985 resultados para DIAGNOdent pen 2190


Relevância:

100.00% 100.00%

Publicador:

Resumo:

Background. The aim of this study was to evaluate the influence of zero-value subtraction on the performance of two laser fluorescence (LF) devices developed to 'detect occlusal caries.Methods. The authors selected 119 permanent molars. Two examiners assessed three areas (cuspal, middle and cervical) of both mesial and distal portions of the buccal surface and one occlusal site using an LF device and an LF pen. For each tooth, the authors subtracted the value measured in the cuspal, middle and cervical areas in the buccal surface from the value measured in the respective occlusal site.Results. The authors observed differences among the readings for both devices in the cuspal, middle and cervical areas in the buccal surface as well as differences for both devices with and without the zero-value subtraction in the occlusal surface. When the authors did not perform the zero-value subtraction, they found statistically significant differences for sensitivity and accuracy far the LF device. When this was done with the LF pen, specificity increased and sensitivity decreased significantly.Conclusions. For the LF device, the zero-value subtraction decreased the sensitivity. For this reason, the authors concluded that clinicians can obtain measures with the LF device effectively without using zero-value subtraction. For the LF pen, however, the absence of the zero-value subtraction changed both the Sensitivity and specificity, and so the authors concluded that clinicians should not eliminate this step from the procedure.Clinical Implications. When using the LF device, clinicians might not need to perform the zero-value subtraction; however, for the LF pen, clinicians should do so.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Background. The aim of this study was to evaluate the influence of zero-value subtraction on the performance of two laser fluorescence (LF) devices developed to detect occlusal caries. Methods. The authors selected 119 permanent molars. Two examiners assessed three areas (cuspal, middle and cervical) of both mesial and distal portions of the buccal surface and one occlusal site using an LF device and an LF pen. For each tooth, the authors subtracted the value measured in the cuspal, middle and cervical areas in the buccal surface from the value measured in the respective occlusal site. Results. The authors observed differences among the readings for both devices in the cuspal, middle and cervical areas in the buccal surface as well as differences for both devices with and without the zero-value subtraction in the occlusal surface. When the authors did not perform the zero-value subtraction, they found statistically significant differences for sensitivity and accuracy for the LF device. When this was done with the LF pen, specificity increased and sensitivity decreased significantly. Conclusions. For the LF device, the zero-value subtraction decreased the sensitivity. For this reason, the authors concluded that clinicians can obtain measures with the LF device effectively without using zero-value subtraction. For the LF pen, however, the absence of the zero-value subtraction changed both the sensitivity and specificity, and so the authors concluded that clinicians should not eliminate this step from the procedure. Clinical Implications. When using the LF device, clinicians might not need to perform the zero-value subtraction; however, for the LF pen, clinicians should do so.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Pós-graduação em Ciências Odontológicas - FOAR

Relevância:

100.00% 100.00%

Publicador:

Resumo:

The aim of this study was to assess the performance of two light-emitting diode (LED)- and two laser fluorescence-based devices in detecting occlusal caries in vitro. Ninety-seven permanent molars were assessed twice by two examiners using two LED- (Midwest Caries - MID and VistaProof - VP) and two laser fluorescence-based (DIAGNOdent 2095 - LF and DIAGNOdent pen 2190 - LFpen) devices. After measuring, the teeth were histologically prepared and classified according to lesion extension. At D1 the specificities were 0.76 (LF and LFpen), 0.94 (MID), and 0.70 (VP); the sensitivities were 0.70 (LF), 0.62 (LFpen), 0.31 (MID), and 0.75 (VP). At D(3) threshold the specificities were 0.88 (LF), 0.87 (LFpen), 0.90 (MID), and 0.70 (VP); the sensitivities were 0.63 (LF and LFpen), 0.70 (MID), and 0.96 (VP). Spearman's rank correlations with histology were 0.56 (LF), 0.51 (LFpen), 0.55 (MID), and 0.58 (VP). Inter- and intraexaminer ICC values were high and varied from 0.83 to 0.90. Both LF devices seemed to be useful auxiliary tools to the conventional methods, presenting good reproducibility and better accuracy at D(3) threshold. MID was not able to differentiate sound surfaces from enamel caries and VP still needs improvement on the cut-off limits for its use.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

This study compared the performance of fluorescence-based methods, radiographic examination, and International Caries Detection and Assessment System (ICDAS) II on occlusal surfaces. One hundred and nineteen permanent human molars were assessed twice by 2 experienced dentists using the laser fluorescence (LF and LFpen) and fluorescence camera (FC) devices, ICDAS II and bitewing radiographs (BW). After measuring, the teeth were histologically prepared and assessed for caries extension. The sensitivities for dentine caries detection were 0.86 (FC), 0.78 (LFpen), 0.73 (ICDAS II), 0.51 (LF) and 0.34 (BW). The specificities were 0.97 (BW), 0.89 (LF), 0.65 (ICDAS II), 0.63 (FC) and 0.56 (LFpen). BW presented the highest values of likelihood ratio (LR)+ (12.47) and LR- (0.68). Rank correlations with histology were 0.53 (LF), 0.52 (LFpen), 0.41 (FC), 0.59 (ICDAS II) and 0.57 (BW). The area under the ROC curve varied from 0.72 to 0.83. Inter- and intraexaminer intraclass correlation values were respectively 0.90 and 0.85 (LF), 0.93 and 0.87 (LFpen) and 0.85 and 0.76 (FC). The ICDAS II kappa values were 0.51 (interexaminer) and 0.61 (intraexaminer). The BW kappa values were 0.50 (interexaminer) and 0.62 (intraexaminer). The Bland and Altman limits of agreement were 46.0 and 38.2 (LF), 55.6 and 40.0 (LFpen) and 1.12 and 0.80 (FC), for intra- and interexaminer reproducibilities. The posttest probability for dentine caries detection was high for BW and LF. In conclusion, LFpen, FC and ICDAS II presented better sensitivity and LF and BW better specificity. ICDAS II combined with BW showed the best performance and is the best combination for detecting caries on occlusal surfaces. Copyright (C) 2008 S. Karger AG, Basel.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)

Relevância:

80.00% 80.00%

Publicador:

Resumo:

Pós-graduação em Ciências Odontológicas - FOAR

Relevância:

80.00% 80.00%

Publicador:

Resumo:

Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)

Relevância:

80.00% 80.00%

Publicador:

Resumo:

The aim of this clinical study was to evaluate and compare the performance of visual exam with use of the Nyvad criteria (visual examination - (VE)), interproximal radiography (BW), laser fluorescence device (DIAGNOdent Pen-DDPen), and their association in the diagnosis of proximal lesions in primary teeth. For this purpose, 45 children (n = 59 surfaces) of both sexes, aged between 5 and 9 years were selected, who presented healthy primary molars or primary molars with signs suggestive of the presence of caries lesions. The surfaces were clinically evaluated and coded according to the Nyvad criteria and immediately afterwards with the DDPen. Radiographic exam was performed only on the surfaces coded with Nyvad scores 2, 3, 5, or 6. Active caries lesions and/or those with discontinuous surfaces were restored, considering the depth of lesion as reference standard. Sensitivity, specificity, accuracy, and area under ROC curve were calculated for each technique and its associations. Visual exam with Nyvad criteria presented the highest specificity, accuracy, and area under ROC curve values. The DDPen presented the highest sensitivity values. Association with one or more methods resulted in an increase in specificity. The performance of visual, radiographic, and DDpen exams and their associations were good; however, the clinical examination with the Nyvad criteria was sufficient for the diagnosis of interproximal lesions in primary teeth.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

The aim of this study was to investigate the influence of interdental spacing on the performance of proximal caries detection methods in primary molars. In addition, aspects related to temporary tooth separation with orthodontic separators were evaluated. The proximal spaces between the posterior primary teeth (n = 344) of 76 children (4-12 years old) were evaluated before and after temporary separation. Stainless steel strips with different standardized thicknesses were used to measure the presence of biological spacing and the spacing obtained after temporary separation with orthodontic rubber rings. First, the presence of proximal caries lesions was assessed by visual inspection, bitewing radiographs and a pen-type laser fluorescence device (DIAGNOdent pen). Visual inspection after temporary separation with separators was the reference standard method in checking the actual presence of caries. Multilevel analyses were performed considering different outcomes: the performance of the methods in detecting caries lesions and the spacing after temporary separation. The spacing did not influence the performance of the caries detection methods. The maximum spacing obtained with temporary tooth separation was 0.80 mm (mean +/- standard deviation = 0.46 +/- 0.13 mm). The temporary separation was more effective in the upper arch and less effective when an initial biological interdental spacing was present. The biological interdental spacing does not influence the performance of proximal caries detection methods in primary molars, and temporary tooth separation provides spacing narrower than 1.0 mm.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

The aim of this study was to determine the influence of thickness and aging on the intrinsic fluorescence of sealing materials and their ability to block fluorescence from the underlying surface as assessed using a laser fluorescence device. Cavities of 0.5 mm and 1 mm depth were drilled into acrylic boards which were placed over two surfaces with different fluorescence properties: a low-fluorescence surface, to assess the intrinsic fluorescence of the sealing materials, and a high-fluorescence surface, to assess the fluorescence-blocking ability of the sealing materials. Ten cavities of each depth were filled with different sealing materials: Adper Scotchbond Multi-Purpose, Adper Single Bond 2, FluroShield, Conseal f and UltraSeal XT Plus. Fluorescence was measured with a DIAGNOdent pen at five different time points: empty cavity, after polymerization, and 1 day, 1 week and 1 month after filling. The individual values after polymerization, as well as the area under the curve for the different periods were submitted to ANOVA and the Tukey test (p < 0.05). At 0.5 mm, Scotchbond, FluroShield and UltraSeal showed insignificant changes in intrinsic fluorescence with aging and lower fluorescence after polymerization than Single Bond and Conseal. At 1 mm, Scotchbond and FluroShield showed the lowest intrinsic fluorescence, but only Scotchbond showed no chagnes in fluorescence with aging. At both depths, Scotchbond blocked significantly less fluorescence. All sealing materials blocked more fluorescence when applied to a depth of 1 mm. At 0.5 mm, fissure sealants blocked more fluorescence than adhesives, and did not show significant changes with aging. Scotchbond had the least affect on the fluorescence from the underlying surface and would probably have the least affect on the monitoring of sealed dental caries by laser fluorescence.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

International Journal of Paediatric Dentistry 2012; 22: 459466 Aim. This in vitro study aimed to test the performance of fluorescence-based methods in detecting occlusal caries lesions in primary molars compared to conventional methods. Design. Two examiners assessed 113 sites on 77 occlusal surfaces of primary molars using three fluorescence devices: DIAGNOdent (LF), DIAGNOdent pen (LFpen), and fluorescence camera (VistaProof-FC). Visual inspection (ICDAS) and radiographic methods were also evaluated. One examiner repeated the evaluations after one month. As reference standard method, the lesion depth was determined after sectioning and evaluation in stereomicroscope. The area under the ROC curve (Az), sensitivity, specificity, and accuracy of the methods were calculated at enamel (D1) and dentine caries (D3) lesions thresholds. The intra and interexaminer reproducibility were calculated using the intraclass correlation coefficient (ICC) and kappa statistics. Results. At D1, visual inspection presented higher sensitivities (0.970.99) but lower specificities (0.180.25). At D3, all the methods demonstrated similar performance (Az values around 0.90). Visual and radiographic methods showed a slightly higher specificity (values higher than 0.96) than the fluorescence based ones (values around 0.88). In general, all methods presented high reproducibility (ICC higher than 0.79). Conclusions. Although fluorescence-based and conventional methods present similar performance in detecting occlusal caries lesions in primary teeth, visual inspection alone seems to be sufficient to be used in clinical practice.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

The aim of this study was to evaluate the influence of dental plaque and moisture on performance of a laser fluorescence device in detecting occlusal and proximal caries lesions in primary teeth. Fifty-five occlusal and 58 proximal sites on primary molars were evaluated using a DIAGNOdent pen (LFpen) device. For the drying time study, the evaluations were performed in: (I) moist teeth; (II) teeth dried for 3 s, or (III) dried for 15 s. For the plaque study, the evaluations were done in sites: (I) without plaque; (II) with plaque, and (III) after cleaning. Evaluation of the teeth sections in stereomicroscope was the reference standard method. LF pen values, sensitivity, specificity, and accuracy were compared. The values obtained after 15 s of drying were higher than those obtained with moist teeth or dried for 3 s at both occlusal and proximal surfaces. However, there was no change in the performance in detecting caries lesions. With regard to the presence of plaque, there was no significant change in the readings of the device, but specificity was decreased in occlusal surfaces with plaque. At proximal surfaces, however, no significant differences were observed. In conclusion, the moisture conditions do not influence significantly the performance of the LFpen, but the presence of plaque can affect its performance in detecting occlusal caries lesions in primary teeth.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

Because discomfort caused by different approximal caries detection methods can influence their performance, the assessment of this discomfort is important. Thus, this study aimed to evaluate the discomfort reported by children after the use of different diagnostic methods to detect approximal caries lesions in primary teeth: visual inspection, bitewing radiography, laser fluorescence (DIAGNOdent pen - LFpen) and temporary separation with orthodontic rubbers. Seventy-six children aged 4 to 12 years were examined using these methods. Their discomfort was assessed using the Wong-Baker scale and compared among the methods. Visual inspection caused less discomfort than did other methods. Radiography and the LFpen presented similar levels of discomfort. Older children reported higher discomfort using temporary separation, whereas younger children reported less discomfort with the LFpen. In conclusion, radiographic, temporary separation and LFpen methods provoke higher discomfort than visual inspection.