BACKGROUND Measurement of HbA1c is the most important parameter to assess glycemic control in diabetic patients. Different point-of-care devices for HbA1c are available. The aim of this study was to evaluate two point-of-care testing (POCT) analyzers (DCA Vantage from Siemens and Afinion from Axis-Shield). We studied the bias and precision as well as interference from carbamylated hemoglobin. METHODS Bias of the POCT analyzers was obtained by measuring 53 blood samples from diabetic patients with a wide range of HbA1c, 4%-14% (20-130 mmol/mol), and comparing the results with those obtained by the laboratory method: HPLC HA 8160 Menarini. Precision was performed by 20 successive determinations of two samples with low 4.2% (22 mmol/mol) and high 9.5% (80 mmol/mol) HbA1c values. The possible interference from carbamylated hemoglobin was studied using 25 samples from patients with chronic renal failure. RESULTS The means of the differences between measurements performed by each POCT analyzer and the laboratory method (95% confidence interval) were: 0.28% (p<0.005) (0.10-0.44) for DCA and 0.27% (p<0.001) (0.19-0.35) for Afinion. Correlation coefficients were: r=0.973 for DCA, and r=0.991 for Afinion. The mean bias observed by using samples from chronic renal failure patients were 0.2 (range -0.4, 0.4) for DCA and 0.2 (-0.2, 0.5) for Afinion. Imprecision results were: CV=3.1% (high HbA1c) and 2.97% (low HbA1c) for DCA, CV=1.95% (high HbA1c) and 2.66% (low HbA1c) for Afinion. CONCLUSIONS Both POCT analyzers for HbA1c show good correlation with the laboratory method and acceptable precision.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Aims: Measurement of glycated hemoglobin (HbA1c) is an important indicator of glucose control over time. Point-of-care (POC) devices allow for rapid and convenient measurement of HbA1c, greatly facilitating diabetes care. We assessed two POC analyzers in the Peruvian Amazon where laboratory-based HbA1c testing is not available.

Methods: Venous blood samples were collected from 203 individuals from six different Amazonian communities with a wide range of HbA1c, 4.4-9.0% (25-75 mmol/mol). The results of the Afinion AS100 and the DCA Vantage POC analyzers were compared to a central laboratory using the Premier Hb9210 high-performance liquid chromatography (HPLC) method. Imprecision was assessed by performing 14 successive tests of a single blood sample.

Results: The correlation coefficient r for POC and HPLC results was 0.92 for the Afinion and 0.93 for the DCA Vantage. The Afinion generated higher HbA1c results than the HPLC (mean difference = +0.56% [+6 mmol/mol]; p < 0.001), as did the DCA Vantage (mean difference = +0.32% [4 mmol/mol]). The bias observed between POC and HPLC did not vary by HbA1c level for the DCA Vantage (p = 0.190), but it did for the Afinion (p < 0.001). Imprecision results were: CV = 1.75% for the Afinion, CV = 4.01% for the DCA Vantage. Sensitivity was 100% for both devices, specificity was 48.3% for the Afinion and 85.1% for the DCA Vantage, positive predictive value (PPV) was 14.4% for the Afinion and 34.9% for the DCA Vantage, and negative predictive value (NPV) for both devices was 100%. The area under the receiver operating characteristic (ROC) curve was 0.966 for the Afinion and 0.982 for the DCA Vantage. Agreement between HPLC and POC in classifying diabetes and prediabetes status was slight for the Afinion (Kappa = 0.12) and significantly different (McNemar’s statistic = 89; p < 0.001), and moderate for the DCA Vantage (Kappa = 0.45) and significantly different (McNemar’s statistic = 28; p < 0.001).

Conclusions: Despite significant variation of HbA1c results between the Afinion and DCA Vantage analyzers compared to HPLC, we conclude that both analyzers should be considered in health clinics in the Peruvian Amazon for therapeutic adjustments if healthcare workers are aware of the differences relative to testing in a clinical laboratory. However, imprecision and bias were not low enough to recommend either device for screening purposes, and the local prevalence of anemia and malaria may interfere with diagnostic determinations for a substantial portion of the population.