13 resultados para implant fracture

em BORIS: Bern Open Repository and Information System - Berna - Suiça


Relevância:

70.00% 70.00%

Publicador:

Resumo:

OBJECTIVE: The aim of this systematic review was to assess the survival rates of short-span implant-supported cantilever fixed dental prostheses (ICFDPs) and the incidence of technical and biological complications after an observation period of at least 5 years. MATERIAL AND METHODS: An electronic MEDLINE search supplemented by manual searching was conducted to identify prospective or retrospective cohort studies reporting data of at least 5 years on ICFDPs. Five- and 10-year estimates for failure and complication rates were calculated using standard or random-effect Poisson regression analysis. RESULTS: The five studies eligible for the meta-analysis yielded an estimated 5- and 10-year ICFDP cumulative survival rate of 94.3% [95 percent confidence interval (95% CI): 84.1-98%] and 88.9% (95% CI: 70.8-96.1%), respectively. Five-year estimates for peri-implantitis were 5.4% (95% CI: 2-14.2%) and 9.4% (95% CI: 3.3-25.4%) at implant and prosthesis levels, respectively. Veneer fracture (5-year estimate: 10.3%; 95% CI: 3.9-26.6%) and screw loosening (5-year estimate: 8.2%; 95% CI: 3.9-17%) represented the most common complications, followed by loss of retention (5-year estimate: 5.7%; 95% CI: 1.9-16.5%) and abutment/screw fracture (5-year estimate: 2.1%; 95% CI: 0.9-5.1%). Implant fracture was rare (5-year estimate: 1.3%; 95% CI: 0.2-8.3%); no framework fracture was reported. Radiographic bone level changes did not yield statistically significant differences either at the prosthesis or at the implant levels when comparing ICFDPs with short-span implant-supported end-abutment fixed dental prostheses. CONCLUSIONS: ICFDPs represent a valid treatment modality; no detrimental effects can be expected on bone levels due to the presence of a cantilever extension per se.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

Purpose: This retrospective study assessed the 10-year outcomes of titanium implants with a sandblasted and acid-etched (SLA) surface in a large cohort of partially edentulous patients. Materials and Methods: Records of patients treated with SLA implants between May 1997 and January 2001 were screened. Eligible patients were contacted and invited to undergo a clinical and radiologic examination. Each implant was classified according to strict success criteria. Results: Three hundred three patients with 511 SLA implants were available for the examination. The mean age of the patients at implant surgery was 48 years. Over the 10-year period, no implant fracture was noted, whereas six implants (1.2%) were lost. Two implants (0.4%) showed signs of suppuration at the 10-year examination, whereas seven implants had a history of peri-implantitis (1.4%) during the 10-year period, but presented with healthy peri-implant soft tissues at examination. The remaining 496 implants fulfilled the success criteria. The mean Plaque Index was 0.65 (±0.64), the mean Sulcus Bleeding Index 1.32 (±0.57), the mean Probing Depth 3.27 mm (±1.06), and the mean distance from the implant shoulder to the mucosal margin value -0.42 mm (±1.27). The radiologic mean distance from the implant shoulder to the first bone-to-implant contact was 3.32 mm (±0.73). Conclusion: The present retrospective analysis resulted in a 10-year implant survival rate of 98.8% and a success rate of 97.0%. In addition, the prevalence of peri-implantitis in this large cohort of orally healthy patients was low with 1.8% during the 10-year period.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

OBJECTIVES: To test the survival rates, and the technical and biological complication rates of customized zirconia and titanium abutments 5 years after crown insertion. MATERIAL AND METHODS: Twenty-two patients with 40 single implants in maxillary and mandibular canine and posterior regions were included. The implant sites were randomly assigned to zirconia abutments supporting all-ceramic crowns or titanium abutments supporting metal-ceramic crowns. Clinical examinations were performed at baseline, and at 6, 12, 36 and 60 months of follow-up. The abutments and reconstructions were examined for technical and/or biological complications. Probing pocket depth (PPD), plaque control record (PCR) and Bleeding on Probing (BOP) were assessed at abutments (test) and analogous contralateral teeth (control). Radiographs of the implants revealed the bone level (BL) on mesial (mBL) and distal sides (dBL). Data were statistically analyzed with nonparametric mixed models provided by Brunner and Langer and STATA (P < 0.05). RESULTS: Eighteen patients with 18 zirconia and 10 titanium abutments were available at a mean follow-up of 5.6 years (range 4.5-6.3 years). No abutment fracture or loss of a reconstruction occurred. Hence, the survival rate was 100% for both. Survival of implants supporting zirconia abutments was 88.9% and 90% for implants supporting titanium abutments. Chipping of the veneering ceramic occurred at three metal-ceramic crowns supported by titanium abutments. No significant differences were found at the zirconia and titanium abutments for PPD (meanPPD(ZrO2) 3.3 ± 0.6 mm, mPPD(T) (i) 3.6 ± 1.1 mm), PCR (mPCR(Z) (rO) (2) 0.1 ± 0.3, mPCR(T) (i) 0.3 ± 0.2) and BOP (mBOP(Z) (rO) (2) 0.5 ± 0.3, mBOP(T) (i) 0.6 ± 0.3). Moreover, the BL was similar at implants supporting zirconia and titanium abutments (mBL(Z) (rO) (2) 1.8 ± 0.5, dBL(Z) (rO) (2) 2.0 ± 0.8; mBL(T) (i) 2.0 ± 0.8, dBL(T) (i) 1.9 ± 0.8). CONCLUSIONS: There were no statistically or clinically relevant differences between the 5-year survival rates, and the technical and biological complication rates of zirconia and titanium abutments in posterior regions.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

OBJECTIVE: To report clinical evaluation of the clamp rod internal fixator 4.5/5.5 (CRIF 4.5/5.5) in bovine long-bone fracture repair. STUDY DESIGN: Retrospective study. ANIMALS: Cattle (n=22) with long-bone fractures. METHODS: Records for cattle with long-bone fractures repaired between 1999 and 2004 with CRIF 4.5/5.5 were reviewed. Quality of fracture repair, fracture healing, and clinical outcome were investigated by means of clinical examination, medical records, radiographs, and telephone questionnaire. RESULTS: Successful long-term outcome was achieved in 18 cattle (82%); 4 were euthanatized 2-14 days postoperatively because of fracture breakdowns. Two cattle had movement of clamps on the rod. Moderate to severe callus formation was evident in 11 cattle 6 months postoperatively. CONCLUSIONS: Movement of clamps on the rod was recognized as implant failure unique to the CRIF. This occurred in cattle with poor fracture stability because of an extensive cortical defect. The CRIF system may not be ideal to treat metacarpal/metatarsal fractures because its voluminous size makes skin closure difficult, thereby increasing the risk of postoperative infections. CLINICAL RELEVANCE: CRIF cannot be recommended for repair of complicated long-bone fractures in cattle.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

OBJECTIVES: The objective of this systematic review was to assess the 5-year survival of implant-supported single crowns (SCs) and to describe the incidence of biological and technical complications. METHODS: An electronic MEDLINE search complemented by manual searching was conducted to identify prospective and retrospective cohort studies on SCs with a mean follow-up time of at least 5 years. Failure and complication rates were analyzed using random-effects Poisson's regression models to obtain summary estimates of 5-year proportions. RESULTS: Twenty-six studies from an initial yield of 3601 titles were finally selected and data were extracted. In a meta-analysis of these studies, survival of implants supporting SCs was 96.8% [95% confidence interval (CI): 95.9-97.6%] after 5 years. The survival rate of SCs supported by implants was 94.5% (95% CI: 92.5-95.9%) after 5 years of function. The survival rate of metal-ceramic crowns, 95.4% (95% CI: 93.6-96.7%), was significantly (P=0.005) higher than the survival rate, 91.2% (95% CI: 86.8-94.2%), of all-ceramic crowns. Peri-implantitis and soft tissue complications occurred adjacent to 9.7% of the SCs and 6.3% of the implants had bone loss exceeding 2 mm over the 5-year observation period. The cumulative incidence of implant fractures after 5 years was 0.14%. After 5 years, the cumulative incidence of screw or abutment loosening was 12.7% and 0.35% for screw or abutment fracture. For supra-structure-related complications, the cumulative incidence of ceramic or veneer fractures was 4.5%. CONCLUSION: It can be concluded that after an observation period of 5 years, high survival rates for implants and implant-supported SCs can be expected. However, biological and particularly technical complications are frequent.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

Surgical navigation systems visualize the positions and orientations of surgical instruments and implants as graphical overlays onto a medical image of the operated anatomy on a computer monitor. The orthopaedic surgical navigation systems could be categorized according to the image modalities that are used for the visualization of surgical action. In the so-called CT-based systems or 'surgeon-defined anatomy' based systems, where a 3D volume or surface representation of the operated anatomy could be constructed from the preoperatively acquired tomographic data or through intraoperatively digitized anatomy landmarks, a photorealistic rendering of the surgical action has been identified to greatly improve usability of these navigation systems. However, this may not hold true when the virtual representation of surgical instruments and implants is superimposed onto 2D projection images in a fluoroscopy-based navigation system due to the so-called image occlusion problem. Image occlusion occurs when the field of view of the fluoroscopic image is occupied by the virtual representation of surgical implants or instruments. In these situations, the surgeon may miss part of the image details, even if transparency and/or wire-frame rendering is used. In this paper, we propose to use non-photorealistic rendering to overcome this difficulty. Laboratory testing results on foamed plastic bones during various computer-assisted fluoroscopybased surgical procedures including total hip arthroplasty and long bone fracture reduction and osteosynthesis are shown.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

OBJECTIVES: The objective of this systematic review was to assess the 5-year survival rates and incidences of complications associated with ceramic abutments and to compare them with those of metal abutments. METHODS: An electronic Medline search complemented by manual searching was conducted to identify randomized-controlled clinical trials, and prospective and retrospective studies providing information on ceramic and metal abutments with a mean follow-up time of at least 3 years. Patients had to have been examined clinically at the follow-up visit. Assessment of the identified studies and data abstraction was performed independently by three reviewers. Failure rates were analyzed using standard and random-effects Poisson regression models to obtain summary estimates of 5-year survival proportions. RESULTS: Twenty-nine clinical and 22 laboratory studies were selected from an initial yield of 7136 titles and data were extracted. The estimated 5-year survival rate of ceramic abutments was 99.1% [95% confidence interval (CI): 93.8-99.9%] and 97.4% (95% CI: 96-98.3%) for metal abutments. The estimated cumulative incidence of technical complications after 5 years was 6.9% (95% CI: 3.5-13.4%) for ceramic abutments and 15.9% (95% CI: 11.6-21.5%) for metal abutments. Abutment screw loosening was the most frequent technical problem, occurring at an estimated cumulative incidence after 5 years of 5.1% (95% CI: 3.3-7.7%). All-ceramic crowns supported by ceramic abutments exhibited similar annual fracture rates as metal-ceramic crowns supported by metal abutments. The cumulative incidence of biological complications after 5 years was estimated at 5.2% (95% CI: 0.4-52%) for ceramic and 7.7% (95% CI: 4.7-12.5%) for metal abutments. Esthetic complications tended to be more frequent at metal abutments. A meta-analysis of the laboratory data was impossible due to the non-standardized test methods of the studies included. CONCLUSION: The 5-year survival rates estimated from annual failure rates appeared to be similar for ceramic and metal abutments. The information included in this review did not provide evidence for differences of the technical and biological outcomes of ceramic and metal abutments. However, the information for ceramic abutments was limited in the number of studies and abutments analyzed as well as the accrued follow-up time. Standardized methods for the analysis of abutment strength are needed.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

INTRODUCTION Stable reconstruction of proximal femoral (PF) fractures is especially challenging due to the peculiarity of the injury patterns and the high load-bearing requirement. Since its introduction in 2007, the PF-locking compression plate (LCP) 4.5/5.0 has improved osteosynthesis for intertrochanteric and subtrochanteric fractures of the femur. This study reports our early results with this implant. METHODS Between January 2008 and June 2010, 19 of 52 patients (12 males, 7 females; mean age 59 years, range 19-96 years) presenting with fractures of the trochanteric region were treated at the authors' level 1 trauma centre with open reduction and internal fixation using PF-LCP. Postoperatively, partial weight bearing was allowed for all 19 patients. Follow-up included a thorough clinical and radiological evaluation at 1.5, 3, 6, 12, 24, 36 and 48 months. Failure analysis was based on conventional radiological and clinical assessment regarding the type of fracture, postoperative repositioning, secondary fracture dislocation in relation to the fracture constellation and postoperative clinical function (Merle d'Aubigné score). RESULTS In 18 patients surgery achieved adequate reduction and stable fixation without intra-operative complications. In one patient an ad latus displacement was observed on postoperative X-rays. At the third month follow-up four patients presented with secondary varus collapse and at the sixth month follow-up two patients had 'cut-outs' of the proximal fragment, with one patient having implant failure due to a broken proximal screw. Revision surgeries were performed in eight patients, one patient receiving a change of one screw, three patients undergoing reosteosynthesis with implantation of a condylar plate and one patient undergoing hardware removal with secondary implantation of a total hip prosthesis. Eight patients suffered from persistent trochanteric pain and three patients underwent hardware removal. CONCLUSIONS Early results for PF-LCP osteosynthesis show major complications in 7 of 19 patients requiring reosteosynthesis or prosthesis implantation due to secondary loss of reduction or hardware removal. Further studies are required to evaluate the limitations of this device.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

Purpose: The objective of this systematic review was to assess and compare the survival and complication rates of implant-supported prostheses reported in studies published in the year 2000 and before, to those reported in studies published after the year 2000. Materials and Methods: Three electronic searches complemented by manual searching were conducted to identify 139 prospective and retrospective studies on implant-supported prostheses. The included studies were divided in two groups: a group of 31 older studies published in the year 2000 or before, and a group of 108 newer studies published after the year 2000. Survival and complication rates were calculated using Poisson regression models, and multivariable robust Poisson regression was used to formally compare the outcomes of older and newer studies. Results: The 5-year survival rate of implant-supported prostheses was significantly increased in newer studies compared with older studies. The overall survival rate increased from 93.5% to 97.1%. The survival rate for cemented prostheses increased from 95.2% to 97.9%; for screw-retained reconstruction, from 77.6% to 96.8%; for implant-supported single crowns, from 92.6% to 97.2%; and for implant-supported fixed dental prostheses (FDPs), from 93.5% to 96.4%. The incidence of esthetic complications decreased in more recent studies compared with older ones, but the incidence of biologic complications was similar. The results for technical complications were inconsistent. There was a significant reduction in abutment or screw loosening by implant-supported FDPs. On the other hand, the total number of technical complications and the incidence of fracture of the veneering material was significantly increased in the newer studies. To explain the increased rate of complications, minor complications are probably reported in more detail in the newer publications. Conclusions: The results of the present systematic review demonstrated a positive learning curve in implant dentistry, represented in higher survival rates and lower complication rates reported in more recent clinical studies. The incidence of esthetic, biologic, and technical complications, however, is still high. Hence, it is important to identify these complications and their etiology to make implant treatment even more predictable in the future.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

PURPOSE Clinical studies related to the long-term outcomes with implant-supported reconstructions are still sparse. The aim of this 10-year retrospective study was to assess the rate of mechanical/technical complications and failures with implant supported fixed dental prostheses (FDPs) and single crowns (SCs) in a large cohort of partially edentulous patients. MATERIALS AND METHODS The comprehensive multidisciplinary examination consisted of a medical/dental history, clinical examination, and a radiographic analysis. Prosthodontic examination evaluated the implant-supported reconstructions for mechanical/technical complications and failures, occlusal analysis, presence/absence of attrition, and location, extension, and retention type. RESULTS Out of three hundred ninety seven fixed reconstructions in three hundred three patients, two hundred sixty eight were SCs and one hundred twenty seven were FDPs. Of these three hundred ninety seven implant-supported reconstructions, 18 had failed, yielding a failure rate of 4.5% and a survival rate of 95.5% after a mean observation period of 10.75 years (range: 8.4-13.5 years). The most frequent complication was ceramic chipping (20.31%) followed by occlusal screw loosening (2.57%) and loss of retention (2.06%). No occlusal screw fracture, one abutment loosening, and two abutment fractures were noted. This resulted in a total mechanical/technical complication rate of 24.7%. The prosthetic success rate over a mean follow-up time of 10.75 years was 70.8%. Generalized attrition and FDPs were associated with statistically significantly higher rates of ceramic fractures when compared with SCs. Cantilever extensions, screw retention, anterior versus posterior, and gender did not influence the chipping rate. CONCLUSIONS After a mean exposure time of 10.75 years, high survival rates for reconstructions supported by Sand-blasted Large-grit Acid-etched implants can be expected. Ceramic chipping was the most frequent complication and was increased in dentitions with attrition and in FDPs compared with SCs.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

PURPOSE To assess the survival outcomes and reported complications of screw- and cement-retained fixed reconstructions supported on dental implants. MATERIALS AND METHODS A Medline (PubMed), Embase, and Cochrane electronic database search from 2000 to September 2012 using MeSH and free-text terms was conducted. Selected inclusion and exclusion criteria guided the search. All studies were first reviewed by abstract and subsequently by full-text reading by two examiners independently. Data were extracted by two examiners and statistically analyzed using a random effects Poisson regression. RESULTS From 4,324 abstracts, 321 full-text articles were reviewed. Seventy-three articles were found to qualify for inclusion. Five-year survival rates of 96.03% (95% confidence interval [CI]: 93.85% to 97.43%) and 95.55% (95% CI: 92.96% to 97.19%) were calculated for cemented and screw-retained reconstructions, respectively (P = .69). Comparison of cement and screw retention showed no difference when grouped as single crowns (I-SC) (P = .10) or fixed partial dentures (I-FDP) (P = .49). The 5-year survival rate for screw-retained full-arch reconstructions was 96.71% (95% CI: 93.66% to 98.31). All-ceramic reconstruction material exhibited a significantly higher failure rate than porcelain-fused-to-metal (PFM) in cemented reconstructions (P = .01) but not when comparing screw-retained reconstructions (P = .66). Technical and biologic complications demonstrating a statistically significant difference included loss of retention (P ≤ .01), abutment loosening (P ≤ .01), porcelain fracture and/or chipping (P = .02), presence of fistula/suppuration (P ≤ .001), total technical events (P = .03), and total biologic events (P = .02). CONCLUSIONS Although no statistical difference was found between cement- and screw-retained reconstructions for survival or failure rates, screw-retained reconstructions exhibited fewer technical and biologic complications overall. There were no statistically significant differences between the failure rates of the different reconstruction types (I-SCs, I-FDPs, full-arch I-FDPs) or abutment materials (titanium, gold, ceramic). The failure rate of cemented reconstructions was not influenced by the choice of a specific cement, though cement type did influence loss of retention.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

BACKGROUND Implant-overdentures supported by rigid bars provide stability in the edentulous atrophic mandible. However, fractures of solder joints and matrices, and loosening of screws and matrices were observed with soldered gold bars (G-bars). Computer-aided designed/computer-assisted manufactured (CAD/CAM) titanium bars (Ti-bars) may reduce technical complications due to enhanced material quality. PURPOSE To compare prosthetic-technical maintenance service of mandibular implant-overdentures supported by CAD/CAM Ti-bar and soldered G-bar. MATERIALS AND METHODS Edentulous patients were consecutively admitted for implant-prosthodontic treatment with a maxillary complete denture and a mandibular implant-overdenture connected to a rigid G-bar or Ti-bar. Maintenance service and problems with the implant-retention device complex and the prosthesis were recorded during minimally 3-4 years. Annual peri-implant crestal bone level changes (ΔBIC) were radiographically assessed. RESULTS Data of 213 edentulous patients (mean age 68 ± 10 years), who had received a total of 477 tapered implants, were available. Ti-bar and G-bar comprised 101 and 112 patients with 231 and 246 implants, respectively. Ti-bar mostly exhibited distal bar extensions (96%) compared to 34% of G-bar (p < .001). Fracture rate of bars extensions (4.7% vs 14.8%, p < .001) and matrices (1% vs 13%, p < .001) was lower for Ti-bar. Matrices activation was required 2.4× less often in Ti-bar. ΔBIC remained stable for both groups. CONCLUSIONS Implant overdentures supported by soldered gold bars or milled CAD/CAM Ti-bars are a successful treatment modality but require regular maintenance service. These short-term observations support the hypothesis that CAD/CAM Ti-bars reduce technical complications. Fracture location indicated that the titanium thickness around the screw-access hole should be increased.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

Technical complications in implant prosthetic cases represent a major challenge in dentistry. This case report describes minimally invasive management to recover an implant with a fractured remnant of a zirconia abutment, including provisional rehabilitation during a sequential treatment protocol in the esthetic zone. A patient was treated with a screw-retained one-piece implant-supported reconstruction made of a customized zirconia abutment with direct ceramic veneering in the maxillary right central incisor position. During the prosthetic try-in, a fracture in the apical portion of the abutment was evident. The first rescue attempt led to fracture of the retrieval instrument. Immediately, an individualized wired construction was applied to bond the existing fractured reconstruction to the neighboring teeth to maintain the peri-implant mucosal architecture. Because the implant screw canal was blocked, a customized round bur had to be manufactured and was placed in the implant axis with a specific bracket tool from the service set to protect the interior implant threads. Then, the drills of the service set were guided by the newly created access to remove the fractured remnants. The implant screw was retapped and the area rinsed with chlorhexidine solution. All remnants were removed without the need for surgical intervention. Neither the implant connection nor the bone-to-implant interface was damaged. The stepwise treatment approach with the customized round bur combined with the system-specific drills of the service set saved the blocked implant so that the patient could be successfully rehabilitated with a new implant reconstruction.