19 resultados para Hearing Aids
em BORIS: Bern Open Repository and Information System - Berna - Suiça
Resumo:
OBJECTIVE: To generate anatomical data on the human middle ear and adjacent structures to serve as a base for the development and optimization of new implantable hearing aid transducers. Implantable middle ear hearing aid transducers, i.e. the equivalent to the loudspeaker in conventional hearing aids, should ideally fit into the majority of adult middle ears and should utilize the limited space optimally to achieve sufficiently high maximal output levels. For several designs, more anatomical data are needed. METHODS: Twenty temporal bones of 10 formalin-fixed adult human heads were scanned by a computed tomography system (CT) using a slide thickness of 0.63 mm. Twelve landmarks were defined and 24 different distances were calculated for each temporal bone. RESULTS: A statistical description of 24 distances in the adult human middle ear which may limit or influence the design of middle ear transducers is presented. Significant inter-individual differences but no significant differences for gender, side, age or degree of pneumatization of the mastoid were found. Distances, which were not analyzed for the first time in this study, were found to be in good agreement with the results of earlier studies. CONCLUSION: A data set describing the adult human middle ear anatomy quantitatively from the point of view of designers of new implantable hearing aid transducers has been generated. In principle, the method employed in this study using standard CT scans could also be used preoperatively to rule out exclusion criteria.
Resumo:
CONCLUSIONS: Speech understanding is better with the Baha Divino than with the Baha Compact in competing noise from the rear. No difference was found for speech understanding in quiet. Subjectively, overall sound quality and speech understanding were rated better for the Baha Divino. OBJECTIVES: To compare speech understanding in quiet and in noise and subjective ratings for two different bone-anchored hearing aids: the recently developed Baha Divino and the Baha Compact. PATIENTS AND METHODS: Seven adults with bilateral conductive or mixed hearing losses who were users of a bone-anchored hearing aid were tested with the Baha Compact in quiet and in noise. Tests were repeated after 3 months of use with the Baha Divino. RESULTS: There was no significant difference between the two types of Baha for speech understanding in quiet when tested with German numbers and monosyllabic words at presentation levels between 50 and 80 dB. For speech understanding in noise, an advantage of 2.3 dB for the Baha Divino vs the Baha Compact was found, if noise was emitted from a loudspeaker to the rear of the listener and the directional microphone noise reduction system was activated. Subjectively, the Baha Divino was rated statistically significantly better in terms of overall sound quality.
Resumo:
OBJECTIVE: To investigate correlations between preoperative hearing thresholds and postoperative aided thresholds and speech understanding of users of Bone-anchored Hearing Aids (BAHA). Such correlations may be useful to estimate the postoperative outcome with BAHA from preoperative data. STUDY DESIGN: Retrospective case review. SETTING: Tertiary referral center. PATIENTS:: Ninety-two adult unilaterally implanted BAHA users in 3 groups: (A) 24 subjects with a unilateral conductive hearing loss, (B) 38 subjects with a bilateral conductive hearing loss, and (C) 30 subjects with single-sided deafness. INTERVENTIONS: Preoperative air-conduction and bone-conduction thresholds and 3-month postoperative aided and unaided sound-field thresholds as well as speech understanding using German 2-digit numbers and monosyllabic words were measured and analyzed. MAIN OUTCOME MEASURES: Correlation between preoperative air-conduction and bone-conduction thresholds of the better and of the poorer ear and postoperative aided thresholds as well as correlations between gain in sound-field threshold and gain in speech understanding. RESULTS: Aided postoperative sound-field thresholds correlate best with BC threshold of the better ear (correlation coefficients, r2 = 0.237 to 0.419, p = 0.0006 to 0.0064, depending on the group of subjects). Improvements in sound-field threshold correspond to improvements in speech understanding. CONCLUSION: When estimating expected postoperative aided sound-field thresholds of BAHA users from preoperative hearing thresholds, the BC threshold of the better ear should be used. For the patient groups considered, speech understanding in quiet can be estimated from the improvement in sound-field thresholds.
Resumo:
OBJECTIVES To establish whether complex signal processing is beneficial for users of bone anchored hearing aids. METHODS Review and analysis of two studies from our own group, each comparing a speech processor with basic digital signal processing (either Baha Divino or Baha Intenso) and a processor with complex digital signal processing (either Baha BP100 or Baha BP110 power). The main differences between basic and complex signal processing are the number of audiologist accessible frequency channels and the availability and complexity of the directional multi-microphone noise reduction and loudness compression systems. RESULTS Both studies show a small, statistically non-significant improvement of speech understanding in quiet with the complex digital signal processing. The average improvement for speech in noise is +0.9 dB, if speech and noise are emitted both from the front of the listener. If noise is emitted from the rear and speech from the front of the listener, the advantage of the devices with complex digital signal processing as opposed to those with basic signal processing increases, on average, to +3.2 dB (range +2.3 … +5.1 dB, p ≤ 0.0032). DISCUSSION Complex digital signal processing does indeed improve speech understanding, especially in noise coming from the rear. This finding has been supported by another study, which has been published recently by a different research group. CONCLUSIONS When compared to basic digital signal processing, complex digital signal processing can increase speech understanding of users of bone anchored hearing aids. The benefit is most significant for speech understanding in noise.
Resumo:
Although non-organic hearing losses are relatively rare, it is important to identify suspicious findings early to be able to administer specific tests, such as objective measurements and specific counseling. In this retrospective study, we searched for findings that were specific ti or typical for non-organic hearing losses. Patient records from a 6 year period (2003-2008) from the University ENT Department of Bern, Switzerland, were reviewed. In this period, 40 subjects were diagnosed with a non-organic hearing loss (22 children, ages 7-16, mean 10.6 years; 18 adults, ages 19-57, mean 39.7 years; 25 females and 15 males). Pure tone audiograms in children and adults showed predominantly sensorineural and frequency-independent hearing losses, mostly in the range of 40-60 dB. In all cases, objective measurements (otoacoustic emissions and/or auditory-evoked potentials) indicated normal or substantially better hearing thresholds than those found in pure tone audiometry. In nine subjects (22.5%; 2 children, 7 adults), hearing aids had been fitted before the first presentation at our center. Six children (27%) had a history of middle ear problems with a transient hearing loss and 11 (50%) knew a person with a hearing loss. Two new and hitherto unreported findings emerged from the analysis: it was observed that a small air-bone gap of 5-20 dB was typical for non-organic hearing losses and that speech audiometry might show considerably poorer results than expected from pure tone audiometry.
Resumo:
Bone Anchored Hearing Implants (BAHI) are routinely used in patients with conductive or mixed hearing loss, e.g. if conventional air conduction hearing aids cannot be used. New sound processors and new fitting software now allow the adjustment of parameters such as loudness compression ratios or maximum power output separately. Today it is unclear, how the choice of these parameters influences aided speech understanding in BAHI users. In this prospective experimental study, the effect of varying the compression ratio and lowering the maximum power output in a BAHI were investigated. Twelve experienced adult subjects with a mixed hearing loss participated in this study. Four different compression ratios (1.0; 1.3; 1.6; 2.0) were tested along with two different maximum power output settings, resulting in a total of eight different programs. Each participant tested each program during two weeks. A blinded Latin square design was used to minimize bias. For each of the eight programs, speech understanding in quiet and in noise was assessed. For speech in quiet, the Freiburg number test and the Freiburg monosyllabic word test at 50, 65, and 80 dB SPL were used. For speech in noise, the Oldenburg sentence test was administered. Speech understanding in quiet and in noise was improved significantly in the aided condition in any program, when compared to the unaided condition. However, no significant differences were found between any of the eight programs. In contrast, on a subjective level there was a significant preference for medium compression ratios of 1.3 to 1.6 and higher maximum power output.
Resumo:
The data of 46 adults with single-sided sensorineural deafness who were candidates for bone-anchored hearing aids (Baha) CROS (contralateral routing of signals) were analyzed. All candidates tested a Baha with a headband in their normal environment. Subsequently, 29 of the candidates chose a permanent Baha CROS fitting, and 17 declined, thus forming the two study groups. No significant difference regarding age, sex or duration of deafness was found between the two groups. Similarly, the transcranial attenuation was not significantly different between those who accepted and declined a Baha. Subjects with some residual hearing in their poorer ear tended to decline a Baha, but the effect was not statistically significant. For a subset of 28 subjects, the Bern Benefit in Single-Sided Deafness questionnaire was administered. The questionnaire consists of 10 visual analogue scales rating the subjectively perceived benefit of the Baha or any other CROS device in different situations. Scores were found to be significantly higher for speech understanding at some distance (p = 0.026), for speech understanding in noise (p = 0.037), for group conversations (p < 0.01), and for the overall benefit (p < 0.01) for those candidates who chose to use a Baha as a CROS device permanently.
Resumo:
To investigate the effect of low-frequency attenuation of Bone-Anchored Hearing Aids (Bahas) in users with single-sided sensorineural deafness (SSD). The underlying notion is that low-frequency sounds up to approximately 1500 Hz reach the contralateral ear without significant attenuation and that Bahas tend to show more distortion at lower frequencies. Furthermore, to transmit low frequencies, higher moving masses are needed when compared with high frequencies.
Resumo:
A life-size mechanical middle ear model and human temporal bones were used to evaluate three different middle ear transducers for implantable hearing aids: the driving rod transducer (DRT), the floating mass transducer (FMT) or vibrant sound bridge, and the contactless transducer (CLT). Results of the experiments with the mechanical model were within the range of the results for human temporal bones. However, results with the mechanical model showed better reproducibility. The handling of the mechanical model was considerably simpler and less time-consuming. Systematic variations of mounting parameters showed that the angle of the rod has virtually no effect on the output of the DRT, the mass loading on the cable of the FMT has a larger impact on the output than does the tightness of crimping, and the output level of the CLT can be increased by 10 dB by optimizing the mounting parameters.
Resumo:
OBJECTIVE: External auditory canal cholesteatoma (EACC) is a rarity. Although there have been numerous case reports, there are only few systematic analyses of case series, and the pathogenesis of idiopathic EACC remains enigmatic. STUDY DESIGN: In a tertiary referral center for a population of 1.5 million inhabitants, 34 patients with 35 EACC (13 idiopathic [1 bilateral] and 22 secondary) who were treated between 1994 and 2006 were included in the study. RESULTS: EACC cardinal symptoms were longstanding otorrhea (65%) and dull otalgia (12%). Focal bone destruction in the external auditory canal with retained squamous debris and an intact tympanic membrane were characteristic. Only 27% of the patients showed conductive hearing loss exceeding 20 dB. Patients with idiopathic EACC had lesions typically located on the floor of the external auditory canal and were older, and the mean smoking intensity was also greater (p < 0.05) compared with patients with secondary EACC. The secondary lesions were assigned to categories (poststenotic [n = 6], postoperative [n = 6], and posttraumatic EACC [n = 4]) and rare categories (radiogenic [n = 2], postinflammatory [n = 1], and postobstructive EACC [n = 1]). In addition, we describe 2 patients with EACC secondary to the complete remission of a Langerhans cell histiocytosis of the external auditory canal. Thirty of 34 patients were treated surgically and became all free of recurrence, even after extensive disease. DISCUSSION: For the development of idiopathic EACC, repeated microtrauma (e.g., microtrauma resulting from cotton-tipped applicator abuse or from hearing aids) and diminished microcirculation (e.g., from smoking) might be risk factors. A location other than in the inferior portion of the external auditory canal indicates a secondary form of the disease, as in the case of 2 patients with atypically located EACC after years of complete remission of Langerhans cell histiocytosis, which we consider as a new posttumorous category and specific late complication of this rare disease.
Resumo:
Users of cochlear implant systems, that is, of auditory aids which stimulate the auditory nerve at the cochlea electrically, often complain about poor speech understanding in noisy environments. Despite the proven advantages of multimicrophone directional noise reduction systems for conventional hearing aids, only one major manufacturer has so far implemented such a system in a product, presumably because of the added power consumption and size. We present a physically small (intermicrophone distance 7 mm) and computationally inexpensive adaptive noise reduction system suitable for behind-the-ear cochlear implant speech processors. Supporting algorithms, which allow the adjustment of the opening angle and the maximum noise suppression, are proposed and evaluated. A portable real-time device for test in real acoustic environments is presented.
Resumo:
The cochlear implant (CI) is one of the most successful neural prostheses developed to date. It offers artificial hearing to individuals with profound sensorineural hearing loss and with insufficient benefit from conventional hearing aids. The first implants available some 30 years ago provided a limited sensation of sound. The benefit for users of these early systems was mostly a facilitation of lip-reading based communication rather than an understanding of speech. Considerable progress has been made since then. Modern, multichannel implant systems feature complex speech processing strategies, high stimulation rates and multiple sites of stimulation in the cochlea. Equipped with such a state-of-the-art system, the majority of recipients today can communicate orally without visual cues and can even use the telephone. The impact of CIs on deaf individuals and on the deaf community has thus been exceptional. To date, more than 300,000 patients worldwide have received CIs. In Switzerland, the first implantation was performed in 1977 and, as of 2012, over 2,000 systems have been implanted with a current rate of around 150 CIs per year. The primary purpose of this article is to provide a contemporary overview of cochlear implantation, emphasising the situation in Switzerland.
Resumo:
INTRODUCTION Sound can reach the inner ear via at least two different pathways: air conduction and bone conduction (BC). BC hearing is used clinically for diagnostic purposes and for BC hearing aids. Research on the motion of the human middle ear in response to BC stimulation is typically conducted using cadaver models. We evaluated middle ear motion of Thiel-embalmed whole-head specimens in terms of linearity, reproducibility, and consistency with the reported middle ear motion of living subjects, fresh cadaveric temporal bones, and whole-heads embalmed with a Non-Thiel solution of salts. METHODS We used laser Doppler vibrometry to measure the displacement of the skull, the umbo, the cochlear promontory, the stapes, and the round window in seven ears from four human whole-head specimens embalmed according to Thiel's method. The ears were stimulated with a Baha(®) implanted behind the auricle. RESULTS The Thiel model shows promontory velocity similar to that reported in the literature for whole-heads embalmed with a Non-Thiel solution of salts (0- to 7-dB difference). The Thiel heads' relative velocity of the stapes with respect to the promontory was similar to that of fresh cadaver temporal bones (0- to 4-dB difference). The velocity of the umbo was comparable in Thiel-embalmed heads and living subjects (0- to 10-dB difference). The skull and all middle ear elements measured responded linearly to different stimulation levels, with an average difference less than 1 dB. The variability of repeated measurements for both short- (2 h; 4 dB) and long-term (4-16 weeks; 6 dB) repetitions in the same ear, and the difference between the two ears of the same donor (approximately 10 dB) were lower than the inter-individual difference (up to 25 dB). CONCLUSION Thiel-embalmed human whole-head specimens can be used as an alternative model for the study of human middle ear mechanics secondary to BC stimulation. At some frequencies, differences from living subjects must be considered.
Resumo:
The bone-anchored port (BAP) is an investigational implant, which is intended to be fixed on the temporal bone and provide vascular access. There are a number of implants taking advantage of the stability and available room in the temporal bone. These devices range from implantable hearing aids to percutaneous ports. During temporal bone surgery, injuring critical anatomical structures must be avoided. Several methods for computer-assisted temporal bone surgery are reported, which typically add an additional procedure for the patient. We propose a surgical guide in the form of a bone-thickness map displaying anatomical landmarks that can be used for planning of the surgery, and for the intra-operative decision of the implant’s location. The retro-auricular region of the temporal and parietal bone was marked on cone-beam computed tomography scans and tridimensional surfaces displaying the bone thickness were created from this space. We compared this method using a thickness map (n = 10) with conventional surgery without assistance (n = 5) in isolated human anatomical whole head specimens. The use of the thickness map reduced the rate of Dura Mater exposition from 100% to 20% and OPEN ACCESS Materials 2013, 6 5292 suppressed sigmoid sinus exposures. The study shows that a bone-thickness map can be used as a low-complexity method to improve patient’s safety during BAP surgery in the temporal bone.
Resumo:
INTRODUCTION We report the first findings of functional magnetic resonance imaging of the auditory cortex in a young woman with a bilateral cochleovestibular deficit as first manifestation of Brown-Vialetto-Van Leare syndrome. The patient had no open speech discrimination, even with hearing aids, and is depending on lip reading for communication. METHODS To evaluate the possible efficiency of a cochlear implantation, we investigated hemodynamic responses within the central auditory pathways using an auditory functional magnetic resonance imaging paradigm. RESULTS Blood oxygen level-dependent correlates were detected bilaterally along the auditory pathways after exposure to intermittent clicking tone stimulation at 2 kHz. CONCLUSION These results suggest integrity of the central auditory pathways and represent a positive argument to propose a cochlear implantation with the aim to restore hearing.