Embedded progressive-three-layered fiber long-period gratings for respiratory monitoring
Data(s) |
01/07/2003
|
---|---|
Resumo |
A long-period grating (LPG) was written into a progressive three-layered single-mode fiber that was embedded into a flexible platform as a curvature sensor. The spectral location and profile of the LPGs were unaltered after implantation in the platform. The curvature sensitivity was 3.747 nm m with a resolution of ±1.1×10–2 m–1. The bend sensor is intended to be part of a respiratory monitoring system and was tested on a resuscitation training manikin. |
Formato |
application/pdf |
Identificador |
Allsop, Thomas D.P.; Earthrowl-Gould, T.; Webb, David J. and Bennion, Ian (2003). Embedded progressive-three-layered fiber long-period gratings for respiratory monitoring. Journal of biomedical optics, 8 (3), pp. 552-558. |
Relação |
http://dx.doi.org/10.1117/1.1578497 http://eprints.aston.ac.uk/6422/ |
Tipo |
Article PeerReviewed |