Resumo:
Accommodating Intraocular Lenses (IOLs), multifocal IOLs (MIOLs) and toric IOLs are designed to provide a greater level of spectacle independency post cataract surgery. All of these IOLs are reliant on the accurate calculation of intraocular lens power determined through reliable ocular biometry. A standardised defocus area metric and reading performance index metric were devised for the evaluation of the range of focus and the reading ability of subjects implanted with presbyopic correcting IOLs. The range of clear vision after implantation of an MIOL is extended by a second focal point; however, this results in the prevalence of dysphotopsia. A bespoke halometer was designed and validated to assess this photopic phenomenon. There is a lack of standardisation in the methods used for determining IOL orientation and thus rotation. A repeatable, objective method was developed to allow the accurate assessment of IOL rotation, which was used to determine the rotational and positional stability of a closed loop haptic IOL. A new commercially available biometry device was validated for use with subjects prior to cataract surgery. The optical low coherence reflectometry instrument proved to be a valid method for assessing ocular biometry and covered a wider range of ocular parameters in comparison with previous instruments. The advantages of MIOLs were shown to include an extended range of clear vision translating into greater reading ability. However, an increased prevalence of dysphotopsia was shown with a bespoke halometer, which was dependent on the MIOL optic design. Implantation of a single optic accommodating IOL did not improve reading ability but achieved high subjective ratings of near vision. The closed-loop haptic IOL displayed excellent rotational stability in the late period but relatively poor rotational stability in the early period post implantation. The orientation error was compounded by the high frequency of positional misalignment leading to an extensive overall misalignment of the IOL. This thesis demonstrates the functionality of new IOL lens designs and the importance of standardised testing methods, thus providing a greater understanding of the consequences of implanting these IOLs. Consequently, the findings of the thesis will influence future designs of IOLs and testing methods.
Resumo:
The aim of this thesis was to develop standards of best practice for the subjective assessment of near visual function in presbyopia. Near visual acuity (VA) is a quick and simple measure but an assessment of the maximum reading speed and the smallest print size that can maintain this are equally important, to gain a better reflection of real world visual function. These metrics are dependent on the amplitude of accommodation (AoA) and often this must be evaluated using subjective techniques. Defocus curves are less susceptible than the push-up/push-down test to the influence of blur tolerance but their implementation must be standardised such that letter sequences and the order of lens presentation are randomised, to avoid memory effects, whilst the AoA should be quantified as the range of defocus for which only the best VA is maintained. In addition to such clinical assessments, subjective questionnaire evaluations are also important, to determine whether at least an individual’s needs are met. The Near Activity Visual Questionnaire (NAVQ) developed in this thesis can be used for this. Using these standardised near vision metrics it is shown that visual performance with monovision and multifocal contact lenses is comparable whilst initial outcomes of single optic ‘accommodating’ intraocular lens implantation are unlikely to be sustained in the long-term.
Resumo:
The tear film, cornea and lens dictate the refractive power of the eye and the retinal image quality is principally defined by diffraction, whole eye wavefront error, scatter, and chromatic aberration. Diffraction and wave aberration are fundamentally pupil diameter dependent; however scatter can be induced by refractive surgery and in the normal ageing eye becomes an increasingly important factor defining retinal image quality. The component of visual quality most affected by the tear film, refractive surgery and multifocal contact and intraocular lenses is the wave aberration of the eye. This body of work demonstrates the effects of each of these anomalies on the visual quality of the eye. When assessing normal or borderline self-diagnosed dry eye subjects using aberrometry, combining lubricating eye drops and spray does not offer any benefit over individual products. However, subjects perceive a difference in comfort for all interventions after one hour. Total higher order aberrations increase after laser assisted sub-epithelial keratectomy performed using a solid-state laser on myopes, but this causes no significant decrease in contrast sensitivity or increase in glare disability. Mean sensitivity and reliability indices for perimetry were comparable to pre-surgery results. Multifocal contact lenses and intraocular lenses are designed to maximise vision when the patient is binocular, so any evaluation of the eyes individually is confounded by reduced individual visual acuity and visual quality. Different designs of aspheric multifocal contact lenses do not provide the same level of visual quality. Multifocal contact lenses adversely affect mean deviation values for perimetry and this should be considered when screening individuals with multifocal contact or intraocular lenses. Photographic image quality obtained through a multifocal contact or intraocular lens appears to be unchanged. Future work should evaluate the effect of these anomalies in combination; with the aim of providing the best visual quality possible and supplying normative data for screening purposes.
Resumo:
Purpose: Evaluating the impact of splitting toric power on patient tolerance to misorientation such as with intraocular lens rotation. Setting: University vision clinic. Methods: Healthy, non astigmats had +1.50D astigmatism induced with spectacle lenses at 90°, 135°, 180° and +3.00D at 90°. Two correcting cylindrical lenses of the opposite sign and half the power each were subsequently added to the trial frame misaligned by 0°, 5° or 10° in a random order and misorientated from the initial axis in a clockwise direction by up to 15° in 5° steps. A second group of adapted astigmats with between 1.00 and 3.00DC had their astigmatism corrected with two toric spectacle lenses of half the power separated by 0°, 5° or 10° and misorientated from the initial axis in both directions by up to 15° in 5° steps. Distance, high contrast visual acuity was measured using a computerised test chart at each lens misalignment and misorientation. Results: Misorientation of the split toric lenses caused a statistically significant drop in visual acuity (F= 70.341; p< 0.001). Comparatively better acuities were observed around 180°, as anticipated (F= 3.775; p= 0.035). Misaligning the split toric power produced no benefit in visual acuity retention with axis misorientation when subjects had astigmatism induced with a low (F= 2.190, p= 0.129) or high cylinder (F= 0.491, p= 0.617) or in the adapted astigmats (F= 0.120, p= 0.887). Conclusion: Misalignment of toric lens power split across the front and back lens surfaces had no beneficial effect on distance visual acuity, but also no negative effect. © 2013 British Contact Lens Association.
Resumo:
As we welcome 2014 we say goodbye to 2013 and I must start with an apology to authors who have submitted papers to CLAE and seen a delay in either the review process or the hard copy publication of their proofed article. The delays were caused by a major hike in the number of submissions to the journal in 2012 that increased further in 2013. In the 12 months leading to the end of October 2011 we had 94 new paper submissions, and for the same period to the end of 2012 the journal had 116 new papers. In 2012 we were awarded an impact factor for the first time and following that the next 12 month period to the end of October 2013 saw a massive increase in submissions with 171 new manuscripts being submitted. This is nearly twice as many papers as 2 years ago and 3 times as many as when I took over as Editor-in-Chief. In addition to this the UK academics will know that 2014 is a REF year (Research Excellence Framework) where universities are judged on their research and one of the major components of this measure remains to be published papers so there is a push to publishing before the REF deadline for counting. The rejection rate at CLAE has gone up too and currently is around 50% (more than double the rejection rate when I took over as Editor-in-Chief). At CLAE the number of pages that we publish each year has remained the same since 2007. When compiling issue 1 for 2014 I chose the papers to be included from the papers that were proofed and ready to go and there were around 200 proofed pages ready, which is enough to fill 3½ issues! At present Elsevier and the BCLA are preparing to increase the number the pages published per issue so that we can clear some of this backlog and remain up to date with the papers published in CLAE. I should add that on line publishing of papers is still available and there may have been review delays but there are no publishing online so authors can still get an epub on line final version of their paper with a DOI (digital object identifier) number enabling the paper to be cited. There are two awards that were made in 2013 that I would like to make special mention of. One was for my good friend Jan Bergmanson, who was awarded an honorary life fellowship of the College of Optometrists. Jan has served on the editorial board of CLAE for many years and in 2013 also celebrated 30 years of his annual ‘Texan Corneal and contact lens meeting’. The other award I wish to mention is Judith Morris, who was the BCLA Gold Medal Award winner in 2013. Judith has had many roles in her career and worked at Moorfields Eye Hospital, the Institute of Optometry and currently at City University. She has been the Europe Middle East and Africa President of IACLE (International Association of Contact Lens Educators) for many years and I think I am correct in saying that Judith is the only person who was President of both the BCLA (1983) and a few years later she was the President College of Optometrists (1989). Judith was also instrumental in introducing Vivien Freeman to the BCLA as they had been friends and Judith suggested that Vivien apply for an administrative job at the BCLA. Fast forward 29 years and in December 2013 Vivien stepped down as Secretary General of the BCLA. I would like to offer my own personal thanks to Vivien for her support of CLAE and of me over the years. The BCLA will not be the same and I wish you well in your future plans. But 2014 brings in a new position to the BCLA – Cheryl Donnelly has been given the new role of Chief Executive Officer. Cheryl was President of the BCLA in 2000 and has previously served on council. I look forward to working with Cheryl and envisage a bright future for the BCLA and CLAE. In this issue we have some great papers including some from authors who have not published with CLAE before. There is a nice paper on contact lens compliance in Nepal which brings home some familiar messages from an emerging market. A paper on how corneal curvature is affected by the use of hydrogel lenses is useful when advising patients how long they should leave their contact lenses out for to avoid seeing changes in refraction or curvature. This is useful information when refracting these patients or pre-laser surgery. There is a useful paper offering tips on fitting bitoric gas permeable lenses post corneal graft and a paper detailing surgery to implant piggyback multifocal intraocular lenses. One fact that I noted from the selection of papers in the current issue is where they were from. In this issue none of the corresponding authors are from the United Kingdom. There are two papers each from the United States, Spain and Iran, and one each from the Netherlands, Ireland, Republic of Korea, Australia and Hong Kong. This is an obvious reflection of the widening interest in CLAE and the BCLA and indicates the new research groups emerging in the field.
Resumo:
Pseudophakic patients are frequently encountered in optometric practice, often the result of cataract extraction but also presbyopia correction. Given advances in technology and surgery, the demand for intraocular lenses for correcting a variety of refractive requirements has increased owing to an ageing population. Based on the patient’s needs, either fixed focus, toric, accommodating or multifocal intraocular lenses (IOLs) may be implanted. During optometric examination, attention should be drawn to a history of IOLs and the potential complications they may cause in order to manage them effectively, particularly where sight is threatened. Although objective and subjective refraction does not differ greatly between phakic and pseudophakic patients, care should be taken to set the patient up correctly and the reflex during retinoscopy observed for posterior sub-capsular opacification. Additional tests such as reading speed, and glare and contrast sensitivity are necessary to determine the outcome of IOL surgery and detect potential problems associated with multifocal and accommodating IOLs. Based upon the results of these tests, refraction, and type of IOL, contact lens or spectacle correction may be required to meet the visual demands of the patient.
Resumo:
Purpose: To evaluate the effect of reducing the number of visual acuity measurements made in a defocus curve on the quality of data quantified. Setting: Midland Eye, Solihull, United Kingdom. Design: Evaluation of a technique. Methods: Defocus curves were constructed by measuring visual acuity on a distance logMAR letter chart, randomizing the test letters between lens presentations. The lens powers evaluated ranged between +1.50 diopters (D) and -5.00 D in 0.50 D steps, which were also presented in a randomized order. Defocus curves were measured binocularly with the Tecnis diffractive, Rezoom refractive, Lentis rotationally asymmetric segmented (+3.00 D addition [add]), and Finevision trifocal multifocal intraocular lenses (IOLs) implanted bilaterally, and also for the diffractive IOL and refractive or rotationally asymmetric segmented (+3.00 D and +1.50 D adds) multifocal IOLs implanted contralaterally. Relative and absolute range of clear-focus metrics and area metrics were calculated for curves fitted using 0.50 D, 1.00 D, and 1.50 D steps and a near add-specific profile (ie, distance, half the near add, and the full near-add powers). Results: A significant difference in simulated results was found in at least 1 of the relative or absolute range of clear-focus or area metrics for each of the multifocal designs examined when the defocus-curve step size was increased (P<.05). Conclusion: Faster methods of capturing defocus curves from multifocal IOL designs appear to distort the metric results and are therefore not valid. Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned. © 2013 ASCRS and ESCRS.
Resumo:
PURPOSE: To determine the utility of a range of clinical and non-clinical indicators to aid the initial selection of the optimum presbyopic contact lens. In addition, to assess whether lens preference was influenced by the visual performance compared to the other designs trialled (intra-subject) or compared to participants who preferred other designs (inter-subject). METHODS: A double-masked randomised crossover trial of Air Optix Aqua multifocal, PureVision 2 for Presbyopia, Acuvue OASYS for Presbyopia, Biofinity multifocal and monovision was conducted on 35 presbyopes (54.3±6.2years). Participant lifestyle, personality, pupil characteristics and aberrometry were assessed prior to lens fitting. After 4 weeks of wear, high and low contrast visual acuity (VA) under photopic and mesopic conditions, reading speed, Near Activity Visual Questionnaire (NAVQ) rating, subjective quality-of-vision scoring, defocus curves, stereopsis, halometry, aberrometry and ocular physiology were quantified. RESULTS: After trialling all the lenses, preference was mixed (n=12 Biofinity, n=10 monovision, n=7 Purevision, n=4 Air Optix Aqua, n=2 Oasys). Lens preference was not dependent on personality (F=1.182, p=0.323) or the hours spent working at near (p=0.535) or intermediate (p=0.759) distances. No intersubject or strong intrasubject relationships emerged between lens preference and reading speed, NAVQ rating, halo size, aberrometry or ocular physiology (p>0.05). CONCLUSIONS: Participant lifestyle and personality, ocular optics, contact lens visual performance and ocular physiology provided poor indicators of the preferred lens type after 4 weeks of wear. This is confounded by the wide range of task visual demands of presbyopes and the limited optical differences between current multifocal contact lens designs.
Resumo:
Cataract surgery is a technique described since recorded history, yet it has greatly evolved only in the latter half of the past century. The development of the intraocular lens and phacoemulsification as a technique for cataract removal could be considered as the two most significant strides that have been made in this surgical field. This review takes a comprehensive look at all aspects of cataract surgery, starting from patient selection through the process of consent, anaesthesia, biometry, lens power calculation, refractive targeting, phacoemulsification, choice of intraocular lens and management of complications, such as posterior capsular opacification, as well as future developments. As the most common ophthalmic surgery and with the expanding range of intraocular lens options, optometrists have an important and growing role in managing patients with cataract.
Resumo:
Presbyopia is an age-related eye condition where one of the signs is the reduction in the amplitude of accommodation, resulting in the loss of ability to change the eye's focus from far to near. It is the most common age-related ailments affecting everyone around their mid-40s. Methods for the correction of presbyopia include contact lens and spectacle options but the surgical correction of presbyopia still remains a significant challenge for refractive surgeons. Surgical strategies for dealing with presbyopia may be extraocular (corneal or scleral) or intraocular (removal and replacement of the crystalline lens or some type of treatment on the crystalline lens itself). There are however a number of limitations and considerations that have limited the widespread acceptance of surgical correction of presbyopia. Each surgical strategy presents its own unique set of advantages and disadvantages. For example, lens removal and replacement with an intraocular lens may not be preferable in a young patient with presbyopia without a refractive error. Similarly treatment on the crystalline lens may not be a suitable choice for a patient with early signs of cataract. This article is a review of the options available and those that are in development stages and are likely to be available in the near future for the surgical correction of presbyopia.
Resumo:
High levels of corneal astigmatism are prevalent in a significant proportion of the population. During cataract surgery pre-existing astigmatism can be corrected using single or paired incisions on the steep axis of the cornea, using relaxing incisions or with the use of a toric intraocular lens. This review provides an overview of the conventional methods of astigmatic correction during cataract surgery and in particular, discusses the various types of toric lenses presently available and the techniques used in determining the correct axis for the placement of such lenses. Furthermore, the potential causes of rotation in toric lenses are identified, along with techniques for assessing and quantifying the amount of rotation and subsequent management options for addressing post-operative rotation.
Resumo:
PURPOSE OF REVIEW: Imaging of the crystalline lens and intraocular lens is becoming increasingly more important to optimize the refractive outcome of cataract surgery, to detect and manage complications and to ascertain advanced intraocular lens performance. This review examines recent advances in anterior segment imaging. RECENT FINDINGS: The main techniques used for imaging the anterior segment are slit-lamp biomicroscopy, ultrasound biomicroscopy, scheimpflug imaging, phakometry, optical coherence tomography and magnetic resonance imaging. They have principally been applied to the assessment of intraocular lens centration, tilt, position relative to the iris and movement with ciliary body contraction. SUMMARY: Despite the advances in anterior chamber imaging technology, there is still the need for a clinical, high-resolution, true anatomical, noninvasive technique to image behind the peripheral iris. © 2007 Lippincott Williams & Wilkins, Inc.
Resumo:
The binding theme of this thesis is the examination of both phakic and pseudophakic accommodation by means of theoretical modelling and the application of a new biometric measuring technique. Anterior Segment Optical Coherence Tomography (AS-OCT) was used to assess phakic accommodative changes in 30 young subjects (19.4 2.0 years; range, 18 to 25 years). A new method of assessing curvature change with this technique was employed with limited success. Changes in axial accommodative spacing, however, proved to be very similar to those of the Scheimpflug-based data. A unique biphasic trend in the position of the posterior crystalline lens surface during accommodation was discovered, which has not been alluded to in the literature. All axial changes with accommodation were statistically significant (p < 0.01) with the exception of corneal thickness (p = 0.81). A two-year follow-up study was undertaken for a cohort of subjects previously implanted with a new accommodating intraocular lens (AIOL) (Lenstec Tetraflex KH3500). All measures of best corrected distance visual acuity (BCDVA; +0.04 0.24 logMAR), distance corrected near visual acuity (DCNVA; +0.61 0.17 logMAR) and contrast sensitivity (+1.35 0.21 log units) were good. The subjective accommodation response quantified with the push-up technique (1.53 0.64 D) and defocus curves (0.77 0.29 D) was greater than the objective stimulus response (0.21 0.19 D). AS-OCT measures with accommodation stimulus revealed a small mean posterior movement of the AIOLs (0.02 0.03 mm for a 4.0 D stimulus); this is contrary to proposed mechanism of the anterior focus-shift principle.
Resumo:
PURPOSE: To assess the clinical outcomes after implantation of a new hydrophobic acrylic toric intraocular lens (IOL) to correct preexisting corneal astigmatism in patients having routine cataract surgery. SETTING: Four hospital eye clinics throughout Europe. DESIGN: Cohort study. METHODS: This study included eyes with at least 0.75 diopter (D) of preexisting corneal astigmatism having routine cataract surgery. Phacoemulsification was performed followed by insertion and alignment of a Tecnis toric IOL. Patients were examined 4 to 8 weeks postoperatively; uncorrected distance visual acuity (UDVA), corrected distance visual acuity, manifest refraction, and keratometry were measured. Individual patient satisfaction with uncorrected vision and the surgeon’s assessment of ease of handling and performance of the IOL were also documented. The cylinder axis of the toric IOL was determined by dilated slitlamp examination. RESULTS: The study enrolled 67 eyes of 60 patients. Four to 8 weeks postoperatively, the mean UDVA was 0.15 logMAR G 0.17 (SD) and the UDVA was 20/40 or better in 88% of eyes. The mean refractive cylinder decreased significantly postoperatively, from -1.91 +/- 1.07 D to -0.67 +/- 0.54 D. No significant change in keratometric cylinder was observed. The mean absolute IOL misalignment from the intended axis was 3.4 degrees (range 0 to 12 degrees). The good UDVA resulted in high levels of patient satisfaction. CONCLUSION: Implantation of the new toric IOL was an effective, safe, and predictable method to manage corneal astigmatism in patients having routine cataract surgery.
Resumo:
PURPOSE: To validate a new miniaturised, open-field wavefront device which has been developed with the capacity to be attached to an ophthalmic surgical microscope or slit-lamp. SETTING: Solihull Hospital and Aston University, Birmingham, UK DESIGN: Comparative non-interventional study. METHODS: The dynamic range of the Aston Aberrometer was assessed using a calibrated model eye. The validity of the Aston Aberrometer was compared to a conventional desk mounted Shack-Hartmann aberrometer (Topcon KR1W) by measuring the refractive error and higher order aberrations of 75 dilated eyes with both instruments in random order. The Aston Aberrometer measurements were repeated five times to assess intra-session repeatability. Data was converted to vector form for analysis. RESULTS: The Aston Aberrometer had a large dynamic range of at least +21.0 D to -25.0 D. It gave similar measurements to a conventional aberrometer for mean spherical equivalent (mean difference ± 95% confidence interval: 0.02 ± 0.49D; correlation: r=0.995, p<0.001), astigmatic components (J0: 0.02 ± 0.15D; r=0.977, p<0.001; J45: 0.03 ± 0.28; r=0.666, p<0.001) and higher order aberrations RMS (0.02 ± 0.20D; r=0.620, p<0.001). Intraclass correlation coefficient assessments of intra-sessional repeatability for the Aston Aberrometer were excellent (spherical equivalent =1.000, p<0.001; astigmatic components J0 =0.998, p<0.001, J45=0.980, p<0.01; higher order aberrations RMS =0.961, p<0.001). CONCLUSIONS: The Aston Aberrometer gives valid and repeatable measures of refractive error and higher order aberrations over a large range. As it is able to measure continuously, it can provide direct feedback to surgeons during intraocular lens implantations and corneal surgery as to the optical status of the visual system.