Purpose: To determine likely errors in estimating retinal shape using partial coherence interferometric instruments when no allowance is made for optical distortion. Method: Errors were estimated using Gullstrand’s No. 1 schematic eye and variants which included a 10 D axial myopic eye, an emmetropic eye with a gradient-index lens, and a 10.9 D accommodating eye with a gradient-index lens. Performance was simulated for two commercial instruments, the IOLMaster (Carl Zeiss Meditec) and the Lenstar LS 900 (Haag-Streit AG). The incident beam was directed towards either the centre of curvature of the anterior cornea (corneal-direction method) or the centre of the entrance pupil (pupil-direction method). Simple trigonometry was used with the corneal intercept and the incident beam angle to estimate retinal contour. Conics were fitted to the estimated contours. Results: The pupil-direction method gave estimates of retinal contour that were much too flat. The cornea-direction method gave similar results for IOLMaster and Lenstar approaches. The steepness of the retinal contour was slightly overestimated, the exact effects varying with the refractive error, gradient index and accommodation. Conclusion: These theoretical results suggest that, for field angles ≤30º, partial coherence interferometric instruments are of use in estimating retinal shape by the corneal-direction method with the assumptions of a regular retinal shape and no optical distortion. It may be possible to improve on these estimates out to larger field angles by using optical modeling to correct for distortion.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Purpose: We provide an account of the relationships between eye shape, retinal shape and peripheral refraction. Recent findings: We discuss how eye and retinal shapes may be described as conicoids, and we describe an axis and section reference system for determining shapes. Explanations are given of how patterns of retinal expansion during the development of myopia may contribute to changing patterns of peripheral refraction, and how pre-existing retinal shape might contribute to the development of myopia. Direct and indirect techniques for determining eye and retinal shape are described, and results are discussed. There is reasonable consistency in the literature of eye length increasing at a greater rate than height and width as the degree of myopia increases, so that eyes may be described as changing from oblate/spherical shapes to prolate shapes. However, one study indicates that the retina itself, while showing the same trend, remains oblate in shape for most eyes (discounting high myopia). Eye shape and retinal shape are not the same and merely describing an eye shape as being prolate or oblate is insufficient without some understanding of the parameters contributing to this; in myopia a prolate eye shape is likely to involve both a steepening retina near the posterior pole combined with a flattening (or a reduction in steepening compared with an emmetrope) away from the pole.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Purpose:Race appears to be associated with myopiogenesis, with East Asians showing high myopia prevalence. Considering structural variations in the eye, it is possible that retinal shapes are different between races. The purpose of this study was to quantify and compare retinal shapes between racial groups using peripheral refraction (PR) and peripheral eye lengths (PEL). Methods:A Shin-Nippon SRW5000 autorefractor and a Haag-Streit Lenstar LS900 biometer measured PR and PEL, respectively, along horizontal (H) and vertical (V) fields out to ±35° in 5° steps in 29 Caucasian (CA), 16 South Asian (SA) and 23 East Asian (EA) young adults (spherical equivalent range +0.75D to –5.00D in all groups). Retinal vertex curvature Rv and asphericity Q were determined from two methods: a) PR (Dunne): The Gullstrand-Emsley eye was modified according to participant’s intraocular lengths and anterior cornea curvature. Ray-tracing was performed at each angle through the stop, altering cornea asphericity until peripheral astigmatism matched experimental measurements. Retinal curvature and hence retinal co-ordinate intersection with the chief ray were altered until sagittal refraction matched its measurement. b) PEL: Ray-tracing was performed at each angle through the anterior corneal centre of curvature of the Gullstrand-Emsley eye. Ignoring lens refraction, retinal co-ordinates relative to the fovea were determined from PEL and trigonometry. From sets of retinal co-ordinates, conic retinal shapes were fitted in terms of Rv and Q. Repeated-measures ANOVA were conducted on Rv and Q, and post hoc t-tests with Bonferroni correction were used to compare races. Results:In all racial groups both methods showed greater Rv for the horizontal than for the vertical meridian and greater Rv for myopes than emmetropes. Rv was greater in EA than in CA (P=0.02), with Rv for SA being intermediate and not significantly different from CA and EA. The PEL method provided larger Rv than the PR method: PEL: EA vs CA 87±13 vs 83±11 m-1 (H), 79±13 vs 72±14 m-1 (V); PR: EA vs CA 79±10 vs 67±10 m-1 (H), 71±17 vs 66±12 m-1 (V). Q did not vary significantly with race. Conclusions:Estimates of Rv, but not of Q, varied significantly with race. The greater Rv found in EA than in CA and the comparatively high prevalence rate of myopia in many Asian countries may be related.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

This study concerned development and validation of a simple and inexpensive method involving partial coherence interferometry for measuring retinal shape, and its use in exploring association between retinal shape and myopia. Retinal shapes estimates using partial coherence interferometry were validated against estimates obtained from magnetic resonance imaging. Steeper retinas were found along the horizontal than along the vertical meridian, in myopes than in emmetropes, and in East Asian myopes than in Caucasian myopes. The racial differences, combined with the high prevalence of myopia in East Asia, suggest that retinal shape may play a role in myopia development.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

To validate a simple partial coherence interferometry (PCI) based retinal shape method, estimates of retinal shape were determined in 60 young adults using off-axis PCI, with three stages of modeling using variants of the Le Grand model eye, and magnetic resonance imaging (MRI). Stage 1 and 2 involved a basic model eye without and with surface ray deviation, respectively and Stage 3 used model with individual ocular biometry and ray deviation at surfaces. Considering the theoretical uncertainty of MRI (12-14%), the results of the study indicate good agreement between MRI and all three stages of PCI modeling with <4% and <7% differences in retinal shapes along horizontal and vertical meridians, respectively. Stage 2 and Stage 3 gave slightly different retinal co-ordinates than Stage 1 and we recommend the intermediate Stage 2 as providing a simple and valid method of determining retinal shape from PCI data.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

Purpose: To assess intrasessional and intersessional repeatability of two commercial partial coherence interferometry instruments for measuring peripheral eye lengths and to investigate the agreement between the two instruments. Methods: Central and peripheral eye lengths were determined with the IOLMaster (Carl-Zeiss Meditec AG, Jena, Germany) and the Lenstar (Haag Streit, Bern, Switzerland) in seven adults. Measurements were performed out to 35° and 30° from fixation for horizontal and vertical visual fields, respectively, in 5° intervals. An external fixation target at optical infinity was used. At least four measurements were taken at each location for each instrument, and measurements were taken at two sessions. Results: The mean intrasessional SDs for the IOLMaster along both the horizontal and vertical visual fields were 0.04 ± 0.04 mm; corresponding results for the Lenstar were 0.02 ± 0.02 mm along both fields. The intersessional SDs for the IOLMaster for the horizontal and vertical visual fields were ±0.11 and ±0.08 mm, respectively; corresponding limits for the Lenstar were ±0.05 and ±0.04 mm. The intrasessional and intersessional variability increased away from fixation. The mean differences between the two instruments were 0.01 ± 0.07 mm and 0.02 ± 0.07 mm in the horizontal and vertical visual fields, but the lengths with the Lenstar became greater than those with the IOLMaster as axial length increased (rate of approximately 0.016 mm/mm). Conclusions: Both the IOLMaster and the Lenstar demonstrated good intrasessional and intersessional repeatability for peripheral eye length measurements, with the Lenstar showing better repeatability. The Lenstar would be expected to give a slightly greater range of eye lengths than the IOLMaster across the visual field.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

Purpose The eye rotation approach for measuring peripheral eye length leads to concern about whether the rotation influences results, such as through pressure exerted by eyelids or extra-ocular muscles. This study investigated whether this approach is valid. Methods Peripheral eye lengths were measured with a Lenstar LS 900 biometer for eye rotation and no-eye rotation conditions (head rotation for horizontal meridian and instrument rotation for vertical meridian). Measurements were made for 23 healthy young adults along the horizontal visual field (±30°) and, for a subset of eight participants along the vertical visual field (±25°). To investigate the influence of the duration of eye rotation, for six participants measurements were made at 0, 60, 120, 180 and 210 s after eye rotation to ±30° along horizontal and vertical visual fields. Results Peripheral eye lengths were not significantly different for the conditions along the vertical meridian (F1,7 = 0.16, p = 0.71). The peripheral eye lengths for the conditions were significantly different along the horizontal meridian (F1,22 = 4.85, p = 0.04), although not at individual positions (p ≥ 0.10) and were not important. There were no apparent differences between the emmetropic and myopic groups. There was no significant change in eye length at any position after maintaining position for 210 s. Conclusion Eye rotation and no-eye rotation conditions were similar for measuring peripheral eye lengths along horizontal and vertical visual field meridians at ±30° and ±25°, respectively. Either condition can be used to estimate retinal shape from peripheral eye lengths.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

This paper presents a prototype tracking system for tracking people in enclosed indoor environments where there is a high rate of occlusions. The system uses a stereo camera for acquisition, and is capable of disambiguating occlusions using a combination of depth map analysis, a two step ellipse fitting people detection process, the use of motion models and Kalman filters and a novel fit metric, based on computationally simple object statistics. Testing shows that our fit metric outperforms commonly used position based metrics and histogram based metrics, resulting in more accurate tracking of people.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Summary Generalized Procrustes analysis and thin plate splines were employed to create an average 3D shape template of the proximal femur that was warped to the size and shape of a single 2D radiographic image of a subject. Mean absolute depth errors are comparable with previous approaches utilising multiple 2D input projections. Introduction Several approaches have been adopted to derive volumetric density (g cm-3) from a conventional 2D representation of areal bone mineral density (BMD, g cm-2). Such approaches have generally aimed at deriving an average depth across the areal projection rather than creating a formal 3D shape of the bone. Methods Generalized Procrustes analysis and thin plate splines were employed to create an average 3D shape template of the proximal femur that was subsequently warped to suit the size and shape of a single 2D radiographic image of a subject. CT scans of excised human femora, 18 and 24 scanned at pixel resolutions of 1.08 mm and 0.674 mm, respectively, were equally split into training (created 3D shape template) and test cohorts. Results The mean absolute depth errors of 3.4 mm and 1.73 mm, respectively, for the two CT pixel sizes are comparable with previous approaches based upon multiple 2D input projections. Conclusions This technique has the potential to derive volumetric density from BMD and to facilitate 3D finite element analysis for prediction of the mechanical integrity of the proximal femur. It may further be applied to other anatomical bone sites such as the distal radius and lumbar spine.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

We developed orthogonal least-squares techniques for fitting crystalline lens shapes, and used the bootstrap method to determine uncertainties associated with the estimated vertex radii of curvature and asphericities of five different models. Three existing models were investigated including one that uses two separate conics for the anterior and posterior surfaces, and two whole lens models based on a modulated hyperbolic cosine function and on a generalized conic function. Two new models were proposed including one that uses two interdependent conics and a polynomial based whole lens model. The models were used to describe the in vitro shape for a data set of twenty human lenses with ages 7–82 years. The two-conic-surface model (7 mm zone diameter) and the interdependent surfaces model had significantly lower merit functions than the other three models for the data set, indicating that most likely they can describe human lens shape over a wide age range better than the other models (although with the two-conic-surfaces model being unable to describe the lens equatorial region). Considerable differences were found between some models regarding estimates of radii of curvature and surface asphericities. The hyperbolic cosine model and the new polynomial based whole lens model had the best precision in determining the radii of curvature and surface asphericities across the five considered models. Most models found significant increase in anterior, but not posterior, radius of curvature with age. Most models found a wide scatter of asphericities, but with the asphericities usually being positive and not significantly related to age. As the interdependent surfaces model had lower merit function than three whole lens models, there is further scope to develop an accurate model of the complete shape of human lenses of all ages. The results highlight the continued difficulty in selecting an appropriate model for the crystalline lens shape.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Insulin-like growth factor binding proteins (IGFBPs) are prime regulators of IGF-action in numerous cell types including the retinal pigment epithelium (RPE). The RPE performs several functions essential for vision, including growth factor secretion and waste removal via a phagocytic process mediated in part by vitronectin (Vn). In the course of studying the effects of IGFBPs on IGF-mediated VEGF secretion and Vn-mediated phagocytosis in the RPE cell line ARPE-19, we have discovered that these cells avidly ingest synthetic microspheres (2.0 μm diameter) coated with IGFBPs. Given the novelty of this finding and the established role for endocytosis in mediating IGFBP actions in other cell types, we have explored the potential role of candidate cell surface receptors. Moreover, we have examined the role of key IGFBP structural motifs, by comparing responses to three members of the IGFBP family (IGFBP-3, IGFBP-4 and IGFBP-5) which display overlapping variations in primary structure and glycosylation status. Coating of microspheres (FluoSpheres®, sulfate modified polystyrene filled with a fluorophore) was conducted at 37 °C for 1 h using 20 μg/mL of test protein, followed by extensive washing. Binding of proteins was confirmed using a microBCA assay. The negative control consisted of microspheres treated with 0.1% bovine serum albumin (BSA), and all test samples were post-treated with BSA in an effort to coat any remaining free protein binding sites, which might otherwise encourage non-specific interactions with the cell surface. Serum-starved cultures of ARPE-19 cells were incubated with microspheres for 24 h, using a ratio of approximately 100 microspheres per cell. Uptake of microspheres was quantified using a fluorometer and was confirmed visually by confocal fluorescence microscopy. The ARPE-19 cells displayed little affinity for BSA-treated microspheres, but avidly ingested large quantities of those pre-treated with Vn (ANOVA; p < 0.001). Strong responses were also observed towards recombinant formulations of non-glycosylated IGFBP-3, glycosylated IGFBP-3 and glycosylated IGFBP-5 (all p < 0.001), while glycosylated IGFBP-4 induced a relatively minor response (p < 0.05). The response to IGFBP-3 was unaffected in the presence of excess soluble IGFBP-3, IGF-I or Vn. Likewise, soluble IGFBP-3 did not induce uptake of BSA-treated microspheres. Antibodies to either the transferrin receptor or type 1 IGF-receptor displayed slight inhibitory effects on responses to IGFBPs and Vn. Heparin abolished responses to Vn, IGFBP-5 and non-glycosylated IGFBP-3, but only partially inhibited the response to glycosylated IGFBP-3. Our results demonstrate for the first time IGFBP-mediated endocytosis in ARPE-19 cells and suggest roles for the IGFBP-heparin-binding domain and glycosylation status. These findings have important implications for understanding the mechanisms of IGFBP actions on the RPE, and in particular suggest a role for IGFBP-endocytosis.

Relevância:

20.00% 20.00%

Publicador:

Resumo:

Purpose: To investigate associations between the diurnal variation in a range of corneal parameters, including anterior and posterior corneal topography, and regional corneal thickness. ----- Methods: Fifteen subjects had their corneas measured using a rotating Scheimpflug camera (Pentacam) every 3-7 hours over a 24-hour period. Anterior and posterior corneal axial curvature, pachymetry and anterior chamber depth were analysed. The best fitting corneal sphero-cylinder from the axial curvature, and the average corneal thickness for a series of different corneal regions were calculated. Intraocular pressure and axial length were also measured at each measurement session. Repeated measures ANOVA were used to investigate diurnal change in these parameters. Analysis of covariance was used to examine associations between the measured ocular parameters. ----- Results: Significant diurnal variation was found to occur in both the anterior and posterior corneal curvature and in the regional corneal thickness. Flattening of the anterior corneal best sphere was observed at the early morning measurement (p < 0.0001). The posterior cornea also underwent a significant steepening (p < 0.0001) and change in astigmatism 90/180° at this time. A significant swelling of the cornea (p < 0.0001) was also found to occur immediately after waking. Highly significant associations were found between the diurnal variation in corneal thickness and the changes in corneal curvature. ----- Conclusions: Significant diurnal variation occurs in the regional thickness and the shape of the anterior and posterior cornea. The largest changes in the cornea were typically evident upon waking. The observed non-uniform regional corneal thickness changes resulted in a steepening of the posterior cornea, and a flattening of the anterior cornea to occur at this time.