24 resultados para Fluorescence induite par laser

em BORIS: Bern Open Repository and Information System - Berna - Suiça


Relevância:

40.00% 40.00%

Publicador:

Resumo:

This study evaluated the performance of the DIAGNOdent pen laser fluorescence device (LFpen) in comparison with visual examination (VE), bitewing radiographs (BW) and visual examination combined with bitewing radiographs (VEBW) in detecting secondary approximal caries associated with composite restorations. In total, 60 approximal surfaces from 43 permanent molars with composite restorations were assessed twice by two examiners using the LFpen, VE, BW and VEBW. After histological preparation and hardness measurements, the sample was assigned to either a crown or root caries group, depending on the location of the lesions as the gold standard. For crown caries at D1, the highest values of specificity and sensitivity were observed for the LFpen at a cutoff value of 18 (1.00) and for the VEBW (0.89). At D3 (cutoff of 30), the LFpen showed the highest values of sensitivity and specificity. For root caries, the LFpen and VEBW showed the highest values of specificity (0.54), sensitivity (0.81) and accuracy (0.69). The Spearman rank correlation coefficients for crown/root caries with histology were 0.54/0.37 (LFpen), 0.29/0.10 (BW), 0.29/0.18 (VE) and 0.23/0.37 (VEBW). For the LFpen, the ICC varied from 0.80 (interexaminer) to 0.97 (intraexaminer B); the kappa value was 0.19 for BW and 0.35 for VE (interexaminer). Intraexaminer kappa values for BW were 0.25 (A) and 0.29 (B), and those for VE were 0.31 (A) and 0.32 (B). The LFpen device exhibited a performance comparable to that of conventional methods but with higher interexaminer reproducibility. Therefore, the LFpen should be considered an auxiliary method for the detection of secondary approximal caries associated with composite restorations.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

OBJECTIVES: The aim of this randomised clinical trial was to investigate if a laser fluorescence device is able to discriminate between sound and carious approximal sites and between enamel and dentinal lesions, as well as to find appropriate cut-off values. METHODS: One hundred and seventeen sound or uncavitated carious sites in permanent molars were visually and radiographically examined, then either opened or not, after which their laser fluorescence was measured. Forty-three lesions were opened, the caries removed and the clinically identified caries depths were registered in addition to the radiographical scoring. Seventy-four sites were radiographically deemed sound or had enamel caries and were not opened. Here, the radiographical scorings were registered. RESULTS: Taking the radiographic scoring as gold standard for all investigated approximal sites, sound sites (D(0), n=40) showed significantly lower laser fluorescence measurements than carious sites (D(1-4), n=77) (Mann-Whitney test, P<0.025) suggesting a cut-off at 7 (sensitivity=0.68, specificity=0.7). Comparing measurements of D(0-2) (n=74) and D(3,4) (n=43), the results were also different by a statistically significant amount (P<0.025) and the cut-off calculated to be 16 (sensitivity=0.6, specificity=0.84). A fair positive correlation between laser fluorescence values and radiographical scoring was found (rho=+0.47, P<0.01). Analysing the 43 opened lesions with their clinically found lesion depths as gold standard, there was a fair positive correlation to the laser fluorescence values (rho=+0.34, P=0.03) and a moderately strong correlation to the radiographic scoring (rho=+0.67, P<0.01). CONCLUSION: The device may be an adjunct tool in the approximal detection of caries along with established procedures.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

The aim of this in vitro study was to compare the performance of two laser fluorescence devices (LF, LFpen), conventional visual criteria (VE), ICDAS and radiographic examination on occlusal surfaces of primary teeth. Thirty-seven primary human molars were selected from a pool of extracted teeth, which were stored frozen at -20°C until use. Teeth were assessed twice by two experienced examiners using laser fluorescence devices (LF and LFpen), conventional visual criteria, ICDAS and bitewing radiographs, with a 2-week interval between measurements. After measurement, the teeth were histologically prepared and assessed for caries extension. The highest sensitivity was observed for ICDAS at D(1) and D(3) thresholds, with no statistically significant difference when compared to the LF devices, except at the D(3) threshold. Bitewing radiographs presented the lowest values of sensitivity. Specificity at D(1) was higher for LFpen (0.90) and for VE at D(3) (0.94). When VE was combined with LFpen the post-test probabilities were the highest (94.0% and 89.2% at D(1) and D(3) thresholds, respectively). High values were observed for the combination of ICDAS and LFpen (92.0% and 80.0%, respectively). LF and LFpen showed the highest values of ICC for interexaminer reproducibility. However, regarding ICDAS, BW and VE, intraexaminer reproducibility was not the same for the two examiners. After primary visual inspection using ICDAS or not, the use of LFpen may aid in the detection of occlusal caries in primary teeth. Bitewing radiographs may be indicated only for approximal caries detection.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

The aim of this study was to assess the performance of two light-emitting diode (LED)- and two laser fluorescence-based devices in detecting occlusal caries in vitro. Ninety-seven permanent molars were assessed twice by two examiners using two LED- (Midwest Caries - MID and VistaProof - VP) and two laser fluorescence-based (DIAGNOdent 2095 - LF and DIAGNOdent pen 2190 - LFpen) devices. After measuring, the teeth were histologically prepared and classified according to lesion extension. At D1 the specificities were 0.76 (LF and LFpen), 0.94 (MID), and 0.70 (VP); the sensitivities were 0.70 (LF), 0.62 (LFpen), 0.31 (MID), and 0.75 (VP). At D(3) threshold the specificities were 0.88 (LF), 0.87 (LFpen), 0.90 (MID), and 0.70 (VP); the sensitivities were 0.63 (LF and LFpen), 0.70 (MID), and 0.96 (VP). Spearman's rank correlations with histology were 0.56 (LF), 0.51 (LFpen), 0.55 (MID), and 0.58 (VP). Inter- and intraexaminer ICC values were high and varied from 0.83 to 0.90. Both LF devices seemed to be useful auxiliary tools to the conventional methods, presenting good reproducibility and better accuracy at D(3) threshold. MID was not able to differentiate sound surfaces from enamel caries and VP still needs improvement on the cut-off limits for its use.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

The aim of this study was to compare the in situ and in vitro performances of a laser fluorescence (LF) device (DIAGNOdent 2095) with visual inspection for the detection of occlusal caries in permanent teeth. Sixty-four sites were selected, and visual inspection and LF assessments were carried out, in vitro, three times by two independent examiners, with a 1-week interval between evaluations. Afterwards, the occlusal surfaces were mounted on the palatal portion of removable acrylic orthodontic appliances and placed in six volunteers. Assessments were repeated and validated by histological analysis of the tooth sections under a stereomicroscope. For both examiners, the highest intra-examiner values were observed for the visual inspection when in vitro and in situ evaluations were compared. The inter-examiner reproducibility varied from 0.61 to 0.64, except for the in vitro assessment using LF, which presented a lower value (0.43). The methods showed high specificity at the D(1) threshold (considering enamel and dentin caries as disease). In vitro evaluations showed the highest values of sensitivity for both methods when compared to the in situ evaluations at D(1) and D(2) (considering only dentinal caries as the disease) thresholds. For both methods, the results of sensitivity (at D(1) and D(2)) and accuracy (at D(1)) showed significant differences between in vitro and in situ conditions. However, the sensitivity (at D(1) and D(2)), specificity and accuracy (both at D(1)) of the methods were not significantly different when the same condition was considered. It can be concluded that visual inspection and LF showed better performance in vitro than in situ.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

A device based on infrared laser fluorescence (IRLF) has become available as an adjunct for the diagnosis of dental caries.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

The aim of this study was to assess the influence on the infrared laser fluorescence response of some storage methods commonly used in dental research. Forty extracted permanent teeth, selected from a pool of frozen teeth, were divided into four groups of 10. Three groups were stored at 4 degrees C in 1% chloramine, 10% formalin or 0.02% thymol solution. The fourth group was stored at -20 degrees C (no storage solution added). Fluorescence measurements were performed at 14, 77, 113, 168, 232, 486 and 737 days. After 2 years, significant decreases in fluorescence (p<0.01) for the samples in formalin (-60%), chloramine (-72%) and thymol (-54%) were observed. The frozen teeth showed a slight but non-significant increase in fluorescence of 5% (p>0.01). Storing solutions have a significant influence on the fluorescence yield. Samples used for in vitro purposes stored frozen do not significantly change their fluorescence response. Thus, cut-off values obtained under the latter conditions could be extrapolated to the in vivo situation.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

The laser device DIAGNOdent developed for the detection of occlusal caries has limited value on approximal surfaces. The aim of this study was to develop and to test a new laser fluorescence (LF) device for the detection of approximal caries. Light with a wavelength of 655 nm was transported to the approximal surface using two different sapphire fibre tips. Seventy-five teeth were selected from a pool of extracted permanent human molars, frozen at -20 degrees C until use. Before being measured, they were defrosted, cleaned and calculus was removed with a scaler. The molars were set in blocks simulating the contact area of adults. Bitewing radiographs were obtained using Kodak Insight films. After two independent assessments with the new LF device, the teeth were histologically prepared, and assessed for caries extension. Using the laser, specificity values for D1 threshold (outer half of enamel), D2 threshold (inner half of enamel), D3 threshold (dentine) ranged between 0.81 and 0.93, sensitivity between 0.84 and 0.92 with no difference between the two tips. Bitewing radiography showed an inferior performance compared to LF (p<0.05). Intraex aminer reproducibility was high (kappa>.74). The new LF system might be a useful additional tool in detecting approximal caries. Because of its good reproducibility, it could be used to monitor caries regression or progression on approximal surfaces.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

The new device DIAGNOdent pen based on red laser light induced fluorescence was introduced for the detection of approximal and occlusal caries. The aim of this study was to test its performance on occlusal surfaces. The new device comes with two different sapphire fibre tips: a cylindrical tip and a conical tip. The two new sapphire fibre tips were used and compared with the tip currently available with DIAGNOdent (DD). METHODS: The teeth were selected from a pool of extracted permanent human molars, which were stored frozen at -20 degrees C, until use. Prior to being measured the teeth were defrosted and cleaned. One hundred and nineteen teeth were selected and measured with the old tip and with the two new tips of the new device by two independent assessments. The teeth were histologically prepared and assessed for caries extension. RESULTS: Specificity values for D(1), D(2) and D(3) ranged between 0.69 and 0.89, sensitivity between 0.78 and 0.96. There were no statistically significant differences obtained between the two tips of the new and the one tip of the old device (p>0.05). Intra-examiner reliability with kappa values of >0.83 was high. CONCLUSIONS: In this study, the new laser fluorescence device performed on occlusal surfaces as well as the available device.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

OBJECTIVES: To determine the clinical performance of a laser fluorescence device (DIAGNOdent pen, KaVo) to discriminate between different occlusal caries depths (D(0)-D(1-4); D(0-2)-D(3,4)) in permanent molars. METHODS: In this prospective, randomized two-centre-study 120 sound/uncavitated carious sites in 120 patients were measured after visual and radiographic caries assessment. In cases of operative intervention (n=86), the lesion depths after caries removal were recorded (reference). In cases of preventive intervention (n=34), the sites were reassessed visually/radiographically after 12 months to verify the status assessed before (reference). The discrimination performance was determined statistically (Mann-Whitney test, Spearman's rho coefficient, and areas under the receiver operating characteristic curves (AUCs)). Sensitivities (SE) and specificities (SP) were plotted as a function of the measured values and cut-off values for the mentioned thresholds suggested. RESULTS: Sound sites (n=13) had significantly minor fluorescence values than carious sites (n=107) (P<0.0001) as had sites with no/enamel caries (n=63) compared to dentinal caries (n=57). The AUCs for the same discriminations were 0.92 and 0.78 (P<0.001). For the D(0)-D(1-4) threshold, a cut-off at a value of 12 (SE: 0.88, SP: 0.85) and for the D(0-2)-D(3,4) threshold at 25 (SE: 0.67, SP: 0.79) can be suggested. A moderate positive correlation between the measurements and the caries depths was calculated (rho=+0.57, P=0.01). CONCLUSION: Within this study, the device's discrimination performance for different caries depths was moderate to very good and it may be recommended as adjunct tool in the diagnosis of occlusal caries.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

BACKGROUND: The aim of this study was to evaluate the influence of zero-value subtraction on the performance of two laser fluorescence (LF) devices developed to detect occlusal caries. METHODS: The authors selected 119 permanent molars. Two examiners assessed three areas (cuspal, middle and cervical) of both mesial and distal portions of the buccal surface and one occlusal site using an LF device and an LF pen. For each tooth, the authors subtracted the value measured in the cuspal, middle and cervical areas in the buccal surface from the value measured in the respective occlusal site. RESULTS: The authors observed differences among the readings for both devices in the cuspal, middle and cervical areas in the buccal surface as well as differences for both devices with and without the zero-value subtraction in the occlusal surface. When the authors did not perform the zero-value subtraction, they found statistically significant differences for sensitivity and accuracy for the LF device. When this was done with the LF pen, specificity increased and sensitivity decreased significantly. CONCLUSIONS: For the LF device, the zero-value subtraction decreased the sensitivity. For this reason, the authors concluded that clinicians can obtain measures with the LF device effectively without using zero-value subtraction. For the LF pen, however, the absence of the zero-value subtraction changed both the sensitivity and specificity, and so the authors concluded that clinicians should not eliminate this step from the procedure. CLINICAL IMPLICATIONS: When using the LF device, clinicians might not need to perform the zero-value subtraction; however, for the LF pen, clinicians should do so.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

The aim of this study was to assess the influence of the zero value subtraction on the performance of laser fluorescence (LFpen) for approximal caries detection. Three areas (cuspal, middle and cervical) of both mesial and distal buccal surfaces of 78 permanent molars were assessed using both wedge-shaped (WDG) and tapered wedge-shaped (TWDG) tips. With the addition of the average, one cut-off value for each area was obtained and the performance was assessed. The areas under the receiver operating characteristics (ROC) curve, specificity, sensitivity and accuracy with and without the zero value subtraction were calculated. The McNemar test revealed a statistically significant difference for specificity at thresholds D(1), D(2) and D(3) (WDG) and D(1) and D(2) (TWDG) when the zero value subtraction was not performed. Influence of the zero value subtraction on the LFpen performance was observed for approximal caries detection. However, when modified cut-off values were used, the zero value subtraction could be eliminated.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

OBJECTIVE: The aim of this study was to determine the influence of polyvinyl chloride (PVC) wrapping on the performance of two laser fluorescence devices (LF and LFpen) by assessing tooth occlusal surfaces. BACKGROUND DATA: Protection of their tips may influence LF measurements. To date there are no studies evaluating the influence of this protection on the performance of the LFpen on permanent teeth, or comparing it to the original LF device. MATERIALS AND METHODS: One hundred nineteen permanent molars were assessed by two experienced dentists using the LF and the LFpen devices, both with and without PVC wrapping. The teeth were histologically prepared and assessed for caries extension. RESULTS: The LF values with and without PVC wrapping were significantly different. For both LF devices, the sensitivity and accuracy were lower when the PVC wrapping was used. The specificity was statistically significantly higher for the LFpen with PVC. No difference was found between the areas under the ROC curves with and without PVC wrapping. The ICC showed excellent interexaminer agreement. The Bland and Altman method showed a range between the upper and the lower limits of agreement of 63.4 and 57.8 units for the LF device, and 49.4 and 74.2 for the LFpen device, with and without PVC wrapping, respectively. CONCLUSIONS: We found an influence of the PVC wrapping on the performance of the LF and LFpen devices. However, since its influence on detection of occlusal caries lesions is considered for, the use of one PVC layer is suggested to avoid cross-contamination in clinical practice.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

The aim of this study was to compare the performance of the DIAGNOdent 2095 with visual examination for occlusal caries detection in permanent and primary molars. The sample comprised 148 permanent human molars and 179 primary human molars. The samples were measured and visually examined three times by two examiners. After measurement, the teeth were histologically prepared and assessed for caries extension. Sensitivity, specificity, accuracy and area under the receiver operating characteristics (ROC) curve were calculated. Intra-class correlation (ICC), unweighted kappa and the Bland and Altman method were used to assess inter- and intra-examiner reproducibility. DIAGNOdent showed higher specificity and lower sensitivity than did visual examination. The ICC values indicated an excellent agreement between the examinations. Kappa values varied from good to excellent for DIAGNOdent but from poor to good for visual examination. In conclusion, the DIAGNOdent may be a useful adjunct to conventional methods for occlusal caries detection.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

The aim of this in vivo study was to evaluate the performance of laser fluorescence (LF) comparing different cut-off limits for occlusal caries detection. One hundred and thirty first permanent molars were selected. Visual examination and LF assessments were performed independently. The extent of caries was assessed after operative intervention. New cut-off limits were established and compared with those proposed by the manufacturer and by Lussi et al. (Eur J Oral Sci 109:14-19, 2001). Similar sensitivity and higher specificity was found at D(2) (considering as disease only dentin caries) when the LF cut-off limits proposed by Lussi et al. and the new one were compared. At the D(3) threshold (considering as disease only deep dentin caries), no statistically significant difference among the cut-off limits for sensitivity was found. However, the new cut-off limits showed higher specificity. The LF device provided good ability to detect dentin caries lesions. Furthermore, the new cut-off limits and the values proposed by Lussi et al. could be suggested for the in vivo detection of occlusal caries.