4 resultados para measurement accuracy
em Biblioteca Digital da Produção Intelectual da Universidade de São Paulo (BDPI/USP)
Resumo:
The purpose of this study was to evaluate ex vivo the accuracy an electronic apex locator during root canal length determination in primary molars. Methods: One calibrated examiner determined the root canal length in 15 primary molars (total=34 root canals) with different stages of root resorption. Root canal length was measured both visually, with the placement of a K-file 1 mm short of the apical foramen or the apical resorption bevel, and electronically using an electronic apex locator (Digital Signal Processing). Data were analyzed statistically using the intraclass correlation (ICC) test. Results: Comparing the actual and electronic root canal length measurements in the primary teeth showed a high correlation (ICC=0.95) Conclusions: The Digital Signal Processing apex locator is useful and accurate for apex foramen location during root canal length measurement in primary molars. (Pediatr Dent 200937:320-2) Received April 75, 2008 vertical bar Lost Revision August 21, 2008 vertical bar Revision Accepted August 22, 2008
Resumo:
Aim To evaluate ex vivo the accuracy of two electronic apex locators during root canal length determination in primary incisor and molar teeth with different stages of physiological root resorption. Methodology One calibrated examiner determined the root canal length in 17 primary incisors and 16 primary molars (total of 57 root canals) with different stages of root resorption based on the actual canal length and using two electronic apex locators. Root canal length was measured both visually, with the placement of a K-file 1 mm short of the apical foramen or the apical resorption bevel, and electronically using two electronic apex locators (Root ZX II - J. Morita Corp. and Mini Apex Locator - SybronEndo) according to the manufacturers` instructions. Data were analysed statistically using the intraclass correlation (ICC) test. Results Comparison of the actual root canal length and the electronic root canal length measurements revealed high correlation (ICC = 0.99), regardless of the tooth type (single-rooted and multi-rooted teeth) or the presence/absence of physiological root resorption. Conclusions Root ZX II and Mini Apex Locator proved useful and accurate for apex foramen location during root canal length measurement in primary incisors and molars.
Resumo:
P>Aim To evaluate ex vivo the accuracy of the iPex multi-frequency electronic apex locator (NSK Ltd, Tokyo, Japan) for working length determination in primary molar teeth. Methodology One calibrated examiner determined the working length in 20 primary molar teeth (total of 33 root canals). Working length was measured both visually, with the placement of a K-file 1 mm short of the apical foramen or the most coronal limit of root resorption, and electronically using the electronic apex locator iPex, according to the manufacturers` instructions. Data were analysed statistically using the intraclass correlation (ICC) test. Results Comparison of the actual and the electronic measurements revealed high correlation (ICC = 0.99) between the methods, regardless of the presence or absence of physiological root resorption. Conclusions In this laboratory study, the iPex accurately identified the apical foramen or the apical opening location for working length measurement in primary molar teeth.
Resumo:
Background: Oxidative modification of low-density lipoprotein (LDL) plays a key role in the pathogenesis of atherosclerosis. LDL(-) is present in blood plasma of healthy subjects and at higher concentrations in diseases with high cardiovascular risk, such as familial hypercholesterolemia or diabetes. Methods: We developed and validated a sandwich ELISA for LDL(-) in human plasma using two monoclonal antibodies against LDL(-) that do not bind to native LDL, extensively copper-oxidized LDL or malondialdehyde-modified LDL. The characteristics of assay performance, such as limits of detection and quantification, accuracy, inter- and intra-assay precision were evaluated. The linearity, interferences and stability tests were also performed. Results: The calibration range of the assay is 0.625-20.0 mU/L at 1: 2000 sample dilution. ELISA validation showed intra- and inter- assay precision and recovery within the required limits for immunoassays. The limits of detection and quantification were 0.423 mU/L and 0.517 mU/L LDL(-), respectively. The intra- and inter- assay coefficient of variation ranged from 9.5% to 11.5% and from 11.3% to 18.9%, respectively. Recovery of LDL(-) ranged from 92.8% to 105.1%. Conclusions: This ELISA represents a very practical tool for measuring LDL(-) in human blood for widespread research and clinical sample use. Clin Chem Lab Med 2008; 46: 1769-75.