1 resultado para 290-1

em Biblioteca Digital da Produção Intelectual da Universidade de São Paulo (BDPI/USP)


Relevância:

30.00% 30.00%

Publicador:

Resumo:

Data obtained during routine diagnosis of human T-cell lymphotropic virus type 1 (HTLV-1) and 2 (HTLV-2) in ""at-risk"" individuals from Sao Paulo, Brazil using signal-to-cutoff (S/C) values obtained by first, second, and third generation enzyme immunoassay (EIA) kits, were compared. The highest S/C values were obtained with third generation EIA kits, but no correlation was detected between these values and specific antibody reactivity to HTLV-1, HTLV-2, or untyped HTLV (p = 0.302). In addition, use of these third generation kits resulted in HTLV-1/2 false-positive samples. In contrast, first and second generation EIA kits showed high specificity, and the second generation EIA kits showed the highest efficiency, despite lower S/C values. Using first and second generation EIA kits, significant differences in specific antibody detection of HTLV-1, relative to HTLV-2 (p = 0.019 for first generation and p < 0.001 for second generation EIA kits) and relative to untyped HTLV (p = 0.025 for first generation EIA kits), were observed. These results were explained by the composition and format of the assays. In addition, using receiver operating characteristics (ROC) analysis, a slight adjustment in cutoff values for third generation EIA kits improved their specificities and should be used when HTLV ""at-risk"" populations from this geographic area are to be evaluated. (C) 2009 Elsevier B.V. All rights reserved.