188 resultados para Vaginal Ring


Relevância:

100.00% 100.00%

Publicador:

Resumo:

Background

Product adherence is a pivotal issue in the development of effective vaginal microbicides to reduce sexual transmission of HIV. To date, the six Phase III studies of vaginal gel products have relied primarily on self-reporting of adherence. Accurate and reliable methods for monitoring user adherence to microbicide-releasing vaginal rings have yet to be established.

Methods

A silicone elastomer vaginal ring prototype containing an embedded, miniature temperature logger has been developed and tested in vitro and in cynomolgus macaques for its potential to continuously monitor environmental temperature and accurately determine episodes of ring insertion and removal.

Results

In vitro studies demonstrated that DST nano-T temperature loggers encapsulated in medical grade silicone elastomer were able to accurately and continuously measure environmental temperature. The devices responded quickly to temperature changes despite being embedded in different thickness of silicone elastomer. Prototype vaginal rings measured higher temperatures compared with a subcutaneously implanted device, showed high sensitivity to diurnal fluctuations in vaginal temperature, and accurately detected periods of ring removal when tested in macaques.

Conclusions

Vaginal rings containing embedded temperature loggers may be useful in the assessment of product adherence in late-stage clinical trials.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Administration of biomacromolecular drugs in effective quantities from conventional vaginal rings is hampered by poor drug permeability in the polymers from which rings are commonly constructed. Here, we report the formulation development and testing of rod insert rings for sustained release of the candidate antiretroviral peptides T-1249 and JNJ54310516-AFP (JNJ peptide), both of which have potential as HIV microbicides. Rod inserts were prepared comprising antiviral peptides T-1249 or JNJ peptide in combination with a hydrophilic excipient (sodium chloride, sodium glutamate, lactose or zinc acetate) dispersed at different loadings within a medical grade silicone elastomer. The inserts were tested for weight change and swelling when immersed in simulated vaginal fluid (SVF). Dye migration into the inserts was also assessed visually over 28 days. In vitro release of T-1249 and JNJ peptide from rings containing various insert types was tested. Weight change and degree of swelling of rods immersed in SVF was dependent on the type and concentration of excipient present. The rods displayed the following rank order in terms of weight change: sodium glutamate > zinc acetate ≈ sodium chloride > lactose. The weight change and degree of swelling of the inserts did not correlate with the level of dye uptake observed. In vitro release of T-1249 was improved through addition of lactose, sodium chloride and sodium glutamate, while release of JNJ peptide was improved through addition of sodium chloride or sodium glutamate. Sustained release of hydrophobic peptides can be achieved using a rod insert ring design formulated to include a hydrophilic excipient. Release rates were dependent upon the type of excipient used. The degree of release improvement with different inserts partially reflects their ability to imbibe surrounding fluid and swell in aqueous environments.

Relevância:

70.00% 70.00%

Publicador:

Resumo:

Silicone elastomer vaginal rings are currently being pursued as a controlled-release strategy for delivering microbicidal substances for the prevention of heterosexual transmission of HIV. Although it is well established that the distribution of drugs in delivery systems influences the release characteristics, in practice the distribution is often difficult to quantify in-situ. Therefore, the aim of this work was to determine whether Raman spectroscopy might provide a rapid, non-contact means of measuring the concentrations of the lead candidate HIV microbicide TMC120 in a silicone elastomer reservoir-type vaginal ring. Vaginal rings loaded with TMC120 were manufactured and sectioned before either Raman mapping an entire ring cross-section (100 µm resolution) or running line scans at appropriate time intervals up to 30 h after manufacture. The results demonstrated that detectable amounts of TMC120, above the silicone elastomer saturation concentration, could be detected up to 1 mm into the sheath, presumably as a consequence of permeation and subsequent reprecipitation. The extent of permeation was found to be similar in rings manufactured at 25 and 80°C.

Relevância:

70.00% 70.00%

Publicador:

Resumo:

Vaginal ring devices capable of providing sustained/controlled release of incorporated actives are already marketed for steroidal contraception and estrogen replacement therapy. In recent years, there has been considerable interest in developing similar ring devices for the administration of microbicidal compounds to prevent vaginal HIV transmission. Intended to be worn continuously, such coitally independent micro- bicide rings are being developed to maintain effective vaginal microbicide concentrations over many weeks or months, thereby overcoming issues around timing of product application, user compliance and acceptability associated with more conventional semi-solid formulations. In this article, an overview of vaginal ring technologies is presented, followed by a review of recent advances and issues pertaining to their application for the delivery of HIV microbicides. This article forms part of a special supplement on presentations covering intravaginal rings, based on the symposium “Trends in Microbicide Formulations”, held on 25 and 26 January 2010, Arlington, VA.

Relevância:

70.00% 70.00%

Publicador:

Resumo:

A major goal in vaccine development is elimination of the ‘cold chain’, the transport and storage system for maintenance and distribution of the vaccine product. This is particularly pertinent to liquid formulation of vaccines. We have previously described the rod-insert vaginal ring (RiR) device, comprising an elastomeric body into which are inserted lyophilised, rod-shaped, solid drug dosage forms, and having potential for sustained mucosal delivery of biomacromolecules, such as HIV envelope protein-based vaccine candidates. Given the solid, lyophilised nature of these insert dosage forms, we hypothesised that antigen stability may be significantly increased compared with more conventional solubilised vaginal gel format. In this study, we prepared and tested vaginal ring devices fitted with lyophilised rod inserts containing the model antigen bovine serum albumin (BSA). Both the RiRs and the gels that were freeze-dried to prepare the inserts were evaluated for BSA stability using PAGE, turbidimetry, microbial load, MALDI-TOF and qualitative precipitate solubility measurements. When stored at 4 oC, but not when stored at 40 oC / 75% RH, the RiR formulation offered protection against structural and conformational changes to BSA. The insert also retained matrix integrity and release characteristics. The results demonstrate that lypophilised gels can provide relative protection against degradation at lower temperatures compared to semi-solid gels. The major mechanism of degradation at 40 oC / 75% RH was shown to be protein aggregation. Finally, in a preliminary study, we found that addition of trehalose to the formulation significantly reduces the rate of BSA degradation as compared to the original formulation when stored at 40 oC /75% RH. Establishing the mechanism of degradation, and finding that degradation is decelerated in the presence of trehalose, will help inform further development of RiRs specifically and polymer based freeze-dried systems in general.

Relevância:

70.00% 70.00%

Publicador:

Resumo:

Following the successful development of long-acting steroid-releasing vaginal ring devices for treatment of menopausal symptoms and contraception, there is now considerable interest in applying similar devices to the controlled release of microbicides against HIV. In this review article, the vaginal ring concept is first considered within the wider context of the early advances in controlled release technology, before describing the various types of ring device available today. The remainder of the article highlights the key developments in HIV microbicide-releasing vaginal rings, with a particular focus on the dapivirine ring that is presently in late-stage clinical testing. © 2012 Malcolm et al, publisher and licensee Dove Medical Press Ltd.

Relevância:

70.00% 70.00%

Publicador:

Resumo:

Objectives: This study measured and compared the pharmacokinetics of CMPD167, a small molecule antiretro- viral CCR5 inhibitor with potential as an HIV microbicide, following vaginal, rectal and oral administration in rhe- sus macaques.
Methods: A vaginal hydroxyethylcellulose (HEC) gel, a rectal HEC gel, a silicone elastomer matrix-type vaginal ring and an oral solution, each containing CMPD167, were prepared and administered to rhesus macaques pretreated with Depo-Provera. CMPD167 concentrations in vaginal fluid, vaginal tissue (ring only), rectal fluid and blood plasma were quantified by HPLC–mass spectrometry.
Results: CMPD167 concentrations measured in rectal fluid, vaginal fluid and blood plasma were highly depend- ent on both the route of administration and the formulation type. Although rectal and vaginal fluid concentra- tions were highest when CMPD167 was administered locally (via either gel or ring), lower concentrations of the drug were also measured in these compartments following administration at the remote mucosal site or orally. CMPD167 levels in the vaginal and rectal fluid following oral administration were relatively low compared with local administration.
Conclusions: The study provides clear evidence for vaginal – rectal and rectal – vaginal drug transfer pathways and suggests that oral pre-exposure prophylaxis with CMPD167 may be less efficacious at preventing sexual trans- mission of HIV-1 than topically applied products.

Relevância:

70.00% 70.00%

Publicador:

Resumo:

Background: Following progress of the dapivirine (DPV)-releasing silicone elastomer (SE) vaginal ring (VR) into Phase III clinical studies, there is now interest in developing next-generation rings that additionally provide contraception. Levonorgestrel (LNG) is a safe and effective progestin that is being widely considered for use as a hormonal contraceptive agent in future multipurpose prevention technology (MPT) products. Although LNG has previously been incorporated into various controlled release SE devices, minimal attention has focused on its propensity to irreversibly react with addition cure SE systems. Here, for the first time, we investigate this LNG binding phenomenon and outline strategies for overcoming it.
Methods: VRs containing various loadings of DPV and LNG were manufactured and in vitro release assessed. Different LNG-only SE samples were also prepared to assess the following parameters: (i) addition cure vs. condensation cure SEs; (ii) different types of addition cure SEs; (iii) mixing time, (iv) cure temperature, (v) cure time; and (vi) LNG particle size. After manufacture, the LNG-only samples were assayed for total drug content using a solvent extraction method. The SE curing reaction and the LNG binding reaction was probed using nuclear magnetic resonance (NMR) spectroscopy. Results:
Under certain drug/formulation/processing conditions, LNG was not recoverable from VRs. Further studies using non-ring samples showed that: (a) the phenomenon was only observed with addition cure SEs (and not condensation cure SEs); (b) the extent of binding was dependent upon the type of addition cure SE; (c) micronised LNG showed significantly greater binding than non-micronised LNG; (d) the extent of binding correlated with increased mixing time, cure time and cure temperature.
Conclusions: Careful control of the API characteristics, the SE composition, and the manufacturing conditions will be necessary to establish a practical VR formulation for controlled release of LNG.

Relevância:

70.00% 70.00%

Publicador:

Resumo:

A matrix-type silicone elastomer vaginal ring providing 28-day continuous release of dapivirine (DPV) - a lead candidate human immunodeficiency virus type 1 (HIV-1) microbicide compound - has recently demonstrated moderate levels of protection in two Phase III clinical studies. Here, next-generation matrix and reservoir-type silicone elastomer vaginal rings are reported for the first time offering simultaneous and continuous in vitro release of DPV and the contraceptive progestin levonorgestrel (LNG) over a period of between 60 and 180days. For matrix-type vaginal rings comprising initial drug loadings of 100, 150 or 200mg DPV and 0, 16 or 32mg LNG, Day 1 daily DPV release values were between 4132 and 6113μg while Day 60 values ranged from 284 to 454μg. Daily LNG release ranged from 129 to 684μg on Day 1 and 2-91μg on Day 60. Core-type rings comprising one or two drug-loaded cores provided extended duration of in vitro release out to 180days, and maintained daily drug release rates within much narrower windows (either 75-131μg/day or 37-66μg/day for DPV, and either 96-150μg/day or 37-57μg/day for LNG, depending on core ring configuration and ignoring initial lag release effect for LNG) compared with matrix-type rings. The data support the continued development of these devices as multi-purpose prevention technologies (MPTs) for HIV prevention and long-acting contraception.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

The in vitro release characteristics of eight low-molecular-weight drugs (clindamycin, 17beta-estradiol, 17beta-estradiol-3-acetate, 17beta-estradiol diacetate, metronidazole, norethisterone, norethisterone acetate and oxybutynin) from silicone matrixtype intravaginal rings of various drug loadings have been evaluated under sink conditions. Through modelling of the release data using the Higuchi equation, and determination of the silicone solubility of the drugs, the apparent silicone elastomer diffusion coefficients of the drugs have been calculated. Furthermore, in an attempt to develop a quantitative model for predicting release rates of new drug substances from these vaginal ring devices, it has been observed that linear relationships exist between the log of the silicone solubility of the drug (mg ml(-1)) and the reciprocal of its melting point (K-1) (y = 3.558x - 9.620, R = 0.77), and also between the log of the diffusion coefficient (cm(2) s(-1)) and the molecular weight of the drug molecule (g mol(-1)) (y = - 0.0068x - 4.0738, R = 0.95). Given that the silicone solubility and silicone diffusion coefficient are the major parameters influencing the permeation of drugs through silicone elastomers, it is now possible to predict through use of the appropriate mathematical equations both matrix-type and reservoir-type intravaginal ring release rates simply from a knowledge of drug melting temperature and molecular weight. (C) 2003 Elsevier Science B.V. All rights reserved.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

BACKGROUND: In vitro release testing of vaginal formulations is usually performed in a one-compartment model (OCM) where the release medium, usually comprising pH-adjusted water, an aqueous surfactant solution or a solvent-water solution, provides sink conditions throughout the release experiment. Although this model is useful in evaluating the effect of formulation parameters upon release, it rarely reflects in vivo conditions. Here we report use of a two-compartment diffusion cell model (TCM, comprising a small volume donor, a large volume receptor, and separated by a model epithelial membrane) to more closely mimic in vivo vaginal release and tissue absorption following administration of a UC781 vaginal ring.

METHODS: Macaque-sized matrix silicone elastomer vaginal rings containing 100mg UC781 were prepared by injection molding, and in vitro release testing performed using both OCM (20mL simulated vaginal fluid, SVF) and TCM (5mL SVF in donor cell and variable quantities of Tween 80; silicone elastomer membrane; 100mL 3:2 ethanol/water in receptor cell). In the TCM, drug levels were measured by HPLC in both donor and receptor cells, representing fluid and tissue levels respectively. Rings containing 100mg UC781 and 10% w/w Tween 80 were also manufactured and tested.

RESULTS: The amount of UC781 released from rings was significantly influenced by the choice of release model. Greatest release (56mg/14 days) was measured in the ethanol/water OCM, compared with no measurable release into SVF only. Increasing the concentration of Tween 80 in the SVF medium (1, 3 and 5% w/w) led to increased UC781 release (11, 16 and 18mg, respectively), demonstrating that vaginal fluid solubility of UC781 may be rate-determining in vivo. In the TCM, UC781 accumulates in the receptor cell medium over time, despite not being measured in the donor medium containing the ring device. Incorporation of Tween 80 directly into the ring provided enhanced release in both donor and receptor cells.

CONCLUSIONS: Release of UC781 was influenced by the choice of release medium and the inclusion of Tween 80 in the ring. Although use of SVF-only in the OCM indicated no measurable UC781 release from rings, data from the TCM confirms that UC781 is not only released but is also capable of penetrating across the model epithelial membrane. The TCM may therefore provide a more representative in vitro release model for mimicking in vivo absorption.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

There is an urgent global need for preventative strategies against HIV-1 infections. Llama heavy-chain antibody fragments (VHH) are a class of molecules recently described as potent cross-clade HIV-1 entry inhibitors. We studied the potential of a VHH-based microbicide in an application-oriented fashion. We show that VHH can be inexpensively produced in high amounts in the GRAS organism S. cerevisiae, resulting in very pure, and endotoxin free product. VHH are very stable under conditions they might encounter during transport, storage or use by women. We developed active formulations of VHH in aqueous gel and compressed and lyophilized tablets for controlled release from an intra vaginal device. The release profile of the VHH from e.g. a vaginal ring suggests sufficient bioavailability and protective concentration of the molecule at the mucosal site at the moment of the infection. The ex vivo penetration kinetics through human tissues show that the VHH diffuse into the mucosal layer and open the possibility to create a second defense layer either by blocking the HIV receptor binding sites or by blocking the receptors of immune cells in the mucosa. In conclusion, our data show that VHH have

Relevância:

60.00% 60.00%

Publicador:

Resumo:

Background: Combination drug products can display thermal behaviour that is more complex than for the corresponding single drug products. For example, the contraceptive vaginal ring (VR) Nuvaring contains a eutectic (lowest melting) composition of etonogestrel (ETN) and ethinyl estradiol. Here we report the predisposition of dapivirine (DPV) to form reduced melting/eutectic mixtures when combined with other contraceptive hormones and antiretrovirals, and discuss the implications for development of combination microbicide and multipurpose prevention technology (MPT) products.
Methods: Binary mixtures of DPV with darunavir (DRV), levonorgestrel (LNG), ETN or maraviroc (MVC) were prepared either by physical mixing or by solvent evaporation. Selected binary mixtures were also incorporated into silicone elastomer (SE) VR devices. Thermal behavior of the mixtures was analyzed using differential scanning calorimetry (DSC) operating in standard heating ramp mode (10 °C/min). DSC data were used to construct two component phase diagrams for each binary system.
Results: Drug mixtures typically showed reduced melting transitions for both drug components, with clear evidence for a eutectic mixture at a well-defined intermediate composition. Eutectic temperatures and compositions for the various mixtures were: 40% DPV / 60% ETN - 170°C; 25% DPV / 75% MVC - 172°C; 65% DPV / 35% LNG - 192°C. In each case, the eutectic composition was also detected when the drug mixtures were incorporated into SE VRs. For the DPV/DRV system, the thermal behaviour is complicated by desolvation from the darunavir ethanolate polymorph.
Conclusions: When DPV is combined with small molecular weight hydrophobic drugs, the melting temperature for both drugs is typically reduced to a degree dependent on the composition of the mixture. At specified compositions, a low melting eutectic system results. The formation of eutectic behavior in binary drug systems needs to be carefully characterised in order to define product performance and drug release.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

TMC 120 (Dapivirine) is a potent non-nucleoside reverse transcriptase inhibitor that is presently being developed as a vaginal HIV microbicide. To date, most vaginal microbicides under clinical investigation have been formulated as single-dose semi-solid gels, designed for application to the vagina before each act of intercourse. However, a clear rationale exists for providing long-term, controlled release of vaginal microbicides in order to afford continuous protection against heterosexually transmitted HIV infection and to improve user compliance. In this study we report on the incorporation of various pharmaceutical excipients into TMC 120 silicone, reservoir-type intravaginal rings (IVRs) in order to modify the controlled release characteristics of the microbicide. The results demonstrate that TMC 120 is released in zero-order fashion from the rings over a 28-day period and that release parameters could be modified by the inclusion of release-modifying excipients in the IVR. The hydrophobic liquid excipient isopropyl myristate had little effect on steady-state daily release rates, but did increase the magnitude and duration of burst release in proportion to excipient loading in the IVR. By comparison, the hydrophobic liquid poly(dimethylsiloxane) had little effect on TMC 120 release parameters. A hydrophilic excipient, lactose, had the surprising effect of decreasing TMC 120 burst release while increasing the apparent steady-state daily release in a concentration-dependent manner. Based on previous cell culture data and vaginal physiology, TMC120 is released from the various ring formulations in amounts potentially capable of maintaining a protective vaginal concentration. It is further predicted that the observed release rates may be maintained for at least a period of 1 year from a single ring device. TMC 120 release profiles and the mechanical properties of rings could be modified by the physicochemical nature of hydrophobic and hydrophilic excipients incorporated into the IVRs.

Relevância:

30.00% 30.00%

Publicador:

Resumo:

BACKGROUND: HIV microbicide trials have emphasized the need to evaluate the safety of topical microbicides and delivery platforms in an animal model prior to conducting clinical efficacy trials. An ideal delivery device should provide sustainable and sufficient concentrations of effective products to prevent HIV transmission while not increasing transmission risk by either local mucosal inflammation and/or disruption of the normal vaginal microflora.

METHODS: Safety analyses of macaque-sized elastomeric silicone and polyurethane intravaginal rings (IVRs) loaded with candidate antiretroviral (ARV) drugs were tested in four studies ranging in duration from 49 to 73 days with retention of the IVR being 28 days in each study. Macaques were assigned to 3 groups; blank IVR, ARV-loaded IVR, and naïve. In sequential studies, the same macaques were used but rotated into different groups. Mucosal and systemic levels of cytokines were measured from vaginal fluids and plasma, respectively, using multiplex technology. Changes in vaginal microflora were also monitored. Statistical analysis (Mann-Whitney test) was used to compare data between two groups of unpaired samples (with and without IVR, and IVR with and without ARV) for the groups collectively, and also for individual macaques.

RESULTS: There were few statistically significant differences in mucosal and systemic cytokine levels measured longitudinally when the ring was present or absent, with or without ARVs. Of the 8 proinflammatory cytokines assayed a significant increase (p = 0.015) was only observed for IL8 in plasma with the blank and ARV loaded IVR (median of 9.2 vs. 5.7 pg/ml in the absence of IVR). There were no significant differences in the prevalence of H2O2-producing lactobacilli or viridans streptococci, or other microorganisms indicative of healthy vaginal microflora. However, there was an increase in the number of anaerobic gram negative rods in the presence of the IVR (p= < 0.0001).

CONCLUSIONS: IVRs with or without ARVs neither significantly induce the majority of potentially harmful proinflammatory cytokines locally or systemically, nor alter the lactobacillus or G. vaginalis levels. The increase in anaerobic gram negative rods alone suggests minimal disruption of normal vaginal microflora. The use of IVRs as a long-term sustained delivery device for ARVs is promising and preclinical studies to demonstrate the prevention of transmission in the HIV/SHIV nonhuman primate model should continue.