The Use of a Pressure-Indicating Sensor Film to Provide Feedback upon Hydrogel-Forming Microneedle Array Self-Application In Vivo


Autoria(s): Vicente-Perez, Eva M.; Quinn, Helen L.; McAlister, Emma; O' Neill, Shannon; Hanna, Lezley-Anne; Barry, Johanne G.; Donnelly, Ryan F.
Data(s)

15/09/2016

Resumo

PURPOSE:<br/>To evaluate the combination of a pressure-indicating sensor film with hydrogel-forming microneedle arrays, as a method of feedback to confirm MN insertion in vivo.<br/>METHODS:<br/>Pilot in vitro insertion studies were conducted using a Texture Analyser to insert MN arrays, coupled with a pressure-indicating sensor film, at varying forces into excised neonatal porcine skin. In vivo studies involved twenty human volunteers, who self-applied two hydrogel-forming MN arrays, one with a pressure-indicating sensor film incorporated and one without. Optical coherence tomography was employed to measure the resulting penetration depth and colorimetric analysis to investigate the associated colour change of the pressure-indicating sensor film.<br/>RESULTS:<br/>Microneedle insertion was achieved in vitro at three different forces, demonstrating the colour change of the pressure-indicating sensor film upon application of increasing pressure. When self-applied in vivo, there was no significant difference in the microneedle penetration depth resulting from each type of array, with a mean depth of 237 μm recorded. When the pressure-indicating sensor film was present, a colour change occurred upon each application, providing evidence of insertion.<br/>CONCLUSIONS:<br/>For the first time, this study shows how the incorporation of a simple, low-cost pressure-indicating sensor film can indicate microneedle insertion in vitro and in vivo, providing visual feedback to assure the user of correct application. Such a strategy may enhance usability of a microneedle device and, hence, assist in the future translation of the technology to widespread clinical use.

Formato

application/pdf

Identificador

http://pure.qub.ac.uk/portal/en/publications/the-use-of-a-pressureindicating-sensor-film-to-provide-feedback-upon-hydrogelforming-microneedle-array-selfapplication-in-vivo(9bb4aa6a-f9fc-458c-95b5-07135979424c).html

http://dx.doi.org/10.1007/s11095-016-2032-z

http://pure.qub.ac.uk/ws/files/102239410/art_3A10.1007_2Fs11095_016_2032_z.pdf

Idioma(s)

eng

Direitos

info:eu-repo/semantics/openAccess

Fonte

Vicente-Perez , E M , Quinn , H L , McAlister , E , O' Neill , S , Hanna , L-A , Barry , J G & Donnelly , R F 2016 , ' The Use of a Pressure-Indicating Sensor Film to Provide Feedback upon Hydrogel-Forming Microneedle Array Self-Application In Vivo ' Pharmaceutical Research . DOI: 10.1007/s11095-016-2032-z

Palavras-Chave #Application #Feedback #Microneedles #Pressure-indicating sensor film #Transdermal
Tipo

article