6 resultados para extracorporeal
em National Center for Biotechnology Information - NCBI
Resumo:
Objective: To compare the resource implications and short term outcomes of extracorporeal membrane oxygenation and conventional management for term babies with severe respiratory failure.
Resumo:
Ischaemic cerebral accidents are frequent following extracorporeal membrane oxygenation (ECMO), especially after fixing the reinjection cannula in the right primitive carotid artery, which leads to an interruption in downstream flow. We describe a rare and unusual symptom of cerebral ischaemic accident that is known as Capgras syndrome. This feature is interesting because it may be documented by computed tomography (CT) scan and particular electroencephalography signals. It appears that our observation represents the first documented case of Capgras syndrome complicating ECMO. This incident emphasizes the potential hazards associated with right common artery ligature for venoarterial extracorporeal membrane oxygenation (VAECMO). In addition, it shows that this psychiatric symptom (that has been interpreted psychodynamically for many years) can have an organic basis, which should be studied.
Resumo:
Efficient and safe heparin anticoagulation has remained a problem for continuous renal replacement therapies and intermittent hemodialysis for patients with acute renal failure. To make heparin therapy safer for the patient with acute renal failure at high risk of bleeding, we have proposed regional heparinization of the circuit via an immobilized heparinase I filter. This study tested a device based on Taylor-Couette flow and simultaneous separation/reaction for efficacy and safety of heparin removal in a sheep model. Heparinase I was immobilized onto agarose beads via cyanogen bromide activation. The device, referred to as a vortex flow plasmapheretic reactor, consisted of two concentric cylinders, a priming volume of 45 ml, a microporous membrane for plasma separation, and an outer compartment where the immobilized heparinase I was fluidized separately from the blood cells. Manual white cell and platelet counts, hematocrit, total protein, and fibrinogen assays were performed. Heparin levels were indirectly measured via whole-blood recalcification times (WBRTs). The vortex flow plasmapheretic reactor maintained significantly higher heparin levels in the extracorporeal circuit than in the sheep (device inlet WBRTs were 1.5 times the device outlet WBRTs) with no hemolysis. The reactor treatment did not effect any physiologically significant changes in complete blood cell counts, platelets, and protein levels for up to 2 hr of operation. Furthermore, gross necropsy and histopathology did not show any significant abnormalities in the kidney, liver, heart, brain, and spleen.