1 resultado para Materials Science, Multidisciplinary

em Université de Lausanne, Switzerland


Relevância:

100.00% 100.00%

Publicador:

Resumo:

The application of two approaches for high-throughput, high-resolution X-ray phase contrast tomographic imaging being used at the tomographic microscopy and coherent radiology experiments (TOMCAT) beamline of the SLS is discussed and illustrated. Differential phase contrast (DPC) imaging, using a grating interferometer and a phase-stepping technique, is integrated into the beamline environment at TOMCAT in terms of the fast acquisition and reconstruction of data and the availability to scan samples within an aqueous environment. A second phase contrast method is a modified transfer of intensity approach that can yield the 3D distribution of the decrement of the refractive index of a weakly absorbing object from a single tomographic dataset. The two methods are complementary to one another: the DPC method is characterised by a higher sensitivity and by moderate resolution with larger samples; the modified transfer of intensity approach is particularly suited for small specimens when high resolution (around 1 mu m) is required. Both are being applied to investigations in the biological and materials science fields.