9 resultados para Ocular Response Analyzer


Relevância:

100.00% 100.00%

Publicador:

Resumo:

PURPOSE: To estimate the relationships between ocular parameters and tonometrically measured intraocular pressure (IOP), to determine the influence of ocular parameters on different instrument measurements of IOP, and to evaluate the association of ocular parameters with a parameter called hysteresis. METHODS: Patients presenting at a glaucoma clinic were recruited for this study. Subjects underwent IOP measurement with the Goldmann applanation tonometer (GAT), the TonoPen, and the Reichert Ocular Response Analyzer (ORA), and also measurements of central corneal thickness (CCT), axial length, corneal curvature, corneal astigmatism, central visual acuity, and refractive error. Chart information was reviewed to determine glaucoma treatment history. The ORA instrument provided a measurement called corneal hysteresis. The association between measured IOP and the other ocular characteristics was estimated using generalized estimating equations. RESULTS: Among 230 patients, IOP measurements from the TonoPen read lowest, and ORA read highest, and GAT measurements were closest to the mean IOP of the 3 instruments. In a multiple regression model adjusting for age, sex, race, and other ocular characteristics, a 10 microm increase in CCT was associated with an increase of 0.79 mm Hg measured IOP in untreated eyes (P<0.0001). Of the 3 tonometers, GAT was the least affected by CCT (0.66 mm Hg/10 mum, P<0.0001). Hysteresis was significantly correlated with CCT with a modest correlation coefficient (r=0.20, P<0.0007). CONCLUSIONS: Among parameters related to measured IOP, features in addition to CCT, such as hysteresis and corneal curvature, may also be important. Tonometric instruments seem to be affected differently by various physiologic characteristics.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

PURPOSE: To determine and compare the corneal biomechanical properties between eyes with primary open angle glaucoma (POAG) and eyes with normal tension glaucoma (NTG). PATIENTS AND METHODS: Prospective cross-sectional study. Consecutive eligible POAG and NTG patients attending the Glaucoma Clinic had assessment of their corneal biomechanical properties-corneal hysteresis (CH) and corneal resistance factor (CRF)-using the Ocular Response Analyzer by an observer masked to the diagnosis. Exclusion criteria included previous intraocular surgery, corneal pathology, inflammatory connective tissue disease, and refraction of 5-dimensional or over. If both eyes were eligible, then the right eye was used for analysis. The main outcome measures were corneal hysteresis and CRF measurements. Data analysis was performed using the t test and general linear model. RESULTS: Eighty-one patients (80 whites) were analyzed. Forty had NTG, whereas 41 had POAG. Thirty-five were females. There was a statistically significant difference in mean CH (NTG 9.6±1.3 mm Hg; POAG 9.0±1.4 mm Hg; P=0.01), but not in mean CRF (NTG 9.9±1.4; POAG 10.8±1.7; P=0.06). The highest recorded Goldmann applanation intraocular pressure (IOP) was statistically significantly associated with lower CH (P=0.01) and higher CRF (P=0.02). CONCLUSIONS: There was a small but statistically significant difference in the mean CH between POAG and NTG (CH was higher in NTG). The highest recorded Goldmann applanation IOP was also statistically significantly correlated with lower CH and higher CRF, suggesting that alterations to the corneal biomechanical properties may occur as a result of chronic raised IOP in POAG. © 2008 by Lippincott Williams & Wilkins.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Objective: Acquired pit-like changes of the optic nerve head (APON) are characteristic of glaucomatous damage and may be a sign of a localized susceptibility of the optic nerve. Thus, it is possible that biomechanical properties of the ocular tissues may play a pressure-independent role in the pathogenesis of glaucoma. Corneal hysteresis (CH) appears to provide information of the biomechanical properties of the ocular hull tissues. The purpose of this study was to compare CH of patients with primary open angle glaucoma (POAG) with and without APON. Methods: A prospective case control study was done. POAG patients with and without APON were measured using the Ocular Response Analyzer by masked investigators. Patients in both groups were matched for sex, age, corneal thickness, and type of glaucoma according to maximal IOP (NTG or POAG). Statistical analysis was done using ANOVA. Results: Corneal hysteresis of 16 glaucomatous eyes with APON and 32 controls (glaucoma without APON) was measured. The mean (±SD) CH in the APON group was 8.89 (±1.53) and 10.2 (±1.05) in the control group. The difference is statistically significant (p = 0.005). Conclusions: Corneal hysteresis in POAG patients with APON was significantly lower than in patients that did not have such structural changes of the optic disc. These findings may reflect pressure-independent mechanisms involved in the pathogenesis of such glaucomatous optic nerve changes. © Springer-Verlag 2007.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

OBJECTIVE: To assess the agreement of tonometers available for clinical practice with the Goldmann applanation tonometer (GAT), the most commonly accepted reference device.

DESIGN: A systematic review and meta-analysis of directly comparative studies assessing the agreement of 1 or more tonometers with the reference tonometer (GAT).

PARTICIPANTS: A total of 11 582 participants (15 525 eyes) were included.

METHODS: Summary 95% limits of agreement (LoA) were produced for each comparison.

MAIN OUTCOME MEASURES: Agreement, recordability, and reliability.

RESULTS: A total of 102 studies, including 130 paired comparisons, were included, representing 8 tonometers: dynamic contour tonometer, noncontact tonometer (NCT), ocular response analyzer, Ocuton S, handheld applanation tonometer (HAT), rebound tonometer, transpalpebral tonometer, and Tono-Pen. The agreement (95% limits) seemed to vary across tonometers: 0.2 mmHg (-3.8 to 4.3 mmHg) for the NCT to 2.7 mmHg (-4.1 to 9.6 mmHg) for the Ocuton S. The estimated proportion within 2 mmHg of the GAT ranged from 33% (Ocuton S) to 66% and 59% (NCT and HAT, respectively). Substantial inter- and intraobserver variability were observed for all tonometers.

CONCLUSIONS: The NCT and HAT seem to achieve a measurement closest to the GAT. However, there was substantial variability in measurements both within and between studies.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

PURPOSE:

We sought to measure the impact of central corneal thickness (CCT), a possible risk factor for glaucoma damage, and corneal hysteresis, a proposed measure of corneal resistance to deformation, on various indicators of glaucoma damage.

DESIGN:

Observational study.

METHODS:

Adult patients of the Wilmer Glaucoma Service underwent measurement of hysteresis on the Reichert Ocular Response Analyzer and measurement of CCT by ultrasonic pachymetry. Two glaucoma specialists (H.A.Q., N.G.C.) reviewed the chart to determine highest known intraocular pressure (IOP), target IOP, diagnosis, years with glaucoma, cup-to-disk ratio (CDR), mean defect (MD), pattern standard deviation (PSD), glaucoma hemifield test (GHT), and presence or absence of visual field progression.

RESULTS:

Among 230 subjects, the mean age was 65 +/- 14 years, 127 (55%) were female, 161 (70%) were white, and 194 (85%) had a diagnosis of primary open-angle glaucoma (POAG) or suspected POAG. In multivariate generalized estimating equation models, lower corneal hysteresis value (P = .03), but not CCT, was associated with visual field progression. When axial length was included in the model, hysteresis was not a significant risk factor (P = .09). A thinner CCT (P = .02), but not hysteresis, was associated with a higher CDR at the most recent examination. Neither CCT nor hysteresis was associated with MD, PSD, or GHT "outside normal limits."

CONCLUSIONS:

Thinner CCT was associated with the state of glaucoma damage as indicated by CDR. Axial length and corneal hysteresis were associated with progressive field worsening.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

PURPOSE: To evaluate the association between corneal hysteresis and axial length/refractive error among rural Chinese secondary school children. DESIGN: Cross-sectional cohort study. METHODS: Refractive error (cycloplegic auto-refraction with subjective refinement), central corneal thickness (CCT) and axial length (ultrasonic measurement), intraocular pressure (IOP), and corneal hysteresis (Reichert Ocular Response Analyzer) were measured on a rural school-based cohort of children. RESULTS: Among 1,233 examined children, the mean age was 14.7 +/- 0.8 years and 699 (56.7%) were girls. The mean spherical equivalent (n = 1,232) was -2.2 +/- 1.6 diopters (D), axial length (n = 643) was 23.7 +/- 1.1 mm, corneal hysteresis (n = 1,153) was 10.7 +/- 1.6 mm Hg, IOP (n = 1,153) was 17.0 +/- 3.4 mm Hg, and CCT (n = 1,226) was 553 +/- 33 microns. In linear regression models, longer axial length was significantly (P < .001 for both) associated with lower corneal hysteresis and higher IOP. Hysteresis in this population was significantly (P < .001) lower than has previously been reported for normal White children (n = 42, 12.3 +/- 1.3 mm Hg), when adjusting for age and gender. This difference did not appear to depend on differences in axial length between the populations, as it persists when only Chinese children with normal uncorrected vision are included. CONCLUSIONS: Prospective studies will be needed to determine if low hysteresis places eyes at risk for axial elongation secondary or if primary elongation results in lower hysteresis.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

BACKGROUND: We sought to determine whether corneal biomechanical parameters are predictive of reduction in axial length after anti-metabolite trabeculectomy. METHODS: Chinese subjects undergoing trabeculectomy with mitomycin C by a single experienced surgeon underwent the following measurements: Corneal hysteresis (CH, Ocular Response Analyzer, Reichert Ophthalmic Instruments), Goldmann intra-ocular pressure (IOP), central corneal thickness (CCT) and axial length (AL, IOLMaster, Carl Zeiss Meditec, Dublin, CA) were measured pre-operatively, and AL, CH and IOP were measured 1 day and 1 week post-operatively. RESULTS: Mean age of 31 subjects was 52.0 ± 15.2 years, and 15 (48.4%) were female. The mean pre-operative IOP of 21.4 ± 9.3 mmHg was reduced to 8.2 ± 4.6 mmHg 1 day and 11.0 ± 4.4 mmHg 1 week post-operatively (p < 0.001). AL declined from 22.99 ± 0.90 to 22.76 ± 0.87 mm at 1 day and 22.74 ± 0.9 mm at 1 week; 30/31 (%) eyes had a decline in AL (p < 0.001, sign test). In multivariate linear regression models including post-operative data from 1 day and 1 week, greater decline in Goldmann IOP (p < 0.0001, greater pre-op axial length (p < 0.001) and lower pre-operative CH (p = 0.03), were all associated with greater decline in post-operative axial length. CONCLUSIONS: Eyes with lesser ability of the ocular coat to absorb energy (lower CH) had significantly greater decrease in axial length after trabeculectomy-induced IOP-lowering.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

We assessed whether quantitative analysis of Doppler flow velocity waveforms is able to identify subclinical microvascular abnormalities in SLE and whether eigenvector analysis can detect changes not detectable using the resistive index (RI). Fifty-four SLE patients with no conventional cardiovascular risk factors, major organ involvement or retinopathy were compared to 32 controls. Flow velocity waveforms were obtained from the ophthalmic artery (OA), central retinal artery (CRA) and common carotid artery (CA). The waveforms were analysed using eigenvector decomposition and compared between groups at each arterial site. The RI was also determined. The RI was comparable between groups. In the OA and CRA, there were significant differences in the lower frequency sinusoidal components (P <0.05 for each component). No differences were apparent in the CA between groups. Eigenvector analysis of Doppler flow waveforms, recorded in proximity of the terminal vascular bed, identified altered ocular microvascular haemodynamics in SLE. Altered waveform structure could not be identified by changes in RI, the traditional measure of downstream vascular resistance. This analytical approach to waveform analysis is more sensitive in detecting preclinical microvascular abnormalities in SLE. It may hold potential as a useful tool for assessing disease activity, response to treatment, and predicting future vascular complications.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

PURPOSE: To consider whether STZ-induced hyperglycemia renders rat retinal function and ocular blood flow more susceptible to acute intraocular pressure (IOP) challenge.

METHODS: Retinal function (electroretinogram, ERG) was measured during acute IOP challenge (10-100 mmHg, 5 mmHg increments, 3 min/step, vitreal cannulation) in adult Long-Evans rats (6-week old, citrate: n=6, STZ: n=10) 4 weeks after citrate buffer or streptozotocin (STZ, 65 mg/kg, blood glucose > 15 mmol/l) injection. At each IOP, dim and bright flash (-4.56, -1.72 log cd.s.m^-2) ERG responses were recorded to measure inner retinal and ON-bipolar cell function, respectively. Ocular blood flow (laser Doppler flowmetry, citrate; n=6, STZ; n=10) was also measured during acute IOP challenge. Retinae were isolated for qPCR analysis of nitric oxide synthase mRNA expression endothelial, eNos; inducible, iNos; neuronal, nNos).

RESULTS: STZ-induced diabetes increased the susceptibility of inner retinal (IOP at 50% response, 60.1, CI: 57.0-62.0 mmHg vs. citrate: 67.5, CI: 62.1-72.4 mmHg) and ON-bipolar cell function (STZ: 60.3, CI: 58.0-62.8 mmHg vs. citrate: 65.1, CI: 58.0-62.78 mmHg) and ocular blood flow (43.9, CI: 40.8-46.8 vs. citrate: 53.4, CI: 50.7-56.1 mmHg) to IOP challenge. Citrate eyes showed elevated eNos mRNA (+49.7%) after IOP stress, an effect not found in STZ-diabetic eyes (-5.7%, P<0.03). No difference was observed for iNos or nNos (P>0.05) following IOP elevation.

CONCLUSIONS: STZ-induced diabetes increased functional susceptibility during acute IOP challenge. This functional vulnerability is associated with a reduced capacity for diabetic eyes to upregulate eNOS expression and to autoregulate blood flow in response to stress.