5 resultados para AK44-5000

em Aston University Research Archive


Relevância:

20.00% 20.00%

Publicador:

Resumo:

A clinical evaluation of the Shin-Nippon SRW-5000 (Japan), a newly released commercial autorefractor, was undertaken to assess its repeatability and validity compared to subjective refraction. Measurements of refractive error were performed on 200 eyes of 100 subjects (aged 24.4±8.0 years) subjectively (non-cycloplegic) by one optometrist and objectively with the SRW-5000 autorefractor by a second. Repeatability was assessed by examining the differences between the seven autorefractor readings taken from each eye and by re-measuring the objective prescription of 50 eyes at a subsequent session. Although the SRW-5000 read slightly more plus than subjective refraction (mean spherical equivalent +0.16±0.44D), it was found to be highly valid (accurate) compared to subjective refraction and repeatable over the prescription range of +6.50 to -15.00D examined. The Shin-Nippon SRW-5000 autorefractor is therefore a valuable complement to subjective refraction and as it offers the advantage of a binocular open field-of-view, has a great potential benefit for accommodation research studies. Copyright © 2001 The College of Optometrists.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

PURPOSE. The purpose of this study was to evaluate the potential of the portable Grand Seiko FR-5000 autorefractor to allow objective, continuous, open-field measurement of accommodation and pupil size for the investigation of the visual response to real-world environments and changes in the optical components of the eye. METHODS. The FR-5000 projects a pair of infrared horizontal and vertical lines on either side of fixation, analyzing the separation of the bars in the reflected image. The measurement bars were turned on permanently and the video output of the FR-5000 fed into a PC for real-time analysis. The calibration between infrared bar separation and the refractive error was assessed over a range of 10.0 D with a model eye. Tolerance to longitudinal instrument head shift was investigated over a ±15 mm range and to eye alignment away from the visual axis over eccentricities up to 25.0°. The minimum pupil size for measurement was determined with a model eye. RESULTS. The separation of the measurement bars changed linearly (r = 0.99), allowing continuous online analysis of the refractive state at 60 Hz temporal and approximately 0.01 D system resolution with pupils >2 mm. The pupil edge could be analyzed on the diagonal axes at the same rate with a system resolution of approximately 0.05 mm. The measurement of accommodation and pupil size were affected by eccentricity of viewing and instrument focusing inaccuracies. CONCLUSIONS. The small size of the instrument together with its resolution and temporal properties and ability to measure through a 2 mm pupil make it useful for the measurement of dynamic accommodation and pupil responses in confined environments, although good eye alignment is important. Copyright © 2006 American Academy of Optometry.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

The internal optics of the recent models of the Shin-Nippon SRW-5000 autorefractor (also marketed as the Grand Seiko WV-500) have been modified by the manufacturer so that the infrared measurement ring has been replaced by pairs of horizontal and vertical infrared bars, on either side of fixation. The binocular, open field-of-view, allowing the accommodative state to be objectively monitored while a natural environment is viewed, has made the SRW-5000 a valuable tool in further understanding the nature of the oculomotor response. It is shown that the root-mean-square of model eye measures was least (0.017 ± 0.002D) when the separation of the horizontal measurement bars were averaged twice. The separation of the horizontal bars changes by 3.59 pixels/dioptre (r2 = 0.99), allowing continuous on-line analysis of the refractive state at up to 60 Hz temporal resolution to an accuracy of <0.001D, with pupils >3 mm. The pupil edge is not obscured in the diagonal axis by the measurement bars, unlike the ring of the original optics, so in the newer model pupil size can be measured simultaneously at the same rate with a resolution of <0.001 mm. The measurements of accommodation and pupil size are relatively unaffected by eccentricity of viewing up to ±10° from the visual axis and instrument focusing inaccuracies over a range of 10 mm towards the eye and 5 mm away from the eye. The resolution and temporal properties of the analysis are therefore ideal for the simultaneous measurement of dynamic accommodation and pupil responses. © 2004 The College of Optometrists.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

A newly released commercial autorefractor, the Shin-Nippon SRW-5000 (Japan), has been found to be valid compared to subjective refraction and repeatable over a wide prescription range. Its binocular open field-of-view allows the accommodative state to be monitored while a natural environment is viewed. In conventional static mode, the device can take up to 45 readings in 1min using digital image analysis of the reflected retinal image of a measurement ring. Continuous on-line analysis of the ring provides high (up to 60Hz) temporal resolution of the refractive state to an accuracy of <0.001D. Pupil size can also be analysed to a resolution of <0.001mm. The measurement of accommodation and pupil size was relatively unaffected by eccentricity of viewing up to ±10° and instrument focusing inaccuracies of ±5mm. The resolution properties of the analysis are shown to be ideal for measurement of dynamic accommodation and pupil responses. Copyright © 2001 The College of Optometrists.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Purpose: The Shin-Nippon SRW-5000 is an open view autorefractor that superseded the Canon R-1 autorefractor in the mid-1990s and has been used widely in optometry and vision science laboratories. It has been used to measure refractive error, accommodation responses both statically and dynamically, off-axis refractive error, and adapted to measure pupil size. This paper presents an overview of the original 2001 clinical evaluation of the SRW-5000 in adults (Mallen et al., Ophthal Physiol Opt 2001; 21: 101) and provides an update on the use and modification of the instrument since the original publication. Recent findings: The SRW-5000 instrument, and the family of devices which followed, have shown excellent validity, repeatability, and utility in clinical and research settings. The instruments have also shown great potential for increased research functionality following a number of modifications. Summary: The SRW-5000 and its derivatives have been, and continue to be, of significant importance in our drive to understand myopia progression, myopia control techniques, and oculomotor function in human vision.