2 resultados para liquid laser
em QSpace: Queen's University - Canada
Resumo:
Short liquid core waveguides (LCWs) were included into a fiber-loop cavity ring-down absorption spectrometer to reduce the detection limit over, both, single pass absorption in a LCW and cavityenhanced absorption using a conventional fiber-loop cavity. LCWs of 5 and 10 cm length were interfaced with a pressure-flow system and a multimode fiber-loop cavity using concave fiber lenses with matching numerical apertures and diameters. Two red dyes, Allura Red AC and Congo Red, were detected with a 532 nm pulsed laser at a 5 nM limit of detection in a detection volume of less than 1 μL, corresponding to a minimal detectable absorbance of less than 4 × 10−4 cm−1 and a minimal detectable change in absorption cross section, σmin = Vdet × ε × CLOD, of about 14 μm2 (Allura Red AC) and 37 μm2 (Congo Red).
Resumo:
Visible and near-infrared laser light pulses were coupled into two different types of optical fiber cavities. One cavity consisted of a short strand of fiber waveguide that contained two identical fiber Bragg gratings. Another cavity was made using a loop of optical fiber. In either cavity ∼ 40 ps laser pulses, which were generated using a custom-built gainswitched diode laser, circulated for a large number of round trips. The optical loss of either cavity was determined from the ring-down times. Cavity ring-down spectroscopy was performed on 200 pL volumes of liquid samples that were injected into the cavities using a 100 μm gap in the fiber loop. A detection limit of 20 ppm of methylene blue dye in aqueous solution, corresponding to a minimum absorptivity of εC < 6 cm−1, was realized.