17 resultados para SNR maximisation
Resumo:
Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES)
Resumo:
Since discovery, computed tomography is a widely used diagnostic modality. However, this modality imparts relatively high doses to the patients and with fast technological advancement, it is necessary optimize the doses used and ensure the quality of the images through a quality assurance program. This work intended to compare Computed Tomography Dose Index (CTDIW) and effective dose with components of image quality: the Contrast-to-Noise Ratio (CNR) and Signal-to-Noise Ratio (SNR), and a quality factor (Q) deduced by the Rose model for two groups of computed tomography units. The first group was composed by equipments with 10 up to 16 slices and the second one presented 40 up to 320 slices detectors, for the protocols of head and abdomen. It was realized a comparison between different selectable parameters in the protocol of a Philips Brilliance 16, too. The results of the first group to CTDIW, effective dose, SNR and CNR showed variations of 28%, 33%, 37% and 32% respectively for head protocol, and 21%, 27%, 43% and 37% respectively for abdomen protocol. The results of the group 2 to CTDIW, effective dose, SNR and CNR showed variations of 15%, 26%, 36% and 34% respectively for the head protocol, and 8%, 13%, 50% and 60% respectively for abdomen protocol. The comparison between both groups demonstrated similar levels of doses impartation to patients though having so many different configurations, if the uncertainties associated with this measurement were considered. The results of the comparison between different parameters in the Philips Brilliance 16 scanner were in agreement with expected