Resumo:

Since its introduction, pulse oximetry has become a conventional clinical measure. Besides being arterial blood oxygen saturation (SpO2) measure, pulse oximeters can be used for other cardiovascular measurements, like heart rate (HR) estimations, derived from its photo plethysmographic (PPG) signals. The temporal coherence of the PPG signals and thereby HR estimates are heavily dependent on its minimal phase variability. A Masimo SET Rad-9TM, Novametrix Oxypleth and a custom designed PPG system were investigated for their relative phase variation. R-R intervals from electro-cardiogram (ECG) were recorded concurrently as reference. PPG signals obtained from the 3 systems were evaluated by comparing their respective beat-to-beat (B-B) intervals with the corresponding R-R estimates during a static test. For their relative B-B comparison to the ECG, Novametrix system differed 0.680.52% (p