946 resultados para LASER FLUORESCENCE DEVICE


Relevância:

100.00% 100.00%

Publicador:

Resumo:

To evaluate the performance of a pen‑type laser fluorescence device (DIAGNOdent 2190; LFpen, KaVo, Germany) and bitewing radiographs (BW) for approximal caries detection in permanent and primary teeth. Materials and Methods: A total of 246 anterior approximal surfaces (102 permanent and 144 primary) were selected. Contact points were simulated using sound teeth. Two examiners assessed all approximal surfaces using LFpen and BW. The teeth were histologically assessed for the reference standard. Optimal cut‑off limits were calculated for LFpen for primary and permanent teeth. Sensitivity, specificity, accuracy and area under the receiver operating characteristic curve (Az) were calculated for D1 (enamel and dentin lesions) and D3 (dentin lesions) thresholds. The reproducibility was assessed by intraclass correlation coefficient (ICC) and Cohen's weighted kappa values. Results: For permanent teeth, the LFpen cut‑off were 0– 27 (sound), 28– 33 (enamel caries) and >33 (dentin caries). For primary teeth, the LFpen cut‑off were 0– 7 (sound), 8– 32 (enamelcaries) and >32 (dentin caries). The LFpen presented higher sensitivity values than BW for primary teeth (0.58 vs. 0.32 at D1 and 0.80 vs. 0.47 at D3) and permanent teeth (0.80 vs. 0.57 at D1 and 0.94 vs. 0.51 at D3). Specificity did not show a significant difference between the methods. Rank correlations with histology were 0.59 and 0.83 (LFpen) and 0.36 and 0.70 (BW) for primary and permanent teeth, respectively, considering all lesions. ICC values for LFpen were 0.71 (inter) and 0.86 (intra) for permanent teeth and 0.94 (inter) and 0.90/0.99 for primary teeth. Kappa values for BW were 0.69 (inter) and 0.68/0.90 (intra) for permanent teeth and 0.64 (inter) and 0.89/0.89 for primary teeth. Conclusion: LFpen presented better reproducibility for primary and permanent teeth and higher accuracy in detecting caries lesions at D1 threshold than BW for permanent teeth. LFpen should be used as an adjunct method for approximal caries detection.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

The aim of this study was to evaluate the influence of dental plaque and moisture on performance of a laser fluorescence device in detecting occlusal and proximal caries lesions in primary teeth. Fifty-five occlusal and 58 proximal sites on primary molars were evaluated using a DIAGNOdent pen (LFpen) device. For the drying time study, the evaluations were performed in: (I) moist teeth; (II) teeth dried for 3 s, or (III) dried for 15 s. For the plaque study, the evaluations were done in sites: (I) without plaque; (II) with plaque, and (III) after cleaning. Evaluation of the teeth sections in stereomicroscope was the reference standard method. LF pen values, sensitivity, specificity, and accuracy were compared. The values obtained after 15 s of drying were higher than those obtained with moist teeth or dried for 3 s at both occlusal and proximal surfaces. However, there was no change in the performance in detecting caries lesions. With regard to the presence of plaque, there was no significant change in the readings of the device, but specificity was decreased in occlusal surfaces with plaque. At proximal surfaces, however, no significant differences were observed. In conclusion, the moisture conditions do not influence significantly the performance of the LFpen, but the presence of plaque can affect its performance in detecting occlusal caries lesions in primary teeth.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

OBJECTIVES: The aim of this randomised clinical trial was to investigate if a laser fluorescence device is able to discriminate between sound and carious approximal sites and between enamel and dentinal lesions, as well as to find appropriate cut-off values. METHODS: One hundred and seventeen sound or uncavitated carious sites in permanent molars were visually and radiographically examined, then either opened or not, after which their laser fluorescence was measured. Forty-three lesions were opened, the caries removed and the clinically identified caries depths were registered in addition to the radiographical scoring. Seventy-four sites were radiographically deemed sound or had enamel caries and were not opened. Here, the radiographical scorings were registered. RESULTS: Taking the radiographic scoring as gold standard for all investigated approximal sites, sound sites (D(0), n=40) showed significantly lower laser fluorescence measurements than carious sites (D(1-4), n=77) (Mann-Whitney test, P<0.025) suggesting a cut-off at 7 (sensitivity=0.68, specificity=0.7). Comparing measurements of D(0-2) (n=74) and D(3,4) (n=43), the results were also different by a statistically significant amount (P<0.025) and the cut-off calculated to be 16 (sensitivity=0.6, specificity=0.84). A fair positive correlation between laser fluorescence values and radiographical scoring was found (rho=+0.47, P<0.01). Analysing the 43 opened lesions with their clinically found lesion depths as gold standard, there was a fair positive correlation to the laser fluorescence values (rho=+0.34, P=0.03) and a moderately strong correlation to the radiographic scoring (rho=+0.67, P<0.01). CONCLUSION: The device may be an adjunct tool in the approximal detection of caries along with established procedures.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

The laser device DIAGNOdent developed for the detection of occlusal caries has limited value on approximal surfaces. The aim of this study was to develop and to test a new laser fluorescence (LF) device for the detection of approximal caries. Light with a wavelength of 655 nm was transported to the approximal surface using two different sapphire fibre tips. Seventy-five teeth were selected from a pool of extracted permanent human molars, frozen at -20 degrees C until use. Before being measured, they were defrosted, cleaned and calculus was removed with a scaler. The molars were set in blocks simulating the contact area of adults. Bitewing radiographs were obtained using Kodak Insight films. After two independent assessments with the new LF device, the teeth were histologically prepared, and assessed for caries extension. Using the laser, specificity values for D1 threshold (outer half of enamel), D2 threshold (inner half of enamel), D3 threshold (dentine) ranged between 0.81 and 0.93, sensitivity between 0.84 and 0.92 with no difference between the two tips. Bitewing radiography showed an inferior performance compared to LF (p<0.05). Intraex aminer reproducibility was high (kappa>.74). The new LF system might be a useful additional tool in detecting approximal caries. Because of its good reproducibility, it could be used to monitor caries regression or progression on approximal surfaces.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

The new device DIAGNOdent pen based on red laser light induced fluorescence was introduced for the detection of approximal and occlusal caries. The aim of this study was to test its performance on occlusal surfaces. The new device comes with two different sapphire fibre tips: a cylindrical tip and a conical tip. The two new sapphire fibre tips were used and compared with the tip currently available with DIAGNOdent (DD). METHODS: The teeth were selected from a pool of extracted permanent human molars, which were stored frozen at -20 degrees C, until use. Prior to being measured the teeth were defrosted and cleaned. One hundred and nineteen teeth were selected and measured with the old tip and with the two new tips of the new device by two independent assessments. The teeth were histologically prepared and assessed for caries extension. RESULTS: Specificity values for D(1), D(2) and D(3) ranged between 0.69 and 0.89, sensitivity between 0.78 and 0.96. There were no statistically significant differences obtained between the two tips of the new and the one tip of the old device (p>0.05). Intra-examiner reliability with kappa values of >0.83 was high. CONCLUSIONS: In this study, the new laser fluorescence device performed on occlusal surfaces as well as the available device.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

OBJECTIVES: To determine the clinical performance of a laser fluorescence device (DIAGNOdent pen, KaVo) to discriminate between different occlusal caries depths (D(0)-D(1-4); D(0-2)-D(3,4)) in permanent molars. METHODS: In this prospective, randomized two-centre-study 120 sound/uncavitated carious sites in 120 patients were measured after visual and radiographic caries assessment. In cases of operative intervention (n=86), the lesion depths after caries removal were recorded (reference). In cases of preventive intervention (n=34), the sites were reassessed visually/radiographically after 12 months to verify the status assessed before (reference). The discrimination performance was determined statistically (Mann-Whitney test, Spearman's rho coefficient, and areas under the receiver operating characteristic curves (AUCs)). Sensitivities (SE) and specificities (SP) were plotted as a function of the measured values and cut-off values for the mentioned thresholds suggested. RESULTS: Sound sites (n=13) had significantly minor fluorescence values than carious sites (n=107) (P<0.0001) as had sites with no/enamel caries (n=63) compared to dentinal caries (n=57). The AUCs for the same discriminations were 0.92 and 0.78 (P<0.001). For the D(0)-D(1-4) threshold, a cut-off at a value of 12 (SE: 0.88, SP: 0.85) and for the D(0-2)-D(3,4) threshold at 25 (SE: 0.67, SP: 0.79) can be suggested. A moderate positive correlation between the measurements and the caries depths was calculated (rho=+0.57, P=0.01). CONCLUSION: Within this study, the device's discrimination performance for different caries depths was moderate to very good and it may be recommended as adjunct tool in the diagnosis of occlusal caries.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

The aim of this study was to assess the influence of the zero value subtraction on the performance of laser fluorescence (LFpen) for approximal caries detection. Three areas (cuspal, middle and cervical) of both mesial and distal buccal surfaces of 78 permanent molars were assessed using both wedge-shaped (WDG) and tapered wedge-shaped (TWDG) tips. With the addition of the average, one cut-off value for each area was obtained and the performance was assessed. The areas under the receiver operating characteristics (ROC) curve, specificity, sensitivity and accuracy with and without the zero value subtraction were calculated. The McNemar test revealed a statistically significant difference for specificity at thresholds D(1), D(2) and D(3) (WDG) and D(1) and D(2) (TWDG) when the zero value subtraction was not performed. Influence of the zero value subtraction on the LFpen performance was observed for approximal caries detection. However, when modified cut-off values were used, the zero value subtraction could be eliminated.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

AIM To evaluate the performance of a pen‑type laser fluorescence device (DIAGNOdent 2190; LFpen, KaVo, Germany) and bitewing radiographs (BW) for approximal caries detection in permanent and primary teeth. MATERIALS AND METHODS A total of 246 anterior approximal surfaces (102 permanent and 144 primary) were selected. Contact points were simulated using sound teeth. Two examiners assessed all approximal surfaces using LFpen and BW. The teeth were histologically assessed for the reference standard. Optimal cut‑off limits were calculated for LFpen for primary and permanent teeth. Sensitivity, specificity, accuracy and area under the receiver operating characteristic curve (Az) were calculated for D1 (enamel and dentin lesions) and D3 (dentin lesions) thresholds. The reproducibility was assessed by intraclass correlation coefficient (ICC) and Cohen's weighted kappa values. RESULTS For permanent teeth, the LFpen cut‑off were 0- 27 (sound), 28- 33 (enamel caries) and >33 (dentin caries). For primary teeth, the LFpen cut‑off were 0- 7 (sound), 8- 32 (enamelcaries) and >32 (dentin caries). The LFpen presented higher sensitivity values than BW for primary teeth (0.58 vs. 0.32 at D1 and 0.80 vs. 0.47 at D3) and permanent teeth (0.80 vs. 0.57 at D1 and 0.94 vs. 0.51 at D3). Specificity did not show a significant difference between the methods. Rank correlations with histology were 0.59 and 0.83 (LFpen) and 0.36 and 0.70 (BW) for primary and permanent teeth, respectively, considering all lesions. ICC values for LFpen were 0.71 (inter) and 0.86 (intra) for permanent teeth and 0.94 (inter) and 0.90/0.99 for primary teeth. Kappa values for BW were 0.69 (inter) and 0.68/0.90 (intra) for permanent teeth and 0.64 (inter) and 0.89/0.89 for primary teeth. CONCLUSION LFpen presented better reproducibility for primary and permanent teeth and higher accuracy in detecting caries lesions at D1 threshold than BW for permanent teeth. LFpen should be used as an adjunct method for approximal caries detection.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Purpose: To compare visual inspection (VI), radiographic examination (RX) and the laser fluorescence device DIAGNOdent (L), as well as their combinations in vitro regarding treatment decisions for occlusal surfaces. Methods: 72 extracted human permanent teeth (molars and premolars) were used. Treatment decisions were recorded by three calibrated examiners, and the options available were fissure sealant and conservative restoration. For validation of treatment decisions, the teeth were sectioned and examined in a stereomicroscope. Thereafter, dental slices were scanned and the images were edited to facilitate classification of existing carious lesions. Intra and inter-examiner reproducibility for the determination of treatment plans were calculated using Cohen's kappa test (95%-CI). Sensitivity, specificity, positive and negative predictive values, and the area under the ROC curve were also calculated. Results: VI and L provided on average the greatest intra- and inter-examiner reproducibility, respectively. Although the combination of diagnostic methods may decrease both intra- and inter examiners reproducibility, combination of VI, L and RX resulted in the greatest sensitivity, being statistically superior to RX and L. There was more inter-examiner agreement for the option of restorative treatment, while the use of sealants as a treatment option yielded the lowest values. Negative predictive values were numerically inferior to positive predictive values, indicating that the examiners preferred not to restore a carious tooth than to proceed operatively in an intact tooth. The combination of the three methods studied showed the best results in determining treatment plans for occlusal surfaces, when compared to the other types of exams. on the other hand, radiographic examination and laser fluorescence were less efficient when used alone.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

The aim of this study was to determine the influence of thickness and aging on the intrinsic fluorescence of sealing materials and their ability to block fluorescence from the underlying surface as assessed using a laser fluorescence device. Cavities of 0.5 mm and 1 mm depth were drilled into acrylic boards which were placed over two surfaces with different fluorescence properties: a low-fluorescence surface, to assess the intrinsic fluorescence of the sealing materials, and a high-fluorescence surface, to assess the fluorescence-blocking ability of the sealing materials. Ten cavities of each depth were filled with different sealing materials: Adper Scotchbond Multi-Purpose, Adper Single Bond 2, FluroShield, Conseal f and UltraSeal XT Plus. Fluorescence was measured with a DIAGNOdent pen at five different time points: empty cavity, after polymerization, and 1 day, 1 week and 1 month after filling. The individual values after polymerization, as well as the area under the curve for the different periods were submitted to ANOVA and the Tukey test (p < 0.05). At 0.5 mm, Scotchbond, FluroShield and UltraSeal showed insignificant changes in intrinsic fluorescence with aging and lower fluorescence after polymerization than Single Bond and Conseal. At 1 mm, Scotchbond and FluroShield showed the lowest intrinsic fluorescence, but only Scotchbond showed no chagnes in fluorescence with aging. At both depths, Scotchbond blocked significantly less fluorescence. All sealing materials blocked more fluorescence when applied to a depth of 1 mm. At 0.5 mm, fissure sealants blocked more fluorescence than adhesives, and did not show significant changes with aging. Scotchbond had the least affect on the fluorescence from the underlying surface and would probably have the least affect on the monitoring of sealed dental caries by laser fluorescence.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

This study evaluated the performance of the DIAGNOdent pen laser fluorescence device (LFpen) in comparison with visual examination (VE), bitewing radiographs (BW) and visual examination combined with bitewing radiographs (VEBW) in detecting secondary approximal caries associated with composite restorations. In total, 60 approximal surfaces from 43 permanent molars with composite restorations were assessed twice by two examiners using the LFpen, VE, BW and VEBW. After histological preparation and hardness measurements, the sample was assigned to either a crown or root caries group, depending on the location of the lesions as the gold standard. For crown caries at D1, the highest values of specificity and sensitivity were observed for the LFpen at a cutoff value of 18 (1.00) and for the VEBW (0.89). At D3 (cutoff of 30), the LFpen showed the highest values of sensitivity and specificity. For root caries, the LFpen and VEBW showed the highest values of specificity (0.54), sensitivity (0.81) and accuracy (0.69). The Spearman rank correlation coefficients for crown/root caries with histology were 0.54/0.37 (LFpen), 0.29/0.10 (BW), 0.29/0.18 (VE) and 0.23/0.37 (VEBW). For the LFpen, the ICC varied from 0.80 (interexaminer) to 0.97 (intraexaminer B); the kappa value was 0.19 for BW and 0.35 for VE (interexaminer). Intraexaminer kappa values for BW were 0.25 (A) and 0.29 (B), and those for VE were 0.31 (A) and 0.32 (B). The LFpen device exhibited a performance comparable to that of conventional methods but with higher interexaminer reproducibility. Therefore, the LFpen should be considered an auxiliary method for the detection of secondary approximal caries associated with composite restorations.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

The aim of this study was to determine the influence of thickness and aging on the intrinsic fluorescence of sealing materials and their ability to block fluorescence from the underlying surface as assessed using a laser fluorescence device. Cavities of 0.5 mm and 1 mm depth were drilled into acrylic boards which were placed over two surfaces with different fluorescence properties: a low-fluorescence surface, to assess the intrinsic fluorescence of the sealing materials, and a high-fluorescence surface, to assess the fluorescence-blocking ability of the sealing materials. Ten cavities of each depth were filled with different sealing materials: Adper Scotchbond Multi-Purpose, Adper Single Bond 2, FluroShield, Conseal f and UltraSeal XT Plus. Fluorescence was measured with a DIAGNOdent pen at five different time points: empty cavity, after polymerization, and 1 day, 1 week and 1 month after filling. The individual values after polymerization, as well as the area under the curve for the different periods were submitted to ANOVA and the Tukey test (p < 0.05). At 0.5 mm, Scotchbond, FluroShield and UltraSeal showed insignificant changes in intrinsic fluorescence with aging and lower fluorescence after polymerization than Single Bond and Conseal. At 1 mm, Scotchbond and FluroShield showed the lowest intrinsic fluorescence, but only Scotchbond showed no chagnes in fluorescence with aging. At both depths, Scotchbond blocked significantly less fluorescence. All sealing materials blocked more fluorescence when applied to a depth of 1 mm. At 0.5 mm, fissure sealants blocked more fluorescence than adhesives, and did not show significant changes with aging. Scotchbond had the least affect on the fluorescence from the underlying surface and would probably have the least affect on the monitoring of sealed dental caries by laser fluorescence.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

The aim of this study was to compare the in situ and in vitro performances of a laser fluorescence (LF) device (DIAGNOdent 2095) with visual inspection for the detection of occlusal caries in permanent teeth. Sixty-four sites were selected, and visual inspection and LF assessments were carried out, in vitro, three times by two independent examiners, with a 1-week interval between evaluations. Afterwards, the occlusal surfaces were mounted on the palatal portion of removable acrylic orthodontic appliances and placed in six volunteers. Assessments were repeated and validated by histological analysis of the tooth sections under a stereomicroscope. For both examiners, the highest intra-examiner values were observed for the visual inspection when in vitro and in situ evaluations were compared. The inter-examiner reproducibility varied from 0.61 to 0.64, except for the in vitro assessment using LF, which presented a lower value (0.43). The methods showed high specificity at the D(1) threshold (considering enamel and dentin caries as disease). In vitro evaluations showed the highest values of sensitivity for both methods when compared to the in situ evaluations at D(1) and D(2) (considering only dentinal caries as the disease) thresholds. For both methods, the results of sensitivity (at D(1) and D(2)) and accuracy (at D(1)) showed significant differences between in vitro and in situ conditions. However, the sensitivity (at D(1) and D(2)), specificity and accuracy (both at D(1)) of the methods were not significantly different when the same condition was considered. It can be concluded that visual inspection and LF showed better performance in vitro than in situ.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

The aim of this in vivo study was to evaluate the performance of laser fluorescence (LF) comparing different cut-off limits for occlusal caries detection. One hundred and thirty first permanent molars were selected. Visual examination and LF assessments were performed independently. The extent of caries was assessed after operative intervention. New cut-off limits were established and compared with those proposed by the manufacturer and by Lussi et al. (Eur J Oral Sci 109:14-19, 2001). Similar sensitivity and higher specificity was found at D(2) (considering as disease only dentin caries) when the LF cut-off limits proposed by Lussi et al. and the new one were compared. At the D(3) threshold (considering as disease only deep dentin caries), no statistically significant difference among the cut-off limits for sensitivity was found. However, the new cut-off limits showed higher specificity. The LF device provided good ability to detect dentin caries lesions. Furthermore, the new cut-off limits and the values proposed by Lussi et al. could be suggested for the in vivo detection of occlusal caries.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Objective: The purpose of this study was to evaluate the efficacy of the laser fluorescence (LF) device in detecting in vitro demineralization and remineralization of smooth surface caries-like lesions. Background Data: The early detection of smooth surface caries-like lesions is important to provide proper management of carious lesions, and allows monitoring of them over time. Also, some authors suggest that LF could be useful in monitoring the caries pathological process. Materials and Methods: Seventy-eight blocks of bovine teeth were obtained, and then submitted to artificial caries lesion induction and to a pH-cycling process. Superficial microhardness (SMH) and laser fluorescence analysis were performed at baseline, after induction of caries-like lesions, and after the pH-cycling regimen to promote remineralization. Results: Friedman's and multiple comparison tests were performed for all variables. SMH analysis showed significant differences (p < 0.05) between baseline (286.77 +/- 1.49 Vickers hardness number [VHN] units), before (38.48 +/- 0.85 VHN), and after remineralization (131.93 +/- 2.63 VHN). Baseline values for LF were extremely low (2.71 +/- 0.05), and a statistically significant difference was observed only after remineralization (3.61 +/- 0.08), as demonstrated by the increase in LF values. Conclusion: The LF device did not show efficacy for monitoring in vitro demineralization and remineralization of smooth enamel surfaces.