37 resultados para Urethral Sphincter

em Université de Lausanne, Switzerland


Relevância:

100.00% 100.00%

Publicador:

Resumo:

OBJECTIVE: To investigate by electromyography (EMG), the presence of complex repetitive discharges (CRDs) and decelerating bursts (DBs) in the striated external urethral sphincter during the menstrual cycle in female volunteers with no urinary symptoms and complete bladder emptying. SUBJECTS AND METHODS: Healthy female volunteers aged 20-40 years, with regular menstrual cycles and no urinary symptoms were recruited. Volunteers completed a menstruation chart, urinary symptom questionnaires, pregnancy test, urine dipstick, urinary free flow and post-void ultrasound bladder scan. Exclusion criteria included current pregnancy, use of hormonal medication or contraception, body mass index of >35 kg/m(2) , incomplete voiding and a history of pelvic surgery. Eligible participants underwent an external urethral sphincter EMG, using a needle electrode in the early follicular phase and the mid-luteal phase of their menstrual cycles. Serum oestradiol and progesterone were measured at each EMG test. RESULTS: In all, 119 women enquired about the research and following screening, 18 were eligible to enter the study phase. Complete results were obtained in 15 women. In all, 30 EMG tests were undertaken in the 15 asymptomatic women. Sphincter EMG was positive for CRDs and DBs at one or both phases of the menstrual cycle in eight (53%) of the women. Three had CRDs and DBs in both early follicular and mid-luteal phases. Five had normal EMG activity in the early follicular phase and CRDs and DBs in the mid-luteal phase. No woman had abnormal EMG activity in the early follicular phase and normal activity in the luteal phase. There was no relationship between EMG activity and age, parity or serum levels of oestradiol and progesterone. CONCLUSIONS: CRDs and DB activity in the external striated urethral sphincter is present in a high proportion of asymptomatic young women. This abnormal EMG activity has been shown for the first time to change during the menstrual cycle in individual women. CRDs and DBs are more commonly found in the luteal phase of the menstrual cycle. The importance of CRDs and DBs in the aetiology of urinary retention in young women remains uncertain. The distribution and or quantity of abnormal EMG activity in the external urethral sphincter may be important. In a woman with urinary retention the finding of CRDs and DBs by needle EMG does not automatically establish Fowler's syndrome as the explanation for the bladder dysfunction. Urethral pressure profilometry may be helpful in establishing a diagnosis. Opiate use and psychological stress should be considered in young women with urinary retention.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

But de l'étude Un enregistrement spécifique EMG du sphincter strié urétral avec décharges répétitives et complexes ainsi que salves de décélération a été décrit comme pathognomonique du syndrome de Fowler, un trouble de la relaxation du sphincter strié urétral chez la femme jeune responsable d'une retention urinaire. Nous avons souhaité étudier la présence de cet enregistrement EMG spécifique chez la femme asymptomatique, ceci à différents moments du cycle menstruel. Matériel et Méthode Nous avons recruté des femmes volontaires saines âgées entre 20 et 40 ans, ayant un cycle hormonal régulier, et ne présentant aucun symptôme urinaire. Les critères d'exclusion étaient la presence d'une dysfonction mictionnelle, d'une infection urinaire, la grossesse, la prise d'une thérapie hormonale ou d'hormone contraceptive, une obésité et des antécédants d'intervention pelvienne. Nous avons procédé à deux enregistrements EMG du sphincter strié urétral des participantes éligibles, utilisant une aiguille concentrique, ceci dans la première phase du cycle (phase folliculaire) et dans la dernière phase du cycle (phase lutéale). Les taux sériques de progestérone et d'oestrogène étaient mesurés à chaque enregistrement. Résultats 15 participantes ont complété l'étude. L' enregistrement EMG du sphincter a été positif avec présence de décharges répétitives et de salves de décélération lors d'une ou des deux phases du cycle menstruel chez 8 participantes (53%). Trois participantes présentaient cet enregistrement spécifique lors des deux phases du cycle et cinq participantes présentaient cet enregistrement spécifique lors de la phase lutéale uniquement. Aucune femme ne présentait cet enregistrement spécifique en début de cycle uniquement. Il n'y avait pas de relation avec l'âge, la parité ou les taux hormonaux. Conclusions L'enregistrement EMG spécifique du sphincter strié urétral, avec décharges répétitives et salves de décélération, se retrouve chez une proportion élevée de femmes asymptomatiques. Il a été montré que ce tracé change lors du cycle menstruel, en étant retrouvé plus fréquemment dans la dernière phase du cycle. L'importance de cet enregistrement EMG dans l'étiologie de la retention urinaire de la femme jeune reste à éclaircir. Nous devons considérer que sa présence ne pose pas automatiquement un diagnostic de syndrome de Fowler chez la femme en rétention urinaire.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

Abstract: The aim of the study was to assess the effects of epidural analgesia on pelvic floor function. Eighty- two primiparous women (group 1, consisting of 41 given an epidural, and group 2 of 41 not given an epidural) were investigated during pregnancy and at 2 and 10 months after delivery by a questionnaire, clinical examination, and assessment of bladder neck behavior, urethral sphincter function and intravaginal/intra-anal pressures. The prevalence of stress urinary incontinence was similar in both groups at 2 months (24% vs. 17%, P = 0.6) and 10 months (22% vs. 7%, P = 0.1), as was the prevalence of decreased sexual vaginal response at 10 months (27% vs. 10%, P= 0.08). Bladder neck behavior, urethral sphincter function and intravaginal and intra-anal pressures showed no significant differences between the two groups. Ten months after spontaneous delivery, there were no significant differences in the prevalence of stress urinary incontinence and decreased sexual vaginal response, or in bladder neck behavior, urethral sphincter function and pelvic floor muscle strength between women who had or had not had epidural analgesia.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

Objectives: The AMS 800TM is the current artificial urinary sphincter (AUS) for incontinence due to intrinsic sphincter deficiency. Despite good clinical results, technical failures inherent to the hydraulic mechanism or urethral ischemic injury contribute to revisions up to 60%. We are developing an electronic AUS, called ARTUS to overcome the rigors of AMS. The objective of this study was to evaluate the technical efficacy and tissue tolerance of the ARTUS system in an animal model.Methods: The ARTUS is composed by three parts: the contractile unit, a series of rings and an integrated microprocessor. The contractile unit is made of Nitinol fibers. The rings are placed around the urethra to control the flow of urine by squeezing the urethra. They work in a sequential alternative mode and are controlled by a microprocessor. In the first phase a three-rings device was used while in the second phase a two-rings ARTUS was used. The device was implanted in 14 sheep divided in two groups of six and eight animals for study purpose. The first group aimed at bladder leak point pressure (BLPP) measurement and validation of the animal model; the second group aimed at verifying mid-term tissue tolerance by explants at twelve weeks. General animal tolerance was also evaluated.Results: The ARTUS system implantation was uneventful. When the system was activated, the BLPP was measured at 1.038±0.044 bar (mean±SD). Urethral tissue analysis did not show significant morphological changes. No infection and no sign of discomfort were noted in animals at 12 weeks.Conclusions: The ARTUS proved to be effective in continence achievement in this study. Histological results support our idea that a sequential alternative mode can avoid urethral atrophy and ischemia. Further technical developments are needed to verify long-term outcome and permit human use.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

WHAT'S KNOWN ON THE SUBJECT? AND WHAT DOES THE STUDY ADD?: The AMS 800 urinary control system is the gold standard for the treatment of urinary incontinence due to sphincter insufficiency. Despite excellent functional outcome and latest technological improvements, the revision rate remains significant. To overcome the shortcomings of the current device, we developed a modern electromechanical artificial urinary sphincter. The results demonstrated that this new sphincter is effective and well tolerated up to 3 months. This preliminary study represents a first step in the clinical application of novel technologies and an alternative compression mechanism to the urethra. OBJECTIVES: To evaluate the effectiveness in continence achievement of a new electromechanical artificial urinary sphincter (emAUS) in an animal model. To assess urethral response and animal general response to short-term and mid-term activation of the emAUS. MATERIALS AND METHODS: The principle of the emAUS is electromechanical induction of alternating compression of successive segments of the urethra by a series of cuffs activated by artificial muscles. Between February 2009 and May 2010 the emAUS was implanted in 17 sheep divided into three groups. The first phase aimed to measure bladder leak point pressure during the activation of the device. The second and third phases aimed to assess tissue response to the presence of the device after 2-9 weeks and after 3 months respectively. Histopathological and immunohistochemistry evaluation of the urethra was performed. RESULTS: Bladder leak point pressure was measured at levels between 1091 ± 30.6 cmH2 O and 1244.1 ± 99 cmH2 O (mean ± standard deviation) depending on the number of cuffs used. At gross examination, the explanted urethra showed no sign of infection, atrophy or stricture. On microscopic examination no significant difference in structure was found between urethral structure surrounded by a cuff and control urethra. In the peripheral tissues, the implanted material elicited a chronic foreign body reaction. Apart from one case, specimens did not show significant presence of lymphocytes, polymorphonuclear leucocytes, necrosis or cell degeneration. Immunohistochemistry confirmed the absence of macrophages in the samples. CONCLUSIONS: This animal study shows that the emAUS can provide continence. This new electronic controlled sequential alternating compression mechanism can avoid damage to urethral vascularity, at least up to 3 months after implantation. After this positive proof of concept, long-term studies are needed before clinical application could be considered.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

Objective: There are only a few established artificial urinary sphincters for treatment of incontinence. We have developed a new device composed by three parts: the actuator, three contractile rings and a control unit. The actuator is made of Nitinol fibers, driven by microprocessor. The fibers are linked to the rings placed around the urethra. They function with alternance in their open and closed position. This concept is called piano concept. With this set-up, the constant compression on the urethra is strongly reduced. Methods: Six male sheep have been used for this study. The sphincter was open each hour for a period of 10 min., to guaranty urination. The bladder was filled with water while one cuff was closed and bladder pressure was monitored. The animals were sacrificed. Two biopsies around two cuffs of each explant and all three cuffs from each explant including urethra were analyzed. Urethra not surrounded by a cuff was taken as control. Results: The pressure exerted by the sphincter around the urethra provided continence. Simulated incontinence occurred at a pressure of 1bar measured on the bladder wall using a pressure probe. The closing force of the cuff was approx. 0·7N. No difference in tissue structure and organization of the urethra with and without artificial sphincter was observed. Conclusions: This device has several advantages compared to other urinary sphincters. It is easy to implant, has no hydraulic nature and reduces ischemic injury of the urethra by the alternance of urethral part compressed. Proof of concept in vivo has been demonstrated. Other studies are planned to determine long-term outcome.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

Introduction and objectives: The AMS 800TM is considered the gold standard for sphincter replacement. However, the one-ring design can erode the urethra and lead to severe complications. A mechanism that could alternatively compress successive segments of the urethra would limit such deleterious outcome. We report 12 weeks animal urethral tissue analysis following implantation of a new modular artificial sphincter. METHODS: The device is composed by three parts: the contractile unit, two rings and an integrated microprocessor. The contractile unit is made of Nitinol fibers. The rings are placed around the urethra to control the flow of urine by squeezing the urethra. They work in a sequential alternative mode and are controlled by a microprocessor connected to an external computer. The computer can reveal specific failure of device components. The device was impkanted in eight male sheep. The rings were positioned around the urethra and the control unit was placed 5cm away. The device was working twenty hours per day; it was open 10min. per hour to allow urination. The animals were sacrificed after 12 weeks. The urethra and the tissues surrounding the control unit were macroscopically and microscopically examined. Two transversal sections crossing the sphincter and two transversal sections crossing the urethra alone were obtained and stained with modified Paragon after resin embedding. Urethra was also embedded in paraffin. The first section was stained with safranin-hematoxylin-eosin, the second section was stained with Masson's Trichrome and the remaining eight sections were available for immunolabelling of the macrophages.Results: The chronic study went uneventful. No clinical infection or pain was observed. The computer registered no specific failure in ring function, Nitinol wires and tube connectors. At explantation, except for a slight grade of lymphocytes in two out of eight specimens, no urethral stricture or atrophy could be observed. Immunohistochemistry confirmed the absence of macrophages. Tissue structure and organization of the urethra with and without artificial sphincter were similar. No migration of the device was observed.Conclusions: The study clearly showed no tissue damage or inflammation of the urethra. Electronic design, preservation of urethral vascularisation and adjustability after implantation are the key ideas to improve the actual AUS. Further studies will be carried out to evaluate this potential.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

Objectives: The AMS 800 is the current artifi cial urinary sphincter (AUS) forincontinence due to intrinsic sphincter defi ciency. Despite good clinical results,technical failures inherent to the hydraulic mechanism or urethral ischemicinjury contribute to revisions up to 60%. We are developing an electronic AUS,called ARTUS to overcome the rigors of AMS. The objective of this study wasto evaluate the technical effi cacy and tissue tolerance of the ARTUS systemin an animal model.Methods: The ARTUS is composed by three parts: thecontractile unit, a series of rings and an integrated microprocessor. The contractileunit is made of Nitinol fi bers. The rings are placed around the urethrato control the fl ow of urine by squeezing the urethra. They work in a sequentialalternative mode and are controlled by a microprocessor. In the fi rst phase athree-rings device was used while in the second phase a two-rings ARTUS wasused. The device was implanted in 14 sheep divided in two groups of six andeight animals for study purpose. The fi rst group aimed at bladder leak pointpressure (BLPP) measurement and validation of the animal model; the secondgroup aimed at verifying midterm tissue tolerance by explants at twelve weeks.General animal tolerance was also evaluated.Results: The ARTUS systemimplantation was uneventful. When the system was activated, the BLPP wasmeasured at 1.038 ± 0.044 bar (mean ± SD). Urethral tissue analysis did notshow signifi cant morphological changes. No infection and no sign of discomfortwere noted in animals at 12 weeks.Conclusions: The ARTUS proved to beeffective in continence achievement in this study. Histological results supportour idea that a sequential alternative mode can avoid urethral atrophy andischemia. Further technical developments are needed to verify long-termoutcome and permit human use.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

INTRODUCTION AND HYPOTHESIS: This study aims to estimate fecal, urinary incontinence, and sexual function 6 years after an obstetrical anal sphincter tear. METHODS: Among 13,213 women who had a vaginal delivery of a cephalic singleton at term, 196 women sustained an anal sphincter tear. They were matched to 588 controls. Validated questionnaires grading fecal and urinary incontinence, and sexual dysfunction were completed by the participants. RESULTS: Severe fecal incontinence was more frequently reported by women who had sustained an anal sphincter tear compared to the controls. Women with an anal sphincter tear had no increased risk of urinary incontinence, but reported significantly more pain, difficulty with vaginal lubrication, and difficulty achieving orgasm compared to the controls. A fetal occiput posterior position during childbirth was an independent risk factor for both severe urinary incontinence and severe sexual dysfunction. CONCLUSIONS: Fecal incontinence is strongly associated with an anal sphincter tear. A fetal occiput posterior position represents a risk factor for urinary incontinence and sexual dysfunction.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Congenital malformations or injuries of the urethra can be treated using existing autologous tissue, but these procedures are sometimes associated with severe complications. Therefore, tissue engineering may be advantageous for generating urethral grafts. We evaluated engineered high-density collagen gel tubes as urethral grafts in 16 male New Zealand white rabbits. The constructs were either acellular or seeded with autologous smooth muscle cells, isolated from an open bladder biopsy. After the formation of a urethral defect by excision, the tissue-engineered grafts were interposed between the remaining urethral ends. No catheter was placed postoperatively. The animals were evaluated at 1 or 3 months by contrast urethrography and histological examination. Comparing the graft caliber to the control urethra at 3 months, a larger caliber was found in the cell-seeded grafts (96.6% of the normal caliber) than in the acellular grafts (42.3%). Histology of acellular and cell-seeded grafts did not show any sign of inflammation, and spontaneous regrowth of urothelium could be demonstrated in all grafts. Urethral fistulae, sometimes associated with stenosis, were observed, which might be prevented by urethral catheter application. High-density collagen gel tubes may be clinically useful as an effective treatment of congenital and acquired urethral pathologies.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

INTRODUCTION: To report acute and late toxicities in patients with intermediate- and high-risk prostate cancer treated with combined high-dose-rate brachytherapy (HDR-B) and intensity-modulated radiation therapy (IMRT). MATERIALS AND METHODS: From March 2003 to September 2005, 64 men were treated with a single implant HDR-B with 21 Gy given in three fractions, followed by 50 Gy IMRT along with organ tracking. Median age was 66.1 years, and risk of recurrence was intermediate in 47% of the patients or high in 53% of the patients. Androgen deprivation therapy was received by 69% of the patients. Toxicity was scored according to the CTCAE version 3.0. Median follow-up was 3.1 years. RESULTS: Acute grade 3 genitourinary (GU) toxicity was observed in 7.8% of the patients, and late grades 3 and 4 GU toxicity was observed in 10.9% and 1.6% of the patients. Acute grade 3 gastrointestinal (GI) toxicity was experienced by 1.6% of the patients, and late grade 3 GI toxicity was absent. The urethral V(120) (urethral volume receiving > or =120% of the prescribed HDR-B dose) was associated with acute (P=.047) and late > or = grade 2 GU toxicities (P=.049). CONCLUSIONS: Late grades 3 and 4GU toxicity occurred in 10.9% and 1.6% of the patients after HDR-B followed by IMRT in association with the irradiated urethral volume. The impact of V(120) on GU toxicity should be validated in further studies.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Fistulas are common complications in Crohn's disease, as their cumulative prevalence reaches up to two thirds of the patients in the long term. Fistulas worsen the overall patient prognosis, with permanent sphincter and perineal tissue destruction as well as professional and personal disabilities. The importance of healing these fistulas has been less well appreciated than mucosal healing for luminal disease. Management should not be left to any specialty alone, but requires an optimal combination of surgery, infection control, and immunosuppression. Outcome of therapy beyond fistula drainage is unclear and the means of assessing healing over a long time period is poorly characterized. Recent studies suggested that a substantial proportion of patients can achieve fistula healing with surgical and medical therapies. However, studies that measure the benefit of integrated approaches, of early intervention and of precise healing assessment are still missing. Such information is particularly needed in this subset of sick patients that undergo substantial physical and emotional distress because of pain, discharge, incontinence, perineal and genital disfigurement. The advent of adequate pelvic imaging, improved surgical outcomes, and potent biological therapies make it timely to develop best-management strategies and appropriateness of care criteria.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

AIMS OF THE STUDY: Analysis of indications and results of paediatric renal transplantation in a single centre, before and after the introduction of cyclosporine A (CSA). METHODS: Historical retrospective study. RESULTS: 19 transplantations were performed in 14 patients (5 second grafts) between 1971 and 1987 (group I). 13 patients were transplanted between 1988 and 1998 (no second transplant) (group II). In group II, all the patients had immunosuppression with CSA, but none in group I. Group II, with CSA, showed better renal survival than patients without CSA. In group I, obstructive uropathies (posterior urethral valves, pyelo-ureteral junction stenosis, vesico-ureteral reflux) represent a common cause (35%) of terminal chronic renal failure (TCRF), whereas in group II they represent only 15% of the causes and chronic glomerulonephritis is the most common cause (69%) of TCRF. Acute and chronic graft rejections were the cause of 9 and 1 graft losses in group I and II respectively. Living related donors account for 14% of all renal transplantations in group I and 46% in group II. CONCLUSIONS: The incidence of paediatric patients referred to Lausanne for TCRF is stable. We have observed a constant and steady decrease in obstructive uropathies leading to TCRF and renal transplantations, whereas glomerulonephritis are increasingly frequent. Graft survival has much improved since the introduction of cyclosporine A, without an increase in morbidity. In carefully selected cases, intrafamilial renal transplantation provides good results and helps to shorten the time spent on dialysis.