3 resultados para Positional Accuracy
Resumo:
Background
]In modern radiotherapy, it is crucial to monitor the performance of all linac components including gantry, collimation system and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method has been introduced in conjunction with an algorithm to investigate the stability of these systems during arc treatments with the aim of ensuring the accuracy of linac mechanical performance.
The Varian EPID sag, gantry sag, changes in source-to-detector distance (SDD), EPID and collimator skewness, EPID tilt, and the sag in MLC carriages as a result of linac rotation were separately investigated by acquisition of EPID images of a simple phantom comprised of 5 ball-bearings during arc delivery. A fast and robust software package was developed for automated analysis of image data. Twelve Varian linacs of different models were investigated.
The average EPID sag was within 1 mm for all tested linacs. All machines showed less than 1 mm gantry sag. Changes in SDD values were within 1.7 mm except for three linacs of one centre which were within 9 mm. Values of EPID skewness and tilt were negligible in all tested linacs. The maximum sag in MLC leaf bank assemblies was around 1 mm. The EPID sag showed a considerable improvement in TrueBeam linacs.
The methodology and software developed in this study provide a simple tool for effective investigation of the behaviour of linac components with gantry rotation. It is reproducible and accurate and can be easily performed as a routine test in clinics.
Resumo:
This work investigated the differences between multileaf collimator (MLC) positioning accuracy determined using either log files or electronic portal imaging devices (EPID) and then assessed the possibility of reducing patient specific quality control (QC) via phantom-less methodologies. In-house software was developed, and validated, to track MLC positional accuracy with the rotational and static gantry picket fence tests using an integrated electronic portal image. This software was used to monitor MLC daily performance over a 1 year period for two Varian TrueBeam linear accelerators, with the results directly compared with MLC positions determined using leaf trajectory log files. This software was validated by introducing known shifts and collimator errors. Skewness of the MLCs was found to be 0.03 ± 0.06° (mean ±1 standard deviation (SD)) and was dependent on whether the collimator was rotated manually or automatically. Trajectory log files, analysed using in-house software, showed average MLC positioning errors with a magnitude of 0.004 ± 0.003 mm (rotational) and 0.004 ± 0.011 mm (static) across two TrueBeam units over 1 year (mean ±1 SD). These ranges, as indicated by the SD, were lower than the related average MLC positioning errors of 0.000 ± 0.025 mm (rotational) and 0.000 ± 0.039 mm (static) that were obtained using the in-house EPID based software. The range of EPID measured MLC positional errors was larger due to the inherent uncertainties of the procedure. Over the duration of the study, multiple MLC positional errors were detected using the EPID based software but these same errors were not detected using the trajectory log files. This work shows the importance of increasing linac specific QC when phantom-less methodologies, such as the use of log files, are used to reduce patient specific QC. Tolerances of 0.25 mm have been created for the MLC positional errors using the EPID-based automated picket fence test. The software allows diagnosis of any specific leaf that needs repair and gives an indication as to the course of action that is required.
Resumo:
This article explores statistical approaches for assessing the relative accuracy of medieval mapping. It focuses on one particular map, the Gough Map of Great Britain. This is an early and remarkable example of a medieval “national” map covering Plantagenet Britain. Conventionally dated to c. 1360, the map shows the position of places in and coastal outline of Great Britain to a considerable degree of spatial accuracy. In this article, aspects of the map's content are subjected to a systematic analysis to identify geographical variations in the map's veracity, or truthfulness. It thus contributes to debates among historical geographers and cartographic historians on the nature of medieval maps and mapping and, in particular, questions of their distortion of geographic space. Based on a newly developed digital version of the Gough Map, several regression-based approaches are used here to explore the degree and nature of spatial distortion in the Gough Map. This demonstrates that not only are there marked variations in the positional accuracy of places shown on the map between regions (i.e., England, Scotland, and Wales), but there are also fine-scale geographical variations in the spatial accuracy of the map within these regions. The article concludes by suggesting that the map was constructed using a range of sources, and that the Gough Map is a composite of multiscale representations of places in Great Britain. The article details a set of approaches that could be transferred to other contexts and add value to historic maps by enhancing understanding of their contents.