263 resultados para Heart-assist devices
Resumo:
Hypoglycaemia is a major cause of neonatal morbidity and may induce long-term developmental sequelae. Clinical signs of hypoglycaemia in neonatal infants are unspecific or even absent, and therefore, precise and accurate methods for the assessment of glycaemia are needed. Glycaemia measurement in newborns has some particularities like a very low limit of normal glucose concentration compared to adults and a large range of normal haematocrit values. Many bedside point-of-care testing (POCT) systems are available, but literature about their accuracy in newborn infants is scarce and not very convincing. In this retrospective study, we identified over a 1-year study period 1,324 paired glycaemia results, one obtained at bedside with one of three different POCT systems (Elite? XL, Ascensia? Contour? and ABL 735) and the other in the central laboratory of the hospital with the hexokinase reference method. All three POCT systems tended to overestimate glycaemia values, and none of them fulfilled the ISO 15197 accuracy criteria. The Elite XL appeared to be more appropriate than Contour to detect hypoglycaemia, however with a low specificity. Contour additionally showed an important inaccuracy with increasing haematocrit. The bench analyzer ABL 735 was the most accurate of the three tested POCT systems. Both of the tested handheld glucometers have important drawbacks in their use as screening tools for hypoglycaemia in newborn infants. ABL 735 could be a valuable alternative, but the blood volume needed is more than 15 times higher than for handheld glucometers. Before daily use in the newborn population, careful clinical evaluation of each new POCT system for glucose measurement is of utmost importance.
Resumo:
High-field (>or=3 T) cardiac MRI is challenged by inhomogeneities of both the static magnetic field (B(0)) and the transmit radiofrequency field (B(1)+). The inhomogeneous B fields not only demand improved shimming methods but also impede the correct determination of the zero-order terms, i.e., the local resonance frequency f(0) and the radiofrequency power to generate the intended local B(1)+ field. In this work, dual echo time B(0)-map and dual flip angle B(1)+-map acquisition methods are combined to acquire multislice B(0)- and B(1)+-maps simultaneously covering the entire heart in a single breath hold of 18 heartbeats. A previously proposed excitation pulse shape dependent slice profile correction is tested and applied to reduce systematic errors of the multislice B(1)+-map. Localized higher-order shim correction values including the zero-order terms for frequency f(0) and radiofrequency power can be determined based on the acquired B(0)- and B(1)+-maps. This method has been tested in 7 healthy adult human subjects at 3 T and improved the B(0) field homogeneity (standard deviation) from 60 Hz to 35 Hz and the average B(1)+ field from 77% to 100% of the desired B(1)+ field when compared to more commonly used preparation methods.
Resumo:
BACKGROUND: Subclinical hypothyroidism has been associated with systolic and diastolic cardiac dysfunction and an elevated cholesterol level, but data on cardiovascular outcomes and death are limited. METHODS: We studied 2730 men and women, aged 70 to 79 years, with baseline thyrotropin (TSH) measurements and 4-year follow-up data to determine whether subclinical hypothyroidism was associated with congestive heart failure (CHF), coronary heart disease, stroke, peripheral arterial disease, and cardiovascular-related and total mortality. After the exclusion of participants with abnormal thyroxine levels, subclinical hypothyroidism was defined as a TSH level of 4.5 mIU/L or greater, and was further classified according to TSH levels (4.5-6.9, 7.0-9.9, and > or = 10.0 mIU/L). RESULTS: Subclinical hypothyroidism was present in 338 (12.4%) of the participants. Compared with euthyroid participants, CHF events occurred more frequently among those with a TSH level of 7.0 mIU/L or greater (35.0 vs 16.5 per 1000 person-years; P = .006), but not among those with TSH levels between 4.5 and 6.9 mIU/L. In multivariate analyses, the risk of CHF was higher among those with high TSH levels (TSH of 7.0-9.9 mIU/L: hazard ratio, 2.58 [95% confidence interval, 1.19-5.60]; and TSH of > or = 10.0 mIU/L: hazard ratio, 3.26 [95% confidence interval, 1.37-7.77]). Among the 2555 participants without CHF at baseline, the hazard ratio for incident CHF events was 2.33 (95% confidence interval, 1.10-4.96; P = .03) in those with a TSH of 7.0 mIU/L or greater. Subclinical hypothyroidism was not associated with increased risk for coronary heart disease, stroke, peripheral arterial disease, or cardiovascular-related or total mortality. CONCLUSIONS: Subclinical hypothyroidism is associated with an increased risk of CHF among older adults with a TSH level of 7.0 mIU/L or greater, but not with other cardiovascular events and mortality. Further investigation is warranted to assess whether subclinical hypothyroidism causes or worsens preexisting heart failure.
Resumo:
Catopril, an inhibitor of angiotensin converting enzyme, was given orally during cardiac catheterisation to 6 normotensive patients with refractory congestive heart-failure. 60--180 minutes after administration of 25 mg captopril, arterial pressure fell by 25%, cardiac index rose by 38%, and left-ventricular pressure and right-atrial pressure fell by 25% and 40% respectively. Plasma-renin activity rose while plasma noradrenaline and aldosterone fell. These data suggest that, in the short term, captopril can reduce both preload and afterload, and improve cardiac function, in refractory congestive heart-failure.
Resumo:
Loss of T-tubules (TT), sarcolemmal invaginations of cardiomyocytes (CMs), was recently identified as a general heart failure (HF) hallmark. However, whether TT per se or the overall sarcolemma is altered during HF process is still unknown. In this study, we directly examined sarcolemmal surface topography and physical properties using Atomic Force Microscopy (AFM) in living CMs from healthy and failing mice hearts. We confirmed the presence of highly organized crests and hollows along myofilaments in isolated healthy CMs. Sarcolemma topography was tightly correlated with elasticity, with crests stiffer than hollows and related to the presence of few packed subsarcolemmal mitochondria (SSM) as evidenced by electron microscopy. Three days after myocardial infarction (MI), CMs already exhibit an overall sarcolemma disorganization with general loss of crests topography thus becoming smooth and correlating with a decreased elasticity while interfibrillar mitochondria (IFM), myofilaments alignment and TT network were unaltered. End-stage post-ischemic condition (15days post-MI) exacerbates overall sarcolemma disorganization with, in addition to general loss of crest/hollow periodicity, a significant increase of cell surface stiffness. Strikingly, electron microscopy revealed the total depletion of SSM while some IFM heaps could be visualized beneath the membrane. Accordingly, mitochondrial Ca(2+) studies showed a heterogeneous pattern between SSM and IFM in healthy CMs which disappeared in HF. In vitro, formamide-induced sarcolemmal stress on healthy CMs phenocopied post-ischemic kinetics abnormalities and revealed initial SSM death and crest/hollow disorganization followed by IFM later disarray which moved toward the cell surface and structured heaps correlating with TT loss. This study demonstrates that the loss of crest/hollow organization of CM surface in HF occurs early and precedes disruption of the TT network. It also highlights a general stiffness increased of the CM surface most likely related to atypical IFM heaps while SSM died during HF process. Overall, these results indicate that initial sarcolemmal stress leading to SSM death could underlie subsequent TT disarray and HF setting.
Resumo:
Prenatal heart valve interventions aiming at the early and systematic correction of congenital cardiac malformations represent a promising treatment option in maternal-fetal care. However, definite fetal valve replacements require growing implants adaptive to fetal and postnatal development. The presented study investigates the fetal implantation of prenatally engineered living autologous cell-based heart valves. Autologous amniotic fluid cells (AFCs) were isolated from pregnant sheep between 122 and 128 days of gestation via transuterine sonographic sampling. Stented trileaflet heart valves were fabricated from biodegradable PGA-P4HB composite matrices (n = 9) and seeded with AFCs in vitro. Within the same intervention, tissue engineered heart valves (TEHVs) and unseeded controls were implanted orthotopically into the pulmonary position using an in-utero closed-heart hybrid approach. The transapical valve deployments were successful in all animals with acute survival of 77.8% of fetuses. TEHV in-vivo functionality was assessed using echocardiography as well as angiography. Fetuses were harvested up to 1 week after implantation representing a birth-relevant gestational age. TEHVs showed in vivo functionality with intact valvular integrity and absence of thrombus formation. The presented approach may serve as an experimental basis for future human prenatal cardiac interventions using fully biodegradable autologous cell-based living materials.
Resumo:
The magnitude of coffee-induced thermogenesis and the influence of coffee ingestion on substrate oxidation were investigated in 10 lean and 10 obese women, over two 24-h periods in a respiratory chamber. On one occasion the subjects consumed caffeinated coffee and on the other occasion, decaffeinated coffee. The magnitude of thermogenesis was smaller in obese (4.9 +/- 2.0%) than in lean subjects (7.6 +/- 1.3%). The thermogeneic response to caffeine was prolonged during the night in lean women only. The coffee-induced stimulation of energy expenditure was mediated by a concomitant increase in lipid and carbohydrate oxidation. During the next day, in postabsorptive basal conditions, the thermogenic effect of coffee had vanished, but a significant increase in lipid oxidation was observed in both groups. The magnitude of this effect was, however, blunted in obese women (lipid oxidation increased by 29 and 10% in lean and obese women, respectively). Caffeine increased urinary epinephrine excretion. Whereas urinary caffeine excretion was similar in both groups, obese women excreted more theobromine, theophylline, and paraxanthine than lean women. Despite the high levels of urinary methylxanthine excretion, thermogenesis and lipid oxidation were less stimulated in obese than in lean subjects.
Resumo:
PURPOSE: To determine if, compared to pressure support (PS), neurally adjusted ventilatory assist (NAVA) reduces patient-ventilator asynchrony in intensive care patients undergoing noninvasive ventilation with an oronasal face mask. METHODS: In this prospective interventional study we compared patient-ventilator synchrony between PS (with ventilator settings determined by the clinician) and NAVA (with the level set so as to obtain the same maximal airway pressure as in PS). Two 20-min recordings of airway pressure, flow and electrical activity of the diaphragm during PS and NAVA were acquired in a randomized order. Trigger delay (T(d)), the patient's neural inspiratory time (T(in)), ventilator pressurization duration (T(iv)), inspiratory time in excess (T(iex)), number of asynchrony events per minute and asynchrony index (AI) were determined. RESULTS: The study included 13 patients, six with COPD, and two with mixed pulmonary disease. T(d) was reduced with NAVA: median 35 ms (IQR 31-53 ms) versus 181 ms (122-208 ms); p = 0.0002. NAVA reduced both premature and delayed cyclings in the majority of patients, but not the median T(iex) value. The total number of asynchrony events tended to be reduced with NAVA: 1.0 events/min (0.5-3.1 events/min) versus 4.4 events/min (0.9-12.1 events/min); p = 0.08. AI was lower with NAVA: 4.9 % (2.5-10.5 %) versus 15.8 % (5.5-49.6 %); p = 0.03. During NAVA, there were no ineffective efforts, or late or premature cyclings. PaO(2) and PaCO(2) were not different between ventilatory modes. CONCLUSION: Compared to PS, NAVA improved patient ventilator synchrony during noninvasive ventilation by reducing T(d) and AI. Moreover, with NAVA, ineffective efforts, and late and premature cyclings were absent.
Resumo:
While mobile technologies can provide great personalized services for mobile users, they also threaten their privacy. Such personalization-privacy paradox are particularly salient for context aware technology based mobile applications where user's behaviors, movement and habits can be associated with a consumer's personal identity. In this thesis, I studied the privacy issues in the mobile context, particularly focus on an adaptive privacy management system design for context-aware mobile devices, and explore the role of personalization and control over user's personal data. This allowed me to make multiple contributions, both theoretical and practical. In the theoretical world, I propose and prototype an adaptive Single-Sign On solution that use user's context information to protect user's private information for smartphone. To validate this solution, I first proved that user's context is a unique user identifier and context awareness technology can increase user's perceived ease of use of the system and service provider's authentication security. I then followed a design science research paradigm and implemented this solution into a mobile application called "Privacy Manager". I evaluated the utility by several focus group interviews, and overall the proposed solution fulfilled the expected function and users expressed their intentions to use this application. To better understand the personalization-privacy paradox, I built on the theoretical foundations of privacy calculus and technology acceptance model to conceptualize the theory of users' mobile privacy management. I also examined the role of personalization and control ability on my model and how these two elements interact with privacy calculus and mobile technology model. In the practical realm, this thesis contributes to the understanding of the tradeoff between the benefit of personalized services and user's privacy concerns it may cause. By pointing out new opportunities to rethink how user's context information can protect private data, it also suggests new elements for privacy related business models.
Resumo:
ABSTRACT: BACKGROUND: Chest pain raises concern for the possibility of coronary heart disease. Scoring methods have been developed to identify coronary heart disease in emergency settings, but not in primary care. METHODS: Data were collected from a multicenter Swiss clinical cohort study including 672 consecutive patients with chest pain, who had visited one of 59 family practitioners' offices. Using delayed diagnosis we derived a prediction rule to rule out coronary heart disease by means of a logistic regression model. Known cardiovascular risk factors, pain characteristics, and physical signs associated with coronary heart disease were explored to develop a clinical score. Patients diagnosed with angina or acute myocardial infarction within the year following their initial visit comprised the coronary heart disease group. RESULTS: The coronary heart disease score was derived from eight variables: age, gender, duration of chest pain from 1 to 60 minutes, substernal chest pain location, pain increases with exertion, absence of tenderness point at palpation, cardiovascular risks factors, and personal history of cardiovascular disease. Area under the receiver operating characteristics curve was of 0.95 with a 95% confidence interval of 0.92; 0.97. From this score, 413 patients were considered as low risk for values of percentile 5 of the coronary heart disease patients. Internal validity was confirmed by bootstrapping. External validation using data from a German cohort (Marburg, n = 774) revealed a receiver operating characteristics curve of 0.75 (95% confidence interval, 0.72; 0.81) with a sensitivity of 85.6% and a specificity of 47.2%. CONCLUSIONS: This score, based only on history and physical examination, is a complementary tool for ruling out coronary heart disease in primary care patients complaining of chest pain.
Resumo:
To assess the role of arginine vasopressin (AVP) in congestive heart failure (CHF), 10 patients with CHF refractory to conventional treatment were studied before and 60 minutes after intravenous administration of 5 micrograms/kg of d(CH2)5Tyr(Me)AVP, a specific antagonist of AVP at the vascular receptor level. Heart rate, systemic arterial pressure, pulmonary arterial pressure, pulmonary capillary wedge pressure, cardiac index by thermodilution and cutaneous blood flow by laser-Doppler technique were measured. In 9 patients with no significant hemodynamic and cutaneous blood flow response to the AVP antagonist, baseline values (mean +/- standard deviation) were: heart rate, 77 +/- 14 beats/min; systemic arterial pressure, 120/79 +/- 18/8 mm Hg; pulmonary arterial pressure, 42/21 +/- 12/8 mm Hg; pulmonary capillary wedge pressure, 19 +/- 7 mm Hg; cardiac index, 2.2 +/- 0.6 liters/min/m2; plasma AVP, 2.3 +/- 0.8 pg/ml; and plasma osmolality, 284 +/- 14 mosm/kg H2O. The tenth patient had the most severe CHF. His plasma AVP level was 55 pg/ml and plasma osmolality was 290 mosm/kg. He responded to the AVP antagonist with a decrease in systemic arterial pressure from 115/61 to 79/41 mm Hg, in pulmonary arterial pressure from 58/31 to 33/13 mm Hg and in pulmonary capillary wedge pressure from 28 to 15 mm Hg. Simultaneously, cardiac index increased from 1.1 to 2.2 liters/min/m2 and heart rate from 113 to 120 beats/min; cutaneous blood flow increased 5-fold.(ABSTRACT TRUNCATED AT 250 WORDS)
Resumo:
High-dose dobutamine magnetic resonance stress testing has been shown to be superior to dobutamine stress echocardiography for diagnosis of coronary artery disease (CAD). We determined the feasibility of quantitative myocardial tagging during low- and high-dose dobutamine stress and tested the ability of global systolic and diastolic quantitative parameters to identify patients with significant CAD. Twenty-five patients suspected of having significant CAD were examined with a standard high-dose dobutamine/atropine stress magnetic resonance protocol (1.5-T scanner, Philips). All patients underwent invasive coronary angiography as the standard of reference for the presence (n = 13) or absence (n = 12) of significant CAD. During low-dose dobutamine stress, systolic (circumferential shortening, systolic rotation, and systolic rotation velocity) and diastolic (velocity of circumferential lengthening and diastolic rotation velocity) parameters changed significantly in patients without CAD (all P < 0.05 vs. rest) but not in patients with CAD. Identification of patients without and with CAD during low-dose stress was possible using the diastolic parameter of "time to peak untwist." At high-dose stress, none of the global systolic or diastolic parameters showed the potential to identify the presence of significant CAD. With myocardial tagging, a quantitative analysis of systolic and diastolic function was feasible during low- and high-dose dobutamine stress. In our study, the diastolic parameter of time to peak untwist as assessed during low-dose dobutamine stress was the most promising global parameter for identification of patients with significant CAD. Thus quantitative myocardial tagging may become a tool that reduces the need for high-dose dobutamine stress.