996 resultados para Dynamic Contour Tonometry


Relevância:

100.00% 100.00%

Publicador:

Resumo:

OBJECTIVES: To test whether dynamic contour tonometry yields ocular pulse amplitude (OPA) measurements that are independent of corneal thickness and curvature, and to assess variables of observer agreement. METHODS: In a multivariate cluster analysis on 223 eyes, the relationship between central corneal thickness, corneal curvature, axial length, anterior chamber depth, intraocular pressure, sex, age, and OPA measurements was assessed. Intraobserver and interobserver variabilities were calculated from repeated measurements obtained from 8 volunteers by 4 observers. RESULTS: The OPA readings were not affected by central corneal thickness (P = .08), corneal curvature (P = .47), anterior chamber depth (P = .80), age (P = .60), or sex (P = .73). There was a positive correlation between OPA and intraocular pressure (0.12 mm Hg/1 mm Hg of intraocular pressure; P<.001) and a negative correlation between OPA and axial length (0.27 mm Hg/1 mm of length; P<.001). Intraobserver and interobserver variabilities were 0.08 and 0.02 mm Hg, respectively, and the intraclass correlation coefficient was 0.89. CONCLUSIONS: The OPA readings obtained with dynamic contour tonometry in healthy subjects are not influenced by the structure of the anterior segment of the eye but are affected by intraocular pressure and axial length. We found a high amount of agreement within and between observers.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

PURPOSE: To compare the performance of dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT) in measuring intraocular pressure in eyes with irregular corneas. METHODS: GAT and DCT measures were taken in 30 keratoconus and 29 postkeratoplasty eyes of 35 patients after pachymetry and corneal topography. Regression and correlation analyses were performed between both tonometry methods and between tonometry methods and corneal parameters. Bland-Altman plots were constructed. RESULTS: DCT values were significantly higher than GAT values in both study groups: +4.1 +/- 2.3 mm Hg (mean +/- SD) in keratoconus and +3.1 +/- 2.5 mm Hg after keratoplasty. In contrast to DCT, GAT values were significantly higher in postkeratoplasty eyes than in keratoconus. The correlation between the 2 tonometry methods was moderate in keratoconus (Kendall correlation coefficient, tau = 0.34) as well in postkeratoplasty eyes (tau = 0.66). The +/-1.96 SD span of the DCT-GAT differences showed a considerable range: -0.42 to +8.70 mm Hg in keratoconus and -1.87 to +7.98 mm Hg in postkeratoplasty eyes. In the keratoconus group, neither DCT nor GAT correlated significantly with any of the corneal parameters. In the postkeratoplasty group, both DCT and GAT measures showed a moderate positive correlation with corneal steepness, but only DCT had a significant negative correlation with the central corneal thickness (tau = -0.33). CONCLUSIONS: DCT measured significantly higher intraocular pressures than GAT in keratoconus and postkeratoplasty eyes. DCT and GAT measures varied considerably, and DCT was not less dependent on biomechanical properties of irregular corneas than GAT.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Aims: To investigate IOP measurements with the dynamic contour tonometer (DCT) and non contact tonometer (NCT) in subjects with keratoconus. Methods: Twenty keratoconic subjects and 20 age-matched control subjects had IOP measurements taken using DCT and NCT instruments. Central and offcentre measures were taken with the DCT in order to highlight any systematic errors associated with corneal biomechanical factors. Measures of anterior and posterior corneal topography and thickness were also taken for each subject. Results: No significant difference was found between the central and off-centre DCT IOP readings for the keratoconics and age-matched controls (p>0.05). The average DCT IOP for the keratoconics was 14.2 ± 1.4 mmHg and for the agematched controls was 14.2 ± 1.6 mmHg. However, the average NCT readings differed significantly (p<0.001) between the keratoconics (9.2 ± 1.5 mmHg) and age-matched controls (12.9 ± 2.4 mmHg). DCT IOP showed no significant (p>0.05) correlation with the severity of keratoconus, as determined through measures of corneal topography and thickness. NCT IOP was correlated significantly with certain measures of corneal curvature and thickness in the keratoconic population. The difference between DCT and NCT IOP was strongly correlated with measures of corneal topography and thickness, with differences increasing for more advanced keratoconus. Conclusions: The measurements from the DCT do not appear to be dependent upon corneal factors, unlike the NCT. The presence or severity of keratoconus was not correlated with DCT IOP values.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Justification: Le glaucome entraîne une perte progressive de la vision causée par la détérioration du nerf optique. Le glaucome est répandu dans le monde et cause la cécité dans environ sept millions de personnes. Le glaucome touche plus de 400 000 Canadiens et sa prévalence augmente avec le vieillissement de la population.1,2 Il s'agit d'une maladie chronique surnoise dont les symptômes se manifestent uniquement lors des stades avancés et qui peuvent mener à la cécité. Présentement, le seul moyen possible d’arrêter la progression du glaucome au stade initial est de diminuer la pression intra-oculaire (PIO). Les analogues de prostaglandines (APG) topiques sont fréquemment utilisées comme traitement de première ligne. Cependant, la recherche démontre que cette classe de médicaments peut changer certaines propriétés de la cornée, et possiblement influencer la mesure de la PIO.3 Objectif: À déterminer si l'utilisation d'APG affecte les propriétés biomécaniques de la cornée. La conclusion sera basée sur l'analyse intégrée des résultats obtenus de l'analyseur Reichert oculaire Réponse (ORA), la tonométrie par applanation de Goldmann (TAG) et la pachymétrie ultrasonographique. Le deuxième objectif potentiel de cette étude est de déterminer la corrélation, le cas échéant, entre les propriétés biomécaniques de la cornée, l'épaisseur de la cornée centrale (ECC) et la PIO chez les patients subissant un traitement d’APG topique. L'hypothèse principale de cette étude est que l’APG influence les propriétés de la cornée telles que l'épaisseur centrale, l'élasticité et la résistance. Patients et méthodes : Soixante-dix yeux de 35 patients, âgés de 50-85 ans, atteints de glaucome à angle ouvert (GAO) et traités avec APG topique ont été examinés. Seulement les sujets avec une réfraction manifeste entre -6,00 D et +4,25 D ont été inclus. Les critères d'exclusion sont: patients avec n'importe quelle autre maladie de la cornée de l’œil, telles que la dystrophie endothéliale de Fuch’s et kératocône, ou tout antécédent de traumatisme ou d'une chirurgie de la cornée, ainsi que le port de lentilles de contact. Nous avons demandé aux patients atteints du glaucome qui ont des paramètres stables et qui utilisent l’APG dans les deux yeux de cesser l’APG dans l'œil moins affecté par la PIO, et de continuer l’utilisation d’APG dans l'œil contralatéral. Le meilleur œil est défini comme celui avec moins de dommage sur le champ visuel (CV) (déviation moyenne (DM), le moins négatif) ou une PIO maximale historique plus basse si la DM est égale ou encore celui avec plus de dommage sur la tomographie par cohérence optique (TCO, Cirrus, CA) ou la tomographie confocale par balayage laser (HRT, Heidelberg, Allemagne). Toutes les mesures ont été prises avant la cessation d’APG et répétées 6 semaines après l’arrêt. Les patients ont ensuite recommencé l’utilisation d’APG et toutes les mesures ont été répétées encore une fois après une période supplémentaire de 6 semaines. Après commencer ou de changer le traitement du glaucome, le patient doit être vu environ 4-6 semaines plus tard pour évaluer l'efficacité de la goutte.4 Pour cette raison, on été décidé d'utiliser 6 semaines d'intervalle. Toutes les mesures ont été effectuées à l’institut du glaucome de Montréal par le même technicien, avec le même équipement et à la même heure de la journée. L'œil contralatéral a servi comme œil contrôle pour les analyses statistiques. La tonométrie par applanation de Goldmann a été utilisée pour mesurer la PIO, la pachymétrie ultrasonographique pour mesurer l'ECC, et l’ORA pour mesurer les propriétés biomécaniques de la cornée, incluant l'hystérèse cornéenne (HC). L’hypothèse de l'absence d'effet de l'arrêt de l’APG sur les propriétés biomécaniques a été examiné par un modèle linéaire à effets mixtes en utilisant le logiciel statistique R. Les effets aléatoires ont été définies à deux niveaux: le patient (niveau 1) et l'œil de chaque patient (niveau 2). Les effets aléatoires ont été ajoutés au modèle pour tenir compte de la variance intra-individuelle. L’âge a également été inclus dans le modèle comme variable. Les contrastes entre les yeux et les temps ont été estimés en utilisant les valeurs p ajustées pour contrôler les taux d'erreur internes de la famille en utilisant multcomp paquet dans R. Résultats: Une augmentation statistiquement significative due l 'HC a été trouvée entre les visites 1 (sur APG) et 2 (aucun APG) dans les yeux de l'étude, avec une moyenne (±erreur standard) des valeurs de 8,98 ± 0,29 mmHg et 10,35 ± 0,29 mmHg, respectivement, correspondant à une augmentation moyenne de 1,37 ± 0,18 mmHg (p <0,001). Une réduction significative de 1,25 ± 0,18 mmHg (p <0,001) a été observée entre les visites 2 et 3, avec une valeur moyenne HC finale de 9,09 ± 0,29 mmHg. En outre, une différence statistiquement significative entre l’oeil d’étude et le contrôle n'a été observée que lors de la visite 2 (1,01 ± 0,23 mmHg, p <0,001) et non lors des visites 1 et 3. Une augmentation statistiquement significative du facteur de résistance conréen (FRC) a été trouvée entre les visites 1 et 2 dans les yeux de l'étude, avec des valeurs moyennes de 10,23 ± 0,34 mmHg et 11,71 ± 0,34 mmHg, respectivement. Le FRC a ensuite été réduit de 1,90 ± 0,21 mmHg (p <0,001) entre les visites 2 et 3, avec une valeur moyenne FRC finale de 9,81 ± 0,34 mmHg. Une différence statistiquement significative entre l’oeil d’étude et le contrôle n'a été observée que lors de la visite 2 (1,46 ± 0,23 mmHg, p <0,001). Une augmentation statistiquement significative de l'ECC a été trouvée entre les visites 1 et 2 dans les yeux de l'étude, avec des valeurs moyennes de 541,83 ± 7,27 µm et 551,91 ± 7,27 µm, respectivement, ce qui correspond à une augmentation moyenne de 10,09 ± 0,94 µm (p <0,001). L'ECC a ensuite diminué de 9,40 ± 0,94 µm (p <0,001) entre les visites 2 et 3, avec une valeur moyenne finale de 542,51 ± 7,27 µm. Une différence entre l’étude et le contrôle des yeux n'a été enregistré que lors de la visite 2 (11,26 ± 1,79 µm, p <0,001). De même, on a observé une augmentation significative de la PIO entre les visites 1 et 2 dans les yeux de l'étude, avec des valeurs moyennes de 15,37 ± 0,54 mmHg et 18,37 ± 0,54 mmHg, respectivement, ce qui correspond à une augmentation moyenne de 3,0 ± 0,49 mmHg (p <0,001). Une réduction significative de 2,83 ± 0,49 mmHg (p <0,001) a été observée entre les visites 2 et 3, avec une valeur moyenne de la PIO finale de 15,54 ± 0,54 mmHg. L’oeil de contrôle et d’étude ne différaient que lors de la visite 2 (1,91 ± 0,49 mmHg, p <0,001), ce qui confirme l'efficacité du traitement de l’APG. Lors de la visite 1, le biais de la PIO (PIOcc - PIO Goldmann) était similaire dans les deux groupes avec des valeurs moyennes de 4,1 ± 0,54 mmHg dans les yeux de contrôles et de 4,8 ± 0,54 mmHg dans les yeux d’études. Lors de la visite 2, après un lavage de 6 semaines d’APG, le biais de la PIO dans l'œil testé a été réduit à 1,6 ± 0,54 mmHg (p <0,001), ce qui signifie que la sous-estimation de la PIO par TAG était significativement moins dans la visite 2 que de la visite 1. La différence en biais PIO moyenne entre l'étude et le contrôle des yeux lors de la visite 2, en revanche, n'a pas atteint la signification statistique (p = 0,124). On a observé une augmentation peu significative de 1,53 ± 0,60 mmHg (p = 0,055) entre les visites 2 et 3 dans les yeux de l'étude, avec une valeur de polarisation finale de la PIO moyenne de 3,10 ± 0,54 mmHg dans les yeux d'études et de 2,8 ± 0,54 mmHg dans les yeux de contrôles. Nous avons ensuite cherché à déterminer si une faible HC a été associée à un stade de glaucome plus avancé chez nos patients atteints du glaucome à angle ouvert traités avec l’APG. Lorsque l'on considère tous les yeux sur l’APG au moment de la première visite, aucune association n'a été trouvée entre les dommages sur le CV et l'HC. Cependant, si l'on considère seulement les yeux avec un glaucome plus avancé, une corrélation positive significative a été observée entre la DM et l'HC (B = 0,65, p = 0,003). Une HC inférieure a été associé à une valeur de DM de champ visuelle plus négative et donc plus de dommages liés au glaucome. Conclusions : Les prostaglandines topiques affectent les propriétés biomécaniques de la cornée. Ils réduisent l'hystérèse cornéenne, le facteur de résistance cornéen et l'épaisseur centrale de la cornée. On doit tenir compte de ces changements lors de l'évaluation des effets d’APG sur la PIO. Plus de recherche devrait être menées pour confirmer nos résultats. De plus, d’autres études pourraient être réalisées en utilisant des médicaments qui diminuent la PIO sans influencer les propriétés biomécaniques de la cornée ou à l'aide de tonomètre dynamique de Pascal ou similaire qui ne dépend pas des propriétés biomécaniques de la cornée. En ce qui concerne l'interaction entre les dommages de glaucome et l'hystérésis de la cornée, nous pouvons conclure qu' une HC inférieure a été associé à une valeur de DM de CV plus négative. Mots Clés glaucome - analogues de prostaglandines - hystérèse cornéenne – l’épaisseur de la cornée centrale - la pression intraoculaire - propriétés biomécaniques de la cornée.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

OBJECTIVE: To assess the agreement of tonometers available for clinical practice with the Goldmann applanation tonometer (GAT), the most commonly accepted reference device.

DESIGN: A systematic review and meta-analysis of directly comparative studies assessing the agreement of 1 or more tonometers with the reference tonometer (GAT).

PARTICIPANTS: A total of 11 582 participants (15 525 eyes) were included.

METHODS: Summary 95% limits of agreement (LoA) were produced for each comparison.

MAIN OUTCOME MEASURES: Agreement, recordability, and reliability.

RESULTS: A total of 102 studies, including 130 paired comparisons, were included, representing 8 tonometers: dynamic contour tonometer, noncontact tonometer (NCT), ocular response analyzer, Ocuton S, handheld applanation tonometer (HAT), rebound tonometer, transpalpebral tonometer, and Tono-Pen. The agreement (95% limits) seemed to vary across tonometers: 0.2 mmHg (-3.8 to 4.3 mmHg) for the NCT to 2.7 mmHg (-4.1 to 9.6 mmHg) for the Ocuton S. The estimated proportion within 2 mmHg of the GAT ranged from 33% (Ocuton S) to 66% and 59% (NCT and HAT, respectively). Substantial inter- and intraobserver variability were observed for all tonometers.

CONCLUSIONS: The NCT and HAT seem to achieve a measurement closest to the GAT. However, there was substantial variability in measurements both within and between studies.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Aim: The aim of this study was to evaluate the practicality and accuracy of tonometers used in routine clinical practice for established keratoconus (KC). Methods: This was a prospective study of 118 normal and 76 keratoconic eyes where intraocular pressure (IOP) was measured in random order using the Goldman applanation tonometer (GAT), Pascal dynamic contour tonometer (DCT), Reichert ocular response analyser (ORA) and TonoPen XL tonometer. Corneal hysteresis (CH) and corneal resistance factor (CRF), as calculated by the ORA, were recorded. Central corneal thickness (CCT) was measured using an ultrasound pachymeter. Results: The difference in IOP values between instruments was highly significant in both study groups (p<0.001). All other IOP measures were significantly higher than those for GAT, except for the Goldmann-correlated IOP (average of the two applanation pressure points) (IOPg) as measured by ORA in the control group and the CH-corrected IOP (corneal-compensated IOP value) (IOPcc) measures in the KC group. CCT, CH and CRF were significantly less in the KC group (p<0.001). Apart from the DCT, all techniques tended to measure IOP higher in eyes with thicker corneas. Conclusion: The DCT and the ORA are currently the most appropriate tonometers to use in KC for the measurement of IOPcc. Corneal factors such as CH and CRT may be of more importance than CCT in causing inaccuracies in applanation tonometry techniques.

Relevância:

90.00% 90.00%

Publicador:

Resumo:

Purpose: To analyse the relationship between measured intraocular pressure (IOP) and central corneal thickness (CCT), corneal hysteresis (CH) and corneal resistance factor (CRF) in ocular hypertension (OHT), primary open-angle (POAG) and normal tension glaucoma (NTG) eyes using multiple tonometry devices. Methods: Right eyes of patients diagnosed with OHT (n=47), normal tension glaucoma (n=17) and POAG (n=50) were assessed, IOP was measured in random order with four devices: Goldmann applanation tonometry (GAT); Pascal(R) dynamic contour tonometer (DCT); Reichert(R) ocular response analyser (ORA); and Tono-Pen(R) XL. CCT was then measured using a hand-held ultrasonic pachymeter. CH and CRF were derived from the air pressure to corneal reflectance relationship of the ORA data. Results: Compared to the GAT, the Tonopen and ORA Goldmann equivalent (IOPg) and corneal compensated (IOPcc) measured higher IOP readings (F=19.351, p<0.001), particularly in NTG (F=12.604, p<0.001). DCT was closest to Goldmann IOP and had the lowest variance. CCT was significantly different (F=8.305, p<0.001) between the 3 conditions as was CH (F=6.854, p=0.002) and CRF (F=19.653, p<0.001). IOPcc measures were not affected by CCT. The DCT was generally not affected by corneal biomechanical factors. Conclusion: This study suggests that as the true pressure of the eye cannot be determined non-invasively, measurements from any tonometer should be interpreted with care, particularly when alterations in the corneal tissue are suspected.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

Aims: To investigate the change that occurs in intraocular pressure (IOP) and ocular pulse amplitude (OPA) with accommodation in young adult myopes and emmetropes. Methods: Fifteen progressing myopic and 17 emmetropic young adult subjects had their IOP and OPA measured using the Pascal dynamic contour tonometer. Measurements were taken initially with accommodation relaxed, and then following 2 min of near fixation (accommodative demand 3 D). Baseline measurements of axial length and corneal thickness were also collected prior to the IOP measures. Results: IOP significantly decreased with accommodation in both the myopic and emmetropic subjects (mean change 1.861.1 mm Hg, p<0.0001). There was no significant difference (p>0.05) between myopes and emmetropes in terms of baseline IOP or the magnitude of change in IOP with accommodation. OPA also decreased significantly with accommodation (mean change for all subjects 0.560.5, p<0.0001). The myopic subjects (baseline OPA 2.060.7 mm Hg) exhibited a significantly lower baseline OPA (p¼0.004) than the emmetropes (baseline OPA 3.261.3 mm Hg),and a significantly lower magnitude of change in OPA with accommodation. Conclusion: IOP decreases significantly with accommodation, and changes similarly in progressing myopic and emmetropic subjects. However, differences found between progressing myopes and emmetropes in the mean OPA levels and the decrease in OPA associated with accommodation suggested some changes in IOP dynamics associated with myopia.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

Conventional cameras have limited dynamic range, and as a result vision-based robots cannot effectively view an environment made up of both sunny outdoor areas and darker indoor areas. This paper presents an approach to extend the effective dynamic range of a camera, achieved by changing the exposure level of the camera in real-time to form a sequence of images which collectively cover a wide range of radiance. Individual control algorithms for each image have been developed to maximize the viewable area across the sequence. Spatial discrepancies between images, caused by the moving robot, are improved by a real-time image registration process. The sequence is then combined by merging color and contour information. By integrating these techniques it becomes possible to operate a vision-based robot in wide radiance range scenes.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

This paper discusses the areawide Dynamic ROad traffic NoisE (DRONE) simulator, and its implementation as a tool for noise abatement policy evaluation. DRONE involves integrating a road traffic noise estimation model with a traffic simulator to estimate road traffic noise in urban networks. An integrated traffic simulation-noise estimation model provides an interface for direct input of traffic flow properties from simulation model to noise estimation model that in turn estimates the noise on a spatial and temporal scale. The output from DRONE is linked with a geographical information system for visual representation of noise levels in the form of noise contour maps.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

In this contribution, a stability analysis for a dynamic voltage restorer (DVR) connected to a weak ac system containing a dynamic load is presented using continuation techniques and bifurcation theory. The system dynamics are explored through the continuation of periodic solutions of the associated dynamic equations. The switching process in the DVR converter is taken into account to trace the stability regions through a suitable mathematical representation of the DVR converter. The stability regions in the Thevenin equivalent plane are computed. In addition, the stability regions in the control gains space, as well as the contour lines for different Floquet multipliers, are computed. Besides, the DVR converter model employed in this contribution avoids the necessity of developing very complicated iterative map approaches as in the conventional bifurcation analysis of converters. The continuation method and the DVR model can take into account dynamics and nonlinear loads and any network topology since the analysis is carried out directly from the state space equations. The bifurcation approach is shown to be both computationally efficient and robust, since it eliminates the need for numerically critical and long-lasting transient simulations.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

The catalytic conversion of adenosine triphosphate (ATP) and adenosine monophosphate (AMP) to adenosine diphosphate (ADP) by adenylate kinase (ADK) involves large amplitude, ligand induced domain motions, involving the opening and the closing of ATP binding domain (LID) and AMP binding domain (NMP) domains, during the repeated catalytic cycle. We discover and analyze an interesting dynamical coupling between the motion of the two domains during the opening, using large scale atomistic molecular dynamics trajectory analysis, covariance analysis, and multidimensional free energy calculations with explicit water. Initially, the LID domain must open by a certain amount before the NMP domain can begin to open. Dynamical correlation map shows interesting cross-peak between LID and NMP domain which suggests the presence of correlated motion between them. This is also reflected in our calculated two-dimensional free energy surface contour diagram which has an interesting elliptic shape, revealing a strong correlation between the opening of the LID domain and that of the NMP domain. Our free energy surface of the LID domain motion is rugged due to interaction with water and the signature of ruggedness is evident in the observed root mean square deviation variation and its fluctuation time correlation functions. We develop a correlated dynamical disorder-type theoretical model to explain the observed dynamic coupling between the motion of the two domains in ADK. Our model correctly reproduces several features of the cross-correlation observed in simulations. (C) 2011 American Institute of Physics. doi:10.1063/1.3516588]

Relevância:

30.00% 30.00%

Publicador:

Resumo:

An improved Boundary Contour System (BCS) and Feature Contour System (FCS) neural network model of preattentive vision is applied to large images containing range data gathered by a synthetic aperture radar (SAR) sensor. The goal of processing is to make structures such as motor vehicles, roads, or buildings more salient and more interpretable to human observers than they are in the original imagery. Early processing by shunting center-surround networks compresses signal dynamic range and performs local contrast enhancement. Subsequent processing by filters sensitive to oriented contrast, including short-range competition and long-range cooperation, segments the image into regions. The segmentation is performed by three "copies" of the BCS and FCS, of small, medium, and large scales, wherein the "short-range" and "long-range" interactions within each scale occur over smaller or larger distances, corresponding to the size of the early filters of each scale. A diffusive filling-in operation within the segmented regions at each scale produces coherent surface representations. The combination of BCS and FCS helps to locate and enhance structure over regions of many pixels, without the resulting blur characteristic of approaches based on low spatial frequency filtering alone.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

An improved Boundary Contour System (BCS) and Feature Contour System (FCS) neural network model of preattentive vision is applied to two large images containing range data gathered by a synthetic aperture radar (SAR) sensor. The goal of processing is to make structures such as motor vehicles, roads, or buildings more salient and more interpretable to human observers than they are in the original imagery. Early processing by shunting center-surround networks compresses signal dynamic range and performs local contrast enhancement. Subsequent processing by filters sensitive to oriented contrast, including short-range competition and long-range cooperation, segments the image into regions. Finally, a diffusive filling-in operation within the segmented regions produces coherent visible structures. The combination of BCS and FCS helps to locate and enhance structure over regions of many pixels, without the resulting blur characteristic of approaches based on low spatial frequency filtering alone.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

Aims: Angiographic ectasias and aneurysms in stented segments have been associated with late stent thrombosis. Using optical coherence tomography (OCT), some stented segments show coronary evaginations reminiscent of ectasias. The purpose of this study was to explore, using computational fluid-dynamic (CFD) simulations, whether OCT-detected coronary evaginations can induce local changes in blood flow. Methods and results: OCT-detected evaginations are defined as outward bulges in the luminal vessel contour between struts, with the depth of the bulge exceeding the actual strut thickness. Evaginations can be characterised cross ectionally by depth and along the stented segment by total length. Assuming an ellipsoid shape, we modelled 3-D evaginations with different sizes by varying the depth from 0.2-1.0 mm, and the length from 1-9 mm. For the flow simulation we used average flow velocity data from non-diseased coronary arteries. The change in flow with varying evagination sizes was assessed using a particle tracing test where the particle transit time within the segment with evagination was compared with that of a control vessel. The presence of the evagination caused a delayed particle transit time which increased with the evagination size. The change in flow consisted locally of recirculation within the evagination, as well as flow deceleration due to a larger lumen - seen as a deflection of flow towards the evagination. Conclusions: CFD simulation of 3-D evaginations and blood flow suggests that evaginations affect flow locally, with a flow disturbance that increases with increasing evagination size.