981 resultados para LUNG MODEL
Resumo:
Background: Noninvasive positive-pressure ventilation (NPPV) modes are currently available on bilevel and ICU ventilators. However, little data comparing the performance of the NPPV modes on these ventilators are available. Methods: In an experimental bench study, the ability of nine ICU ventilators to function in the presence of leaks was compared with a bilevel ventilator using the IngMar ASL5000 lung simulator (IngMar Medical; Pittsburgh, PA) set at a compliance of 60 mL/cm H(2)O, an inspiratory resistance of 10 cm H(2)O/L/s, an expiratory resistance of 20 cm H(2)O/L/s, and a respiratory rate of 15 breaths/min. All of the ventilators were set at 12 cm H(2)O pressure support and 5 cm H(2)O positive end-expiratory pressure. The data were collected at baseline and at three customized leaks. Main results: At baseline, all of the ventilators were able to deliver adequate tidal volumes, to maintain airway pressure, and to synchronize with the simulator, without missed efforts or auto-triggering. As the leak was increased, all of the ventilators (except the Vision [Respironics; Murrysville, PA] and Servo I [Maquet; Solna, Sweden]) needed adjustment of sensitivity or cycling criteria to maintain adequate ventilation, and some transitioned to backup ventilation. Significant differences in triggering and cycling were observed between the Servo I and the Vision ventilators. Conclusions: The Vision and Servo I were the only ventilators that required no adjustments as they adapted to increasing leaks. There were differences in performance between these two ventilators, although the clinical significance of these differences is unclear. Clinicians should be aware that in the presence of leaks, most ICU ventilators require adjustments to maintain an adequate tidal volume. (CHEST 2009; 136:448-456)
Resumo:
OBJECTIVE: : To determine the influence of nebulizer types and nebulization modes on bronchodilator delivery in a mechanically ventilated pediatric lung model. DESIGN: : In vitro, laboratory study. SETTING: : Research laboratory of a university hospital. INTERVENTIONS: : Using albuterol as a marker, three nebulizer types (jet nebulizer, ultrasonic nebulizer, and vibrating-mesh nebulizer) were tested in three nebulization modes in a nonhumidified bench model mimicking the ventilatory pattern of a 10-kg infant. The amounts of albuterol deposited on the inspiratory filters (inhaled drug) at the end of the endotracheal tube, on the expiratory filters, and remaining in the nebulizers or in the ventilator circuit were determined. Particle size distribution of the nebulizers was also measured. MEASUREMENTS AND MAIN RESULTS: : The inhaled drug was 2.8% ± 0.5% for the jet nebulizer, 10.5% ± 2.3% for the ultrasonic nebulizer, and 5.4% ± 2.7% for the vibrating-mesh nebulizer in intermittent nebulization during the inspiratory phase (p < 0.01). The most efficient nebulizer was the vibrating-mesh nebulizer in continuous nebulization (13.3% ± 4.6%, p < 0.01). Depending on the nebulizers, a variable but important part of albuterol was observed as remaining in the nebulizers (jet and ultrasonic nebulizers), or being expired or lost in the ventilator circuit (all nebulizers). Only small particles (range 2.39-2.70 µm) reached the end of the endotracheal tube. CONCLUSIONS: : Important differences between nebulizer types and nebulization modes were seen for albuterol deposition at the end of the endotracheal tube in an in vitro pediatric ventilator-lung model. New aerosol devices, such as ultrasonic and vibrating-mesh nebulizers, were more efficient than the jet nebulizer.
Resumo:
OBJECTIVE: Euro-Collins solution (EC) is routinely used in lung transplantation. The high potassium of EC, however, may damage the vascular endothelium, thereby contributing to postischemic reperfusion injury. To assess the influence of the potassium concentration on lung preservation, we evaluated the effect of a "low potassium Euro-Collins solution" (LPEC), in which the sodium and potassium concentrations were reversed. METHODS: In an extracorporeal rat heart-lung model lungs were preserved with EC and LPEC. The heart-lung blocks (HLB) were perfused with Krebs-Henseleit solution containing washed bovine red blood cells and ventilated with room air. The lungs were perfused via the working right ventricle with deoxygenated perfusate. Oxygenation and pulmonary vascular resistance (PVR) were monitored. After baseline measurements, hearts were arrested with St. Thomas' solution and the lungs were perfused with EC or LPEC, or were not perfused (controls). The HLBs were stored for 5 min or 2 h ischemic time at 4 degrees C. Reperfusion and ventilation was performed for 40 min. At the end of the trial the wet/dry ratio of the lungs was calculated and light microscopic assessment of the degree of edema was performed. RESULTS: After 5 min of ischemia oxygenation was significantly better in both preserved groups compared to the controls. Pulmonary vascular resistance was elevated in all three groups after 30 min reperfusion at both ischemic times. After 2 h of ischemia PVR of the group preserved with LPEC was significantly lower than those of the EC and controls (LPEC-5 min: 184 +/- 65 dynes * sec * cm-5, EC-5 min: 275 +/- 119 dynes * sec * cm * cm-5, LPEC-2 h: 324 +/- 47 dynes * sec * m-5, EC-2 h: 507 +/- 83 dynes * sec * cm-5). Oxygenation after 2 h of ischemia and 30 min reperfusion was significantly better in the LPEC group compared to EC and controls (LPEC: 70 +/- 17 mmHg, EC: 44 +/- 3 mmHg). The wet/dry ratio was significantly lower in the two preserved groups compared to controls (LPEC-5 min: 5.7 +/- 0.7, EC-5 min: 5.8 +/- 1.2, controls-5 min: 7.5 +/- 1.8, LPEC-2 h: 6.7 +/- 0.4, EC: 6.9 +/- 0.4, controls-2 h: 7.3 +/- 0.4). CONCLUSIONS: We thus conclude that LPEC results in better oxygenation and lower PVR in this lung preservation model. A low potassium concentration in lung preservation solutions may help in reducing the incidence of early graft dysfunction following lung transplantation.
Resumo:
Patients suffering from cystic fibrosis (CF) show thick secretions, mucus plugging and bronchiectasis in bronchial and alveolar ducts. This results in substantial structural changes of the airway morphology and heterogeneous ventilation. Disease progression and treatment effects are monitored by so-called gas washout tests, where the change in concentration of an inert gas is measured over a single or multiple breaths. The result of the tests based on the profile of the measured concentration is a marker for the severity of the ventilation inhomogeneity strongly affected by the airway morphology. However, it is hard to localize underlying obstructions to specific parts of the airways, especially if occurring in the lung periphery. In order to support the analysis of lung function tests (e.g. multi-breath washout), we developed a numerical model of the entire airway tree, coupling a lumped parameter model for the lung ventilation with a 4th-order accurate finite difference model of a 1D advection-diffusion equation for the transport of an inert gas. The boundary conditions for the flow problem comprise the pressure and flow profile at the mouth, which is typically known from clinical washout tests. The natural asymmetry of the lung morphology is approximated by a generic, fractal, asymmetric branching scheme which we applied for the conducting airways. A conducting airway ends when its dimension falls below a predefined limit. A model acinus is then connected to each terminal airway. The morphology of an acinus unit comprises a network of expandable cells. A regional, linear constitutive law describes the pressure-volume relation between the pleural gap and the acinus. The cyclic expansion (breathing) of each acinus unit depends on the resistance of the feeding airway and on the flow resistance and stiffness of the cells themselves. Special care was taken in the development of a conservative numerical scheme for the gas transport across bifurcations, handling spatially and temporally varying advective and diffusive fluxes over a wide range of scales. Implicit time integration was applied to account for the numerical stiffness resulting from the discretized transport equation. Local or regional modification of the airway dimension, resistance or tissue stiffness are introduced to mimic pathological airway restrictions typical for CF. This leads to a more heterogeneous ventilation of the model lung. As a result the concentration in some distal parts of the lung model remains increased for a longer duration. The inert gas concentration at the mouth towards the end of the expirations is composed of gas from regions with very different washout efficiency. This results in a steeper slope of the corresponding part of the washout profile.
Resumo:
PURPOSE: The objective of this experiment is to establish a continuous postmortem circulation in the vascular system of porcine lungs and to evaluate the pulmonary distribution of the perfusate. This research is performed in the bigger scope of a revascularization project of Thiel embalmed specimens. This technique enables teaching anatomy, practicing surgical procedures and doing research under lifelike circumstances. METHODS: After cannulation of the pulmonary trunk and the left atrium, the vascular system was flushed with paraffinum perliquidum (PP) through a heart-lung machine. A continuous circulation was then established using red PP, during which perfusion parameters were measured. The distribution of contrast-containing PP in the pulmonary circulation was visualized on computed tomography. Finally, the amount of leak from the vascular system was calculated. RESULTS: A reperfusion of the vascular system was initiated for 37 min. The flow rate ranged between 80 and 130 ml/min throughout the experiment with acceptable perfusion pressures (range: 37-78 mm Hg). Computed tomography imaging and 3D reconstruction revealed a diffuse vascular distribution of PP and a decreasing vascularization ratio in cranial direction. A self-limiting leak (i.e. 66.8% of the circulating volume) towards the tracheobronchial tree due to vessel rupture was also measured. CONCLUSIONS: PP enables circulation in an isolated porcine lung model with an acceptable pressure-flow relationship resulting in an excellent recruitment of the vascular system. Despite these promising results, rupture of vessel walls may cause leaks. Further exploration of the perfusion capacities of PP in other organs is necessary. Eventually, this could lead to the development of reperfused Thiel embalmed human bodies, which have several applications.
Resumo:
Objective: To compare the triggering performance of mid-level ICU mechanical ventilators with a standard ICU mechanical ventilator. Design: Experimental bench study. Setting: The respiratory care laboratory of a university-affiliated teaching hospital. Subject: A computerized mechanical lung model, the IngMar ASL5000. Interventions: Ten mid-level ICU ventilators were compared to an ICU ventilator at two levels of lung model effort, three combinations of respiratory mechanics (normal, COPD and ARDS) and two modes of ventilation, volume and pressure assist/control. A total of 12 conditions were compared. Measurements and main results: Performance varied widely among ventilators. Mean inspiratory trigger time was < 100 ms for only half of the tested ventilators. The mean inspiratory delay time (time from initiation of the breath to return of airway pressure to baseline) was longer than that for the ICU ventilator for all tested ventilators except one. The pressure drop during triggering (Ptrig) was comparable with that of the ICU ventilator for only two ventilators. Expiratory Settling Time (time for pressure to return to baseline) had the greatest variability among ventilators. Conclusions: Triggering differences among these mid-level ICU ventilators and with the ICU ventilator were identified. Some of these ventilators had a much poorer triggering response with high inspiratory effort than the ICU ventilator. These ventilators do not perform as well as ICU ventilators in patients with high ventilatory demand.
Resumo:
INTRODUCTION. The role of turbine-based NIV ventilators (TBV) versus ICU ventilators with NIV mode activated (ICUV) to deliver NIV in case of severe respiratory failure remains debated. OBJECTIVES. To compare the response time and pressurization capacity of TBV and ICUV during simulated NIV with normal and increased respiratory demand, in condition of normal and obstructive respiratory mechanics. METHODS. In a two-chamber lung model, a ventilator simulated normal (P0.1 = 2 mbar, respiratory rate RR = 15/min) or increased (P0.1 = 6 mbar, RR = 25/min) respiratory demand. NIV was simulated by connecting the lung model (compliance 100 ml/mbar; resistance 5 or 20 l/mbar) to a dummy head equipped with a naso-buccal mask. Connections allowed intentional leaks (29 ± 5 % of insufflated volume). Ventilators to test: Servo-i (Maquet), V60 and Vision (Philips Respironics) were connected via a standard circuit to the mask. Applied pressure support levels (PSL) were 7 mbar for normal and 14 mbar for increased demand. Airway pressure and flow were measured in the ventilator circuit and in the simulated airway. Ventilator performance was assessed by determining trigger delay (Td, ms), pressure time product at 300 ms (PTP300, mbar s) and inspiratory tidal volume (VT, ml) and compared by three-way ANOVA for the effect of inspiratory effort, resistance and the ventilator. Differences between ventilators for each condition were tested by oneway ANOVA and contrast (JMP 8.0.1, p\0.05). RESULTS. Inspiratory demand and resistance had a significant effect throughout all comparisons. Ventilator data figure in Table 1 (normal demand) and 2 (increased demand): (a) different from Servo-i, (b) different from V60.CONCLUSION. In this NIV bench study, with leaks, trigger delay was shorter for TBV with normal respiratory demand. By contrast, it was shorter for ICUV when respiratory demand was high. ICUV afforded better pressurization (PTP 300) with increased demand and PSL, particularly with increased resistance. TBV provided a higher inspiratory VT (i.e., downstream from the leaks) with normal demand, and a significantly (although minimally) lower VT with increased demand and PSL.
Resumo:
OBJECTIVE: To assess the suitability of a hot-wire anemometer infant monitoring system (Florian, Acutronic Medical Systems AG, Hirzel, Switzerland) for measuring flow and tidal volume (Vt) proximal to the endotracheal tube during high-frequency oscillatory ventilation. DESIGN: In vitro model study. SETTING: Respiratory research laboratory. SUBJECT: In vitro lung model simulating moderate to severe respiratory distress. INTERVENTION: The lung model was ventilated with a SensorMedics 3100A ventilator. Vt was recorded from the monitor display (Vt-disp) and compared with the gold standard (Vt-adiab), which was calculated using the adiabatic gas equation from pressure changes inside the model. MEASUREMENTS AND MAIN RESULTS: A range of Vt (1-10 mL), frequencies (5-15 Hz), pressure amplitudes (10-90 cm H2O), inspiratory times (30% to 50%), and Fio2 (0.21-1.0) was used. Accuracy was determined by using modified Bland-Altman plots (95% limits of agreement). An exponential decrease in Vt was observed with increasing oscillatory frequency. Mean DeltaVt-disp was 0.6 mL (limits of agreement, -1.0 to 2.1) with a linear frequency dependence. Mean DeltaVt-disp was -0.2 mL (limits of agreement, -0.5 to 0.1) with increasing pressure amplitude and -0.2 mL (limits of agreement, -0.3 to -0.1) with increasing inspiratory time. Humidity and heating did not affect error, whereas increasing Fio2 from 0.21 to 1.0 increased mean error by 6.3% (+/-2.5%). CONCLUSIONS: The Florian infant hot-wire flowmeter and monitoring system provides reliable measurements of Vt at the airway opening during high-frequency oscillatory ventilation when employed at frequencies of 8-13 Hz. The bedside application could improve monitoring of patients receiving high-frequency oscillatory ventilation, favor a better understanding of the physiologic consequences of different high-frequency oscillatory ventilation strategies, and therefore optimize treatment.
Resumo:
The purpose of breathing remained an enigma for a long time. The Hippocratic school described breathing patterns but did not associate breathing with the lungs. Empedocles and Plato postulated that breathing was linked to the passage of air through pores of the skin. This was refuted by Aristotle who believed that the role of breathing was to cool the heart. In Alexandria, breakthroughs were accomplished in the anatomy and physiology of the respiratory system. Later, Galen proposed an accurate description of the respiratory muscles and the mechanics of breathing. However, his heart-lung model was hampered by the traditional view of two non-communicating vascular systems - veins and arteries. After a period of stagnation in the Middle Ages, knowledge progressed with the discovery of pulmonary circulation. The comprehension of the purpose of breathing progressed by steps thanks to Boyle and Mayow among others, and culminated with the contribution of Priestley and the discovery of oxygen by Lavoisier. Only then was breathing recognized as fulfilling the purpose of respiration, or gas exchange. A century later, a controversy emerged concerning the active or passive transfer of oxygen from alveoli to the blood. August and Marie Krogh settled the dispute, showing that passive diffusion was sufficient to meet the oxygen needs. © 2014 S. Karger AG, Basel.
Resumo:
Les travaux effectués dans le cadre de cette thèse de doctorat avaient pour but de mettre au point des nouvelles formulations d’antifongiques sous forme de nanoparticules polymériques (NP) en vue d’améliorer l’efficacité et la spécificité des traitements antifongiques sur des souches sensibles ou résistantes de Candida spp, d’Aspergillus spp et des souches de Candida albicans formant du biofilm. Dans la première partie de ce travail, nous avons synthétisé et caractérisé un polymère à base de polyester-co-polyéther branché avec du poly(éthylène glycol) (PEG-g-PLA). En plus d’être original et innovant, ce co-polymère a l’avantage d’être non-toxique et de posséder des caractéristiques de libération prolongée. Trois antifongiques couramment utilisés en clinique et présentant une biodisponibilité non optimale ont été choisis, soient deux azolés, le voriconazole (VRZ) et l’itraconazole (ITZ) et un polyène, l’amphotéricine B (AMB). Ces principes actifs (PA), en plus des problèmes d’administration, présentent aussi d’importants problèmes de toxicité. Des NP polymériques encapsulant ces PA ont été préparées par une technique d’émulsion huile-dans-l’eau (H/E) suivie d’évaporation de solvant. Une fois fabriquées, les NP ont été caractérisées et des particules de d’environ 200 nm de diamètre ont été obtenues. Les NP ont été conçues pour avoir une structure coeur/couronne avec un coeur constitué de polymère hydrophobe (PLA) et une couronne hydrophile de PEG. Une faible efficacité de chargement (1,3% m/m) a été obtenue pour la formulation VRZ encapsulé dans des NP (NP/VRZ). Toutefois, la formulation AMB encapsulée dans des NP (NP/AMB) a montré des taux de chargement satisfaisants (25,3% m/m). En effet, le caractère hydrophobe du PLA a assuré une bonne affinité avec les PA hydrophobes, particulièrement l’AMB qui est le plus hydrophobe des agents sélectionnés. Les études de libération contrôlée ont montré un relargage des PA sur plusieurs jours. La formulation NP/AMB a été testée sur un impacteur en cascade, un modèle in vitro de poumon et a permis de démontrer le potentiel de cette formulation à être administrée efficacement par voie pulmonaire. En effet, les résultats sur l’impacteur en cascade ont montré que la majorité de la formulation s’est retrouvée à l’étage de collecte correspondant au niveau bronchique, endroit où se situent majoritairement les infections fongiques pulmonaires. Dans la deuxième partie de ces travaux, nous avons testé les nouvelles formulations d’antifongiques sur des souches planctoniques de Candida spp., d’Aspergillus spp. et des souches de Candida albicans formant du biofilm selon les procédures standardisées du National Committee for Clinical Laboratory Standards (NCCLS). Les souches choisies ont démontré des résistances aux azolés et aux polyènes. Les études d’efficacité in vitro ont permis de prouver hors de tout doute que les nouvelles formulations offrent une efficacité nettement améliorée comparée à l’agent antifongique libre. Pour mettre en lumière si l’amélioration de l’efficacité antifongique était due à une internalisation des NP, nous avons évalué le comportement des NP avec les cellules de champignons. Nous avons procédé à des études qualitatives de microscopie de fluorescence sur des NP marquées avec de la rhodamine (Rh). Tel qu’attendu, les NP ont montré une localisation intracellulaire. Pour exclure la possibilité d’une simple adhésion des NP à la surface des levures, nous avons aussi confirmé leur internalisation en microscopie confocale de fluorescence. Il est important de noter que peu d’études à ce jour ont mis l’accent sur l’élaboration de nouvelles formulations d’antifongiques à base de polymères non toxiques destinées aux traitements des mycoses, donnant ainsi une grande valeur et originalité aux travaux effectués dans cette thèse. Les résultats probants obtenus ouvrent la voie vers une nouvelle approche pour contourner les problèmes de résistances fongiques, un problème de plus en plus important dans le domaine de l’infectiologie.
Resumo:
Evaluar si el Heliox reduce la resistencia en la vía aérea en niños y adolescentes con patología bronquial obstructiva que requieren ventilación mecánica. Materiales y Métodos: Estudio prospectivo observacional descriptivo en niños y adolescentes con patología bronquial obstructiva y ventilación mecánica con Fi02 ≤ 0,5. Medición de variables: resistencia, presión pico, presión media de la vía aérea, presión meseta, volumen corriente, autoPEEP, distensibilidad, PetCO2, ventilación de espacio muerto antes de inicio de heliox y a los 30 minutos, 2, 4, 6, 12, 18 y 24 horas y diariamente hasta suspenderlo por extubación o FiO2 > 0,5. Resultados: Resultados parciales, incluyó 9 pacientes encontrando descenso significativo de resistencia espiratoria a los 30 minutos (51,2 vs 32,3; p=0,0008 ), 2 horas ( 51,2 vs 33,4; p=0,0019) y 4 horas (51,2 vs 30,7; p=0,0012) así como de la resistencia inspiratoria a la hora 2 (48,6 vs 36,2; p = 0,013) y hora 4 (48,6 vs 30 ; p=0,004). Se observó tendencia al descenso de la PetCO2 que no fue significativa (52,3 vs 34,3: p=0,06). No se evidenció cambios en las variables; autoPEEP, presión pico, presión media de la vía aérea, distensibilidad, ventilación de espacio muerto, presión meseta y volumen corriente antes y después del inicio del Heliox. Conclusión: La ventilación mecánica con Heliox en niños con patología bronquial obstructiva parece ser que reduce de manera significativa la resistencia de la vía aérea, con tendencia al descenso de la PetC02. Se necesitan estudios prospectivos al menos observacionales analíticos que corroboren estos hallazgos.
Resumo:
Este es un estudio observacional descriptivo, longitudinal y prospectivo realizado con el fin de describir la evolución de de la mecánica ventilatoria, la gasometría y los tiempos de ventilación en los pacientes en ventilación mecánica en modo ASV en la UCI del Hospital San Rafael de Tunja durante los meses de Agosto a Diciembre de 2014 y encontrando que es un método seguro y eficiente para el manejo de la ventilación en pacientes sin mayores comorbilidades ni compromiso orgánico múltiple que bien podría ser utilizado desde el inicio de la ventilación hasta el destete de la misma con el beneficio adicional de menores requerimientos de sedación durante la ventilación pero teniendo en cuenta que factores como hipotensión sostenida, hiperlactatemia, falla renal e hipoxia severa pueden indicar la necesidad de cambio de modo ventilatorio.
Resumo:
Background For reliable assessment of ventilation inhomogeneity, multiple-breath washout (MBW) systems should be realistically validated. We describe a new lung model for in vitro validation under physiological conditions and the assessment of a new nitrogen (N2)MBW system. Methods The N2MBW setup indirectly measures the N2 fraction (FN2) from main-stream carbon dioxide (CO2) and side-stream oxygen (O2) signals: FN2 = 1−FO2−FCO2−FArgon. For in vitro N2MBW, a double chamber plastic lung model was filled with water, heated to 37°C, and ventilated at various lung volumes, respiratory rates, and FCO2. In vivo N2MBW was undertaken in triplets on two occasions in 30 healthy adults. Primary N2MBW outcome was functional residual capacity (FRC). We assessed in vitro error (√[difference]2) between measured and model FRC (100–4174 mL), and error between tests of in vivo FRC, lung clearance index (LCI), and normalized phase III slope indices (Sacin and Scond). Results The model generated 145 FRCs under BTPS conditions and various breathing patterns. Mean (SD) error was 2.3 (1.7)%. In 500 to 4174 mL FRCs, 121 (98%) of FRCs were within 5%. In 100 to 400 mL FRCs, the error was better than 7%. In vivo FRC error between tests was 10.1 (8.2)%. LCI was the most reproducible ventilation inhomogeneity index. Conclusion The lung model generates lung volumes under the conditions encountered during clinical MBW testing and enables realistic validation of MBW systems. The new N2MBW system reliably measures lung volumes and delivers reproducible LCI values.
Resumo:
INTRODUCTION The new ATS/ERS consensus report recommends in vitro validation of multiple-breath inert gas washout (MBW) equipment based on a lung model with simulated physiologic conditions. We aimed to assess accuracy of two MBW setups for infants and young children using this model, and to compare functional residual capacity (FRC) from helium MBW (FRCMBW ) with FRC from plethysmography (FRCpleth ) in vivo. METHODS The MBW setups were based on ultrasonic flow meter technology. Sulfur hexafluoride and helium were used as tracer gases. We measured FRC in vitro for specific model settings with and without carbon dioxide and calculated differences of measured to generated FRC. For in vivo evaluation, difference between FRCMBW and FRCpleth was calculated in 20 healthy children, median age 6.1 years. Coefficient of variation (CV) was calculated per FRC. RESULTS In the infant model (51 runs, FRC 80-300 ml), mean (SD) relative difference between generated and measured FRCs was 0.7 (4.7) %, median CV was 4.4% for measured FRCs. In the young child model, one setting (8 runs, FRC 400 ml) showed a relative difference of up to 13%. For the remaining FRCs (42 runs, FRC 600-1,400 ml), mean (SD) relative difference was -2.0 (3.4) %; median CV was 1.4% for measured FRCs. In vivo FRCpleth exceeded FRCMBW values by 37% on average. CONCLUSIONS Both setups measure lung volumes in the intended age group reliably and reproducibly. Characteristics of different techniques should be considered when measuring lung volumes in vivo. Pediatr Pulmonol. © 2014 Wiley Periodicals, Inc.