80 resultados para DIAGNOdent pen 2190

em BORIS: Bern Open Repository and Information System - Berna - Suiça


Relevância:

100.00% 100.00%

Publicador:

Resumo:

The aim of this study was to assess the performance of two light-emitting diode (LED)- and two laser fluorescence-based devices in detecting occlusal caries in vitro. Ninety-seven permanent molars were assessed twice by two examiners using two LED- (Midwest Caries - MID and VistaProof - VP) and two laser fluorescence-based (DIAGNOdent 2095 - LF and DIAGNOdent pen 2190 - LFpen) devices. After measuring, the teeth were histologically prepared and classified according to lesion extension. At D1 the specificities were 0.76 (LF and LFpen), 0.94 (MID), and 0.70 (VP); the sensitivities were 0.70 (LF), 0.62 (LFpen), 0.31 (MID), and 0.75 (VP). At D(3) threshold the specificities were 0.88 (LF), 0.87 (LFpen), 0.90 (MID), and 0.70 (VP); the sensitivities were 0.63 (LF and LFpen), 0.70 (MID), and 0.96 (VP). Spearman's rank correlations with histology were 0.56 (LF), 0.51 (LFpen), 0.55 (MID), and 0.58 (VP). Inter- and intraexaminer ICC values were high and varied from 0.83 to 0.90. Both LF devices seemed to be useful auxiliary tools to the conventional methods, presenting good reproducibility and better accuracy at D(3) threshold. MID was not able to differentiate sound surfaces from enamel caries and VP still needs improvement on the cut-off limits for its use.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

This study evaluated the performance of the DIAGNOdent pen laser fluorescence device (LFpen) in comparison with visual examination (VE), bitewing radiographs (BW) and visual examination combined with bitewing radiographs (VEBW) in detecting secondary approximal caries associated with composite restorations. In total, 60 approximal surfaces from 43 permanent molars with composite restorations were assessed twice by two examiners using the LFpen, VE, BW and VEBW. After histological preparation and hardness measurements, the sample was assigned to either a crown or root caries group, depending on the location of the lesions as the gold standard. For crown caries at D1, the highest values of specificity and sensitivity were observed for the LFpen at a cutoff value of 18 (1.00) and for the VEBW (0.89). At D3 (cutoff of 30), the LFpen showed the highest values of sensitivity and specificity. For root caries, the LFpen and VEBW showed the highest values of specificity (0.54), sensitivity (0.81) and accuracy (0.69). The Spearman rank correlation coefficients for crown/root caries with histology were 0.54/0.37 (LFpen), 0.29/0.10 (BW), 0.29/0.18 (VE) and 0.23/0.37 (VEBW). For the LFpen, the ICC varied from 0.80 (interexaminer) to 0.97 (intraexaminer B); the kappa value was 0.19 for BW and 0.35 for VE (interexaminer). Intraexaminer kappa values for BW were 0.25 (A) and 0.29 (B), and those for VE were 0.31 (A) and 0.32 (B). The LFpen device exhibited a performance comparable to that of conventional methods but with higher interexaminer reproducibility. Therefore, the LFpen should be considered an auxiliary method for the detection of secondary approximal caries associated with composite restorations.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

To compare the performance of LFpen (DIAGNOdent pen) with two different wedge-shaped tips to conventional bitewing radiography (BW) for detecting proximal secondary caries at the cervical margin of amalgam restorations in vitro.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

The new device DIAGNOdent pen based on red laser light induced fluorescence was introduced for the detection of approximal and occlusal caries. The aim of this study was to test its performance on occlusal surfaces. The new device comes with two different sapphire fibre tips: a cylindrical tip and a conical tip. The two new sapphire fibre tips were used and compared with the tip currently available with DIAGNOdent (DD). METHODS: The teeth were selected from a pool of extracted permanent human molars, which were stored frozen at -20 degrees C, until use. Prior to being measured the teeth were defrosted and cleaned. One hundred and nineteen teeth were selected and measured with the old tip and with the two new tips of the new device by two independent assessments. The teeth were histologically prepared and assessed for caries extension. RESULTS: Specificity values for D(1), D(2) and D(3) ranged between 0.69 and 0.89, sensitivity between 0.78 and 0.96. There were no statistically significant differences obtained between the two tips of the new and the one tip of the old device (p>0.05). Intra-examiner reliability with kappa values of >0.83 was high. CONCLUSIONS: In this study, the new laser fluorescence device performed on occlusal surfaces as well as the available device.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

OBJECTIVES: To determine the clinical performance of a laser fluorescence device (DIAGNOdent pen, KaVo) to discriminate between different occlusal caries depths (D(0)-D(1-4); D(0-2)-D(3,4)) in permanent molars. METHODS: In this prospective, randomized two-centre-study 120 sound/uncavitated carious sites in 120 patients were measured after visual and radiographic caries assessment. In cases of operative intervention (n=86), the lesion depths after caries removal were recorded (reference). In cases of preventive intervention (n=34), the sites were reassessed visually/radiographically after 12 months to verify the status assessed before (reference). The discrimination performance was determined statistically (Mann-Whitney test, Spearman's rho coefficient, and areas under the receiver operating characteristic curves (AUCs)). Sensitivities (SE) and specificities (SP) were plotted as a function of the measured values and cut-off values for the mentioned thresholds suggested. RESULTS: Sound sites (n=13) had significantly minor fluorescence values than carious sites (n=107) (P<0.0001) as had sites with no/enamel caries (n=63) compared to dentinal caries (n=57). The AUCs for the same discriminations were 0.92 and 0.78 (P<0.001). For the D(0)-D(1-4) threshold, a cut-off at a value of 12 (SE: 0.88, SP: 0.85) and for the D(0-2)-D(3,4) threshold at 25 (SE: 0.67, SP: 0.79) can be suggested. A moderate positive correlation between the measurements and the caries depths was calculated (rho=+0.57, P=0.01). CONCLUSION: Within this study, the device's discrimination performance for different caries depths was moderate to very good and it may be recommended as adjunct tool in the diagnosis of occlusal caries.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

This study aimed to evaluate the effectiveness of fluorescence-based methods (DIAGNOdent, LF; DIAGNOdent pen, LFpen, and VistaProof fluorescence camera, FC) in detecting demineralization and remineralization on smooth surfaces in situ. Ten volunteers wore acrylic palatal appliances, each containing 6 enamel blocks that were demineralized for 14 days by exposure to a 20% sucrose solution and 3 of them were remineralized for 7 days with fluoride dentifrice. Sixty enamel blocks were evaluated at baseline, after demineralization and 30 blocks after remineralization by two examiners using LF, LFpen and FC. They were submitted to surface microhardness (SMH) and cross-sectional microhardness analysis. The integrated loss of surface hardness (ΔKHN) was calculated. The intraclass correlation coefficient for interexaminer reproducibility ranged from 0.21 (FC) to 0.86 (LFpen). SMH, LF and LFpen values presented significant differences among the three phases. However, FC fluorescence values showed no significant differences between the demineralization and remineralization phases. Fluorescence values for baseline, demineralized and remineralized enamel were, respectively, 5.4 ± 1.0, 9.2 ± 2.2 and 7.0 ± 1.5 for LF; 10.5 ± 2.0, 15.0 ± 3.2 and 12.5 ± 2.9 for LFpen, and 1.0 ± 0.0, 1.0 ± 0.1 and 1.0 ± 0.1 for FC. SMH and ΔKHN showed significant differences between demineralization and remineralization phases. There was a negative and significant correlation between SMH and LF and LFpen in the remineralization phase. In conclusion, LF and LFpen devices were effective in detecting demineralization and remineralization on smooth surfaces provoked in situ.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

This in vivo study aimed to evaluate the influence of contact points on the approximal caries detection in primary molars, by comparing the performance of the DIAGNOdent pen and visual-tactile examination after tooth separation to bitewing radiography (BW). A total of 112 children were examined and 33 children were selected. In three periods (a, b, and c), 209 approximal surfaces were examined: (a) examiner 1 performed visual-tactile examination using the Nyvad criteria (EX1); examiner 2 used DIAGNOdent pen (LF1) and took BW; (b) 1 week later, after tooth separation, examiner 1 performed the second visual-tactile examination (EX2) and examiner 2 used DIAGNOdent again (LF2); (c) after tooth exfoliation, surfaces were directly examined using DIAGNOdent (LF3). Teeth were examined by computed microtomography as a reference standard. Analyses were based on diagnostic thresholds: D1: D 0 = health, D 1 –D 4 = disease; D2: D 0 , D 1 = health, D 2 –D 4 = disease; D3: D 0 –D 2 = health, D 3 , D 4 = disease. At D1, the highest sensitivity/specificity were observed for EX1 (1.00)/LF3 (0.68), respectively. At D2, the highest sensitivity/ specificity were observed for LF3 (0.69)/BW (1.00), respectively. At D3, the highest sensitivity/specificity were observed for LF3 (0.78)/EX1, EX2 and BW (1.00). EX1 showed higher accuracy values than LF1, and EX2 showed similar values to LF2. We concluded that the visual-tactile examination showed better results in detecting sound surfaces and approximal caries lesions without tooth separation. However, the effectiveness of approximal caries lesion detection of both methods was increased by the absence of contact points. Therefore, regardless of the method of detection, orthodontic separating elastics should be used as a complementary tool for the diagnosis of approximal noncavitated lesions in primary molars.

Relevância:

80.00% 80.00%

Publicador:

Resumo:

The aim of this study was to determine the influence of thickness and aging on the intrinsic fluorescence of sealing materials and their ability to block fluorescence from the underlying surface as assessed using a laser fluorescence device. Cavities of 0.5 mm and 1 mm depth were drilled into acrylic boards which were placed over two surfaces with different fluorescence properties: a low-fluorescence surface, to assess the intrinsic fluorescence of the sealing materials, and a high-fluorescence surface, to assess the fluorescence-blocking ability of the sealing materials. Ten cavities of each depth were filled with different sealing materials: Adper Scotchbond Multi-Purpose, Adper Single Bond 2, FluroShield, Conseal f and UltraSeal XT Plus. Fluorescence was measured with a DIAGNOdent pen at five different time points: empty cavity, after polymerization, and 1 day, 1 week and 1 month after filling. The individual values after polymerization, as well as the area under the curve for the different periods were submitted to ANOVA and the Tukey test (p < 0.05). At 0.5 mm, Scotchbond, FluroShield and UltraSeal showed insignificant changes in intrinsic fluorescence with aging and lower fluorescence after polymerization than Single Bond and Conseal. At 1 mm, Scotchbond and FluroShield showed the lowest intrinsic fluorescence, but only Scotchbond showed no chagnes in fluorescence with aging. At both depths, Scotchbond blocked significantly less fluorescence. All sealing materials blocked more fluorescence when applied to a depth of 1 mm. At 0.5 mm, fissure sealants blocked more fluorescence than adhesives, and did not show significant changes with aging. Scotchbond had the least affect on the fluorescence from the underlying surface and would probably have the least affect on the monitoring of sealed dental caries by laser fluorescence.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

AIM To evaluate the performance of a pen‑type laser fluorescence device (DIAGNOdent 2190; LFpen, KaVo, Germany) and bitewing radiographs (BW) for approximal caries detection in permanent and primary teeth. MATERIALS AND METHODS A total of 246 anterior approximal surfaces (102 permanent and 144 primary) were selected. Contact points were simulated using sound teeth. Two examiners assessed all approximal surfaces using LFpen and BW. The teeth were histologically assessed for the reference standard. Optimal cut‑off limits were calculated for LFpen for primary and permanent teeth. Sensitivity, specificity, accuracy and area under the receiver operating characteristic curve (Az) were calculated for D1 (enamel and dentin lesions) and D3 (dentin lesions) thresholds. The reproducibility was assessed by intraclass correlation coefficient (ICC) and Cohen's weighted kappa values. RESULTS For permanent teeth, the LFpen cut‑off were 0- 27 (sound), 28- 33 (enamel caries) and >33 (dentin caries). For primary teeth, the LFpen cut‑off were 0- 7 (sound), 8- 32 (enamelcaries) and >32 (dentin caries). The LFpen presented higher sensitivity values than BW for primary teeth (0.58 vs. 0.32 at D1 and 0.80 vs. 0.47 at D3) and permanent teeth (0.80 vs. 0.57 at D1 and 0.94 vs. 0.51 at D3). Specificity did not show a significant difference between the methods. Rank correlations with histology were 0.59 and 0.83 (LFpen) and 0.36 and 0.70 (BW) for primary and permanent teeth, respectively, considering all lesions. ICC values for LFpen were 0.71 (inter) and 0.86 (intra) for permanent teeth and 0.94 (inter) and 0.90/0.99 for primary teeth. Kappa values for BW were 0.69 (inter) and 0.68/0.90 (intra) for permanent teeth and 0.64 (inter) and 0.89/0.89 for primary teeth. CONCLUSION LFpen presented better reproducibility for primary and permanent teeth and higher accuracy in detecting caries lesions at D1 threshold than BW for permanent teeth. LFpen should be used as an adjunct method for approximal caries detection.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

Although there has been a significant decrease in caries prevalence in developed countries, the slower progression of dental caries requires methods capable of detecting and quantifying lesions at an early stage. The aim of this study was to evaluate the effectiveness of fluorescence-based methods (DIAGNOdent 2095 laser fluorescence device [LF], DIAGNOdent 2190 pen [LFpen], and VistaProof fluorescence camera [FC]) in monitoring the progression of noncavitated caries-like lesions on smooth surfaces. Caries-like lesions were developed in 60 blocks of bovine enamel using a bacterial model of Streptococcus mutans and Lactobacillus acidophilus . Enamel blocks were evaluated by two independent examiners at baseline (phase I), after the first cariogenic challenge (eight days) (phase II), and after the second cariogenic challenge (a further eight days) (phase III) by two independent examiners using the LF, LFpen, and FC. Blocks were submitted to surface microhardness (SMH) and cross-sectional microhardness analyses. The intraclass correlation coefficient for intra- and interexaminer reproducibility ranged from 0.49 (FC) to 0.94 (LF/LFpen). SMH values decreased and fluorescence values increased significantly among the three phases. Higher values for sensitivity, specificity, and area under the receiver operating characteristic curve were observed for FC (phase II) and LFpen (phase III). A significant correlation was found between fluorescence values and SMH in all phases and integrated loss of surface hardness (ΔKHN) in phase III. In conclusion, fluorescence-based methods were effective in monitoring noncavitated caries-like lesions on smooth surfaces, with moderate correlation with SMH, allowing differentiation between sound and demineralized enamel.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

The aim of this in vitro study was to evaluate the relationship between laser fluorescence values and sealant penetration depth on occlusal fissures. One hundred and sixty-six permanent molars were selected and divided into four groups, which were each treated using a different sealant (two clear and two opaque). The teeth were independently measured twice by two experienced dentists using two laser fluorescence devices-DIAGNOdent (LF and LFpen)-before and after sealing, and then thermoclycled. After measuring, the teeth were histologically prepared and assessed for caries extension. Digital photographs of the cut sealed sites were assessed, and the sealant penetration depth was measured. All 166 sites were measured by one of the examiners taking as limits the outer and inner surface of the sealant into the fissure. For each device (LF and LFpen) and each group, the difference between the values at baseline and after sealing was plotted against the sealant penetration depth and scatter plots were provided. It could be observed that most of the points were concentrated around the zero line, for both LF and LFpen in the four groups. In conclusion, there is no relation between changes in DIAGNOdent values and increasing of depth sealant penetration within the occlusal fissures.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

This study aimed to evaluate the influence of professional prophylactic methods on the DIAGNOdent 2095, DIAGNOdent 2190 and VistaProof performance in detecting occlusal caries. Assessments were performed in 110 permanent teeth at baseline and after bicarbonate jet or prophylactic paste and rinsing. Performance in terms of sensitivity improved after rinsing of the occlusal surfaces when the prophylactic paste was used. However, the sodium bicarbonate jet did not significantly influence the performance of the fluorescence-based methods. It can be concluded that different professional prophylactic methods can significantly influence the performance of fluorescence-based methods for occlusal caries detection.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Regrouping female rabbits in group-housing systems is common management practice in rabbit breeding, which may, however, induce agonistic interactions resulting in social stress and severe injuries. Here we compared two methods of regrouping female rabbits with respect to their effects on behaviour, stress and injuries. Thus, we introduced two unfamiliar rabbits into a group of rabbits either in the group's familiar pen (HOME) or in a novel disinfected pen (NOVEL), and assessed the effects of these treatments on general activity, number and duration of agonistic interactions, number and severity of injuries and body temperature as a measure of stress. General activities were not affected by the method of regrouping. Also, treatment had no effect on the number and duration of agonistic interactions. However, the numbers of injuries (P=0.030) as well as body temperature on the first clay after regrouping (p=0.0036) were increased in rabbits regrouped in a novel clean pen. These findings question the recommendation to introduce unfamiliar does into established groups in a neutral environment and indicate that regrouping in the group's home pen may decrease the risk of severe injuries and social stress. (C) 2011 Elsevier B.V. All rights reserved.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

This study aimed to assess the performance of International Caries Detection and Assessment System (ICDAS), radiographic examination, and fluorescence-based methods for detecting occlusal caries in primary teeth. One occlusal site on each of 79 primary molars was assessed twice by two examiners using ICDAS, bitewing radiography (BW), DIAGNOdent 2095 (LF), DIAGNOdent 2190 (LFpen), and VistaProof fluorescence camera (FC). The teeth were histologically prepared and assessed for caries extent. Optimal cutoff limits were calculated for LF, LFpen, and FC. At the D (1) threshold (enamel and dentin lesions), ICDAS and FC presented higher sensitivity values (0.75 and 0.73, respectively), while BW showed higher specificity (1.00). At the D (2) threshold (inner enamel and dentin lesions), ICDAS presented higher sensitivity (0.83) and statistically significantly lower specificity (0.70). At the D(3) threshold (dentin lesions), LFpen and FC showed higher sensitivity (1.00 and 0.91, respectively), while higher specificity was presented by FC (0.95), ICDAS (0.94), BW (0.94), and LF (0.92). The area under the receiver operating characteristic (ROC) curve (Az) varied from 0.780 (BW) to 0.941 (LF). Spearman correlation coefficients with histology were 0.72 (ICDAS), 0.64 (BW), 0.71 (LF), 0.65 (LFpen), and 0.74 (FC). Inter- and intraexaminer intraclass correlation values varied from 0.772 to 0.963 and unweighted kappa values ranged from 0.462 to 0.750. In conclusion, ICDAS and FC exhibited better accuracy in detecting enamel and dentin caries lesions, whereas ICDAS, LF, LFpen, and FC were more appropriate for detecting dentin lesions on occlusal surfaces in primary teeth, with no statistically significant difference among them. All methods presented good to excellent reproducibility.