983 resultados para Water rights.
Resumo:
Water samples were collected from pre-dawn CTD casts at 5 depths corresponding to 55%, 20%, 7%, 5% and 1% of surface irradiance. 1 litre water samples wrapped with optical filters to replicate light levels. Spiked with 200 µL of 13C labelled sodium bicarbonate. After 24 hourse filtered through ashed 25mm GF/F (Whatman) filters, rinsed with HCl solution (1-2%) and stored at -20oC. On shore encapsulated in tin capsules and analysed for 13C isotopic enrichment. Carbon uptake rates calculated using the equations of Fernandez et al. (2005).
Resumo:
Analysis for micro-molar concentrations of nitrate and nitrite, nitrite, phosphate, silicate and ammonia was undertaken on a SEAL Analytical UK Ltd, AA3 segmented flow autoanalyser following methods described by Kirkwood (1996). Samples were drawn from Niskin bottles on the CTD into 15ml polycarbonate centrifuge tubes and kept refrigerated at approximately 4oC until analysis, which generally commenced within 30 minutes. Overall 23 runs with 597 samples were analysed. This is a total of 502 CTD samples, 69 underway samples and 26 from other sources. An artificial seawater matrix (ASW) of 40g/litre sodium chloride was used as the inter-sample wash and standard matrix. The nutrient free status of this solution was checked by running Ocean Scientific International (OSI) low nutrient seawater (LNS) on every run. A single set of mixed standards were made up by diluting 5mM solutions made from weighed dried salts in 1litre of ASW into plastic 250ml volumetric flasks that had been cleaned by washing in MilliQ water (MQ). Data processing was undertaken using SEAL Analytical UK Ltd proprietary software (AACE 6.07) and was performed within a few hours of the run being finished. The sample time was 60 seconds and the wash time was 30 seconds. The lines were washed daily with wash solutions specific for each chemistry, but comprised of MQ, MQ and SDS, MQ and Triton-X, or MQ and Brij-35. Three times during the cruise the phosphate and silicate channels were washed with a weak sodium hypochlorite solution.