35 resultados para transfemoral osseointegration


Relevância:

10.00% 10.00%

Publicador:

Resumo:

Background: This prospective and controlled histologic study evaluates the impact of smoking on bone-to-implant contact, the bone density in the threaded area, and the bone density outside the threaded area around microimplants with anodized surface retrieved from human jaws. Methods: A total of 24 subjects (mean age 51.32 +/- 7.5 years) were divided in two groups: smokers (n = 13 subjects) and non-smokers (n = 11 subjects). Each subject received one microimplant with oxidized surface during conventional mandible or maxilla implant surgery. After 8 weeks, the microimplants and the surrounding tissue were removed and prepared for histomorphometric analysis. Results: Three microimplants placed in smokers showed no osseointegration. The newly formed bone showed early stages of maturation, mainly in the non-smokers. Marginal bone loss, gap, and fibrous tissue were present around implants retrieved from smokers. Histometric evaluation indicated that the mean bone-to-implant contact ranged between 25.97% +/- 9.02% and 40.01% +/- 12.98% for smokers and non-smokers, respectively (P <0.001). Smokers presented 28.17% +/- 10.32% of bone density in the threaded area, whereas non-smokers showed 46.34% +/- 19.12%. The mean of bone density outside the threaded area ranged between 18.76% and 25.11% for smokers and non-smokers, respectively (P>0.05). Conclusion: The present data obtained in human subjects confirm that smoking has a detrimental effect on early bone tissue response around oxidized implant surfaces. J Periodontol 2010;81:575-583.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Purpose: Nonpassive fit frameworks are believed to lead to implant overload and consequently loss of osseointegration. This is one of the most commonly reported failures of implant prostheses. In an ideal situation of passive fit, when torque is applied to bring the abutment-cylinder interface together some amount of deformation can be expected, and it should be homogeneous along the periphery of the abutment. The aim of this study was to verify the amount of abutment deformation that can be expected when a free-standing cylinder is screwed into place. This could give insight into what should be accepted as passive fit. Materials and Methods: Strain gauges were bonded to the sides of five standard abutments that had machined palladium-silver cylinders or cobalt-chromium cast cylinders screwed into place. Measurements were taken to verify the deformation at each site. Results: Values of abutment deformation after abutment screw tightening ranged from -127.70 to -590.27 mu epsilon. The deformation recorded for palladium-silver prosthetic cylinder tightening ranged from 56.905 to -381.50 mu epsilon (mean: 173.298 mu epsilon) and from -5.62638 to -383.86 mu epsilon ( mean: 200.474 mu epsilon) for cobalt-chromium cylinders. There was no statistically significant difference among the two groups. Conclusion: Both abutment screw tightening and prosthetic cylinder screw tightening result in abutment deformation, which is compressive most of the time. Int J Prosthodont 2009; 22: 391-395.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Clinical feasibility of mandibular implant overdenture retainers submitted to immediate load Introduction: Millions of people around the world do not have access to the benefits of osseointegration. Treatments involving oral rehabilitation with overdentures have been widely used by specialists in the oral medicine field. This is an alternative therapy for retention and stability achievement in total prosthesis with conventional treatment, and two implants are enough to establish a satisfactory overdenture. Objective: The objectives of the study were to evaluate 16 patients of both sexes, with an average age of 47.4 +/- 4 years, using electromyographic analysis of masseter and temporal muscles and analyse the increase of incisive and molar maximal bite force with their existing complete dentures and following mandibular implant overdenture therapy to assess the benefits of this treatment. Materials and methods: For these tests, the Myosystem-BR1 electromyograph and the IDDK Kratos dynamometer were used. Statistical analysis was performed using the repeated measures test (SPSS 17.0). Results: A decrease in electromyographic activity during the rest, lateral and protrusion movements and increase of the maximal incisive and molar bite force after 15 months with a mandibular implant overdenture was observed. Conclusion: All the patients in this study reported a considerable improvement in the masticatory function and prostheses stability following treatment. It is possible to propose that the use of mandibular implants overdenture should become the selected treatment for totally edentulous patients to facilitate oral function and quality of life.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Objective: This study aimed at investigating the influence of the porous titanium (Ti) structure on the osteogenic cell behaviour. Materials and methods: Porous Ti discs were fabricated by the powder metallurgy process with the pore size typically between 50 and 400 mm and a porosity of 60%. Osteogenic cells obtained from human alveolar bone were cultured until subconfluence and subcultured on dense Ti (control) and porous Ti for periods of up to 17 days. Results: Cultures grown on porous Ti exhibited increased cell proliferation and total protein content, and lower levels of alkaline phosphatase (ALP) activity than on dense Ti. In general, gene expression of osteoblastic markers-runt-related transcription factor 2, collagen type I, alkaline phosphatase, bone morphogenetic protein-7, and osteocalcin was lower at day 7 and higher at day 17 in cultures grown on porous Ti compared with dense Ti, a finding consistent with the enhanced growth rate for such cultures. The amount of mineralized matrix was greater on porous Ti compared with the dense one. Conclusion: These results indicate that the porous Ti is an appropriate substrate for osteogenic cell adhesion, proliferation, and production of a mineralized matrix. Because of the three-dimensional environment it provides, porous Ti should be considered an advantageous substrate for promoting desirable implant surface-bone interactions.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

The aim of the present study was to evaluate the in vitro osteogenic potential of subcultured human osteoblastic cells derived from alveolar bone on a titanium (Ti) surface produced by an anodized alkali treatment (BSP-AK). Primary osteoblastic cells were subcultured on BSP-AK and machined Ti discs (control) and grown for periods of up to 21 days under osteogenic conditions. Morphologic and biochemical methods were used to assess important parameters of in vitro bone-like tissue formation. Although no major differences were observed between the BSP-AK and the control Ti surface in terms of cell attachment and mineralized matrix formation, a significant increase in cell population, ALP activity, and collagen content was detected in cultures on BSP-AK surface. Our results demonstrate that human osteoblastic cells are sensitive to the BSP-AK-modified Ti surface during the transitional stage between the end of the proliferative phase and the onset of the differentiation /matrix maturation ones. Together with the good mechanical properties exhibited by the Ca- and P- coating, our findings suggest that BSP-AK treatment could be useful for the development of a new surface for dental and orthopedic implants. (c) 2008 Wiley Periodicals, Inc.J Biomed Mater Res 88A: 841-848, 2009

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Despite wide clinical application, the efficacy of platelet-rich plasma (PRP) for repairing bone defects and enhancing osseointegration of metal implants is still subject of debate. This study aimed to evaluate the effects of a well-defined PRP-like mixture containing platelet-derived growth factor-BB, transforming growth factor (TGF)-beta 1, TGF-beta 2, albumin, fibronectin, and thrombospondin [growth factors (GFs) + proteins] on the development of the osteogenic phenotype on titanium (Ti) in vitro. Human alveolar bone-derived osteoblastic cells were subcultured on Ti discs and exposed during the first 7 days to osteogenic medium supplemented with GFs + proteins and to osteogenic medium alone thereafter up to 14 days. Control cultures were exposed to only osteogenic medium. Dose-response experiments were carried out using rat primary calvarial cells exposed to GFs + proteins and 1:10 or 1:100 dilutions of the mixture. Treated human-derived cell cultures exhibited a significantly higher number of cycling cells at days 1 and 4 and of total cells at days 4 and 7, significantly reduced alkaline phosphatase (ALP) activity at days 4, 7, and 10, and no Alizarin red-stained areas (calcium deposits) at day 14, indicating an impairment in osteoblast differentiation. Although the 1:10 and 1:100 dilutions of the mixture restored the proliferative activity of rat-derived osteogenic cells to control levels and promoted a significant increase in ALP activity at day 10 compared with GFs + proteins, mineralized nodule formation was only observed with the 1:100 dilution (similar to 50% of the control). These results showed that a PRP-like protein mixture inhibits development of the osteogenic phenotype in both human and rat osteoblastic cell cultures grown on Ti. (J Histochem Cytochem 57:265-276, 2009)

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Background: Stability of pen-implant crestal bone plays a relevant role relative to the presence or absence of interdental papilla. Several factors can contribute to the crestal bone resorption observed around two-piece implants, such as the presence of a microgap at the level of the implant abutment junction, the type of connection between implant and prosthetic components, the implant positioning relative to the alveolar crest, and the interimplant distance. Subcrestal positioning of dental implants has been proposed to decrease the risk of exposure of the metal of the top of the implant or of the abutment margin, and to get enough space in a vertical dimension to create a harmoniously esthetic emergence profile. Methods: The present retrospective histologic study was performed to evaluate dental implants retrieved from human jaws that had been inserted in an equicrestal or subcrestal position. A total of nine implants were evaluated: five of these had been inserted in an equicrestal position, whereas the other four had been positioned subcrestally (1 to 3 mm). Results: In all subcrestally placed implants, preexisting and newly formed bone was found over the implant shoulder. In the equicrestal implants, crestal bone resorption (0.5 to 1.5 mm) was present around all implants. Conclusion: The subcrestal position of the implants resulted in bone located above the implant shoulder. J Periodontol 2011;82:708-715.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Introduction: New reconstructive and less invasive methods have been searched to optimize bone formation and osseointegration of dental implants in maxillary sinus augmentation. Purpose: The aim of the presented ovine split-mouth study was to compare bovine bone mineral (BBM) alone and in combination with mesenchymal stem cells (MSCs) regarding their potential in sinus augmentation. Material and Methods: Bilateral sinus floor augmentations were performed in six adult sheep. BBM and MSCs were placed into the test side and only BBM in the contra-lateral control side of each sheep. Animals were sacrificed after 8 and 16 weeks. Augmentation sites were analyzed by computed tomography, histology, and histomorphometry. Results: The initial volumes of both sides were similar and did not change significantly with time. A tight connection between the particles of BBM and the new bone was observed histologically. Bone formation was significantly (p = 0.027) faster by 49% in the test sides. Conclusion: The combination of BBM and MSCs accelerated new bone formation in this model of maxillary sinus augmentation. This could allow early placement of implants.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Objectives To evaluate the influence of implant size and configuration on osseointegration in implants immediately placed into extraction sockets. Material and methods Implants were installed immediately into extraction sockets in the mandibles of six Labrador dogs. In the control sites, cylindrical transmucosal implants (3.3 mm diameter) were installed, while in the test sites, larger and conical (root formed, 5 mm diameter) implants were installed. After 4 months of healing, the resorptive patterns of the alveolar crest were evaluated histomorphometrically. Results With one exception, all implants were integrated in mineralized bone, mainly composed of mature lamellar bone. The alveolar crest underwent resorption at the control as well as at the test implants. This resorption was more pronounced at the buccal aspects and significantly greater at the test (2.7 +/- 0.4 mm) than at the control implants (1.5 +/- 0.6 mm). However, the control implants were associated with residual defects that were deeper at the lingual than at the buccal aspects, while these defects were virtually absent at test implants. Conclusions The installment of root formed wide implants immediately into extraction sockets will not prevent the resorption of the alveolar crest. In contrast, this resorption is more marked both at the buccal and lingual aspects of root formed wide than at standard cylindrical implants. To cite this article:Caneva M, Salata LA, de Souza SS, Bressan E, Botticelli D, Lang NP. Hard tissue formation adjacent to implants of various size and configuration immediately placed into extraction sockets: an experimental study in dogs.Clin. Oral Impl. Res. 21, 2010; 885-895.doi: 10.1111/j.1600-0501.2010.01931.x.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Aim To evaluate the influence of resorbable membranes on hard tissue alterations and osseointegration at implants placed into extraction sockets in a dog model. Material and methods In the mandibular premolar region, implants were installed immediately into the extraction sockets of six Labrador dogs. Collagen-resorbable membranes were placed at the test sites, while the control sites were left uncovered. Implants were intended to heal in a submerged mode. After 4 months of healing, the animals were sacrificed, and ground sections were obtained for histomorphometric evaluation. Results After 4 months of healing, a control implant was not integrated (n=5). Both at the test and at the control sites, bone resorption occurred. While the most coronal bone-to-implant contact was similar between the test and the control sites, the alveolar bone crest outline was maintained to a higher degree at the buccal aspect of the test sites (loss: 1.7 mm) compared with the control sites (loss: 2.2 mm). Conclusions The use of collagen-resorbable membranes at implants immediately placed into extraction sockets contributed to a partial (23%) preservation of the buccal outline of the alveolar process. To cite this article:Caneva M, Botticelli D, Salata LA, Souza SLS, Carvalho Cardoso L, Lang NP. Collagen membranes at immediate implants: a histomorphometric study in dogs.Clin. Oral Impl. Res. 21, 2010; 891-897.doi: 10.1111/j.1600-0501.2010.01946.x.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Aim To evaluate the influence of magnesium-enriched hydroxyapatite (MHA) (SintLife (R)) on bone contour preservation and osseointegration at implants placed immediately into extraction sockets. Material and methods In the mandibular pre-molar region, implants were installed immediately into extraction sockets of six Labrador dogs. MHA was placed at test sites, while the control sites did not receive augmentation materials. Implants were intended to heal in a submerged mode. After 4 months of healing, the animals were sacrificed, and ground sections were obtained for histomorphometric evaluation. Results After 4 months of healing, one control implant was not integrated leaving n=5 test and control implants for evaluation. Both at the test and the control sites, bone resorption occurred. While the most coronal bone-to-implant contact was similar between test and control sites, the alveolar bony crest outline was maintained to a higher degree at the buccal aspect of the test sites (loss: 0.7 mm) compared with the control sites (loss: 1.2 mm), even though this difference did not reach statistical significance. Conclusions The use of MHA to fill the defect around implants placed into the alveolus immediately after tooth extraction did not contribute significantly to the maintenance of the contours of the buccal alveolar bone crest. To cite this article:Caneva M, Botticelli D, Stellini E, Souza SLS, Salata LA, Lang NP. Magnesium-enriched hydroxyapatite at immediate implants: a histomorphometric study in dogs.Clin. Oral Impl. Res. 22, 2011; 512-517doi: 10.1111/j.1600-0501.2010.02040.x.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Background: The purpose of this study was to compare and evaluate bone and soft tissue levels between immediately placed, immediately restored implants positioned in the esthetic anterior region with different interimplant distances (IID). Methods: Forty-nine patients requiring multiple implant restorations in the anterior regions received 152 implants, which were restored immediately. Periapical radiographs and digital images of 99 interimplant sites were taken at the regular follow-up examinations at 0, 6, 12, and 24 months after surgery. They were digitally recorded and analyzed. The presence of the interproximal papilla was assessed and compared to the distances between the bone crest and the contact point between the natural teeth and the restoration crowns. Results: Implants with an IID <2 mm seemed to lose less bone laterally. When the IID was <2 mm, vertical crestal bone loss was significantly greater than in the group with IID >4 mm. The percentage of the interproximal papilla presence decreased when the distance between the bone crest and the contact point between the two restoration crowns was >6 mm and when two implants were placed at a distance >= 4 mm. Conclusions: To guarantee a better esthetic result in immediately placed, immediately restored implants, the contact point between the two prosthetic crowns should be placed at 3 to 4 mm, and never >6 mm, from the bone peak. Two adjacent implants should be placed at a distance >2 and <4 mm.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

The microstructure of the crestal alveolar bone is important for both the maintenance of osseointegration and the location of the gingival soft tissues. The aim of this study was to evaluate and compare the bone microstructure of the alveolar bone and of the interimplant bone in implants inserted at different interimplant distances. The mandibular bilateral premolars of six dogs were extracted, and after 12 weeks, each dog received eight implants, for a total of 48 implants. Two pairs of implants, one for each hemiarch, were separated by 2 mm (group 1) and by 3 mm (group 2). After 12 weeks, the implants received temporary acrylic prostheses. After four more weeks, metallic crowns substituted the temporary prostheses. After an additional 8 weeks the animals were sacrificed and the hemimandibles were removed, dissected, and processed. The longitudinal collagen fiber orientation was 43.2% for the alveolar bone; it was 30.3% for the 2-mm group and 43.9% for the 3-mm group. There was a statistically significant difference between the 2-mm and 3-mm groups (p < .05). The orientation of transverse collagen fibers was 47.8% for the alveolar bone; it was 37.3% for the 2-mm group and 56.3% for the 3-mm group. There was a statistically significant difference between the 2-mm and 3-mm groups (p < .05). The marrow spaces were 34.87% for the alveolar bone, 52.3% for the 2-mm group, and 59.9% for the 3-mm group. There was a statistically significant difference between the alveolar bone and the 3-mm group (p < .05). The low mineral density index was 36.29 for the alveolar bone, 46.76 for the 2-mm group, and 17.91 for the 3-mm group. There was a statistically significant difference between the 2-mm and 3-mm groups (p < .05). The high mineral density was 87.57 for the alveolar bone, 72.58 for the 2-mm group, and 84.91 for the 3-mm group. There was a statistically significant difference between the alveolar bone and the 2-mm group (p < .05). The collagen fiber orientation resulted in statistically significant differences in both the 2-mm and 3-mm groups compared with the alveolar bone. The marrow spaces appeared significantly increased in the 3-mm group compared with the alveolar bone. The low mineral density index was significantly higher in the 2-mm group, while the high mineral density index was significantly higher in the alveolar bone. In conclusion, the interimplant distance should not be less than 3 mm.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Aim To compare the remodeling of the alveolar process at implants installed immediately into extraction sockets by applying a flap or a ""flapless"" surgical approach in a dog model. Material and methods Implants were installed immediately into the distal alveoli of the second mandibular premolars of six Labrador dogs. In one side of the mandible, a full-thickness mucoperiosteal flap was elevated (control site), while contra-laterally, the mucosa was gently dislocated, but not elevated (test site) to disclose the alveolar crest. After 4 months of healing, the animals were sacrificed, ground sections were obtained and a histomorphometric analysis was performed. Results After 4 months of healing, all implants were integrated (n=6). Both at the test and at the control sites, bone resorption occurred with similar outcomes. The buccal bony crest resorption was 1.7 and 1.5 mm at the control and the test sites, respectively. Conclusions ""Flapless"" implant placement into extraction sockets did not result in the prevention of alveolar bone resorption and did not affect the dimensional changes of the alveolar process following tooth extraction when compared with the usual placement of implants raising mucoperiosteal flaps. To cite this article:Caneva M, Botticelli D, Salata LA, Souza SLS, Bressan E, Lang NP. Flap vs. ""flapless"" surgical approach at immediate implants: a histomorphometric study in dogs.Clin. Oral Impl. Res. 21, 2010; 1314-1319.doi: 10.1111/j.1600-0501.2009.01959.x.

Relevância:

10.00% 10.00%

Publicador:

Resumo:

Purpose: The aim of this study was to evaluate, through histomorphometric analysis, the effect that different loading times would have on the bone response around implants. Materials and Methods: Three Replace Select implants were placed on each side of the mandible in eight dogs (n = 48 implants). One pair of implants was selected for an immediate loading protocol (IL). After 7 days, the second pair of implants received prostheses for an early loading protocol (EL). Fourteen days after implant placement, the third pair of implants received prostheses for advanced early loading (AEL). Following 12 weeks of prosthetics, counted following the positioning of the metallic crowns for the AEL group, the animals were sacrificed and the specimens were prepared for histomorphometric analysis. The differences between loading time in the following parameters were evaluated through analysis of variance: bone-to-implant contact, bone density, and crestal bone loss. Results: The mean percentage of bone-to-implant contact for IL was 77.9% +/- 1.71%, for EL it was 79.25% +/- 2.11%, and for AEL it was 79.42% +/- 1.49%. The mean percentage of bone density for IL was 69.97% +/- 3.81%, for EL it was 69.23% +/- 5.68%, and for AEL it was 69.19% +/- 2.90%. Mean crestal bone loss was 1.57 +/- 0.22 mm for IL, 1.23 +/- 0.19 mm for EL, and 1.17 +/- 0.32 mm for AEL. There was no statistical difference for any of the parameters evaluated (P > .05). Conclusion: Different early loading times did not seem to significantly affect the bone response around dental implants. INT J ORAL MAXILLOFAC IMPLANTS 2010;25:473-481