Embedded progressive-three-layered fiber long-period gratings for respiratory monitoring


Autoria(s): Allsop, Thomas D.P.; Earthrowl-Gould, T.; Webb, David J.; Bennion, Ian
Data(s)

01/07/2003

Resumo

A long-period grating (LPG) was written into a progressive three-layered single-mode fiber that was embedded into a flexible platform as a curvature sensor. The spectral location and profile of the LPGs were unaltered after implantation in the platform. The curvature sensitivity was 3.747 nm m with a resolution of ±1.1×10–2 m–1. The bend sensor is intended to be part of a respiratory monitoring system and was tested on a resuscitation training manikin.

Formato

application/pdf

Identificador

http://eprints.aston.ac.uk/6422/1/Embedded_progressive-three-layered_fiber_long-period_gratings_for_respiratory_monitoring.pdf

Allsop, Thomas D.P.; Earthrowl-Gould, T.; Webb, David J. and Bennion, Ian (2003). Embedded progressive-three-layered fiber long-period gratings for respiratory monitoring. Journal of biomedical optics, 8 (3), pp. 552-558.