217 resultados para DIAGNOdent® laser

em Biblioteca Digital da Produção Intelectual da Universidade de São Paulo (BDPI/USP)


Relevância:

30.00% 30.00%

Publicador:

Resumo:

The aim of this study was to verify the drying effect on the reproducibility of DIAGNOdent (Dd) devices to detect caries-like lesions. Three areas were created in each of the 34 bovine incisors: sound (S), demineralized (DE) and remineralized (RE). One examiner measured each area with two Dd devices (denominated X and Y), twice under humid, and twice under dry condition. Intra-rater agreement according each device and inter-device agreement were estimated by kappa statistics (k). Intra-rater agreement for device Y was substantial under humid (k DE=0.68 and k RE+S=0.68) and dry condition (k DE=0.64 and k RE+S=0.67). For device X, it was substantial under humid condition (k DE=0.57 and k RE+S=0.49), and it was almost perfect after air drying (k DE=1.0 and kRE+S=1.0). Inter-device agreement was slight (k =0.17) under humid condition, and it was substantial under dry condition (k =0.62). As reproducibility increased under dry condition, drying is advised to detect caries-like lesions on free smooth surfaces when different devices are used.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

An in vivo study was conducted to verify the ability of laser fluorescence (LF) to assess the activity status of occlusal caries in primary teeth, using different air-drying times. Occlusal sites (707) were examined using LF (DIAGNOdent) after air-drying for 3 s and 15 s, and the difference between readings (DIF15 s-3 s) was calculated. For concurrent validation of LF, visual criteria-Nyvad (NY) and Lesion Activity Assessment associated with the International Caries Detection and Assessment System (LAA-ICDAS)-were the reference standards for lesion activity. Histological exam using a pH-indicator dye (0.1% methyl red) was performed in 46 exfoliated/extracted teeth for criterion validation. LF readings and DIF15 s-3 s were compared using Kruskall-Wallis and Mann-Whitney tests. Receiver operating characteristic analyses were performed and validity parameters calculated, considering the caries activity assessment. Using NY, active lesions (3 s: 30.0 +/- 29.3; 15 s: 34.2 +/- 30.6) presented higher LF readings than inactive lesions (3 s: 17.0 +/- 16.3; 15 s: 19.2 +/- 17.3; p <0.05), different from LAA-ICDAS. Active cavitated caries resulted in higher LF readings (3 s: 50.3 +/- 3.5; 15 s: 54.7 +/- 30.2) than inactive cavitated caries (3 s: 19.9 +/- 16.3; 15 s: 22.8 +/- 16.8). Therefore, LF can distinguish cavitated active and inactive lesions classified by NY, but not by LAA-ICDAS; however, this difference might be related to the visual system rather than to LF. The air-drying time could be an alternative to improve the caries activity assessment; however, longer air-drying time is suggested to be tested subsequently. (C) 2010 Society of Photo-Optical Instrumentation Engineers. [DOI: 10.1117/1.3463007]

Relevância:

30.00% 30.00%

Publicador:

Resumo:

The purposes of this study were to evaluate in vitro the influence of different frequencies of Er:YAG laser on the human dentin caries removal capacity. Thirty fragments obtained from third molars were randomly assigned into three groups (n = 10) according to the laser frequency used: 4, 6, and 10 Hz. The caries lesion (+/-1 mm deep) was induced before the irradiation by S. mutans cultures for 6 weeks. The specimens of all groups were irradiated with 200 mJ of energy in noncontact and focused mode under constant refrigeration (water flow: 2.5 mL/min). Quantitative analysis of the caries removal was performed by DIAGNOdent (TM) and the Axion Vision (TM) software. Qualitative analysis was performed by Scanning electron microscope (SEM) and light microscope (LM). Data were analyzed by ANOVA and Fishers` tests. The DIAGNOdent (TM) revealed that the caries removal was similar with 4 and 6 Hz and was superior with 10 Hz (P < 0.05). The analysis with Axion Vision (TM) software revealed that the caries removal was similar with 6 and 10 Hz and the 4 Hz group promoted the lowest caries removal. Through SEM morphologic analysis, some specimens irradiated with 4 Hz presented, under the demineralized dentin, a disorganized collagenous matrix. The LM images revealed that all frequencies used promoted irregular caries removal, being observed over preparations with 6 and 10 Hz. It can be concluded that the increase of Er:YAG laser frequency provided a higher dentin caries removal without selectivity to the disorganized dentin. Microsc. Res. Tech. 74:281-286, 2011. (C) 2010 Wiley-Liss, Inc.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

Purpose: To compare visual inspection (VI), radiographic examination (RX) and the laser fluorescence device DIAGNOdent (L), as well as their combinations in vitro regarding treatment decisions for occlusal surfaces. Methods: 72 extracted human permanent teeth (molars and premolars) were used. Treatment decisions were recorded by three calibrated examiners, and the options available were fissure sealant and conservative restoration. For validation of treatment decisions, the teeth were sectioned and examined in a stereomicroscope. Thereafter, dental slices were scanned and the images were edited to facilitate classification of existing carious lesions. Intra and inter-examiner reproducibility for the determination of treatment plans were calculated using Cohen`s kappa test (95%-CI). Sensitivity, specificity, positive and negative predictive values, and the area under the ROC curve were also calculated. Results: VI and L provided on average the greatest intra- and inter-examiner reproducibility, respectively. Although the combination of diagnostic methods may decrease both intra- and inter examiners reproducibility, combination of VI, L and RX resulted in the greatest sensitivity, being statistically superior to RX and L. There was more inter-examiner agreement for the option of restorative treatment, while the use of sealants as a treatment option yielded the lowest values. Negative predictive values were numerically inferior to positive predictive values, indicating that the examiners preferred not to restore a carious tooth than to proceed operatively in an intact tooth. The combination of the three methods studied showed the best results in determining treatment plans for occlusal surfaces, when compared to the other types of exams. On the other hand, radiographic examination and laser fluorescence were less efficient when used alone.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Noncarious cervical lesions (NCCLs) are considered to be of multifactorial origin, normally associated with inadequate brushing. This study assessed the influence in vitro of simulated brushing on NCCL formation. Fifteen human premolars were submitted to brushing in the cementoenamel junction region, using hard-, medium- and soft-bristled brushes associated with a toothpaste of medium abrasiveness under a 200 g load, at a speed of 356 rpm for 100 minutes. The surface topography of the region was analyzed before and after brushing, by means of a laser interferometer, using "cut-off" values of 0.25 and considering roughness values in mm. The initial roughness (mm) results for dentin (D / bristle consistency: 1 - soft, 2 - medium and 3 - hard) were as follows: (D1) 1.25 ± 0.45; (D2) 1.12 ± 0.44; (D3) 1.05 ± 0.41. For enamel (E / bristle consistency: 1 - soft, 2 - medium and 3 - hard), the initial results were: (E1) 1.18 ± 0.35; (E2) 1.32 ± 0.25; (E3) 1.50 ± 0.38. After brushing, the following were the values for dentin: (D1) 2.32 ± 1.99; (D2) 3.30 ± 0.96; (D3) Over 500. For enamel, the values after brushing were: (E1) 1.37 ± 0.31; (E2) 2.15 ± 0.90; (E3) 1.22 ± 0.47. Based on the results of the ANOVA and Tukey statistical analyses (a = .05) it was concluded that soft, medium and hard brushes are not capable of abrading enamel, whereas dentin showed changes in surface roughness by the action of medium- and hard-bristled brushes.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

The aim of this study was to analyze the shear bond strength between commercially pure titanium, with and without laser welding, after airbone-particle abrasion (Al2O3) and 2 indirect composites. Sixty-four specimens were cast and divided into 2 groups with and without laser welding. Each group was divided in 4 subgroups, related to Al2O3 grain size: A - 250 µm; B - 180 µm; C- 110 µm; and D - 50 µm. Composite rings were formed around the rods and light polymerized using UniXS unit. Specimens were invested and their shear bond strength at failure was measured with a universal testing machine at a crosshead speed of 2.0 mm/min. Statistical analysis was carried out with ANOVA and Tukey's test (α=0.05). The highest bond strength means were recorded in 250 µm group without laser welding. The lowest shear bond strength means were recorded in 50 µm group with laser welding. Statistically significant differences (p<0.05) were found between all groups. In conclusion, airborne particle abrasion yielded significantly lower bond strength as the Al2O3 particle size decreased. Shear bond strength decreased in the laser welded specimens.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

The aim of this study was to evaluate the effectiveness of low-level laser therapy (LLLT) on the improvement of the mandibular movements and painful symptoms in individuals with temporomandibular disorders (TMD). Forty patients were randomly divided into two groups (n=20): Group 1 received the effective dose (GaAlAs laser ? 830 nm, 40 mW, 5J/cm2) and Group 2 received the placebo application (0 J/cm2), in continuous mode on the affected condyle lateral pole: superior, anterior, posterior, and posterior-inferior, twice a week during 4 weeks. Four evaluations were performed: E1 (before laser application), E2 (right after the last application), E3 (one week after the last application) and E4 (30 days after the last application). The Kruskal-Wallis test showed significant more improvements (p<0.01) in painful symptoms in the treated group than in the placebo group. A significant improvement in the range of mandibular movements was observed when the results were compared between the groups at E4. Laser application can be a supportive therapy in the treatment of TMD, since it resulted in the immediate decrease of painful symptoms and increased range of mandibular movements in the treated group. The same results were not observed in the placebo group.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

This in vitro study aimed to analyze the effect of different parameters of phototherapy with low intensity laser on the viability of human dental pulp fibroblasts under the effect of substances released by bleaching gel. Cells were seeded into 96 wells plates (1 x 10³ cells/well) and placed in contact with culture medium conditioned by a 35 % hydrogen peroxide bleaching gel for 40 minutes, simulating the clinical condition of the in-office bleaching treatment. Cells cultured in ideal growth conditions served as positive control group (PC), and the cells grown in conditioned medium and non-irradiated served as negative control group (NC). Cells grown in conditioned medium were submitted to a single irradiation with a diode laser (40 mW, 0.04 cm²) emitting at visible red (660 nm; RL) or near infrared (780 nm; NIR) using punctual technique, in contact mode and energy densities of 4, 6 or 10 J/cm². The cell viability was analyzed through the MTT reduction assay immediately and 24 hours after the irradiation. The data was compared by ANOVA followed by the Tukey's test (p < 0.05). The cell viability increased significantly in 24 hours within each group. The PC presented cell viability significantly higher than NC in both experimental times. Only the NIR/10 J/cm² group presented cell viability similar to that of PC in 24 hours. The phototherapy with low intensity laser in defined parameters is able to compensate the cytotoxic effects of substances released by 35 % hydrogen peroxide bleaching gel.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Secondary caries has been reported as the main reason for restoration replacement. The aim of this in vitro study was to evaluate the performance of different methods - visual inspection, laser fluorescence (DIAGNOdent), radiography and tactile examination - for secondary caries detection in primary molars restored with amalgam. Fifty-four primary molars were photographed and 73 suspect sites adjacent to amalgam restorations were selected. Two examiners evaluated independently these sites using all methods. Agreement between examiners was assessed by the Kappa test. To validate the methods, a caries-detector dye was used after restoration removal. The best cut-off points for the sample were found by a Receiver Operator Characteristic (ROC) analysis, and the area under the ROC curve (Az), and the sensitivity, specificity and accuracy of the methods were calculated for enamel (D2) and dentine (D3) thresholds. These parameters were found for each method and then compared by the McNemar test. The tactile examination and visual inspection presented the highest inter-examiner agreement for the D2 and D3 thresholds, respectively. The visual inspection also showed better performance than the other methods for both thresholds (Az = 0.861 and Az = 0.841, respectively). In conclusion, the visual inspection presented the best performance for detecting enamel and dentin secondary caries in primary teeth restored with amalgam.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

The purpose of this in vitro study was to evaluate the effect of neodymium:yttrium-aluminum-garnet (Nd:YAG) laser irradiation on intracanal dentin surface by SEM analysis and its interference in the apical seal of filled canals. After endodontic treatment procedures, 34 maxillary human incisors were randomly assigned to 2 groups. In the negative control group (n=17), no additional treatment was performed and teeth were filled with vertically condensed gutta-percha; in the laser-treated group (n=17), the root canals were irradiated with Nd:YAG laser (1.5 W, 100 mJ, 15 Hz) before filling as described for the control group. Two specimens of each group were prepared for SEM analysis to evaluate the presence and extent of morphological changes and removal of debris; the other specimens were immersed in 0.5% methylene blue dye (pH 7.2) for 24 h for evaluation of the linear dye leakage at the apical third. SEM analysis of the laser-treated group showed dentin fusion and resolidification without smear layer or debris. The Student’s t-test showed that the laser-treated group had significantly less leakage in apical third than the control group. Within the limitations of this study, it may be concluded that the morphological changes on the apical intraradicular dentin surface caused by Nd:YAG laser resulted in less linear dye apical leakage.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Er:YAG laser has been studied as a potential tool for restorative dentistry due to its ability to selectively remove oral hard tissue with minimal or no thermal damage to the surrounding tissues. The purpose of this study was to evaluate in vitro the tensile bond strength (TBS) of an adhesive/composite resin system to human enamel surfaces treated with 37% phosphoric acid, Er:YAG laser (lambda=2.94 mum) with a total energy of 16 J (80 mJ/pulse, 2Hz, 200 pulses, 250 ms pulse width), and Er:YAG laser followed by phosphoric acid etching. Analysis of the treated surfaces was performed by scanning electron microscopy (SEM) to assess morphological differences among the groups. TBS means (in MPa) were as follows: Er:YAG laser + acid (11.7 MPa) > acid (8.2 MPa) > Er:YAG laser (6.1 MPa), with the group treated with laser+acid being significantly from the other groups (p=0.0006 and p= 0.00019, respectively). The groups treated with acid alone and laser alone were significantly different from each other (p=0.0003). The SEM analysis revealed morphological changes that corroborate the TBS results, suggesting that the differences in TBS means among the groups are related to the different etching patterns produced by each type of surface treatment. The findings of this study indicate that the association between Er:YAG laser and phosphoric acid can be used as a valuable resource to increase bond strength to laser-prepared enamel.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

PURPOSE: To evaluate the mitochondrial function of the remnant liver (RL) in the early phase of liver regeneration in rats after 70% partial hepatectomy (PH). METHODS: Sixty male Wistar rats (200-250g) submitted to 70% PH were divided into five groups according to the time of euthanasia and application or not of laser light: C = Control, time zero; 2 minutes, 4, 6 and 24 hours after PH. The dose of laser radiation was 22.5 J/cm², wavelength of 660 nm (visible/red), in the remnant liver. We studied the respiration activated by ADP (state 3), basal mitochondrial respiration (state 4), respiratory control ratio (RCR) and mitochondrial membrane potential (MMP). RESULTS: The mitochondrial function of RL changed at 4 and 6 hours after PH, with a significant increase in state 3 and a concomitant increase in state 4 and with maintenance of RCR. MMP differed significantly between the groups biostimulated with laser radiation and the control group 4 hours after HP, with a substantial reduction in the non-laser groups. CONCLUSION: The laser light at the dose used in this study did not induce additional damage to the RL and seems to have delayed the hepatocellular metabolic overload of the remnant liver.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

The aim of this work was to evaluate the performance of femtosecond laser-induced breakdown spectroscopy (fs-LIBS) for the determination of elements in animal tissues. Sample pellets were prepared from certified reference materials, such as liver, kidney, muscle, hepatopancreas, and oyster, after cryogenic grinding assisted homogenization. Individual samples were placed in a two-axis computer-controlled translation stage that moved in the plane orthogonal to a beam originating from a Ti:Sapphire chirped-pulse amplification (CPA) laser system operating at 800 mu and producing a train of 840 mu J and 40 fs pulses at 90 Hz. The plasma emission was coupled into the optical fiber of a high-resolution intensified charge-coupled device (ICCD)-echelle spectrometer. Time-resolved characteristics of the laser-produced plasmas showed that the best results were obtained with delay times between 80 and 120 ns. Data obtained indicate both that it is a matrix-independent sampling process and that fs-LIBS can be used for the determination of Ca, Cu, Fe, K, Mg, Na, and P, but efforts must be made to obtain more appropriate detection limits for Al, Sr, and Zn.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Purpose: The purpose of this study was to evaluate the effectiveness of combined ureteroscopic holmium YAG lithotripsy for renal calculi associated with ipsilateral ureteral stones. Materials and Methods: Between August 2002 and March 2007, retrograde flexible ureteroscopic stone treatment was attempted in 351 cases. Indication for treatment was concurrent symptomatic ureteral stones in 63 patients (group I). Additional operative time and perioperative complication rates were compared to a group of 39 patients submitted to ureteroscopic treatment for ureteral calculi exclusively (group II). Results: Mean ureteral stone size was 8.0 +/- 2.6 mm and 8.1 +/- 3.4 mm for groups I and II, respectively. Mean operative time for group I was 67.9 +/- 29.5 minutes and for group 2 was 49.3 +/- 13.2 minutes (p < 0.001). Flexible ureteroscopic therapy for renal calculi increased 18 minutes in the mean operative time. The overall complication rate was 3.1% and 2.5% for groups I and II, respectively (p = 0.87). Mean renal stone size was 10.7 +/- 6.4 mm, overall stone free rate in group I was 81%. However, considering only patients with renal stones smaller than 15 mm, the stone free rate was 88%. Successful treatment occurred in 81% of patients presenting lower pole stones, but only 76% of patients with multiple renal stones became stone free. As expected, stone free rate showed a significant negative correlation with renal stone size (p = 0.03; r = -0.36). Logistic regression model indicated an independent association of renal stones smaller than 15 mm and stone free rate (OR = 13.5; p = 0.01). Conclusion: Combined ureteroscopic treatment for ureteral and ipsilateral renal calculi is a safe and attractive option for patients presenting for symptomatic ureteral stone and ipsilateral renal calculi smaller than 15 mm.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Objective: This study investigated the effects of low-level laser therapy (LLLT) and electrical stimulation (ES) on bone loss in spinal cord-injured rats. Materials and Methods: Thirty-seven male Wistar rats were divided into four groups: standard control group (CG); spinal cord-injured control (SC); spinal cord-injured treated with laser (SCL; GaAlAs, 830 nm, CW, 30mW/cm, 250 J/cm(2)); and spinal cord-injured treated with electrical field stimulation (SCE; 1.5 MHz, 1: 4 duty cycles, 30 mW, 20 min). Biomechanical, densitometric, and morphometric analyses were performed. Results: SC rats showed a significant decrease in bone mass, biomechanical properties, and morphometric parameters (versus CG). SCE rats showed significantly higher values of inner diameter and internal and external areas of tibia diaphyses; and the SCL group showed a trend toward the same result (versus SC). No increase was found in either mechanical or densitometric parameters. Conclusion: We conclude that the mentioned treatments were able to initiate a positive bone-tissue response, maybe through stimulation of osteoblasts, which was able to determine the observed morphometric modifications. However, the evoked tissue response could not determine either biomechanical or densitometric modifications.