2 resultados para correlation effect
em Universidade Complutense de Madrid
Resumo:
We study theoretically the effect of a new type of blocklike positional disorder on the effective electromagnetic properties of one-dimensional chains of resonant, high-permittivity dielectric particles, where particles are arranged into perfectly well-ordered blocks whose relative position is a random variable. This creates a finite order correlation length that mimics the situation encountered in metamaterials fabricated through self-assembled techniques, whose structures often display short-range order between near neighbors but long-range disorder, due to stacking defects. Using a spectral theory approach combined with a principal component statistical analysis, we study, in the long-wavelength regime, the evolution of the electromagnetic response when the composite filling fraction and the block size are changed. Modifications in key features of the resonant response (amplitude, width, etc.) are investigated, showing a regime transition for a filling fraction around 50%.
Resumo:
Purpose: to determine whether pupil dilation affects biometric measurements and intraocular lens (IOL) power calculation made using the new swept-source optical coherence tomography-based optical biometer (IOLMaster 700©; Carl Zeiss Meditec, Jena, Germany). Procedures: eighty-one eyes of 81 patients evaluated for cataract surgery were prospectively examined using the IOLMaster 700© before and after pupil dilation with tropicamide 1%. The measurements made were: axial length (AL), central corneal thickness (CCT), aqueous chamber depth (ACD), lens thickness (LT), mean keratometry (MK), white-to-white distance (WTW) and pupil diameter (PD). Holladay II and SRK/T formulas were used to calculate IOL power. Agreement between measurement modes (with and without dilation) was assessed through intraclass correlation coefficients (ICC) and Bland-Altman plots. Results: mean patient age was 75.17 ± 7.54 years (range: 57–92). Of the variables determined, CCT, ACD, LT and WTW varied significantly according to pupil dilation. Excellent intraobserver correlation was observed between measurements made before and after pupil dilation. Mean IOL power calculation using the Holladay 2 and SRK/T formulas were unmodified by pupil dilation. Conclusions: the use of pupil dilation produces statistical yet not clinically significant differences in some IOLMaster 700© measurements. However, it does not affect mean IOL power calculation.