280 resultados para Electronic Portal Imaging

em QUB Research Portal - Research Directory and Institutional Repository for Queen's University Belfast


Relevância:

100.00% 100.00%

Publicador:

Resumo:

The aim of this study is to compare the positioning accuracy at different gantry angles of two electronic portal imaging devices (EPIDs) support arm systems by using EPID difference images as a measure for displacement. This work presents a comparison of the mechanical performance of eight Varian aS500 (Varian Medical Systems, Palo Alto, CA) EPIDs, mounted using either the Varian Exact-arm or R-arm.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

This study was carried out to investigate whether the electronic portal imaging (EPI) acquisition process could be optimized, and as a result tolerance and action levels be set for the PIPSPro QC-3V phantom image quality assessment. The aim of the optimization process was to reduce the dose delivered to the patient while maintaining a clinically acceptable image quality. This is of interest when images are acquired in addition to the planned patient treatment, rather than images being acquired using the treatment field during a patient's treatment. A series of phantoms were used to assess image quality for different acquisition settings relative to the baseline values obtained following acceptance testing. Eight Varian aS500 EPID systems on four matched Varian 600C/D linacs and four matched Varian 2100C/D linacs were compared for consistency of performance and images were acquired at the four main orthogonal gantry angles. Images were acquired using a 6 MV beam operating at 100 MU min(-1) and the low-dose acquisition mode. Doses used in the comparison were measured using a Farmer ionization chamber placed at d(max) in solid water. The results demonstrated that the number of reset frames did not have any influence on the image contrast, but the number of frame averages did. The expected increase in noise with corresponding decrease in contrast was also observed when reducing the number of frame averages. The optimal settings for the low-dose acquisition mode with respect to image quality and dose were found to be one reset frame and three frame averages. All patients at the Northern Ireland Cancer Centre are now imaged using one reset frame and three frame averages in the 6 MV 100 MU min(-1) low-dose acquisition mode. Routine EPID QC contrast tolerance (+/-10) and action (+/-20) levels using the PIPSPro phantom based around expected values of 190 (Varian 600C/D) and 225 (Varian 2100C/D) have been introduced. The dose at dmax from electronic portal imaging has been reduced by approximately 28%, and while the image quality has been reduced, the images produced are still clinically acceptable.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

BACKGROUND AND PURPOSE: To describe the clinical implementation of dynamic multileaf collimation (DMLC). Custom compensated four-field treatments of carcinoma of the bladder have been used as a simple test site for the introduction of intensity modulated radiotherapy.MATERIALS AND METHODS: Compensating intensity modulations are calculated from computed tomography (CT) data, accounting for scattered, as well as primary radiation. Modulations are converted to multileaf collimator (MLC) leaf and jaw settings for dynamic delivery on a linear accelerator. A full dose calculation is carried out, accounting for dynamic leaf and jaw motion and transmission through these components. Before treatment, a test run of the delivery is performed and an absolute dose measurement made in a water or solid water phantom. Treatments are verified by in vivo diode measurements and real-time electronic portal imaging. RESULTS: Seven patients have been treated using DMLC. The technique improves dose homogeneity within the target volume, reducing high dose areas and compensating for loss of scatter at the beam edge. A typical total treatment time is 20 min. CONCLUSIONS: Compensated bladder treatments have proven an effective test site for DMLC in an extremely busy clinic.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Objective: Cardiac irradiation during left-sided breast radiotherapy may lead to
deleterious cardiac side effects. Using image guided radiotherapy, it is possible
to exclude the heart from treatment fields and monitor reproducibility of virtual simulation (VS) fields at treatment delivery using electronic portal imaging (EPI). Retrospectively, we evaluate the incidence of cardiac irradiation at VS and subsequent unintended cardiac irradiation during treatment.

Methods: Patients receiving left-sided radiotherapy to the breast or chest wall,
treated with a glancing photon field technique during a four-month period, were
included. VS images and EPIs during radiotherapy delivery were visually assessed.
The presence of any portion of the heart within the treatment field at VS or during treatment was recorded. Central lung distance and maximum heart distance were recorded.

Results: Of 128 patients, 45 (35.1%) had any portion of the heart within the
planned treatment field. Of these, inclusion of the heart was clinically unavoidable in 25 (55.6%). Of those with no heart included in the treatment fields at VS, 41 (49.4%) had presence of the heart as assessed on EPI during treatment.

Conclusion: Unintended cardiac irradiation during left-sided breast radiotherapy treatment occurs in a sizeable proportion of patients.

Advances in knowledge: Despite the use of three-dimensional computed tomography simulation and cardiac shielding, sizeable proportions of patients receiving left-sided breast cancer radiotherapy have unintended cardiac irradiation.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

This work investigated the differences between multileaf collimator (MLC) positioning accuracy determined using either log files or electronic portal imaging devices (EPID) and then assessed the possibility of reducing patient specific quality control (QC) via phantom-less methodologies. In-house software was developed, and validated, to track MLC positional accuracy with the rotational and static gantry picket fence tests using an integrated electronic portal image. This software was used to monitor MLC daily performance over a 1 year period for two Varian TrueBeam linear accelerators, with the results directly compared with MLC positions determined using leaf trajectory log files. This software was validated by introducing known shifts and collimator errors. Skewness of the MLCs was found to be 0.03 ± 0.06° (mean ±1 standard deviation (SD)) and was dependent on whether the collimator was rotated manually or automatically. Trajectory log files, analysed using in-house software, showed average MLC positioning errors with a magnitude of 0.004 ± 0.003 mm (rotational) and 0.004 ± 0.011 mm (static) across two TrueBeam units over 1 year (mean ±1 SD). These ranges, as indicated by the SD, were lower than the related average MLC positioning errors of 0.000 ± 0.025 mm (rotational) and 0.000 ± 0.039 mm (static) that were obtained using the in-house EPID based software. The range of EPID measured MLC positional errors was larger due to the inherent uncertainties of the procedure. Over the duration of the study, multiple MLC positional errors were detected using the EPID based software but these same errors were not detected using the trajectory log files. This work shows the importance of increasing linac specific QC when phantom-less methodologies, such as the use of log files, are used to reduce patient specific QC. Tolerances of 0.25 mm have been created for the MLC positional errors using the EPID-based automated picket fence test. The software allows diagnosis of any specific leaf that needs repair and gives an indication as to the course of action that is required.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Background

]In modern radiotherapy, it is crucial to monitor the performance of all linac components including gantry, collimation system and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method has been introduced in conjunction with an algorithm to investigate the stability of these systems during arc treatments with the aim of ensuring the accuracy of linac mechanical performance.


Methods

The Varian EPID sag, gantry sag, changes in source-to-detector distance (SDD), EPID and collimator skewness, EPID tilt, and the sag in MLC carriages as a result of linac rotation were separately investigated by acquisition of EPID images of a simple phantom comprised of 5 ball-bearings during arc delivery. A fast and robust software package was developed for automated analysis of image data. Twelve Varian linacs of different models were investigated.


Results

The average EPID sag was within 1 mm for all tested linacs. All machines showed less than 1 mm gantry sag. Changes in SDD values were within 1.7 mm except for three linacs of one centre which were within 9 mm. Values of EPID skewness and tilt were negligible in all tested linacs. The maximum sag in MLC leaf bank assemblies was around 1 mm. The EPID sag showed a considerable improvement in TrueBeam linacs.


Conclusion

The methodology and software developed in this study provide a simple tool for effective investigation of the behaviour of linac components with gantry rotation. It is reproducible and accurate and can be easily performed as a routine test in clinics.




Relevância:

100.00% 100.00%

Publicador:

Resumo:

The authors discuss the imaging properties and transfer of amplitude and phase-modulated signals through a phase conjugating lens (PCL). The authors outline the mechanisms of the near-field and far-field subwavelength imaging of Hertzian dipole sources using PCL, particularly the authors show that one-dimensional subwavelength resolution of multiple sources is possible in the far-field using a PCL augmented with specially designed scatterers located in both the adjacent vicinity of the sources and in the mirror symmetric positions in the image plane. These scatterers enable evanescent-to-propagating spectrum and its dual, propagating-to-evanescent, field conversion. Thus, the subwavelength information encoded into propagating waves on the source side can be extracted on the image side. Next, for the first time the transfer of amplitude and phase modulated signals through a PCL augmented with evanescent-to-propagating spectrum conversion is discussed and it has been demonstrated that multiple amplitude or phase modulated dipole sources can be distinguished in the far-field with subwavelength resolution without the necessity for numerical post-processing of the received data. From the study conducted here, it is concluded that a system of transmitters/receivers augmented with a PCL and appropriate scatterers operates without the need for any numerical processing of the receive data in order to separate channel information from very close proximity stations.

Relevância:

90.00% 90.00%

Publicador:

Resumo:

Margins are used in radiotherapy to assist in the calculation of planning target volumes. These margins can be determined by analysing the geometric uncertainties inherent to the radiotherapy planning and delivery process. An important part of this process is the study of electronic portal images collected throughout the course of treatment. Set-up uncertainties were determined for prostate radiotherapy treatments at our previous site and the new purpose-built centre, with margins determined using a number of different methods. In addition, the potential effect of reducing the action level from 5 mm to 3 mm for changing a patient set-up, based on off-line bony anatomy-based portal image analysis, was studied. Margins generated using different methodologies were comparable. It was found that set-up errors were reduced following relocation to the new centre. Although a significant increase in the number of corrections to a patient's set-up was predicted if the action level was reduced from 5 mm to 3 mm, minimal reduction in patient set-up uncertainties would be seen as a consequence. Prescriptive geometric uncertainty analysis not only supports calculation and justification of the margins used clinically to generate planning target volumes, but may also best be used to monitor trends in clinical practice or audit changes introduced by new equipment, technology or practice. Simulations on existing data showed that a 3 mm rather than a 5 mm action level during off-line, bony anatomy-based portal imaging would have had a minimal benefit for the patients studied in this work.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

As the state of the art for high power laser systems increases from terawatt to petawatt level and beyond, a crucial parameter for routinely monitoring high intensity performance is laser spot size on a solid target during an intense interaction in the tight focus regime ( 10(19) Wcm(-2) is demonstrated experimentally and shown to provide the basis for an effective focus diagnostic. Importantly, this technique is also shown to allow in-situ diagnosis of focal spot quality achieved after reflection from a double plasma mirror setup for very intense high contrast interactions (> 10(20) Wcm(-2)) an important application for the field of high laser contrast interaction science.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

Passive equipments operating in the 30-300 GHZ (millimeter wave) band are compared to those in the 300 GHz-3 THz (submillimeter band). Equipments operating in the submillimeter band can measure distance and also spectral information and have been used to address new opportunities in security. Solid state spectral information is available in the submillimeter region making it possible to identify materials, whereas in millimeter region bulk optical properties determine the image contrast. The optical properties in the region from 30 GHz to 3 THz are discussed for some typical inorganic and organic solids. in the millimeter-wave region of the spectrum, obscurants such as poor weather, dust, and smoke can be penetrated and useful imagery generated for surveillance. in the 30 GHZ-3 THZ region dielectrics such as plastic and cloth are also transparent and the detection of contraband hidden under clothing is possible. A passive millimeter-wave imaging concept based on a folded Schmidt camera has been developed and applied to poor weather navigation and security. The optical design uses a rotating mirror and is folded using polarization techniques. The design is very well corrected over a wide field of view making it ideal for surveillance, and security. This produces a relatively compact imager which minimizes the receiver count.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

Charge transfer is a subfemtosecond process in molecules that creates chemical and electronic structure changes. At the quantum level the process can be coherently controlled by ultrashort light pulses. We show how the charge transfer process can be manipulated using a combination of dynamic and static fields and predict how this can be observed experimentally by imaging with photoionization.

Relevância:

40.00% 40.00%

Publicador:

Resumo:

X-ray backscatter imaging can be used for a wide range of imaging applications, in particular for industrial inspection and portal security. Currently, the application of this imaging technique to the detection of landmines is limited due to the surrounding sand or soil strongly attenuating the 10s to 100s of keV X-rays required for backscatter imaging. Here, we introduce a new approach involving a 140 MeV short-pulse (< 100 fs) electron beam generated by laser wakefield acceleration to probe the sample, which produces Bremsstrahlung X-rays within the sample enabling greater depths to be imaged. A variety of detector and scintillator configurations are examined, with the best time response seen from an absorptive coated BaF2 scintillator with a bandpass filter to remove the slow scintillation emission components. An X-ray backscatter image of an array of different density and atomic number items is demonstrated. The use of a compact laser wakefield accelerator to generate the electron source, combined with the rapid development of more compact, efficient and higher repetition rate high power laser systems will make this system feasible for applications in the field.