1 resultado para Trabalho médico

em Universidade Federal de Uberlândia


Relevância:

30.00% 30.00%

Publicador:

Resumo:

This study aims to evaluate the uncertainty associated with measurements made by aneroid sphygmomanometer, neonatal electronic balance and electrocautery. Therefore, were performing repeatability tests on all devices for the subsequent execution of normality tests using Shapiro-Wilk; identification of influencing factors that affect the measurement result of each measurement; proposition of mathematical models to calculate the measurement uncertainty associated with measuring evaluated for all equipament and calibration for neonatal electronic balance; evaluation of the measurement uncertainty; and development of a computer program in Java language to systematize the calibration uncertainty of estimates and measurement uncertainty. It was proposed and carried out 23 factorial design for aneroid sphygmomanometer order to investigate the effect of temperature factors, patient and operator and another 32 planning for electrocautery, where it investigated the effects of temperature factors and output electrical power. The expanded uncertainty associated with the measurement of blood pressure significantly reduced the extent of the patient classification tracks. In turn, the expanded uncertainty associated with the mass measurement with neonatal balance indicated a variation of about 1% in the dosage of medication to neonates. Analysis of variance (ANOVA) and the Turkey test indicated significant and indirectly proportional effects of temperature factor in cutting power values and clotting indicated by electrocautery and no significant effect of factors investigated for aneroid sphygmomanometer.