A magnetization mapping approach for passive shim design in MRI
Contribuinte(s) |
A. Dhawan |
---|---|
Data(s) |
01/01/2006
|
Resumo |
A new passive shim design method is presented which is based on a magnetization mapping approach. Well defined regions with similar magnetization values define the optimal number of passive shims, their shape and position. The new design method is applied in a shimming process without prior-axial shim localization; this reduces the possibility of introducing new errors. The new shim design methodology reduces the number of iterations and the quantity of material required to shim a magnet. Only a few iterations (1-5) are required to shim a whole body horizontal bore magnet with a manufacturing error tolerance larger than 0.1 mm and smaller than 0.5 mm. One numerical example is presented |
Identificador | |
Idioma(s) |
eng |
Publicador |
IEEE |
Palavras-Chave | #Design engineering #Magnetisation #Magnets #Manufacturing processes #Shims #E1 #291500 Biomedical Engineering #671402 Medical instrumentation #0903 Biomedical Engineering |
Tipo |
Conference Paper |