinol (THC) and two of its metabolites Δ9-(11-OH)-tetrahydrocannabinol (11-OH-THC) and 11-nor-Δ9-carboxy-tetrahydrocannabinol (THC-COOH) in human whole blood was developed and validated for use in forensic toxicology. Protein precipitation, integrated solid phase extraction and on-line enrichment followed by high-performance liquid chromatography separation and detection with a triple quadrupole mass spectrometer were combined. The linear ranges used for the three cannabinoids were from 0.5 to 20 ng/mL for THC and 11-OH-THC and from 2.5 to 100 ng/mL for THC-COOH, therefore covering the requirements for forensic use. Correlation coefficients of 0.9980 or better were achieved for all three analytes. No relevant hydrolysis was observed for THC-COOH glucuronide with this procedure--in contrast to our previous GC-MS procedure, which obviously lead to an artificial increase of the THC-COOH concentration due to the hydrolysis of the glucuronide-conjugate occurring at high pH during the phase-transfer catalyzed methylation step.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

This purpose of this project was to investigate the collision-induced dissociation of dipeptides in negative ion electrospray ionization tandem mass spectrometry, with a focus on the mechanism of the production of imidazole-type fragments not previously reported from the fragmentation of the dipeptides being studied. The majority of the dipeptides studied were alanine N-terminal or serine C-terminal dipeptides. All dipeptides were dissolved in 50:50 methanol:water, 3 mM ammonium formate. Collision-induced dissociation in the collision cell of a triple quadrupole mass spectrometer was used to fragment [M-H]- precursor ions. Accurate mass measurements confirmed the molecular formula of the imidazole-type fragments. Further MS/MS studies were performed to provide information about the fragmentation mechanism for the formation of the imidazole-type fragments. The m/z values of intermediate ions in the formation of the imidazole-type fragments were confirmed through second-generation product ion scans and precursor ion scans. More sophisticated instrumentation will be required to further probe the structure of the intermediate ions.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

The quadrupole mass spectrometer (QMS) has over 30 years of spaceflight heritage in making important neutral gas and low energy ion observations. Given their geometrical constraints, these instruments are currently operated at the extreme limit of their capabilities. However, a technique called higher order auxiliary excitation provides a set of novel, robust, electronics-based solutions for improving the performance of these sensors. By driving the quadrupole rods with an additional frequency nearly twice that of the normal RF operating frequency, substantially increased abundance sensitivity, maximum attainable mass resolution, and peak stability can be achieved through operation of voltage scan lines through the center of formed upper stability islands. Such improvements are modeled using numerical simulations of ion trajectories in a quadrupole field with and without applied higher order auxiliary excitation. When compared to a traditional QMS with a mass range up to 500Da, sensors can be designed with the same precision electronics to have expected mass ranges beyond 1500Da with a power increase of less than twice that of its heritage implementations.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Purpose: To develop and validate a simple, efficient and reliable Liquid chromatographic-mass spectrometric (LC-MS/MS) method for the quantitative determination of two dermatological drugs, Lamisil® (terbinafine) and Proscar® (finasteride), in split tablet dosage form. Methods: Thirty tablets each of the 2 studied medications were randomly selected. Tablets were weighed and divided into 3 groups. Ten tablets of each drug were kept intact, another group of 10 tablets were manually split into halves using a tablet cutter and weighed with an analytical balance; a third group were split into quarters and weighed. All intact and split tablets were individually dissolved in a water: methanol mixture (4:1), sonicated, filtered and further diluted with mobile phase. Optimal chromatographic separation and mass spectrometric detection were achieved using an Agilent 1200 HPLC system coupled with an Agilent 6410 triple quadrupole mass spectrometer. Analytes were eluted through an Agilent eclipse plus C8 analytical column (150 mm × 4.6 mm, 5 μm) with a mobile phase composed of solvent A (water) containing 0.1% formic acid and 5mM ammonium formate pH 7.5, and solvent B (acetonitrile mixed with water in a ratio A:B 55:45) at a flow rate of 0.8 mL min-1 with a total run time of 12 min. Mass spectrometric detection was carried out using positive ionization mode with analyte quantitation monitored by multiple reaction monitoring (MRM) mode. Results: The proposed analytical method proved to be specific, robust and adequately sensitive. The results showed a good linear fit over the concentration range of 20 - 100 ng mL-1 for both analytes, with a correlation coefficient (r2) ≥ 0.999 and 0.998 for finasteride and terbinafine, respectively. Following tablet splitting, the drug content of the split tablets fell outside of the proxy USP specification for at least 14 halves (70 %) and 34 quarters (85 %) of FIN, as well as 16 halves (80 %) and 37 quarters (92.5 %) of TBN. Mean weight loss, after splitting, was 0.58 and 2.22 % for FIN half- and quarter tablets, respectively, and 3.96 and 4.09 % for TBN half- and quarter tablets,respectively. Conclusion: The proposed LC-MS/MS method has successfully been used to provide precise drug content uniformity of split tablets of FIN and TBN. Unequal distribution of the drug on the split tablets is indicated by the high standard deviation beyond the accepted value. Hence, it is recommended not to split non-scored tablets especially, for those medications with significant toxicity

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Cyclosporine A (CsA) has been demonstrated to be effective for the treatment of a variety of ophthalmological conditions, including ocular surface disorders such as the dry eye disease (DED). Since CsA is characterised by its low water solubility, the development of a topical ophthalmic formulation represents an interesting pharmaceutical question. In the present study, two different strategies to address this challenge were studied and compared: (i) a water-soluble CsA prodrug formulated within an aqueous solution and (ii) a CsA oil-in-water emulsion (Restasis, Allergan Inc., Irvine, CA). First, the prodrug formulation was shown to have an excellent ocular tolerance as well as no influence on the basal tear production; maintaining the ocular surface properties remained unchanged. Then, in order to allow in vivo investigations, a specific analytical method based on ultra high pressure liquid chromatography coupled with triple quadrupole mass spectrometer (UHPLC-MS/MS) was developed and optimised to quantify CsA in ocular tissues and fluids. The CsA ocular kinetics in lachrymal fluid for both formulations were found to be similar between 15 min and 48 h. The CsA ocular distribution study evidenced the ability of the prodrug formulation to penetrate into the eye, achieving therapeutically active CsA levels in tissues of both the anterior and posterior segments. In addition, the detailed analysis of the in vivo data using a bicompartmental model pointed out a higher bioavailability and lower elimination rate for CsA when it is generated from the prodrug than after direct application as an emulsion. The interesting in vivo properties displayed by the prodrug solution make it a safe and suitable option for the treatment of DED.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

This article describes the application of a recently developed general unknown screening (GUS) strategy based on LC coupled to a hybrid linear IT-triple quadrupole mass spectrometer (LC-MS/MS-LIT) for the simultaneous detection and identification of drug metabolites following in vitro incubation with human liver microsomes. The histamine H1 receptor antagonist loratadine was chosen as a model compound to demonstrate the interest of such approach, because of its previously described complex and extensive metabolism. Detection and mass spectral characterization were based on data-dependent acquisition, switching between a survey scan acquired in the ion-trapping Q3 scan mode with dynamic subtraction of background noise, and a dependent scan in the ion-trapping product ion scan mode of automatically selected parent ions. In addition, the MS(3) mode was used in a second step to confirm the structure of a few fragment ions. The sensitivity of the ion-trapping modes combined with the selectivity of the triple quadrupole modes allowed, with only one injection, the detection and identification of 17 phase I metabolites of loratadine. The GUS procedure used in this study may be applicable as a generic technique for the characterization of drug metabolites after in vitro incubation, as well as probably in vivo experiments.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Les agents anti-infectieux sont utilisés pour traiter ou prévenir les infections chez les humains, les animaux, les insectes et les plantes. L’apparition de traces de ces substances dans les eaux usées, les eaux naturelles et même l’eau potable dans plusieurs pays du monde soulève l’inquiétude de la communauté scientifique surtout à cause de leur activité biologique. Le but de ces travaux de recherche a été d’étudier la présence d’anti-infectieux dans les eaux environnementales contaminées (c.-à-d. eaux usées, eaux naturelles et eau potable) ainsi que de développer de nouvelles méthodes analytiques capables de quantifier et confirmer leur présence dans ces matrices. Une méta-analyse sur l’occurrence des anti-infectieux dans les eaux environnementales contaminées a démontré qu’au moins 68 composés et 10 de leurs produits de transformation ont été quantifiés à ce jour. Les concentrations environnementales varient entre 0.1 ng/L et 1 mg/L, selon le composé, la matrice et la source de contamination. D’après cette étude, les effets nuisibles des anti-infectieux sur le biote aquatique sont possibles et ces substances peuvent aussi avoir un effet indirect sur la santé humaine à cause de sa possible contribution à la dissémination de la résistance aux anti-infecteiux chez les bactéries. Les premiers tests préliminaires de développement d’une méthode de détermination des anti-infectieux dans les eaux usées ont montré les difficultés à surmonter lors de l’extraction sur phase solide (SPE) ainsi que l’importance de la sélectivité du détecteur. On a décrit une nouvelle méthode de quantification des anti-infectieux utilisant la SPE en tandem dans le mode manuel et la chromatographie liquide couplée à la spectrométrie de masse en tandem (LC-MS/MS). Les six anti-infectieux ciblés (sulfaméthoxazole, triméthoprime, ciprofloxacin, levofloxacin, clarithromycin et azithromycin) ont été quantifiés à des concentrations entre 39 et 276 ng/L dans les échantillons d’affluent et d’effluent provenant d’une station d’épuration appliquant un traitement primaire et physico- chimique. Les concentrations retrouvées dans les effluents indiquent que la masse moyenne totale de ces substances, déversées hebdomadairement dans le fleuve St. Laurent, était de ~ 2 kg. En vue de réduire le temps total d’analyse et simplifier les manipulations, on a travaillé sur une nouvelle méthode de SPE couplée-LC-MS/MS. Cette méthode a utilisé une technique de permutation de colonnes pour préconcentrer 1.00 mL d’échantillon dans une colonne de SPE couplée. La performance analytique de la méthode a permis la quantification des six anti-infectieux dans les eaux usées municipales et les limites de détection étaient du même ordre de grandeur (13-60 ng/L) que les méthodes basées sur la SPE manuelle. Ensuite, l’application des colonnes de SPE couplée de chromatographie à débit turbulent pour la préconcentration de six anti-infectieux dans les eaux usées a été explorée pour diminuer les effets de matrice. Les résultats obtenus ont indiqué que ces colonnes sont une solution de réchange intéressante aux colonnes de SPE couplée traditionnelles. Finalement, en vue de permettre l’analyse des anti-infectieux dans les eaux de surface et l’eau potable, une méthode SPE couplée-LC-MS/MS utilisant des injections de grand volume (10 mL) a été développée. Le volume de fuite de plusieurs colonnes de SPE couplée a été estimé et la colonne ayant la meilleure rétention a été choisie. Les limites de détection et de confirmation de la méthode ont été entre 1 à 6 ng/L. L’analyse des échantillons réels a démontré que la concentration des trois anti-infectieux ciblés (sulfaméthoxazole, triméthoprime et clarithromycine) était au dessous de la limite de détection de la méthode. La mesure des masses exactes par spectrométrie de masse à temps d’envol et les spectres des ions produits utilisant une pente d’énergie de collision inverse dans un spectromètre de masse à triple quadripôle ont été explorés comme des méthodes de confirmation possibles.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Herbs and spices have long been used to improve the flavour of food without being considered as nutritionally significant ingredients. However, the bioactive phenolic content of these plant-based products is currently attracting interest.In the present work, liquid chromatography coupled to high-resolution/accurate mass measurement LTQ-Orbitrap mass spectrometry was applied for the comprehensive identification of phenolic constituents of six of the most widely used culinary herbs (rosemary, thyme, oregano and bay) and spices (cinnamon and cumin). In this way, up to 52 compounds were identified in these culinary ingredients, some of them, as far as we know, for the first time. In order to establish the phenolic profiles of the different herbs and spices, accurate quantification of the major phenolics was performed by multiple reaction monitoring in a triple quadrupole mass spectrometer. Multivariate statistical treatment of the results allowed the assessment of distinctive features among the studied herbs and spices. (C) 2014 Elsevier Ltd. All rights reserved.