1 resultado para perimeter

em Université de Lausanne, Switzerland


Relevância:

10.00% 10.00%

Publicador:

Resumo:

PURPOSE: To investigate the effect of intraocular straylight (IOS) induced by white opacity filters (WOF) on threshold measurements for stimuli employed in three perimeters: standard automated perimetry (SAP), pulsar perimetry (PP) and the Moorfields motion displacement test (MDT).¦METHODS: Four healthy young (24-28 years old) observers were tested six times with each perimeter, each time with one of five different WOFs and once without, inducing various levels of IOS (from 10% to 200%). An increase in IOS was measured with a straylight meter. The change in sensitivity from baseline was normalized, allowing comparison of standardized (z) scores (change divided by the SD of normative values) for each instrument.¦RESULTS: SAP and PP thresholds were significantly affected (P < 0.001) by moderate to large increases in IOS (50%-200%). The drop in motion displacement (MD) from baseline with WOF 5, was approximately 5 dB, in both SAP and PP which represents a clinically significant loss; in contrast the change in MD with MDT was on average 1 minute of arc, which is not likely to indicate a clinically significant loss.¦CONCLUSIONS: The Moorfields MDT is more robust to the effects of additional straylight in comparison with SAP or PP.