4 resultados para Stone, John Osgood, 1813-1873.

em Duke University


Relevância:

20.00% 20.00%

Publicador:

Resumo:

INTRODUCTION: The characterization of urinary calculi using noninvasive methods has the potential to affect clinical management. CT remains the gold standard for diagnosis of urinary calculi, but has not reliably differentiated varying stone compositions. Dual-energy CT (DECT) has emerged as a technology to improve CT characterization of anatomic structures. This study aims to assess the ability of DECT to accurately discriminate between different types of urinary calculi in an in vitro model using novel postimage acquisition data processing techniques. METHODS: Fifty urinary calculi were assessed, of which 44 had >or=60% composition of one component. DECT was performed utilizing 64-slice multidetector CT. The attenuation profiles of the lower-energy (DECT-Low) and higher-energy (DECT-High) datasets were used to investigate whether differences could be seen between different stone compositions. RESULTS: Postimage acquisition processing allowed for identification of the main different chemical compositions of urinary calculi: brushite, calcium oxalate-calcium phosphate, struvite, cystine, and uric acid. Statistical analysis demonstrated that this processing identified all stone compositions without obvious graphical overlap. CONCLUSION: Dual-energy multidetector CT with postprocessing techniques allows for accurate discrimination among the main different subtypes of urinary calculi in an in vitro model. The ability to better detect stone composition may have implications in determining the optimum clinical treatment modality for urinary calculi from noninvasive, preprocedure radiological assessment.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Using a reflector insert, the original HM-3 lithotripter field at 20 kV was altered significantly with the peak positive pressure (p(+)) in the focal plane increased from 49 to 87 MPa while the -6 dB focal width decreased concomitantly from 11 to 4 mm. Using the original reflector, p(+) of 33 MPa with a -6 dB focal width of 18 mm were measured in a pre-focal plane 15-mm proximal to the lithotripter focus. However, the acoustic pulse energy delivered to a 28-mm diameter area around the lithotripter axis was comparable ( approximately 120 mJ). For all three exposure conditions, similar stone comminution ( approximately 70%) was produced in a mesh holder of 15 mm after 250 shocks. In contrast, stone comminution produced by the modified reflector either in a 15-mm finger cot (45%) or in a 30-mm membrane holder (14%) was significantly reduced from the corresponding values (56% and 26%) produced by the original reflector (no statistically significant differences were observed between the focal and pre-focal planes). These observations suggest that a low-pressure/broad focal width lithotripter field will produce better stone comminution than its counterpart with high-pressure/narrow focal width under clinically relevant in vitro comminution conditions.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

BACKGROUND AND PURPOSE: Previous studies have demonstrated that treatment strategy plays a critical role in ensuring maximum stone fragmentation during shockwave lithotripsy (SWL). We aimed to develop an optimal treatment strategy in SWL to produce maximum stone fragmentation. MATERIALS AND METHODS: Four treatment strategies were evaluated using an in-vitro experimental setup that mimics stone fragmentation in the renal pelvis. Spherical stone phantoms were exposed to 2100 shocks using the Siemens Modularis (electromagnetic) lithotripter. The treatment strategies included increasing output voltage with 100 shocks at 12.3 kV, 400 shocks at 14.8 kV, and 1600 shocks at 15.8 kV, and decreasing output voltage with 1600 shocks at 15.8 kV, 400 shocks at 14.8 kV, and 100 shocks at 12.3 kV. Both increasing and decreasing voltages models were run at a pulse repetition frequency (PRF) of 1 and 2 Hz. Fragmentation efficiency was determined using a sequential sieving method to isolate fragments less than 2 mm. A fiberoptic probe hydrophone was used to characterize the pressure waveforms at different output voltage and frequency settings. In addition, a high-speed camera was used to assess cavitation activity in the lithotripter field that was produced by different treatment strategies. RESULTS: The increasing output voltage strategy at 1 Hz PRF produced the best stone fragmentation efficiency. This result was significantly better than the decreasing voltage strategy at 1 Hz PFR (85.8% vs 80.8%, P=0.017) and over the same strategy at 2 Hz PRF (85.8% vs 79.59%, P=0.0078). CONCLUSIONS: A pretreatment dose of 100 low-voltage output shockwaves (SWs) at 60 SWs/min before increasing to a higher voltage output produces the best overall stone fragmentation in vitro. These findings could lead to increased fragmentation efficiency in vivo and higher success rates clinically.